검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Split Viewer

Review Article

Progress in Medical Physics 2019; 30(4): 75-88

Published online December 31, 2019 https://doi.org/10.14316/pmp.2019.30.4.75

Copyright © Korean Society of Medical Physics.

Review of National Diagnostic Reference Levels for Interventional Procedures

Min Young Lee , Jae Kwon , Gang Woo Ryu , Ki Hoon Kim , Hyung Woo Nam , Kwang Pyo Kim

Department of Nuclear Engineering, Kyung Hee University, Yongin, Korea

Correspondence to:Kwang Pyo Kim (kpkim@khu.ac.kr)
Tel: 82-31-201-2560 Fax: 82-31-202-8106

Received: December 6, 2019; Revised: December 19, 2019; Accepted: December 23, 2019

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Diagnostic reference level (DRL) is employed to optimize the radiation doses of patients. The objective of this study is to review the DRLs for interventional procedures in Korea and abroad. Literature review was performed to investigate radiation dose index and measurement methodology commonly used in DRL determination. Dose area product (DAP) and fluoroscopy time within each major procedure category were systematically abstracted and analyzed. A wide variation was found in the radiation dose. The DAP values and fluoroscopy times ranged 0.01–3,081 Gy·cm2 and 2–16,878 seconds for all the interventional procedures, 8.5–1,679 Gy·cm2 and 32–5,775 seconds for the transcatheter arterial chemoembolization (TACE), and 0.1–686 Gy·cm2 and 16–6,636 seconds for the transfemoral cerebral angiography (TFCA), respectively. The DRL values of the DAP and fluoroscopy time were 238 Gy·cm2 and 1,224 seconds for the TACE and 189 Gy·cm2 and 686 seconds for the TFCA, respectively. Generally, the DRLs of Korea were lower than those of other developed countries, except for the percutaneous transluminal angioplasty with stent in arteries of the lower extremity (LE PTA and stent), aneurysm coil embolization, and Hickman insertion procedures. The wide variation in the radiation doses of the different procedures suggests that more attention must be paid to reduce unnecessary radiation exposure from medical imaging. Furthermore, periodic nationwide survey of medical radiation exposures is necessary to optimize the patient dose for radiation protection, which will ultimately contribute to patient dose reduction and radiological safety.

KeywordsInterventional radiology, Radiation protection, Diagnostic reference level, Patient dose, Radiation exposure

Interventional radiology is a medical specialization that involves performing a range of imaging procedures to obtain images of the inside of the body of a patient. Interventional radiologists conduct various procedures, such as treating tumors, taking organ biopsies, or placing stents, by inserting miniature instruments and thin catheters into the body via an artery or a vein. Interventional radiology has the advantage of reducing medical expenses as well as the pain from invasive procedures (e.g., surgery). Currently, the use of interventional radiology is rapidly increasing with the increase in average life expectancy and technological development.1)

Fig. 1 shows the number of interventional radiology procedures performed during the period of 2007–2011, as reported in a 2013 domestic study. The use of interventional radiology increased by approximately 12% per year, from 22,000 in 2007 to approximately 35,000 in 2011. These results included only the procedures involving imaging tests performed during interventional radiology in that period. Therefore, the actual number of procedures is expected to be larger than the above value.2) In addition, interventional radiology requires a longer time and relatively higher radiation doses than other diagnostic radiology examinations. Considering its increased usage and high radiation dose, currently, radiation protection and patient dose management are important in interventional radiology. Further, because interventional radiology is an important procedure directly related to the survival of a patient, it is inappropriate to apply extremely strict measures on the radiation exposure received by the patient. However, efforts to optimize the procedure and measures for patient radiation protection are needed.

Guidance or recommendations for radiation protection in diagnostic radiology are provided by the International Commission on Radiological Protection (ICRP). The ICRP uses the following approach for radiation protection of patients. Generally, radiation exposure of a patient during radiology, such as interventional radiology, is higher than that in daily life; however, the patient also receives the medical benefits from the procedure. The ICRP states that radiation in these areas of medical care must be used for the benefit of the patient and the dose should be optimized.3) Summarizing, the basic principle of patient protection in medical exposure is to maximize the medical benefits of radiation use while keeping the dose as low as reasonably achievable considering economic and social factors.46)

The measured or irradiated radiation dose at a radiological examination facility differs according to the radiologist, patient, and hospital, even if the same procedure is used. Therefore, the necessity of setting a reference radiation dose for a patient has emerged. As interventional radiology is a procedure that is directly related to the life of the patient, it is inappropriate to apply extremely strict controls. Therefore, the equipment or institutions that administer extremely high or extremely low radiation doses need to be identified, and the doses should be optimized.

A reference level was introduced to implement radiation dose reduction and establish a reference standard. This concept was first introduced in ICRP Publication 60.3) Subsequently, the ICRP recommended the use of diagnostic reference levels (DRLs) for medical purposes in ICRP Publication 73, 8 and ICRP Supporting Guidance 2.4,7)

The International Atomic Energy Agency (IAEA) also endorses establishing DRLs, the recommended radiation dose levels for patients in accordance with the medical conditions, in each country.8) According to these guidelines, various countries, such as the United Kingdom, the United States, and European countries, are conducting radiation exposure status surveys and DRL studies on various diagnostic procedures, to optimize patient dose management.

In Korea, the Ministry of Food and Drug Safety (MFDS) and the Center for Disease Control and Prevention have conducted DRL determination studies on various diagnostic procedures by measurements or surveys.9,10) The scope of the investigation for DRL determination ranges from hospital units to regional and national levels. Considering the increased usage of interventional radiology and high radiation doses, it is necessary to investigate and analyze studies on DRL determination for domestic and international interventional radiology in terms of patient radiation protection and patient dose management.

The purpose of this study was to review the DRLs for interventional radiology in Korea and abroad. To this end, we first examined the methodology of DRL determination. The radiation dose index and measurement or irradiation methodology commonly used in DRL determination for interventional radiology were investigated. In addition, studies on the DRL determination for domestic and international interventional procedures were examined, and the results were compared and analyzed.

