검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Split Viewer

Original Article

Progress in Medical Physics 2017; 28(3): 83-91

Published online September 30, 2017

https://doi.org/10.14316/pmp.2017.28.3.83

Copyright © Korean Society of Medical Physics.

Numerical Modeling and Experiment for Single Grid-Based Phase-Contrast X-Ray Imaging

Hyunwoo Lim*, Hunwoo Lee*, Hyosung Cho*, Changwoo Seo*, Sooyeul Lee, Byunggyu Chae

*Department of Radiation Convergence Engineering, Yonsei University, Wonju, Bio-Medical IT Convergence Research Division, ETRI, Daejeon, Korea

Correspondence to:Hyosung Cho

Received: July 31, 2017; Revised: September 4, 2017; Accepted: September 5, 2017

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

In this work, we investigated the recently proposed phase-contrast x-ray imaging (PCXI) technique, the so-called single grid-based PCXI, which has great simplicity and minimal requirements on the setup alignment. It allows for imaging of smaller features and variations in the examined sample than conventional attenuation-based x-ray imaging with lower x-ray dose. We performed a systematic simulation using a simulation platform developed by us to investigate the image characteristics. We also performed a preliminary PCXI experiment using an established a table-top setup to demonstrate the performance of the simulation platform. The system consists of an x-ray tube (50 kVp, 5 mAs), a focused-linear grid (200-lines/inch), and a flat-panel detector (48-μm pixel size). According to our results, the simulated contrast of phase images was much enhanced, compared to that of the absorption images. The scattering length scale estimated for a given simulation condition was about 117 nm. It was very similar, at least qualitatively, to the experimental contrast, which demonstrates the performance of the simulation platform. We also found that the level of the phase gradient of oriented structures strongly depended on the orientation of the structure relative to that of linear grids.

KeywordsPhase-contrast x-ray image, Simulation platform, X-ray grid, Image contrast

Conventional x-ray imaging techniques have widely been used for both medical and industrial imaging applications and have in common attenuation-based contrast which arises from differences in elemental composition, thickness, and density of the examined sample. However, they are often limited by low image contrast especially in imaging materials of low atomic number.1) One possible solution to the problem of limited contrast inherent to attenuation-based radiography is the application of phase-contrast x-ray imaging (PCXI) technique that utilizes the phase shift of the x-ray wavefront introduced by the sample under investigation to the transmitted x-rays. Because the variation in phase of x-rays is much larger than that in intensity due to attenuation, it can detect small features and variations in the sample that would be invisible in conventional attenuation-based radiography. Several techniques have been proposed to measure the phase shift, including analyzer-based imaging,2) propagation-based imaging,3) grating-based imaging,4,5)etc. However, most of the techniques put higher demands on the equipment, which has limited the widespread use of these techniques into many related applications.

In this work, we investigated the recently developed technique by Wen et al.,6,7) the so-called single grid-based PCXI, which has great simplicity and minimal requirements on the setup alignment. The information of the phase shift can be extracted by using Fourier processing.6) We developed a useful simulation platform for PCXI and performed a systematic simulation to investigate the image characteristics. We also performed a preliminary PCXI experiment using an established a table-top setup to demonstrate the performance of the simulation platform. In the following sections, we briefly describe the numerical modeling of the PCXI used in the simulation platform and present the results.

1. Numerical modeling for single grid-based PCXI

Fig. 1 shows the schematic illustration of a single grid-based PCXI setup in which an x-ray grid is placed midway between the x-ray source and the detector and a sample is placed ahead of the grid. As illustrated in Fig. 1, when x-rays from the source pass through a sample, the wavefront of the transmitted x-rays is distorted by the refraction of the x-rays due to the difference in the refractive indexes of the sample structures and its intensity is modulated by the periodic x-ray grid strips.

In x-ray physics, image contrast is generated due to the difference in complex refractive index n of the sample and described as follows:

n(x,y,z)=1-δ(x,y,z)+iβ(x,y,z),

where δ is the decrement of the real part of the refractive index responsible for phase shift of the x-rays and the imaginary part β describes the absorption index. When x-rays pass through a sample, not only their amplitude but their phase is altered as well. Because x-rays are a form of high-energy lights, they can be treated as electromagnetic waves and their propagation is described by the Helmholtz equation, assuming that monochromatic x-rays are propagated through free space:

2ψ(r)+k2ψ(r)=0,k=ωc,

where ∇2 is the Laplacian operator, ψ is a scalar wave function, κ is the wave number, ω is the angular frequency, and χ is the speed of light in vacuum. Fig. 2 shows the schematic illustration of a geometry for the free-space propagation of x-rays. The wavefront ψ (x,y,z) in the detector plane is calculated using the Huygens-Fresnel principle:8)

ψ(x,y,z)=1iλ-+ψ(x,y,0)eikrrcosθdxdy,ψ(x,y,0)=Ae-kβ(x,y,z)dz-ikδ(x,y,z)dz,

where λ is the wavelength of the x-rays, r is the distance between points in the sample and detector planes, θ is the angle illustrated in Fig. 2, and A is the wave amplitude. Assuming small diffraction angles (i.e., cos θ≈1) and, rzEq. (3) can be written as:

ψ(x,y,z)=eikziλz-+ψ(x,y,0)eik(x-x)2+(y-y)22zdxdy.

