검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Split Viewer

Original Article

Progress in Medical Physics 2017; 28(1): 27-32

Published online March 31, 2017

https://doi.org/10.14316/pmp.2017.28.1.27

Copyright © Korean Society of Medical Physics.

Dosimetric Effects of Low Dose 4D CT Using a Commercial Iterative Reconstruction on Dose Calculation in Radiation Treatment Planning: A Phantom Study

Hee Jung Kim, Sung Yong Park, Young Hee Park, Ah Ram Chang

Department of Radiation Oncology, Soonchunhyang University Hospital, Seoul, Korea

Correspondence to:

Ah Ram Chang (changaram@schmc.ac.kr)
Tel: 82-2-709-9412
Fax: 82-2-709-9092

Received: March 24, 2017; Revised: March 28, 2017; Accepted: March 28, 2017

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

We investigated the effect of a commercial iterative reconstruction technique (iDose, Philips) on the image quality and the dose calculation for the treatment plan. Using the electron density phantom, the 3D CT images with five different protocols (50, 100, 200, 350 and 400 mAs) were obtained. Additionally, the acquired data was reconstructed using the iDose with level 5. A lung phantom was used to acquire the 4D CT with the default protocol as a reference and the low dose (one third of the default protocol) 4D CT using the iDose for the spine and lung plans. When applying the iDose at the same mAs, the mean HU value was changed up to 85 HU. Although the 1 SD was increased with reducing the CT dose, it was decreased up to 4 HU due to the use of iDose. When using the low dose 4D CT with iDose, the dose change relative to the reference was less than 0.5% for the target and OARs in the spine plan. It was also less than 1.1% in the lung plan. Therefore, our results suggests that this dose reduction technique is applicable to the 4D CT image acquisition for the radiation treatment planning.

KeywordsLow dose 4D CT, Iterative reconstruction, Treatment planning, CT dose reduction

In recent years, many studies considering the respiratory motion during radiation treatment has been published.13) Since the normal breathing can cause the motion artifact when scanning the simulation CT and patients are hard to maintain a stable breath-hold, the most effective method is to generate the CT volume images for the entire respiratory cycle of free-breathing. The most common method is referred to as 4-dimensional CT (4D CT) that is correlated to the respiratory signal acquiring by the external surrogate such as Varian Real-time Position Management (RPM) system (Varian Medical Systems Inc, Palo Alto, CA, USA) or Philips Bellows belt system (Philips Medical Systems, Cleveland, OH, USA). The 4D CT is based on a very slow scanning and retrospectively the acquired data is binned for each phase. Due to this acquisition process of 4D CT, the use of higher CT imaging dose or CT dose is necessary than the conventional CT.47)

To adhere to ALARA (as low as reasonably achievable) principle in CT imaging, several novel reconstruction algorithms have been introduced in order to reduce the dose from CT scanning and to improve the image quality.810) As one of them, Philips provides a 4th generation iterative reconstruction algorithm, iDose (Philips Medical Systems, Cleveland, OH, USA).1113) Although the image contrast increment by using the retrospective reconstruction algorithm is beneficial for a diagnostic CT, the verification of the HU changes is required to use for the dose calculation in treatment planning using the correlation table of HU to material density.

The 4D CT has been commonly used for the radiation treatment planning. Thus, the purpose of this study is to quantify the image quality change due to the low dose 4D CT scanning with the commercial iterative reconstruction algorithm using a phantom. The effect on the dose distribution was also evaluated.

The CIRS electron density phantom (Computerized Imaging Reference Systems, Inc., Norfolk, VA, USA) with various density materials (e.g., lung 0.2 g/cc, lung 0.5 g/cc, adipose 0.96 g/cc, breast 0.99 g/cc, muscle 1.06 g/cc, liver 1.07 g/cc, bone 1.16 g/cc, and bone core 1.53 g/cc) and the CIRS Xsight® lung tracking phantom were used. All CT images for this study was acquired by Philips Brilliance Big Bore CT scanner (Philips Medical Systems, Cleveland, OH, USA). The scanner default protocol provided by a manufacturer is the use of 120 kVp and 350 mAs. To compare the HU values depending on the CT imaging dose, five different protocols (50 mAs, 100 mAs, 200 mAs, 350 mAs and 400 mAs) was applied in order to acquire the CT images of the electron density phantom. After that, the iDose with level 5 was applied to the acquired data.