1. Diagnostic reference level

In this study, we investigated the DRL determination methodology and the radiation dose indicator used to determine a DRL. A DRL may be used as an indicator for evaluating how much radiation dose is used for a patient. DRL determination can be conducted by large-scale radiation dose surveys or measurements across hospitals, regions, or a country. A DRL is generally determined as the 75th percentile level of the measured or surveyed radiation dose distribution. It can be applied in various diagnostic procedures, such as computed tomography (CT), nuclear medicine, and interventional radiology. Fig. 2 presents an example of the radiation dose distribution and DRL. From Fig. 2, it can be seen that although most radiation doses are low, some are extremely high. The identification and corrective actions of facilities or equipment that administer abnormally high radiation doses can significantly reduce the radiation dose of a patient. DRL determination should not end as a one-time event; the DRL needs to be measured and examined periodically and updated. In the UK, where early studies of DRL determination have been conducted, it has been found that the radiation dose decreases over time.11)

2. Radiation dose for interventional procedure diagnostic reference level

The measurement or survey methodology for the radiation dose indicator was investigated. The ICRP, International Commission on Radiation Units and Measurements (ICRU), and IAEA have proposed radiation dose units for DRLs.6,12,13) The ICRP proposes as conditions of the patient radiation dose indicator that the assessment of the amount of ionizing radiation used in a procedure and easy measurement or determination of the indicator.6) The ICRU and IAEA recommend using a physical quantity that can be represented by air kerma as the unit of the radiation dose of a DRL.12,13) In general, the dose area product (DAP) and reference air kerma (RAK) are used for the measurement of the radiation dose of a patient in an interventional procedure, because they are easy to be calculated and measured. Further, the fluoroscopy time is employed as an additional indicator.

The concept of DAP is displayed in Fig. 3. The DAP is defined as the product of the absorbed dose that does not include the backscattering effects in air measured at the same distance from the focal point and the area of the X-ray irradiation field. The DAP can provide information about the irradiation area in which a patient is actually exposed. The intensity of the radiation decreases in inverse proportion to the square of the distance on going further from the focal point. In contrast, the irradiation area of the radiation increases in proportion to the square of the distance; thus, the above two effects cancel each other. Therefore, the DAP can be defined at any position on the beam axis between the X-ray focal point and the patient. It can be derived from direct measurements or calculations. Direct measurement of the DAP employs an area dosimeter at the outlet of the X-ray tube. Recently used interventional procedure equipment utilize the area dosimeter at their outlets or use a calculation formula to display the DAP value employed for the patient during the procedure on a screen.

The conceptual diagram of the RAK measurement is shown in Fig. 4. The RAK is a value indicating the radiation dose incident on the skin surface of a patient during an interventional procedure or fluoroscopy. It is measured at a point in space located at a certain distance from the focal point. In general, in the case of a C-arm, it is measured at a patient incident reference point 15 cm away from the center of rotation on the X-ray beam axis in the focal point direction.14) This incident reference point is similar to the point where the skin of the patient is located, assuming a standard sized patient is lying on the bed. Therefore, the RAK can be used to indirectly determine the skin dose received by a patient. Various terms are used in different organizations to indicate the RAK, such as reference air kerma, cumulative air kerma, reference point air kerma, and cumulative dose.

The fluoroscopy time is the total time the fluoroscopy was performed during the interventional procedure. Because the radiation dose is directly proportional to the time of use, the fluoroscopy time is a significant contributor to the exposure dose. In addition, it has the advantage of easily collecting information because it does not require a separate measuring equipment. However, the fluoroscopy time can be used only as a reference because it does not reflect the intensity of the injected X-ray, performance of the device, or irradiation area. Other studies suggest that the fluoroscopy time alone is not a useful indicator of the dose of a patient during a procedure and that we should not examine it separately for patient dose investigation.1519)

1. UK

In the UK, DRL determination studies on diagnostic radiology are actively conducted by the Health Protection Agency (HPA). The UK Department of Health and Social Security periodically collects exposure dose information regarding diagnostic radiology and determines the DRL based on the collected data. A DRL determination and exposure studies for interventional procedures in the UK were conducted in 2005 and 2010.11,20) In a 2012 study, data on various diagnostic procedures, including interventional procedures and diagnostic radiation, were collected from January 2006 to December 2010 at a total of 320 medical institutions nationwide for patient dose investigation. This number represents approximately 25% of the hospitals with diagnostic radiography equipment in the UK. In the UK, data on six major procedures, such as biliary intervention and facet joint injection, were collected. Data were collected not only on the DAP and the fluoroscopy time, but also on patient information, such as age and weight. The information collection for the DAP was based on approximately 15,000 and 14,000 cases of patients, respectively. Each exposure dose information was collected by filling out questionnaires, and the data verification was performed by a statistical analysis program. Table 1 summarizes the DRL and survey results for the DAP and the fluoroscopy time during the interventional procedures in the UK as of 2010.11)

2. USA

In the US, medical radiation exposure status surveys were conducted driven by the Food and Drug Administration (FDA) and the Society of Interventional Radiology (SIR).21) From 2003 to 2009, these organizations performed a study of the patient dose measurement and evaluation of 21 interventional procedures called “Radiation Doses in Interventional Radiology Procedures (RAD-IR) Study”. To this end, the radiation dose data were collected from a total of approximately 2,100 procedures of the medical institutions participating in the study. The radiation dose data were examined for the DAP, RAK, fluoroscopy time, and number of images. The age, height, and weight of a patient were recorded for each procedure performed, and test cases for each procedure were selected to reduce the deviation caused by the size of the patient. Thus, approximately, 1,960 out of the 2,100 cases of procedures were used for the statistical analysis of the data. Table 2 lists the third quartile DRL values of the DAP and fluoroscopy time for each interventional procedure analyzed in the study in the US. These third quartile values are determined as the national patient DRLs for US interventional procedures.22)

3. Spain

DRL studies in Spain are conducted by the National Society of Interventional Radiology (NSIR) and the Spanish Nuclear Safety Council (SNSC). In Spain, patient dose surveys and DRL determination studies for interventional radiology are conducted under the European Medical Exposures Directive, which recommends DRL determination.23)

In a 2008 study by the NSIR, a survey was conducted with interventional radiologists and medical physicists of ten public hospitals nationwide.24) The survey was performed by seven procedures with the highest frequency, including the biliary drainage and hepatic chemoembolization. Thus, approximately 1,400 cases of the procedures were investigated. Information was collected on the DAP, fluoroscopy time, patient gender, age, size, and weight. In addition, the information on the level of Complexity of the procedures (CI) to determine its difficulty was also classified into three levels (low, medium, and high). However, the study determined that the number of samples collected was not sufficient, and thus, the complexity level was not used. In the investigation, the overall DAP distribution of interventional procedures was 0.6–1136 Gy·cm2. The fluoroscopy time distribution ranged from 0.3 to 80 minutes. The number of digital subtraction angiography (DSA) image distributions ranged from 0 to 893 images. The DRL was set as the third quartile of the DAP, fluoroscopy time, and number of DSA images. Table 3 summarizes the results of determining the interventional procedure DRL in Spain in 2008.

In 2016, the SNSC conducted a study to reestablish the DRL.25) In this study, senior interventionists and medical physicists at eight university hospitals were surveyed. Seven procedures, including biliary drainage and hepatic chemoembolization, were performed, and approximately 1,600 cases were surveyed over three years from 2010 to 2013. The DRL was reset to the third quartile of the surveyed DAP, fluoroscopy time, and DSA image number distribution. Table 4 lists the results of determining the interventional radiology DRL in Spain in 2016.

4. Germany

In Germany, research on medical exposure status and DRL determination has been conducted, with focus on the Federal Office for Radiation Protection (BfS). The BfS revised the radiation-related regulations in 2003 to recommend the establishment of DRLs in diagnostic radiographs. Since then, periodic investigations and DRL determination of various medical radiation have been performed. The DRL for interventional radiology was set based on numerous databases collected nationally from the “ärztliche Stellen” (the Quality Assurance Commission), German Institute of Interventional Radiology (DeGIR), and AQUA Institute. In Germany, the DRLs were determined for interventional radiology in 2003, 2016, and 2019 based on the collected data.2628) In 2003, two procedures of the percutaneous transluminal angioplasty (PTA) and percutaneous transluminal coronary angioplasty were set. In 2016 and 2019, 12 and 11 procedures were announced, including the percutaneous coronary intervention, transcatheter aortic valve implantation, transcatheter arterial chemoembolization, endovascular treatment of abdominal aortic aneurysms, and PTA. Table 5 and 6 summarize Germany’s interventional radiology DRL as of 2019.28)

5. Switzerland

Swiss medical radiation survey research is conducted by the Swiss Federal Office of Public Health. In the early 2000s, the Federal Office of Public Health, in collaboration with two university radiation physics laboratories formed a working group for the radiation protection optimization of X-ray examination. The purpose of this group was to generate awareness of the radiation risks to physicians and radiologists to optimize dose-intensive radiographs. In 2007, DRL determination was performed for interventional radiology and diagnostic fluoroscopy.29) In this study, approximately 1,000 procedures were investigated in the radiology and cardiac disease departments of nine national and private hospitals for two years. Twelve procedures were investigated in total, and among these, six types of interventional radiology were examined, including hepatic embolization, biliary drainage, and stent insertion. The radiation dose information was examined for the DAP, fluoroscopy time, and number of images. Table 7 presents the DRL for Switzerland in 2007.29)

6. Finland

Research on medical radiation in Finland is conducted by the Radiation and Nuclear Safety Authority (STUK). The STUK measures or evaluates the patient doses every three years to establish the interventional radiology DRL. Finland has presented a DRL for interventional radiology in 2016.30) In Finland, in interventional radiology and cardiology procedures, the difficulty of the procedure may have a larger impact on the radiation dose than on the procedure steps or procedure equipment. Thus, it is recommended to use the radiation doses of numerous patients to set the DRL and consider the difficulties in the procedure, if necessary. Table 8 presents the DRL for interventional radiology in Finland conducted in 2016.

7. United Nations Scientific Committee on the Effects of Atomic Radiation

The United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR) is an international organization established by the United Nations (UN) for the analysis of the effects and risks of radiation on the human body. UNSCEAR collects information on the medical radiation exposure doses in all the countries to understand the current status of medical radiation safety management. The information is gathered by collecting questionnaires from representatives of each country or by researching national research reports and literatures. The UNSCEAR recently published a report on the results of worldwide medical radiation exposure surveys obtained from countries and institutions during 1997–2007.1) In this study, although the interventional radiology DRL was not set, the distributions and differences in the DAPs for different countries at a particular procedure were analyzed by the UNSCEAR. The UNSCEAR collected the results of the DAP and effective dose evaluation for 30 procedures regarding interventional radiology according to five major categories of biopsy, biliary tract, and urinary system. In the 2010 report, the DAP collected from each country was presented in each literature. Table 9 lists the distribution of the DAP according to the interventional radiology reported by the UNSCEAR.

8. Korea

Survey research on the actual conditions of patient dose in the interventional radiation procedures in Korea is actively conducted with the MFDS. The MFDS set a DRL based on the collected information of the radiation dose of the interventional radiology performed in 2007 and 2012. In the 2012 study, 11 procedures were selected from among the interventional radiology procedures performed in Korea, which were relatively numerous procedures or were of high importance in relation to the radiation protection, including the TACE, arteriovenous fistula (AVF), percutaneous transluminal angioplasty with stent in arteries of the lower extremity (LE PTA and stent). The study collected patient radiation dose data from the eleven interventional radiology procedures selected from 24 major hospitals nationwide. The data surveys were performed according to the DAP and fluoroscopy time. In addition, 8,415 domestic data were obtained for each procedure, and the DRL was presented as the third quartile of the survey results.10) In addition, the accuracy of the built-in DAP meter of the IR equipment installed in Korea was evaluated, and the actual effects of the dose reduction parameters were evaluated by phantom measurement experiments.

Table 10 lists the results of the 2012 Korean interventional radiology DRL study.10) The DAP distribution for each procedure is found to be 0.01–3081 Gy·cm2 overall, with 8.5–1679 Gy·cm2 for the TACE and 0.1–686 Gy·cm2 for the transfemoral cerebral angiography (TFCA) procedure reprehensively. The fluoroscopy time distribution is overall 2–16,878 seconds, with 32–5,775 seconds for the TACE and 16–6,636 seconds for the TFCA procedure reprehensively. Even for the same procedure, the difference is more than several hundred times for the performer, patient, and hospital. The DRL is set to the third quartile of the DAP and fluoroscopy time distributions for each procedure. Representatively, the DRLs for the TACE and TFCA are 238 and 189 Gy·cm2 for the DAP and 1,224 and 686 seconds for the fluoroscopy time, respectively.

Fig. 5 and 6 display the comparison of the DAPs and fluoroscopy times in domestic and international interventional procedures. The domestic DRL is generally low compared to the DRLs in other countries. Comparing the DRL for the domestic DAP with the average of that in other countries, the domestic DRL is found to be about two times lower for the percutaneous transhepatic biliary drainage (PTBD) and biliary stent procedures. Further, for the TACE and percutaneous nephrostomy (PCN) procedures, it is approximately 55% and 18% lower, respectively. For the LE PTA and stent, aneurysm coil embolization, and Hickman insertion procedures, it is approximately 30%, 8%, and 30% higher, respectively. Further, when comparing the DRL for the fluoroscopy time with the average of the other countries, the domestic DRL is lower, except for the LE PTA and stent procedure, which is at a similar level. It is about approximately thrice lower for the PTBD and PCN procedures and approximately 50% lower for the biliary stent and Hickman insertion procedures. Moreover, it is approximately 30% lower for the TACE and aneurysm coil embolization procedures, and is similar to that of the other countries for the LE PTA and stent procedures.

The use of interventional radiology is rapidly increasing owing to the increase in its demand arising from the average lifespan increase and technology development. It is known that interventional radiology takes longer than other radiographies and uses relatively high radiation doses. Considering the increased usage and high radiation dose of these interventional procedures, the radiation protection of a patient and patient dose management are important during interventional procedures. Under these circumstances, we investigated the current status of the DRL setting in domestic and foreign interventional radiology procedures. Further, they were compared and analyzed to understand the radiation dose exposure status and patient dose management in interventional radiology for the Korean population.

In Korea, a study on setting the interventional radiology DRL was conducted in 2012 by the Ministry of Food and Drug Safety. In this study, DRLs were set for the DAPs and fluoroscopy times of eleven procedures, which relatively have numerous procedures or are of high importance in relation to radiation protection. They included the TACE, AVF, and LE PTA and stent, among interventional radiology performed in Korea. Representatively, the DRLs for the TACE and TFCA were 237.7 and 188.5 Gy·cm2, respectively, and for the fluoroscopy time, 1,224 and 686 seconds, respectively. When comparing the domestic interventional radiology DRL with those of other countries, the domestic DRL was found to be low, in general. When comparing the DRL for the domestic DAP with the average values of the other countries, for the PTBD, biliary stent, TACE, and PCN procedures, it was found to be as low as 20% to as high as nearly twice the average values of the other countries. Further, for the LE PTA and stent, aneurysm coil embolization, and Hickman insertion procedures, it was approximately 30%, 8%, and 30% higher, respectively. In addition, when comparing the DRL for the fluoroscopy time with the average of the other countries, the domestic DRL was lower, except for the LE PTA and stent procedures, which were at similar levels.

Interventional radiology, as medical radiation exposure, benefits a patient and is an important procedure that is directly related to the life of a patient; thus, it is not appropriate to perform extremely strict management of the radiation exposure that a patient receives without considering the overall circumstances. However, to optimize the procedure and reduce patient dose, it is necessary to understand the current status of domestic patient dose management by investigating the patient exposure level and DRL setting during the interventional procedure. In the future, if the nationwide survey is continuously conducted for medical radiation safety management and patient protection, it will ultimately contribute to maximizing the benefit to patients from interventional radiology while minimizing the radiation dose of the patient.

This work was supported by the Nuclear Safety Research Program through the Korea Foundation of Nuclear Safety (KoFONS) using the financial resource granted by the Nuclear Safety and Security Commission (NSSC) of the Republic of Korea (No. 1803013).

Dose area products (DAPs) and fluoroscopy times for interventional procedures in the UK (2012)

Radiation dose unitProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)Biliary intervention102227933611143
Facet joint injection20302,72040.2166
Hickman line insertion213782920.1103
Nephrostomy243146480.52413
Oesophageal stent15241991337613
Pacemaker (permanent)31785,06270.2347
PTCA 1 stent14395,80534128140
Fluoroscopy time (s)Biliary intervention9212766871471,116849
Facet joint injection17282,64161411884
Hickman line insertion183469675451087
Nephrostomy2025332399571,630404
Oesophageal stent122116524951696298
Pacemaker (permanent)24634,475307103670358
PTCA 1 stent10355,44451684882675

Diagnostic reference levels for interventional procedures in the US (2012)

ProcedureNumberDose (Gy)DAP (Gy·cm2)Fluoroscopy time (s)
Transjugular intrahepatic portosystemic shunt creation30035253,600
Biliary drainage201.41001,800
Nephrostomy for obstruction120.440900
Nephrostomy for stone access140.7601,500
Pulmonary angiography2150.5110600
Inferior vena cava filter placement400.2560240
Renal or visceral angioplasty without stent21022001,200
Renal or visceral angioplasty with stent2002.32501,800
Iliac angioplasty without stent3001.252501,200
Iliac angioplasty with stent3501.93001,500
Bronchial artery embolization45022403,000
Hepatic chemoembolization3001.94001,500
Uterine fibroid embolization4503.64502,160
Other tumor embolization3252.63902,100
Gastrointestinal hemorrhage localization and treatment4253.85202,100
Embolization in the head for AVM1,50065508,100
Embolization in the head for aneurysm1,3504.753605,400
Embolization in the head for tumor1,7006.255012,000
Vertebroplasty12021201,260
Pelvic artery embolization for trauma or tumor5502.55502,100
Embolization in the spine for AVM or tumor1,50089507,800

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2008)

Dose IndexProcedureNumberMaximumMinimumMeanDRL
DAP (Gy·cm2)Fistulography180830.6912
Lower limb arteriography6856082.45873
Renal arteriography553977.77589
Biliary drainage2054280.66180
Hepatic chemoembolization15183027.4216289
Iliac stent701,13614.19194
Uterine embolization4567726.6170236
Fluoroscopy time (s)Fistulography1803,12018150150
Lower limb arteriography6853,40830180198
Renal arteriography551,65018336516
Biliary drainage2054,764368701,188
Hepatic chemoembolization1514,8001681,1881,458
Iliac stent704,02678552654
Uterine embolization452,670181,4281,818

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2016)

ProcedureNumberMean DAP (Gy·cm2)3th Q DAP (Gy·cm2)3th Q fluoroscopy time (s)
Lower extremity arteriography784--240
Renal arteriography37--336
Transjugular hepatic biopsies3030.845792
Biliary drainage31430.8301,038
Uterine fibroid embolization56189.62141,860
Colon endoprostheses31120.21692,790
Hepatic chemoembolization269216.13031,578
Femoropopliteal revascularization6283.91191,800
Iliac stent66105.71701,284

Dose area products (DAPs) for interventional procedures in Germany (2019)

ProcedureDAP (Gy·cm2)Deff (mSv)

25th50th75thDRL
Thrombus aspiration after stroke (recanalization of cerebral arteries)519115818011
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)7412119225016
PCI203449489
Combined CA and PCI2840555510
TAVI2549828015
EVAR
 For thoracic aorta4711420323028
 For infrarenal abdominal aorta5510820332
 For suprarenal abdominal aorta479521836
TACE62121224230392
PTA
 For pelvis2244879023
 For thigh and knee815354010
 For lower leg and food61020256

Fluoroscopy times for interventional procedures in Germany (2019)

ProcedureFluoroscopy time (s)

25th50th75th
Thrombus aspiration after stroke (recanalization of cerebral arteries)7201,2602,100
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)1,2602,0403,240
PCI312558780
Combined CA and PCI354570780
TAVI4747201,080
EVAR
 For thoracic aorta4207201,140
 For infrarenal abdominal aorta8401,2601,980
 For suprarenal abdominal aorta8401,5603,120
TACE6601,0201,500
PTA
 For pelvis4206001,020
 For thigh and knee4206601,080
 For lower leg and food5401,0201,860

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Switzerland (2007)

CategoryProcedureFluoroscopy time (min)Frame (n)DAP (Gy·cm2)
DiagnosticBarium meal74060
Lower limb and iliac angiography8150210
Cerebral angiography154801,215
Barium enema1025150
Coronary angiography71,40080
Electrophysiology148010
InterventionalHepatic embolization30160620
Biliary drainage and stent insertion2530240
Cerebral embolization50800440
Iliac dilatation and stent insertion25200460
Percutaneous coronary intervention201,500110
Cardiac thermo-ablation20800140
CombinedCoronary angiography+percutaneous coronary intervention202,800260
Electrophysiology+cardiac thermo-ablation300280

Diagnostic reference levels (DRLs) for interventional procedures in Finland (2016)

ProcedureDAP (Gy·cm2)Fluoroscopy time (min)
Coronary artery examination using a contrast medium304
Percutaneous coronary intervention*7515
Pacemaker installation (excluding CRT pacemaker installation)3.55
Transcatheter aortic valve implantation9019
Electrophysiological treatment of atrial fibrillation2512
CRT pacemaker installation22-
Electrophysiological treatment of atrial flutter16-
Electrophysiological treatment of atrioventricular nodal reentrant tachycardia6-

DAP, dose area products; CRT, cardiac resynchronization therapy.

*Including potential angiography (ad hoc percutaneous coronary intervention).

Dose distributions for interventional procedures in the United Nations Scientific Committee on the Effects of Atomic Radiation (2010)

CategoryProcedureDAP (Gy·cm2)Effective dose (mSv)
BiopsyPathological specimen-1.6
Biopsy61.6
Small bowel biopsy10.26
Venous sampling-0.4
Biliary and urinary systemsBile duct drainage38–1509.9–38
Bile duct dilation/Stenting43–5411–14
Biliary intervention5414
Biliary duct stone extraction277
Lithotripsy51.3
Nephtostomy8–562.1–15
Ureteric stenting184.7
Kidney stent insertion4912.7
CardiovascularEmboliztion75–11420–30
Management of caricocele50.8–1316.4–38
Neuroembolization81–3202.3–11
Thrombolysis13.53.5
TIPS77–52420–87
Valvuloplasty16229
Vascular stenting40–425.8–10
Pelvic vein embolization-60
Insertion of caval filters4813
Removal of foreings bodies-7
Coronary angiography12.7–1473.1–16
PTCA11.8–2055.4–41
Stent procedures41–1667–13
Pacemaker insertion8.46–19-
Uterine fibroid embolization GastroinstinalUterine fibroid embolization30.6–2988–78
Feeding tube133.4
Gastrostomy133.4
Diliation/stenting oesophagus151.5
Dilation pyloric stenosis277
Colonic stent-7
Nerve injection1.70.2

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Korea (2012)

Radiation dose unitsProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)TACE22353,139191.18.51,678.8237.7
AVF91423414.10.2113.117.3
LE PTA & Stent131627885.41.3571.0114.1
TFCA18292,083149.60.1685.5188.5
Aneurysm coil1518320316.935.83,080.5383.5
PTBD152556636.10.31,290.837.5
Biliary Stent213611847.40.7289.464.6
PCN1315219190.5189.122.4
Hickman152051940.0375.34.3
Chemoport20244792.10.0131.72.8
Perm-cath17174603.70.0237.44.4
Fluoroscopy time (s)TACE22353,139935.632.05,775.01,224.0
AVF914234649.830.03,715.0786.1
LE PTA & Stent13162781,090.842.05,897.01,460.4
TFCA18292,083599.916.06,636.0686.0
Aneurysm Coil15183202,325.5250.016,878.02,975.8
PTBD1525566379.316.06,777.0427.5
Biliary Stent2136118606.530.02,493.0759.8
PCN1315219222.98.01,643.0236.0
Hickman152051951.35.0987.054.0
Chemoport202447934.15.0428.443.0
Perm-cath171746056.32.0507.067.8
  1. UNSCEAR 2008 Report. Sources and effects of ionizing radiation. UNSCEAR Report. Vienna: UNSCEAR 2010:2008.
  2. Study on the radiation exposure of the workers of the interventional radiology. Seoul: Ministry of Food and Drug Safety 2013.
  3. ICRP Publication 60: 1990 Recommendations of the International Commission on Radiological Protection. Ottawa: International Commission on Radiological Protection 1991.
  4. ICRP Publication 73: Radiological protection and safety in medicine. Ottawa: International Commission on Radiological Protection 1991.
  5. ICRP Publication 105: Radiological protection in medicine. Ottawa: International Commission on Radiological Protection 2007.
  6. ICRP Publication 135: Diagnostic reference levels in medical imaging. Ottawa: International Commission on Radiological Protection 2017.
  7. ICRP Supporting Guidance 2: Radiation and your patient: a guide for medical practitioners. Ottawa: International Commission on Radiological Protection 2002.
  8. IAEA Safety Series No. 115: International basic safety standards for protection against ionizing radiation and for the safety of radiation sources. Vienna: International Commission on Radiological Protection 1996.
  9. Evaluation of patient dose in interventional radiology. Seoul: Ministry of Food and Drug Safety 2007.
  10. Evaluation of patient doses from interventional radiology procedures. Seoul: Ministry of Food and Drug Safety 2012.
  11. HPA-CRCE-034: Doses to patients from radiographic and fluoroscopic x-ray imaging procedures in the UK - 2010 review. London: Health Protection Agency 2012.
  12. ICRU Report 74: Patient dosimetry for X rays used in medical imaging. Bethesda: International Commission in Radiation Units and Measurement 2005.
  13. IAEA Technical Report Series No. 457: Dosimetry in diagnostic radiology: an international code of practice. Vienna: International Atomic Energy Agency 2007.
  14. IEC 60532: Radiation protection instrumentation - installed dose rate meters, warning assemblies and monitors - X and gamma radiation of energy between 50 keV and 7 MeV. Geneva: International Electrotechnical Commission 2010.
  15. ICRP Publication 120: Radiological protection in cardiology. Ottawa: International Commission on Radiological Protection 2013.
  16. Fletcher DW, Miller DL, Balter S, and Taylor MA. Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 2002;13:391-397.
    Pubmed CrossRef
  17. Chida K, Saito H, Otani H, Kohzuki M, Takahashi S, and Yamada S, et al. Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. AJR Am J Roentgenol 2006;186:774-778.
    Pubmed CrossRef
  18. NCRP Report No. 168: Radiation dose management for fluoroscopically-guided interventional medical procedures. Bethesda: National Council on Radiation Protection & Measurements 2010.
  19. Chambers CE, Fetterly KA, Holzer R, Lin PJ, Blankenship JC, and Balter S, et al. Radiation safety program for the cardiac catheterization laboratory. Catheter Cardiovasc Interv 2011;77:546-556.
    Pubmed CrossRef
  20. HPA-RPD-029: Doses to patients from radiographic and fluoroscopic X-ray imaging procedures in the UK-2005 review. London: Health Protection Agency 2007.
  21. Miller DL, Kwon D, and Bonavia GH. Reference levels for patient radiation doses in interventional radiology: proposed initial values for U.S. practice. Radiology 2009;253:753-764.
    Pubmed KoreaMed CrossRef
  22. NCRP Report 172: Reference levels and achievable doses in medical and dental imaging: recommendations for the United States. Bethesda: National Council on Radiation Protection & Measurements 2012.
  23. 97/43/ EURATOM: The European directive on health protection of individuals against the dangers of ionising radiation in relation to medical exposures. Rome: EURopean ATOMic Energy Community 1997.
  24. Vano E, Sanchez R, Fernandez JM, Gallego JJ, Verdu JF, and de Garay MG, et al. Patient dose reference levels for interventional radiology: a national approach. Cardiovasc Intervent Radiol 2009;32:19-24.
    Pubmed CrossRef
  25. Ruiz-Cruces R, Vano E, Carrera-Magariño F, Moreno-Rodriguez F, Soler-Cantos MM, and Canis-Lopez M, et al. Diagnostic reference levels and complexity indices in interventional radiology: a national programme. Eur Radiol 2016;26:4268-4276.
    Pubmed CrossRef
  26. Bundesamt für Strahlenschutz: Bekanntmachung der aktualisierten diagnostischen Referenzwerte für nuklear-medizinische Untersuchunge. Salzgitter: Bundesamt für Strahlenschutz 2003.
  27. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für diagnostische und interventionelle Röntgenuntersuchungen. Salzgitter: Bundesamt für Strahlenschutz 2016.
  28. Schegerer AA, Loose R, Heuser L, and Brix G. Diagnostic reference levels for diagnostic and interventional X-ray procedures in Germany: update and handling. Rofo 2019;191:739-751.
    Pubmed CrossRef
  29. Aroua A, Rickli H, Stauffer JC, Schnyder P, Trueb PR, and Valley JF, et al. How to set up and apply reference levels in fluoroscopy at a national level. Eur Radiol 2007;17:1621-1633.
    Pubmed CrossRef
  30. Reference levels for the patient’s radiation exposure in cardiology. Helsinki: Radiation and Nuclear Safety Authority 2016.

Article

Review Article

Progress in Medical Physics 2019; 30(4): 75-88

Published online December 31, 2019 https://doi.org/10.14316/pmp.2019.30.4.75

Copyright © Korean Society of Medical Physics.

Review of National Diagnostic Reference Levels for Interventional Procedures

Min Young Lee , Jae Kwon , Gang Woo Ryu , Ki Hoon Kim , Hyung Woo Nam , Kwang Pyo Kim

Department of Nuclear Engineering, Kyung Hee University, Yongin, Korea

Correspondence to:Kwang Pyo Kim (kpkim@khu.ac.kr)
Tel: 82-31-201-2560 Fax: 82-31-202-8106

Received: December 6, 2019; Revised: December 19, 2019; Accepted: December 23, 2019

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Diagnostic reference level (DRL) is employed to optimize the radiation doses of patients. The objective of this study is to review the DRLs for interventional procedures in Korea and abroad. Literature review was performed to investigate radiation dose index and measurement methodology commonly used in DRL determination. Dose area product (DAP) and fluoroscopy time within each major procedure category were systematically abstracted and analyzed. A wide variation was found in the radiation dose. The DAP values and fluoroscopy times ranged 0.01–3,081 Gy·cm2 and 2–16,878 seconds for all the interventional procedures, 8.5–1,679 Gy·cm2 and 32–5,775 seconds for the transcatheter arterial chemoembolization (TACE), and 0.1–686 Gy·cm2 and 16–6,636 seconds for the transfemoral cerebral angiography (TFCA), respectively. The DRL values of the DAP and fluoroscopy time were 238 Gy·cm2 and 1,224 seconds for the TACE and 189 Gy·cm2 and 686 seconds for the TFCA, respectively. Generally, the DRLs of Korea were lower than those of other developed countries, except for the percutaneous transluminal angioplasty with stent in arteries of the lower extremity (LE PTA and stent), aneurysm coil embolization, and Hickman insertion procedures. The wide variation in the radiation doses of the different procedures suggests that more attention must be paid to reduce unnecessary radiation exposure from medical imaging. Furthermore, periodic nationwide survey of medical radiation exposures is necessary to optimize the patient dose for radiation protection, which will ultimately contribute to patient dose reduction and radiological safety.

Keywords: Interventional radiology, Radiation protection, Diagnostic reference level, Patient dose, Radiation exposure

Introduction

Interventional radiology is a medical specialization that involves performing a range of imaging procedures to obtain images of the inside of the body of a patient. Interventional radiologists conduct various procedures, such as treating tumors, taking organ biopsies, or placing stents, by inserting miniature instruments and thin catheters into the body via an artery or a vein. Interventional radiology has the advantage of reducing medical expenses as well as the pain from invasive procedures (e.g., surgery). Currently, the use of interventional radiology is rapidly increasing with the increase in average life expectancy and technological development.1)

Fig. 1 shows the number of interventional radiology procedures performed during the period of 2007–2011, as reported in a 2013 domestic study. The use of interventional radiology increased by approximately 12% per year, from 22,000 in 2007 to approximately 35,000 in 2011. These results included only the procedures involving imaging tests performed during interventional radiology in that period. Therefore, the actual number of procedures is expected to be larger than the above value.2) In addition, interventional radiology requires a longer time and relatively higher radiation doses than other diagnostic radiology examinations. Considering its increased usage and high radiation dose, currently, radiation protection and patient dose management are important in interventional radiology. Further, because interventional radiology is an important procedure directly related to the survival of a patient, it is inappropriate to apply extremely strict measures on the radiation exposure received by the patient. However, efforts to optimize the procedure and measures for patient radiation protection are needed.

Guidance or recommendations for radiation protection in diagnostic radiology are provided by the International Commission on Radiological Protection (ICRP). The ICRP uses the following approach for radiation protection of patients. Generally, radiation exposure of a patient during radiology, such as interventional radiology, is higher than that in daily life; however, the patient also receives the medical benefits from the procedure. The ICRP states that radiation in these areas of medical care must be used for the benefit of the patient and the dose should be optimized.3) Summarizing, the basic principle of patient protection in medical exposure is to maximize the medical benefits of radiation use while keeping the dose as low as reasonably achievable considering economic and social factors.46)

The measured or irradiated radiation dose at a radiological examination facility differs according to the radiologist, patient, and hospital, even if the same procedure is used. Therefore, the necessity of setting a reference radiation dose for a patient has emerged. As interventional radiology is a procedure that is directly related to the life of the patient, it is inappropriate to apply extremely strict controls. Therefore, the equipment or institutions that administer extremely high or extremely low radiation doses need to be identified, and the doses should be optimized.

A reference level was introduced to implement radiation dose reduction and establish a reference standard. This concept was first introduced in ICRP Publication 60.3) Subsequently, the ICRP recommended the use of diagnostic reference levels (DRLs) for medical purposes in ICRP Publication 73, 8 and ICRP Supporting Guidance 2.4,7)

The International Atomic Energy Agency (IAEA) also endorses establishing DRLs, the recommended radiation dose levels for patients in accordance with the medical conditions, in each country.8) According to these guidelines, various countries, such as the United Kingdom, the United States, and European countries, are conducting radiation exposure status surveys and DRL studies on various diagnostic procedures, to optimize patient dose management.

In Korea, the Ministry of Food and Drug Safety (MFDS) and the Center for Disease Control and Prevention have conducted DRL determination studies on various diagnostic procedures by measurements or surveys.9,10) The scope of the investigation for DRL determination ranges from hospital units to regional and national levels. Considering the increased usage of interventional radiology and high radiation doses, it is necessary to investigate and analyze studies on DRL determination for domestic and international interventional radiology in terms of patient radiation protection and patient dose management.

The purpose of this study was to review the DRLs for interventional radiology in Korea and abroad. To this end, we first examined the methodology of DRL determination. The radiation dose index and measurement or irradiation methodology commonly used in DRL determination for interventional radiology were investigated. In addition, studies on the DRL determination for domestic and international interventional procedures were examined, and the results were compared and analyzed.

Diagnostic Reference Level and Radiation Dose for a Diagnostic Reference Level

1. Diagnostic reference level

In this study, we investigated the DRL determination methodology and the radiation dose indicator used to determine a DRL. A DRL may be used as an indicator for evaluating how much radiation dose is used for a patient. DRL determination can be conducted by large-scale radiation dose surveys or measurements across hospitals, regions, or a country. A DRL is generally determined as the 75th percentile level of the measured or surveyed radiation dose distribution. It can be applied in various diagnostic procedures, such as computed tomography (CT), nuclear medicine, and interventional radiology. Fig. 2 presents an example of the radiation dose distribution and DRL. From Fig. 2, it can be seen that although most radiation doses are low, some are extremely high. The identification and corrective actions of facilities or equipment that administer abnormally high radiation doses can significantly reduce the radiation dose of a patient. DRL determination should not end as a one-time event; the DRL needs to be measured and examined periodically and updated. In the UK, where early studies of DRL determination have been conducted, it has been found that the radiation dose decreases over time.11)

2. Radiation dose for interventional procedure diagnostic reference level

The measurement or survey methodology for the radiation dose indicator was investigated. The ICRP, International Commission on Radiation Units and Measurements (ICRU), and IAEA have proposed radiation dose units for DRLs.6,12,13) The ICRP proposes as conditions of the patient radiation dose indicator that the assessment of the amount of ionizing radiation used in a procedure and easy measurement or determination of the indicator.6) The ICRU and IAEA recommend using a physical quantity that can be represented by air kerma as the unit of the radiation dose of a DRL.12,13) In general, the dose area product (DAP) and reference air kerma (RAK) are used for the measurement of the radiation dose of a patient in an interventional procedure, because they are easy to be calculated and measured. Further, the fluoroscopy time is employed as an additional indicator.

The concept of DAP is displayed in Fig. 3. The DAP is defined as the product of the absorbed dose that does not include the backscattering effects in air measured at the same distance from the focal point and the area of the X-ray irradiation field. The DAP can provide information about the irradiation area in which a patient is actually exposed. The intensity of the radiation decreases in inverse proportion to the square of the distance on going further from the focal point. In contrast, the irradiation area of the radiation increases in proportion to the square of the distance; thus, the above two effects cancel each other. Therefore, the DAP can be defined at any position on the beam axis between the X-ray focal point and the patient. It can be derived from direct measurements or calculations. Direct measurement of the DAP employs an area dosimeter at the outlet of the X-ray tube. Recently used interventional procedure equipment utilize the area dosimeter at their outlets or use a calculation formula to display the DAP value employed for the patient during the procedure on a screen.

The conceptual diagram of the RAK measurement is shown in Fig. 4. The RAK is a value indicating the radiation dose incident on the skin surface of a patient during an interventional procedure or fluoroscopy. It is measured at a point in space located at a certain distance from the focal point. In general, in the case of a C-arm, it is measured at a patient incident reference point 15 cm away from the center of rotation on the X-ray beam axis in the focal point direction.14) This incident reference point is similar to the point where the skin of the patient is located, assuming a standard sized patient is lying on the bed. Therefore, the RAK can be used to indirectly determine the skin dose received by a patient. Various terms are used in different organizations to indicate the RAK, such as reference air kerma, cumulative air kerma, reference point air kerma, and cumulative dose.

The fluoroscopy time is the total time the fluoroscopy was performed during the interventional procedure. Because the radiation dose is directly proportional to the time of use, the fluoroscopy time is a significant contributor to the exposure dose. In addition, it has the advantage of easily collecting information because it does not require a separate measuring equipment. However, the fluoroscopy time can be used only as a reference because it does not reflect the intensity of the injected X-ray, performance of the device, or irradiation area. Other studies suggest that the fluoroscopy time alone is not a useful indicator of the dose of a patient during a procedure and that we should not examine it separately for patient dose investigation.1519)

Nationwide Diagnostic Reference Level for Interventional Procedures

1. UK

In the UK, DRL determination studies on diagnostic radiology are actively conducted by the Health Protection Agency (HPA). The UK Department of Health and Social Security periodically collects exposure dose information regarding diagnostic radiology and determines the DRL based on the collected data. A DRL determination and exposure studies for interventional procedures in the UK were conducted in 2005 and 2010.11,20) In a 2012 study, data on various diagnostic procedures, including interventional procedures and diagnostic radiation, were collected from January 2006 to December 2010 at a total of 320 medical institutions nationwide for patient dose investigation. This number represents approximately 25% of the hospitals with diagnostic radiography equipment in the UK. In the UK, data on six major procedures, such as biliary intervention and facet joint injection, were collected. Data were collected not only on the DAP and the fluoroscopy time, but also on patient information, such as age and weight. The information collection for the DAP was based on approximately 15,000 and 14,000 cases of patients, respectively. Each exposure dose information was collected by filling out questionnaires, and the data verification was performed by a statistical analysis program. Table 1 summarizes the DRL and survey results for the DAP and the fluoroscopy time during the interventional procedures in the UK as of 2010.11)

2. USA

In the US, medical radiation exposure status surveys were conducted driven by the Food and Drug Administration (FDA) and the Society of Interventional Radiology (SIR).21) From 2003 to 2009, these organizations performed a study of the patient dose measurement and evaluation of 21 interventional procedures called “Radiation Doses in Interventional Radiology Procedures (RAD-IR) Study”. To this end, the radiation dose data were collected from a total of approximately 2,100 procedures of the medical institutions participating in the study. The radiation dose data were examined for the DAP, RAK, fluoroscopy time, and number of images. The age, height, and weight of a patient were recorded for each procedure performed, and test cases for each procedure were selected to reduce the deviation caused by the size of the patient. Thus, approximately, 1,960 out of the 2,100 cases of procedures were used for the statistical analysis of the data. Table 2 lists the third quartile DRL values of the DAP and fluoroscopy time for each interventional procedure analyzed in the study in the US. These third quartile values are determined as the national patient DRLs for US interventional procedures.22)

3. Spain

DRL studies in Spain are conducted by the National Society of Interventional Radiology (NSIR) and the Spanish Nuclear Safety Council (SNSC). In Spain, patient dose surveys and DRL determination studies for interventional radiology are conducted under the European Medical Exposures Directive, which recommends DRL determination.23)

In a 2008 study by the NSIR, a survey was conducted with interventional radiologists and medical physicists of ten public hospitals nationwide.24) The survey was performed by seven procedures with the highest frequency, including the biliary drainage and hepatic chemoembolization. