Table. 3.

Table. 3.

Evaluation results of linear accelerator 6, 10, 15 MV

Item Energy (MV)

6 10 15
Average (%diff.) –0.48 –0.16 –0.59
Min. value (%diff.) –4.99 –4.71 –3.43
Max value (%diff.) 3.86 3.50 3.93
Standard deviation (%) 1.81 1.93 1.82
Progress in Medical Physics 2020;31:20-8 https://doi.org/10.14316/pmp.2020.31.2.20