To perform the treatment planning for lung and spine target, the CT images of the lung phantom were acquired using the default 4D CT scan protocol (600 mAs) without applying iDose and the low dose protocol (200 mAs) with applying iDose with level 5. The iDose algorithm was applied retrospectively after CT scanning.

For the spine and lung planning, the Pinnacle v9.6 treatment planning system (Philips Healthcare, Fitchburg, WI, USA) and the 6 MV photon beam were used. The 2-dimensional (2D) parallel-opposite anterior-posterior (AP-PA) plan and the volumetric modulated arc therapy (VMAT) plan were performed for the spine target. The 3-dimensional conformal radiotherapy (3D-CRT) plan and VMAT plan were also performed for the lung target. The prescription dose was 30 Gy in 10 fractions for the spine plans and 60 Gy in 30 fractions for the lung plans.

The maximum, minimum, mean and 1 standard deviation (1 SD) of HU were evaluated and the HU changes on the CT images after applying iDose were calculated. To delineate the region of interests inside the various density materials and to calculate that HUs, the MIM software (MIM Software Inc., Cleveland, OH, USA) was used.

Overall, as the tube current-time product (mAs) increased, the maximum HU was increased and the minimum HU was decreased. As shown in Fig. 1, the mean HU was changed due to the CT dose reduction and the use of iDose. The amount of the change had a range from −21 HU to 87 HU. When using the same current (mAs), the mean change had a range from −20 HU to 85HU. The HU had no change (0 HU) in the muscle 1.06 g/cc, but it had a maximum change (85 HU) in the bone core 1.53 g/cc. In the lowest density material (lung 0.2 g/cc), it showed 19 HU change on average.

Fig. 2 shows that as the CT dose decreases, the 1 SD increased by up to 20 HU but it significantly decreased by up to 4 HU due to the use of iDose (p-value <0.001).

Two 4D CT images of the lung phantom were acquired using the default scan parameter (600 mAs) without iDose and the low dose scan parameter (200 mAs) with iDose (Fig. 3). The dose-volume histograms (DVHs) are shown in Fig. 4, 5, 6 and 7. Table 1 shows the dose-volumetric parameters from DVHs for PTV and OARs. For the spine 2D and VMAT plan, the maximum dose differences relative to the prescription dose were less than 0.5%. For the lung 3D conformal and VMAT plan, they were less than 1.1%.

The CT radiation dose depends on the tube current (mA) and slice scan time and it is proportioned to the tube current-time product (mAs). In the result of this study, the 1 SD referred to as the image noise was increased with decreasing the CT dose. However, the use of a commercial iterative reconstruction algorithm, iDose, changed the HU values. From decreasing the 1 SD, the image noise was improved. This result is similar as the literatures.1114) Although the HU was increased in the high density material like bone, the dose change in the plans for the spine target was less than 0.5%. For the lung plans considering the low density around the target (known as a heterogeneity), the change due to the iDose was less than 1.1%. The above results demonstrate that by using the iDose the clinically acceptable quality 4D CT images could be obtained even if the CT dose is reduced to one third. Future studies using the patient 4D CT images would be helpful to investigate the optimal tube current-time product that might be changed depending on the CT scanning area of the body.

Although the iDose has a level from 1 to 7, the level 5 of the iDose was used in this study. As the iDose level increases, the strength of noise decreases independent of the CT dose. However, the noise is increased associated with a dose reduction. The net change in noise is determined from a combination of the dose reduction and the iDose level. For that reason, in this study, the one third dose reduction and iDose level of 5 was selected by using the relevant multiplication factor table for the estimation of the net change in noise. Since the image quality related to the noise strength depends on the iDose level, a detailed investigation for the different iDose level would be necessary if the dose is reduced differently comparing to this study (one third of the default protocol).

In this study, a single CT-to-density table was used for dose calculation on both images with and without iDose. For example, like pediatric patients, some cases should use the low energy X-ray. In fact, since the attenuation coefficient is dependent on the X-ray energy and the material density, the HU is significantly changed as shown in Fig. 6. The effect of the use of iDose could be different depending on the X-ray energy. In this study, as a simple test, the 90 kVp CT was compared to the default protocol 120 kVp CT. Fig. 6 shows that the mean and 1 SD change in HU due to the use of iDose were increased with decreasing the energy from 120 kVp to 90 kVp. Compared to the default protocol, the HU was changed up to 347 HU depending on the kVp and mAs. It was also changed up to 109 HU due to the use of iDose. Thus, when a single CT-to-density table is applied to the dose calculation for the different energy CT scan, the calculated dose can be inaccurate more than the result of this study. After applying the CT-to-density table depending on the X-ray energy, the optimal tube current-time product when using the iDose should be determined.