Neglecting the constant phase term eikz, Eq. (4) can also be expressed as the convolution with the propagation kernel, h(x,y), as follows:

ψ(x,y)=ψ(x,y,0)h(x,y),h(x,y)1iλzeik2z(x2+y2),

where ⊗⊗ indicates the two-dimensional (2D) convolution operator. In addition, considering the system response function (SRF), the image intensity of the sample, I, can be expressed as:

I=|ψ(x,y)|2SRF.

To model an x-ray grid, we considered an ideal linear grid in which the absorption by any interspace material is ignored. Fig. 3 shows the primary transmission, tgrid, of an ideal linear grid oriented in the vertical direction. It can be expressed mathematically as a one-dimensional (1D) square function using a Fourier series:9)

tgrid(x)=n=-+dPsinc(πdnP)e2πxnPi,

where P is the grid pitch and d is the width of the lead strips. Thus, the image intensity of the sample with a linear grid, fsg, can be expressed as:

fsg=I×tgrid(x).

The analysis of the phase shift in PCXI is described in detail in Ref.10,11) Two raw images of the sample with grid (fsg) and the bare grid (fg) are acquired separately and Fourier-transformed. In the Fourier domain, the areas surrounding the primary and the first harmonic peaks are selected using a band-pass filter separately and then inverse Fourier-transformed to yield the primary image (f0,s), i.e., retrieved absorption image, and the first harmonic image (f1,s) of the sample by normalizing with the bare grid images as:

f¯0,s=f0,sgf0,g,f¯1,s=f1,sgf1,g,

where f0,sg and f1,sg are the primary and first harmonic images of the sample with grid, respectively, while f0,g and f1,g are those of bare grid. To extract gradient phase information, the ratio of f1,s and I0,s is taken as:

f=f¯1,sf¯0,s.

Fig. 4 shows the simplified Fourier processing in the single grid-based PCXI to extract absorption image and differential phase image from the two raw images of the sample with grid (fsg) and the bare grid (fg).

The intensity of the phase image depends on the x-ray wavelength (λ), the grid period (p), and the system geometry. The size of the sample structure whose phase information can be appreciably detected is known to be limited to an upper threshold size,10) which is defined as the scattering length scale (L):

L=(SDD-SOD)·(SDD-GDD)SDDλp,

where SDD, SOD, and GDD is the source-to-detector distance, the source-to-object distance, and the grid-to-detector distance, respectively. The larger the scattering length scale, the larger the phase signal.

2. Simulation and experimental setup

We developed a useful simulation platform based on the above descriptions for PCXI study. Fig. 5 shows (a) the three-dimensional (3D) numerical chest phantom (478×258×434 voxels) in anterior-posterior (AP) positioning (left) and lateral (LA) positioning (right) and (b) the 3D Shepp-Logan phantom12) (400×400×400 voxels) used in the simulation. The chest phantom, originally produced by authors at the University of Pernambuco in Brazil,13) was defined by us with proper complex refractive indexes by segmentation into several compartments representing skin, muscle, bone, lung, fat, water, soft tissue, etc.14) In the simulation, monochromatic x-rays of 50 keV were used and the strip density of the used grid was 200 lines/inch. The corresponding wavelength of the x-rays and the grid period were about λ=0.25 Å and p=127 μm, respectively. SDD=200 cm, SOD=80 cm, and GDD=100 cm were used. The scattering length scale estimated for the given simulation condition was about 117 nm. The detailed conditions used in the simulation and experiment are listed in Table 1.