In this study, the HU change and the dosimetric change on the 4D CT due to the use of the commercial iterative reconstruction algorithm (iDose) were analyzed. Although, in general, the 4D CT needs the radiation exposure much more than 3D CT, the iDose could reduce the imaging dose less than the conventional 3D CT dose with the default protocol, preserving the image quality. The effect of HU change due to the use of the iDose for the low dose 4D CT on the dose calculation was clinically acceptable.

All relevant data are within the paper and its supporting information files.

Dose-volumetric parameters of target (PTV) and normal tissues (left and right lung, spinal cord) for spine and lung plan.

SpineLung


2D (AP-PA)VMAT3D CRTVMAT




DefaultiDoseDefaultiDoseDefaultiDoseDefaultiDose
PTV
 D95% (Gy)30.130.029.929.860.360.060.259.6
 D5% (Gy)31.030.932.132.064.464.265.164.5
 Minimum (Gy)29.929.824.224.258.257.658.357.7
 Maximum (Gy)31.331.232.932.864.764.565.765.1
 Mean (Gy)30.530.431.331.262.662.363.262.6
Left lung
 V20Gy (%)0.00.00.00.010.610.69.39.2
 Mean (Gy)0.30.30.90.96.56.56.06.0
Right lung
 V20Gy (%)0.00.00.00.00.00.00.00.0
 Mean (Gy)0.30.30.90.90.90.91.31.3
Spinal cord
 V13.5Gy (%)27.827.88.88.80.20.20.00.0
 Maximum (Gy)31.631.528.628.513.713.79.59.5
  1. Balter JM, Ten Haken RK, Lawrence TS, Lam KL, and Robertson JM. Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing. Int J Radiat Oncol Biol Phys 1996;36:167-174.
    Pubmed CrossRef
  2. Keall PJ, Starkschall G, Shukla H, Forster KM, Ortiz V, and Stevens CW, et al. Acquiring 4D thoracic CT scans using a multislice helical method. Phys Med Biol 2004;49:2053-2067.
    Pubmed CrossRef
  3. Ohara K, Okumura T, Akisada M, Inada T, Mori T, and Yokota H, et al. Irradiation synchronized with respiration gate. Int J Radiat Oncol Biol Phys 1989;17:853-857.
    Pubmed CrossRef
  4. Hubbard P, Callahan J, Cramb J, Budd R, and Kron T. Audit of radiation dose delivered in time-resolved four-dimensional computed tomography in a radiotherapy department. J Med Imaging Radiat Oncol 2015;59:346-352.
    Pubmed CrossRef
  5. Low DA, Nystrom M, Kalinin E, Parikh P, Dempsey JF, and Bradley JD, et al. A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. Med Phys 2003;30:1254-1263.
    Pubmed CrossRef
  6. Matsuzaki Y, Fujii K, Kumagai M, Tsuruoka I, and Mori S. Effective and organ doses using helical 4DCT for thoracic and abdominal therapies. J Radiat Res 2013;54:962-970.
    Pubmed KoreaMed CrossRef
  7. Mori S, Ko S, Ishii T, and Nishizawa K. Effective doses in four-dimensional computed tomography for lung radiotherapy planning. Med Dosim 2009;34:87-90.
    Pubmed CrossRef
  8. Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, and Pavlicek W. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 2009;193:764-771.
    Pubmed CrossRef
  9. Kim Y, Kim YK, Lee BE, Lee SJ, Ryu YJ, and Lee JH, et al. Ultra-low-dose CT of the thorax using iterative reconstruction: evaluation of image quality and radiation dose reduction. American Journal of Roentgenology 2015;204:1197-1202.
    Pubmed CrossRef
  10. Raman SP, Johnson PT, Deshmukh S, Mahesh M, Grant KL, and Fishman EK. CT dose reduction applications: available tools on the latest generation of CT scanners. Journal of the American College of Radiology 2013;10:37-41.
    Pubmed CrossRef
  11. Arapakis I, Efstathopoulos E, Tsitsia V, Kordolaimi S, Economopoulos N, and Argentos S, et al. Using “iDose4” iterative reconstruction algorithm in adults’ chest–abdomen–pelvis CT examinations: effect on image quality in relation to patient radiation exposure. The British journal of radiology 2014;87:20130613.
    Pubmed KoreaMed CrossRef
  12. Mehta D, Thompson R, Morton T, Dhanantwari A, and Shefer E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2013;2:147-155.
  13. Olsson M-L, and Norrgren K. An investigation of the iterative reconstruction method iDose4 on a Philips CT Brilliance 64 using a Catphan 600 phantom. SPIE Medical Imaging 2012:831348.
  14. Li H, Lee D, Low D, Gay H, Michalski J, and Mutic S. SU-E-J-167: Iterative Reconstruction Techniques for Radiation Therapy CT Simulations: A Phantom Study. Medical Physics 2013;40:189-189.
    CrossRef

Article

Original Article

Progress in Medical Physics 2017; 28(1): 27-32

Published online March 31, 2017 https://doi.org/10.14316/pmp.2017.28.1.27

Copyright © Korean Society of Medical Physics.

Dosimetric Effects of Low Dose 4D CT Using a Commercial Iterative Reconstruction on Dose Calculation in Radiation Treatment Planning: A Phantom Study

Hee Jung Kim, Sung Yong Park, Young Hee Park, Ah Ram Chang

Department of Radiation Oncology, Soonchunhyang University Hospital, Seoul, Korea

Correspondence to:

Ah Ram Chang (changaram@schmc.ac.kr)
Tel: 82-2-709-9412
Fax: 82-2-709-9092

Received: March 24, 2017; Revised: March 28, 2017; Accepted: March 28, 2017

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

We investigated the effect of a commercial iterative reconstruction technique (iDose, Philips) on the image quality and the dose calculation for the treatment plan. Using the electron density phantom, the 3D CT images with five different protocols (50, 100, 200, 350 and 400 mAs) were obtained. Additionally, the acquired data was reconstructed using the iDose with level 5. A lung phantom was used to acquire the 4D CT with the default protocol as a reference and the low dose (one third of the default protocol) 4D CT using the iDose for the spine and lung plans. When applying the iDose at the same mAs, the mean HU value was changed up to 85 HU. Although the 1 SD was increased with reducing the CT dose, it was decreased up to 4 HU due to the use of iDose. When using the low dose 4D CT with iDose, the dose change relative to the reference was less than 0.5% for the target and OARs in the spine plan. It was also less than 1.1% in the lung plan. Therefore, our results suggests that this dose reduction technique is applicable to the 4D CT image acquisition for the radiation treatment planning.

Keywords: Low dose 4D CT, Iterative reconstruction, Treatment planning, CT dose reduction

Introduction

In recent years, many studies considering the respiratory motion during radiation treatment has been published.13) Since the normal breathing can cause the motion artifact when scanning the simulation CT and patients are hard to maintain a stable breath-hold, the most effective method is to generate the CT volume images for the entire respiratory cycle of free-breathing. The most common method is referred to as 4-dimensional CT (4D CT) that is correlated to the respiratory signal acquiring by the external surrogate such as Varian Real-time Position Management (RPM) system (Varian Medical Systems Inc, Palo Alto, CA, USA) or Philips Bellows belt system (Philips Medical Systems, Cleveland, OH, USA). The 4D CT is based on a very slow scanning and retrospectively the acquired data is binned for each phase. Due to this acquisition process of 4D CT, the use of higher CT imaging dose or CT dose is necessary than the conventional CT.47)

To adhere to ALARA (as low as reasonably achievable) principle in CT imaging, several novel reconstruction algorithms have been introduced in order to reduce the dose from CT scanning and to improve the image quality.810) As one of them, Philips provides a 4th generation iterative reconstruction algorithm, iDose (Philips Medical Systems, Cleveland, OH, USA).1113) Although the image contrast increment by using the retrospective reconstruction algorithm is beneficial for a diagnostic CT, the verification of the HU changes is required to use for the dose calculation in treatment planning using the correlation table of HU to material density.

The 4D CT has been commonly used for the radiation treatment planning. Thus, the purpose of this study is to quantify the image quality change due to the low dose 4D CT scanning with the commercial iterative reconstruction algorithm using a phantom. The effect on the dose distribution was also evaluated.

Materials and Methods

The CIRS electron density phantom (Computerized Imaging Reference Systems, Inc., Norfolk, VA, USA) with various density materials (e.g., lung 0.2 g/cc, lung 0.5 g/cc, adipose 0.96 g/cc, breast 0.99 g/cc, muscle 1.06 g/cc, liver 1.07 g/cc, bone 1.16 g/cc, and bone core 1.53 g/cc) and the CIRS Xsight® lung tracking phantom were used. All CT images for this study was acquired by Philips Brilliance Big Bore CT scanner (Philips Medical Systems, Cleveland, OH, USA). The scanner default protocol provided by a manufacturer is the use of 120 kVp and 350 mAs. To compare the HU values depending on the CT imaging dose, five different protocols (50 mAs, 100 mAs, 200 mAs, 350 mAs and 400 mAs) was applied in order to acquire the CT images of the electron density phantom. After that, the iDose with level 5 was applied to the acquired data.

To perform the treatment planning for lung and spine target, the CT images of the lung phantom were acquired using the default 4D CT scan protocol (600 mAs) without applying iDose and the low dose protocol (200 mAs) with applying iDose with level 5. The iDose algorithm was applied retrospectively after CT scanning.

For the spine and lung planning, the Pinnacle v9.6 treatment planning system (Philips Healthcare, Fitchburg, WI, USA) and the 6 MV photon beam were used. The 2-dimensional (2D) parallel-opposite anterior-posterior (AP-PA) plan and the volumetric modulated arc therapy (VMAT) plan were performed for the spine target. The 3-dimensional conformal radiotherapy (3D-CRT) plan and VMAT plan were also performed for the lung target. The prescription dose was 30 Gy in 10 fractions for the spine plans and 60 Gy in 30 fractions for the lung plans.

The maximum, minimum, mean and 1 standard deviation (1 SD) of HU were evaluated and the HU changes on the CT images after applying iDose were calculated. To delineate the region of interests inside the various density materials and to calculate that HUs, the MIM software (MIM Software Inc., Cleveland, OH, USA) was used.

Results

Overall, as the tube current-time product (mAs) increased, the maximum HU was increased and the minimum HU was decreased. As shown in Fig. 1, the mean HU was changed due to the CT dose reduction and the use of iDose. The amount of the change had a range from −21 HU to 87 HU. When using the same current (mAs), the mean change had a range from −20 HU to 85HU. The HU had no change (0 HU) in the muscle 1.06 g/cc, but it had a maximum change (85 HU) in the bone core 1.53 g/cc. In the lowest density material (lung 0.2 g/cc), it showed 19 HU change on average.

Fig. 2 shows that as the CT dose decreases, the 1 SD increased by up to 20 HU but it significantly decreased by up to 4 HU due to the use of iDose (p-value <0.001).

Two 4D CT images of the lung phantom were acquired using the default scan parameter (600 mAs) without iDose and the low dose scan parameter (200 mAs) with iDose (Fig. 3). The dose-volume histograms (DVHs) are shown in Fig. 4, 5, 6 and 7. Table 1 shows the dose-volumetric parameters from DVHs for PTV and OARs. For the spine 2D and VMAT plan, the maximum dose differences relative to the prescription dose were less than 0.5%. For the lung 3D conformal and VMAT plan, they were less than 1.1%.

Discussion

The CT radiation dose depends on the tube current (mA) and slice scan time and it is proportioned to the tube current-time product (mAs). In the result of this study, the 1 SD referred to as the image noise was increased with decreasing the CT dose. However, the use of a commercial iterative reconstruction algorithm, iDose, changed the HU values. From decreasing the 1 SD, the image noise was improved. This result is similar as the literatures.1114) Although the HU was increased in the high density material like bone, the dose change in the plans for the spine target was less than 0.5%. For the lung plans considering the low density around the target (known as a heterogeneity), the change due to the iDose was less than 1.1%. The above results demonstrate that by using the iDose the clinically acceptable quality 4D CT images could be obtained even if the CT dose is reduced to one third. Future studies using the patient 4D CT images would be helpful to investigate the optimal tube current-time product that might be changed depending on the CT scanning area of the body.

Although the iDose has a level from 1 to 7, the level 5 of the iDose was used in this study. As the iDose level increases, the strength of noise decreases independent of the CT dose. However, the noise is increased associated with a dose reduction. The net change in noise is determined from a combination of the dose reduction and the iDose level. For that reason, in this study, the one third dose reduction and iDose level of 5 was selected by using the relevant multiplication factor table for the estimation of the net change in noise. Since the image quality related to the noise strength depends on the iDose level, a detailed investigation for the different iDose level would be necessary if the dose is reduced differently comparing to this study (one third of the default protocol).

In this study, a single CT-to-density table was used for dose calculation on both images with and without iDose. For example, like pediatric patients, some cases should use the low energy X-ray. In fact, since the attenuation coefficient is dependent on the X-ray energy and the material density, the HU is significantly changed as shown in Fig. 6. The effect of the use of iDose could be different depending on the X-ray energy. In this study, as a simple test, the 90 kVp CT was compared to the default protocol 120 kVp CT. Fig. 6 shows that the mean and 1 SD change in HU due to the use of iDose were increased with decreasing the energy from 120 kVp to 90 kVp. Compared to the default protocol, the HU was changed up to 347 HU depending on the kVp and mAs. It was also changed up to 109 HU due to the use of iDose. Thus, when a single CT-to-density table is applied to the dose calculation for the different energy CT scan, the calculated dose can be inaccurate more than the result of this study. After applying the CT-to-density table depending on the X-ray energy, the optimal tube current-time product when using the iDose should be determined.

Conclusion

In this study, the HU change and the dosimetric change on the 4D CT due to the use of the commercial iterative reconstruction algorithm (iDose) were analyzed. Although, in general, the 4D CT needs the radiation exposure much more than 3D CT, the iDose could reduce the imaging dose less than the conventional 3D CT dose with the default protocol, preserving the image quality. The effect of HU change due to the use of the iDose for the low dose 4D CT on the dose calculation was clinically acceptable.

Conflicts of Interest

The authors have nothing to disclose.

Availability of Data and Materials

All relevant data are within the paper and its supporting information files.

Tables

Dose-volumetric parameters of target (PTV) and normal tissues (left and right lung, spinal cord) for spine and lung plan.

SpineLung


2D (AP-PA)VMAT3D CRTVMAT




DefaultiDoseDefaultiDoseDefaultiDoseDefaultiDose
PTV
 D95% (Gy)30.130.029.929.860.360.060.259.6
 D5% (Gy)31.030.932.132.064.464.265.164.5
 Minimum (Gy)29.929.824.224.258.257.658.357.7
 Maximum (Gy)31.331.232.932.864.764.565.765.1
 Mean (Gy)30.530.431.331.262.662.363.262.6
Left lung
 V20Gy (%)0.00.00.00.010.610.69.39.2
 Mean (Gy)0.30.30.90.96.56.56.06.0
Right lung
 V20Gy (%)0.00.00.00.00.00.00.00.0
 Mean (Gy)0.30.30.90.90.90.91.31.3
Spinal cord
 V13.5Gy (%)27.827.88.88.80.20.20.00.0
 Maximum (Gy)31.631.528.628.513.713.79.59.5

Fig 1.

Figure 1.Mean change in HU depending on the CT scan parameter for different density materials relative to the default scan parameter.
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Fig 2.

Figure 2.1 Standard deviation change in HU depending on the CT scan parameter for different density materials relative to the default scan parameter.
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Fig 3.

Figure 3.4D CT images of the lung phantom acquired using the default scan parameter (600 mAs) without iDose (left) and the low dose scan parameter (200 mAs) with iDose (right).
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Fig 4.

Figure 4.DVH comparison between the dose calculations on the CT images reconstructed with and without using iDose for the Spine 2D parallel-opposite (AP-PA) and VMAT plan.
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Fig 5.

Figure 5.DVH comparison between the dose calculations on the CT images reconstructed with and without using iDose for the Lung 3D conformal and VMAT plan.
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Fig 6.

Figure 6.Mean and 1 standard deviation changes in HU when using the x-ray energy of 90 kVp.
Progress in Medical Physics 2017; 28: 27-32https://doi.org/10.14316/pmp.2017.28.1.27

Table 1 Dose-volumetric parameters of target (PTV) and normal tissues (left and right lung, spinal cord) for spine and lung plan.

SpineLung


2D (AP-PA)VMAT3D CRTVMAT




DefaultiDoseDefaultiDoseDefaultiDoseDefaultiDose
PTV
 D95% (Gy)30.130.029.929.860.360.060.259.6
 D5% (Gy)31.030.932.132.064.464.265.164.5
 Minimum (Gy)29.929.824.224.258.257.658.357.7
 Maximum (Gy)31.331.232.932.864.764.565.765.1
 Mean (Gy)30.530.431.331.262.662.363.262.6
Left lung
 V20Gy (%)0.00.00.00.010.610.69.39.2
 Mean (Gy)0.30.30.90.96.56.56.06.0
Right lung
 V20Gy (%)0.00.00.00.00.00.00.00.0
 Mean (Gy)0.30.30.90.90.90.91.31.3
Spinal cord
 V13.5Gy (%)27.827.88.88.80.20.20.00.0
 Maximum (Gy)31.631.528.628.513.713.79.59.5

References

  1. Balter JM, Ten Haken RK, Lawrence TS, Lam KL, and Robertson JM. Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing. Int J Radiat Oncol Biol Phys 1996;36:167-174.
    Pubmed CrossRef
  2. Keall PJ, Starkschall G, Shukla H, Forster KM, Ortiz V, and Stevens CW, et al. Acquiring 4D thoracic CT scans using a multislice helical method. Phys Med Biol 2004;49:2053-2067.
    Pubmed CrossRef
  3. Ohara K, Okumura T, Akisada M, Inada T, Mori T, and Yokota H, et al. Irradiation synchronized with respiration gate. Int J Radiat Oncol Biol Phys 1989;17:853-857.
    Pubmed CrossRef
  4. Hubbard P, Callahan J, Cramb J, Budd R, and Kron T. Audit of radiation dose delivered in time-resolved four-dimensional computed tomography in a radiotherapy department. J Med Imaging Radiat Oncol 2015;59:346-352.
    Pubmed CrossRef
  5. Low DA, Nystrom M, Kalinin E, Parikh P, Dempsey JF, and Bradley JD, et al. A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing. Med Phys 2003;30:1254-1263.
    Pubmed CrossRef
  6. Matsuzaki Y, Fujii K, Kumagai M, Tsuruoka I, and Mori S. Effective and organ doses using helical 4DCT for thoracic and abdominal therapies. J Radiat Res 2013;54:962-970.
    Pubmed KoreaMed CrossRef
  7. Mori S, Ko S, Ishii T, and Nishizawa K. Effective doses in four-dimensional computed tomography for lung radiotherapy planning. Med Dosim 2009;34:87-90.
    Pubmed CrossRef
  8. Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, and Pavlicek W. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 2009;193:764-771.
    Pubmed CrossRef
  9. Kim Y, Kim YK, Lee BE, Lee SJ, Ryu YJ, and Lee JH, et al. Ultra-low-dose CT of the thorax using iterative reconstruction: evaluation of image quality and radiation dose reduction. American Journal of Roentgenology 2015;204:1197-1202.
    Pubmed CrossRef
  10. Raman SP, Johnson PT, Deshmukh S, Mahesh M, Grant KL, and Fishman EK. CT dose reduction applications: available tools on the latest generation of CT scanners. Journal of the American College of Radiology 2013;10:37-41.
    Pubmed CrossRef
  11. Arapakis I, Efstathopoulos E, Tsitsia V, Kordolaimi S, Economopoulos N, and Argentos S, et al. Using “iDose4” iterative reconstruction algorithm in adults’ chest–abdomen–pelvis CT examinations: effect on image quality in relation to patient radiation exposure. The British journal of radiology 2014;87:20130613.
    Pubmed KoreaMed CrossRef
  12. Mehta D, Thompson R, Morton T, Dhanantwari A, and Shefer E. Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int J 2013;2:147-155.
  13. Olsson M-L, and Norrgren K. An investigation of the iterative reconstruction method iDose4 on a Philips CT Brilliance 64 using a Catphan 600 phantom. SPIE Medical Imaging 2012:831348.
  14. Li H, Lee D, Low D, Gay H, Michalski J, and Mutic S. SU-E-J-167: Iterative Reconstruction Techniques for Radiation Therapy CT Simulations: A Phantom Study. Medical Physics 2013;40:189-189.
    CrossRef
Korean Society of Medical Physics

Vol.35 No.3
September 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line