Fig. 6 shows the table-top setup established for the PCXI experiment. It consists of an x-ray tube (100-μm focal spot size, Oxford Ins., TF5011), a focused-linear grid (200-lines/inch strip density, JPI Healthcare Corp.), and a CMOS-type flat-panel detector having an active area of 14.5 cm×11.6 cm (48-μm pixel size, Rayence Corp., Xmaru1215). The same system geometry used in the simulation was applied in the experiment. More details of the experimental procedure can be found in our previous paper.15)

Fig. 7 shows the differential phase images of the chest phantom in AP positioning simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). The retrieved absorption image (top left) is also indicated as the reference. As indicated in Fig. 7, the contrast of the phase images was, as expected, much higher than that of the absorption images. In addition, a linear grid can detect phase gradient in only one direction perpendicular to the grid strips. Note that the segments of the structure boundary perpendicular to the grid direction are more highlighted in the differential phase images (see the spine and the clavicle for vertical grid and horizontal grid marked by arrows in Fig. 7, and vice versa). The level of the phase gradient of oriented structures depends on the orientation of the structure relative to that of linear grids.

One possible solution to the orientation problem of a linear grid is to employ more sophisticated grids such as crossed grids that consist of square cells to enable imaging of phase gradient in multiple directions with a single exposure. Kottler et al.16) showed that measurements in two orthogonal directions are often necessary to improve visibility and reduce artifacts in the image. Wen et al.17) recently described an x-ray differential phase-contrast imaging method based on 2D transmission gratings that were directly resolved by an x-ray detector and quantified the effects of x-ray refraction and diffraction in the sample through spatial harmonic analysis. The use of 2D gratings allows differential phase contrast in several directions to be obtained from a single image, which obviates the need for multiple exposures and separate measurements for different directions. In the study, instead, as indicated in Fig. 7, we simply summed the two retrieved differential phase images with a vertical grid and a horizontal grid, which gives phase gradient in a direction at 45 degrees to both the x and y directions.18) See the spine and the clavicle in the combined differential image marked by arrows in Fig. 7. Fig. 8 shows similar phase images of the chest phantom in LA positioning; see also the body of humerus and the ribcage indicated by arrows in Fig. 8.

For more quantitative analysis of the image characteristics of PCXI, we repeated the same simulation procedure using the 3D Shepp-Logan phantom. Fig. 9 shows the differential phase images of the Shepp-Logan phantom simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). Fig. 10 shows the intensity profiles measured along the line segments AB indicated in Fig. 9 for the differential phase image and the absorption image. As indicated in Fig. 10, the vertical grid emphasizes vertical edges by a horizontal gradient, detecting small features and variations in the sample that was not clearly visible in the absorption image (see the intensity variations marked by arrows).

Fig. 11 shows complete sets of the PCXI results retrieved from a single raw image of (a) animal bone and (b) chicken wing with a 200-lines/inch vertical grid obtained at the given x-ray tube conditions of 50 kVp and 5 mAs. The image contrast of the phase images was much enhanced, compared to that of the absorption images, and was similar, at least qualitatively, to the simulated contrast, indicating the performance of the developed simulation platform.

We successfully obtained phase-contrast x-ray images of much enhanced contrast, compared to conventional attenuation-based images, by using the single grid-based technique from both the simulation and experiment. The simulated contrast of the phase images was similar, at least qualitatively, to the experimental contrast, which demonstrates the performance of the developed simulation platform. The scattering length scale estimated for a given simulation condition was about 117 nm. Consequently, the simulation platform worked properly and demonstrated that the single grid-based approach seemed a useful method for PCXI with great simplicity and minimal requirements on the setup alignment. We expect that the simulation platform developed in this work will be useful for designing optimal PCXI systems. More quantitative evaluation of the image characteristics will be performed soon.

This work was supported by Institute for Information & Communications Technology Promotion (IITP) grant funded by the Korea government (MSIT) (2017-0-00049, Study on biomedical imaging and recognition-sensors for acquisition and analysis of high quality bio-information).

Imaging acquisition conditions used in the simulation and the experiment.

ParameterDimension
Source-to-object distance (SOD)80 cm
Object-to-grid distance (OGD)20 cm
Grid-to-detector distance (GDD)100 cm
Grid strip density200 lines/inch
Grid focal distance100 cm
Detector pixel size48 μm
Focal spot size0.1 mm
Tube voltage50 kV (monochromatic in simulation)
50 kVp (polychromatic in experiment)
SampleChest, Shepp-Logan (simulation)
Animal bone, chicken wing (experiment)
  1. Lewis R. Medical phase contrast x-ray imaging: current status and future prospects. Phys Med Biol 2004;49:3573-3583.
    Pubmed CrossRef
  2. Parham C, Zhong Z, Connor D, Chapman L, and Pisano E. Design and implementation of a compact low-dose diffraction enhanced medical imaging system. Acad Radiol 2009;16:911-917.
    Pubmed CrossRef
  3. Burvall A, Lundström U, Takman P, Larsson D, and Hertz H. Phase retrieval in X-ray phase-contrast imaging suitable for tomography. Opt Express 2011;19:10359-10376.
    Pubmed CrossRef
  4. Pfeiffer F, Bech M, Bunk O, Kraft P, Eikenberry E, and Brönnimann C, et al. Hard-X-ray dark-field imaging using a grating interferometer. Nat Mater 2008;7:134-137.
    Pubmed CrossRef
  5. Pfeiffer F, Herzen J, Willner M, Chabior M, Auweter S, and Reiser M, et al. Grating-based x-ray phase contrast for biomedical imaging applications. Z Med Phys 2013;23:176-185.
    Pubmed CrossRef
  6. Wen H, Bennett E, Kopace R, Stein A, and Pai V. Single-shot x-ray differential phase-contrast and diffraction imaging using two-dimensional transmission gratings. Opt Lett 2010;35:1932-1934.
    Pubmed KoreaMed CrossRef
  7. Wen H, Bennett E, Hegedus M, and Carroll S. Spatial harmonic imaging of x-ray Scattering-Initial results. IEEE Trans on Med Imaging 2008;27:997-1002.
    Pubmed KoreaMed CrossRef
  8. Goodman J. Introduction to Fourier optics. 3rd ed. Englewood: Roberts and Company Publishers 2005.
  9. Gauntt D, and Barnes G. Grid line artifact formation: A comprehensive theory. Med Phys 2006;33:1668-1677.
    Pubmed CrossRef
  10. Wen H, Bennett E, Hegedus M, and Rapacchi S. Fourier x-ray Scattering radiography yields bone structural information. Radiology 2009;251:910-918.
    Pubmed KoreaMed CrossRef
  11. Pfeiffer F, Bunk O, David C, Bech M, Duc G, and Bravin A, et al. High-resolution brain tumor visualization using three-dimensional x-ray phase contrast tomography. Phys Med Biol 2007;52:6923-6930.
    Pubmed CrossRef
  12. Shepp L, and Logan B. The Fourier reconstruction of a head section. IEEE Trans on Nucl Sci 1974;21:21-43.
    CrossRef
  13. Kramer R, Cassola V, Khoury H, Vieira J, Lima V, and Brown K. FASH and MASH: female and male adult human phantoms based on polygon mesh surfaces: II. Dosimetric calculations. Phys Med Biol 2010;55:163-189.
    Pubmed CrossRef
  14. Park Y, Cho H, Lim H, Je U, Park C, and Cho H, et al. Experimental Study for Phase-contrast X-ray Imaging Based on a Single Antiscatter Grid and a Polychromatic X-ray Source. Prog in Med Phys 2015;26:1-8.
    CrossRef
  15. Kottler C, David C, Pfeiffer F, and Bunk O. A two-directional approach for grating based differential phase contrast imaging using hard x-rays. Opt Express 2007;15:1175-1181.
    Pubmed CrossRef
  16. Wen H, Bennett E, Kopace R, Stein A, and Pai V. Single-shot x-ray differential phase contrast and diffraction imaging using two-dimensional transmission gratings. Opt Lett 2010;35:1932-1934.
    Pubmed KoreaMed CrossRef
  17. Tahir S. Grid-Based Fourier Transform Phase Contrast Imaging. PhD thesis: State University of New York; 2015.

Article

Original Article

Progress in Medical Physics 2017; 28(3): 83-91

Published online September 30, 2017 https://doi.org/10.14316/pmp.2017.28.3.83

Copyright © Korean Society of Medical Physics.

Numerical Modeling and Experiment for Single Grid-Based Phase-Contrast X-Ray Imaging

Hyunwoo Lim*, Hunwoo Lee*, Hyosung Cho*, Changwoo Seo*, Sooyeul Lee, Byunggyu Chae

*Department of Radiation Convergence Engineering, Yonsei University, Wonju, Bio-Medical IT Convergence Research Division, ETRI, Daejeon, Korea

Correspondence to:Hyosung Cho

Received: July 31, 2017; Revised: September 4, 2017; Accepted: September 5, 2017

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

In this work, we investigated the recently proposed phase-contrast x-ray imaging (PCXI) technique, the so-called single grid-based PCXI, which has great simplicity and minimal requirements on the setup alignment. It allows for imaging of smaller features and variations in the examined sample than conventional attenuation-based x-ray imaging with lower x-ray dose. We performed a systematic simulation using a simulation platform developed by us to investigate the image characteristics. We also performed a preliminary PCXI experiment using an established a table-top setup to demonstrate the performance of the simulation platform. The system consists of an x-ray tube (50 kVp, 5 mAs), a focused-linear grid (200-lines/inch), and a flat-panel detector (48-μm pixel size). According to our results, the simulated contrast of phase images was much enhanced, compared to that of the absorption images. The scattering length scale estimated for a given simulation condition was about 117 nm. It was very similar, at least qualitatively, to the experimental contrast, which demonstrates the performance of the simulation platform. We also found that the level of the phase gradient of oriented structures strongly depended on the orientation of the structure relative to that of linear grids.

Keywords: Phase-contrast x-ray image, Simulation platform, X-ray grid, Image contrast

Introduction

Conventional x-ray imaging techniques have widely been used for both medical and industrial imaging applications and have in common attenuation-based contrast which arises from differences in elemental composition, thickness, and density of the examined sample. However, they are often limited by low image contrast especially in imaging materials of low atomic number.1) One possible solution to the problem of limited contrast inherent to attenuation-based radiography is the application of phase-contrast x-ray imaging (PCXI) technique that utilizes the phase shift of the x-ray wavefront introduced by the sample under investigation to the transmitted x-rays. Because the variation in phase of x-rays is much larger than that in intensity due to attenuation, it can detect small features and variations in the sample that would be invisible in conventional attenuation-based radiography. Several techniques have been proposed to measure the phase shift, including analyzer-based imaging,2) propagation-based imaging,3) grating-based imaging,4,5)etc. However, most of the techniques put higher demands on the equipment, which has limited the widespread use of these techniques into many related applications.

In this work, we investigated the recently developed technique by Wen et al.,6,7) the so-called single grid-based PCXI, which has great simplicity and minimal requirements on the setup alignment. The information of the phase shift can be extracted by using Fourier processing.6) We developed a useful simulation platform for PCXI and performed a systematic simulation to investigate the image characteristics. We also performed a preliminary PCXI experiment using an established a table-top setup to demonstrate the performance of the simulation platform. In the following sections, we briefly describe the numerical modeling of the PCXI used in the simulation platform and present the results.

Materials and Methods

1. Numerical modeling for single grid-based PCXI

Fig. 1 shows the schematic illustration of a single grid-based PCXI setup in which an x-ray grid is placed midway between the x-ray source and the detector and a sample is placed ahead of the grid. As illustrated in Fig. 1, when x-rays from the source pass through a sample, the wavefront of the transmitted x-rays is distorted by the refraction of the x-rays due to the difference in the refractive indexes of the sample structures and its intensity is modulated by the periodic x-ray grid strips.

In x-ray physics, image contrast is generated due to the difference in complex refractive index n of the sample and described as follows:

n(x,y,z)=1-δ(x,y,z)+iβ(x,y,z),

where δ is the decrement of the real part of the refractive index responsible for phase shift of the x-rays and the imaginary part β describes the absorption index. When x-rays pass through a sample, not only their amplitude but their phase is altered as well. Because x-rays are a form of high-energy lights, they can be treated as electromagnetic waves and their propagation is described by the Helmholtz equation, assuming that monochromatic x-rays are propagated through free space:

2ψ(r)+k2ψ(r)=0,k=ωc,

where ∇2 is the Laplacian operator, ψ is a scalar wave function, κ is the wave number, ω is the angular frequency, and χ is the speed of light in vacuum. Fig. 2 shows the schematic illustration of a geometry for the free-space propagation of x-rays. The wavefront ψ (x,y,z) in the detector plane is calculated using the Huygens-Fresnel principle:8)

ψ(x,y,z)=1iλ-+ψ(x,y,0)eikrrcosθdxdy,ψ(x,y,0)=Ae-kβ(x,y,z)dz-ikδ(x,y,z)dz,

where λ is the wavelength of the x-rays, r is the distance between points in the sample and detector planes, θ is the angle illustrated in Fig. 2, and A is the wave amplitude. Assuming small diffraction angles (i.e., cos θ≈1) and, rzEq. (3) can be written as:

ψ(x,y,z)=eikziλz-+ψ(x,y,0)eik(x-x)2+(y-y)22zdxdy.

Neglecting the constant phase term eikz, Eq. (4) can also be expressed as the convolution with the propagation kernel, h(x,y), as follows:

ψ(x,y)=ψ(x,y,0)h(x,y),h(x,y)1iλzeik2z(x2+y2),

where ⊗⊗ indicates the two-dimensional (2D) convolution operator. In addition, considering the system response function (SRF), the image intensity of the sample, I, can be expressed as:

I=|ψ(x,y)|2SRF.

To model an x-ray grid, we considered an ideal linear grid in which the absorption by any interspace material is ignored. Fig. 3 shows the primary transmission, tgrid, of an ideal linear grid oriented in the vertical direction. It can be expressed mathematically as a one-dimensional (1D) square function using a Fourier series:9)

tgrid(x)=n=-+dPsinc(πdnP)e2πxnPi,

where P is the grid pitch and d is the width of the lead strips. Thus, the image intensity of the sample with a linear grid, fsg, can be expressed as:

fsg=I×tgrid(x).

The analysis of the phase shift in PCXI is described in detail in Ref.10,11) Two raw images of the sample with grid (fsg) and the bare grid (fg) are acquired separately and Fourier-transformed. In the Fourier domain, the areas surrounding the primary and the first harmonic peaks are selected using a band-pass filter separately and then inverse Fourier-transformed to yield the primary image (f0,s), i.e., retrieved absorption image, and the first harmonic image (f1,s) of the sample by normalizing with the bare grid images as:

f¯0,s=f0,sgf0,g,f¯1,s=f1,sgf1,g,

where f0,sg and f1,sg are the primary and first harmonic images of the sample with grid, respectively, while f0,g and f1,g are those of bare grid. To extract gradient phase information, the ratio of f1,s and I0,s is taken as:

f=f¯1,sf¯0,s.

Fig. 4 shows the simplified Fourier processing in the single grid-based PCXI to extract absorption image and differential phase image from the two raw images of the sample with grid (fsg) and the bare grid (fg).

The intensity of the phase image depends on the x-ray wavelength (λ), the grid period (p), and the system geometry. The size of the sample structure whose phase information can be appreciably detected is known to be limited to an upper threshold size,10) which is defined as the scattering length scale (L):

L=(SDD-SOD)·(SDD-GDD)SDDλp,

where SDD, SOD, and GDD is the source-to-detector distance, the source-to-object distance, and the grid-to-detector distance, respectively. The larger the scattering length scale, the larger the phase signal.

2. Simulation and experimental setup

We developed a useful simulation platform based on the above descriptions for PCXI study. Fig. 5 shows (a) the three-dimensional (3D) numerical chest phantom (478×258×434 voxels) in anterior-posterior (AP) positioning (left) and lateral (LA) positioning (right) and (b) the 3D Shepp-Logan phantom12) (400×400×400 voxels) used in the simulation. The chest phantom, originally produced by authors at the University of Pernambuco in Brazil,13) was defined by us with proper complex refractive indexes by segmentation into several compartments representing skin, muscle, bone, lung, fat, water, soft tissue, etc.14) In the simulation, monochromatic x-rays of 50 keV were used and the strip density of the used grid was 200 lines/inch. The corresponding wavelength of the x-rays and the grid period were about λ=0.25 Å and p=127 μm, respectively. SDD=200 cm, SOD=80 cm, and GDD=100 cm were used. The scattering length scale estimated for the given simulation condition was about 117 nm. The detailed conditions used in the simulation and experiment are listed in Table 1.

Fig. 6 shows the table-top setup established for the PCXI experiment. It consists of an x-ray tube (100-μm focal spot size, Oxford Ins., TF5011), a focused-linear grid (200-lines/inch strip density, JPI Healthcare Corp.), and a CMOS-type flat-panel detector having an active area of 14.5 cm×11.6 cm (48-μm pixel size, Rayence Corp., Xmaru1215). The same system geometry used in the simulation was applied in the experiment. More details of the experimental procedure can be found in our previous paper.15)

Results and Discussion

Fig. 7 shows the differential phase images of the chest phantom in AP positioning simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). The retrieved absorption image (top left) is also indicated as the reference. As indicated in Fig. 7, the contrast of the phase images was, as expected, much higher than that of the absorption images. In addition, a linear grid can detect phase gradient in only one direction perpendicular to the grid strips. Note that the segments of the structure boundary perpendicular to the grid direction are more highlighted in the differential phase images (see the spine and the clavicle for vertical grid and horizontal grid marked by arrows in Fig. 7, and vice versa). The level of the phase gradient of oriented structures depends on the orientation of the structure relative to that of linear grids.

One possible solution to the orientation problem of a linear grid is to employ more sophisticated grids such as crossed grids that consist of square cells to enable imaging of phase gradient in multiple directions with a single exposure. Kottler et al.16) showed that measurements in two orthogonal directions are often necessary to improve visibility and reduce artifacts in the image. Wen et al.17) recently described an x-ray differential phase-contrast imaging method based on 2D transmission gratings that were directly resolved by an x-ray detector and quantified the effects of x-ray refraction and diffraction in the sample through spatial harmonic analysis. The use of 2D gratings allows differential phase contrast in several directions to be obtained from a single image, which obviates the need for multiple exposures and separate measurements for different directions. In the study, instead, as indicated in Fig. 7, we simply summed the two retrieved differential phase images with a vertical grid and a horizontal grid, which gives phase gradient in a direction at 45 degrees to both the x and y directions.18) See the spine and the clavicle in the combined differential image marked by arrows in Fig. 7. Fig. 8 shows similar phase images of the chest phantom in LA positioning; see also the body of humerus and the ribcage indicated by arrows in Fig. 8.

For more quantitative analysis of the image characteristics of PCXI, we repeated the same simulation procedure using the 3D Shepp-Logan phantom. Fig. 9 shows the differential phase images of the Shepp-Logan phantom simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). Fig. 10 shows the intensity profiles measured along the line segments AB indicated in Fig. 9 for the differential phase image and the absorption image. As indicated in Fig. 10, the vertical grid emphasizes vertical edges by a horizontal gradient, detecting small features and variations in the sample that was not clearly visible in the absorption image (see the intensity variations marked by arrows).

Fig. 11 shows complete sets of the PCXI results retrieved from a single raw image of (a) animal bone and (b) chicken wing with a 200-lines/inch vertical grid obtained at the given x-ray tube conditions of 50 kVp and 5 mAs. The image contrast of the phase images was much enhanced, compared to that of the absorption images, and was similar, at least qualitatively, to the simulated contrast, indicating the performance of the developed simulation platform.

Conclusion

We successfully obtained phase-contrast x-ray images of much enhanced contrast, compared to conventional attenuation-based images, by using the single grid-based technique from both the simulation and experiment. The simulated contrast of the phase images was similar, at least qualitatively, to the experimental contrast, which demonstrates the performance of the developed simulation platform. The scattering length scale estimated for a given simulation condition was about 117 nm. Consequently, the simulation platform worked properly and demonstrated that the single grid-based approach seemed a useful method for PCXI with great simplicity and minimal requirements on the setup alignment. We expect that the simulation platform developed in this work will be useful for designing optimal PCXI systems. More quantitative evaluation of the image characteristics will be performed soon.

Acknowledgements

This work was supported by Institute for Information & Communications Technology Promotion (IITP) grant funded by the Korea government (MSIT) (2017-0-00049, Study on biomedical imaging and recognition-sensors for acquisition and analysis of high quality bio-information).

Conflicts of Interest

The authors have nothing to disclose.

Availability of Data and Materials

All relevant data are within the paper and its Supporting Information files.

Tables

Imaging acquisition conditions used in the simulation and the experiment.

ParameterDimension
Source-to-object distance (SOD)80 cm
Object-to-grid distance (OGD)20 cm
Grid-to-detector distance (GDD)100 cm
Grid strip density200 lines/inch
Grid focal distance100 cm
Detector pixel size48 μm
Focal spot size0.1 mm
Tube voltage50 kV (monochromatic in simulation)
50 kVp (polychromatic in experiment)
SampleChest, Shepp-Logan (simulation)
Animal bone, chicken wing (experiment)

Fig 1.

Figure 1.Schematic illustration of the single grid-based PCXI setup. Here the x-ray grid is placed midway between the x-ray source and the detector and an object to be examined is placed ahead of the grid.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 2.

Figure 2.Schematic illustration of a geometry for the free-space propagation of x-rays. The wavefront ψ (x,y,z) in the detector plane that is calculated using the Huygens-Fresnel principle.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 3.

Figure 3.The primary transmission, tgrid, of an ideal linear grid oriented in the vertical direction (i.e., in x). Here, d and D are the width of the lead strips and the distance between them, respectively, and P is the grid pitch.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 4.

Figure 4.The simplified Fourier processing in the single grid-based PCXI to extract absorption image and differential phase image from the two raw images of the sample with a linear grid (fsg) and the bare grid (fg).
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 5.

Figure 5.(a) The 3D numerical chest phantom (478×258×434 voxels) in AP positioning (left) and LA (right) positioning and (b) the 3D Shepp-Logan phantom (400×400×400 voxels) used in the simulation.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 6.

Figure 6.Table-top setup established for the experiment. It consists of an x-ray tube (100-μm focal spot size), a focused-linear grid (200-lines/inch strip density), and a CMOS-type flat-panel detector (14.5 cm×11.6 cm active area, 48-μm pixel size).
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 7.

Figure 7.The differential phase images of the chest phantom in AP positioning simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). The retrieved absorption image (top left) is also indicated as the reference.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 8.

Figure 8.The differential phase images of the chest phantom in LA positioning simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). The retrieved absorption image (top left) is also indicated as the reference.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 9.

Figure 9.The differential phase images of the Shepp-Logan phantom simulated with a vertical grid (top right) and a horizontal grid (bottom left) and their combined image (bottom right). The retrieved absorption image (top left) is also indicated as the reference.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 10.

Figure 10.Intensity profiles measured along the line segments AB indicated in for the differential phase image and the absorption image of the Shepp-Logan phantom.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Fig 11.

Figure 11.Complete sets of the PCXI results retrieved from a single raw image of (a) animal bone and (b) chicken wing with a 200-lines/inch vertical grid obtained at the given x-ray tube conditions of 50 kVp and 5 mAs.
Progress in Medical Physics 2017; 28: 83-91https://doi.org/10.14316/pmp.2017.28.3.83

Table 1 Imaging acquisition conditions used in the simulation and the experiment.

ParameterDimension
Source-to-object distance (SOD)80 cm
Object-to-grid distance (OGD)20 cm
Grid-to-detector distance (GDD)100 cm
Grid strip density200 lines/inch
Grid focal distance100 cm
Detector pixel size48 μm
Focal spot size0.1 mm
Tube voltage50 kV (monochromatic in simulation)
50 kVp (polychromatic in experiment)
SampleChest, Shepp-Logan (simulation)
Animal bone, chicken wing (experiment)

References

  1. Lewis R. Medical phase contrast x-ray imaging: current status and future prospects. Phys Med Biol 2004;49:3573-3583.
    Pubmed CrossRef
  2. Parham C, Zhong Z, Connor D, Chapman L, and Pisano E. Design and implementation of a compact low-dose diffraction enhanced medical imaging system. Acad Radiol 2009;16:911-917.
    Pubmed CrossRef
  3. Burvall A, Lundström U, Takman P, Larsson D, and Hertz H. Phase retrieval in X-ray phase-contrast imaging suitable for tomography. Opt Express 2011;19:10359-10376.
    Pubmed CrossRef
  4. Pfeiffer F, Bech M, Bunk O, Kraft P, Eikenberry E, and Brönnimann C, et al. Hard-X-ray dark-field imaging using a grating interferometer. Nat Mater 2008;7:134-137.
    Pubmed CrossRef
  5. Pfeiffer F, Herzen J, Willner M, Chabior M, Auweter S, and Reiser M, et al. Grating-based x-ray phase contrast for biomedical imaging applications. Z Med Phys 2013;23:176-185.
    Pubmed CrossRef
  6. Wen H, Bennett E, Kopace R, Stein A, and Pai V. Single-shot x-ray differential phase-contrast and diffraction imaging using two-dimensional transmission gratings. Opt Lett 2010;35:1932-1934.
    Pubmed KoreaMed CrossRef
  7. Wen H, Bennett E, Hegedus M, and Carroll S. Spatial harmonic imaging of x-ray Scattering-Initial results. IEEE Trans on Med Imaging 2008;27:997-1002.
    Pubmed KoreaMed CrossRef
  8. Goodman J. Introduction to Fourier optics. 3rd ed. Englewood: Roberts and Company Publishers 2005.
  9. Gauntt D, and Barnes G. Grid line artifact formation: A comprehensive theory. Med Phys 2006;33:1668-1677.
    Pubmed CrossRef
  10. Wen H, Bennett E, Hegedus M, and Rapacchi S. Fourier x-ray Scattering radiography yields bone structural information. Radiology 2009;251:910-918.
    Pubmed KoreaMed CrossRef
  11. Pfeiffer F, Bunk O, David C, Bech M, Duc G, and Bravin A, et al. High-resolution brain tumor visualization using three-dimensional x-ray phase contrast tomography. Phys Med Biol 2007;52:6923-6930.
    Pubmed CrossRef
  12. Shepp L, and Logan B. The Fourier reconstruction of a head section. IEEE Trans on Nucl Sci 1974;21:21-43.
    CrossRef
  13. Kramer R, Cassola V, Khoury H, Vieira J, Lima V, and Brown K. FASH and MASH: female and male adult human phantoms based on polygon mesh surfaces: II. Dosimetric calculations. Phys Med Biol 2010;55:163-189.
    Pubmed CrossRef
  14. Park Y, Cho H, Lim H, Je U, Park C, and Cho H, et al. Experimental Study for Phase-contrast X-ray Imaging Based on a Single Antiscatter Grid and a Polychromatic X-ray Source. Prog in Med Phys 2015;26:1-8.
    CrossRef
  15. Kottler C, David C, Pfeiffer F, and Bunk O. A two-directional approach for grating based differential phase contrast imaging using hard x-rays. Opt Express 2007;15:1175-1181.
    Pubmed CrossRef
  16. Wen H, Bennett E, Kopace R, Stein A, and Pai V. Single-shot x-ray differential phase contrast and diffraction imaging using two-dimensional transmission gratings. Opt Lett 2010;35:1932-1934.
    Pubmed KoreaMed CrossRef
  17. Tahir S. Grid-Based Fourier Transform Phase Contrast Imaging. PhD thesis: State University of New York; 2015.
Korean Society of Medical Physics

Vol.35 No.3
September 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line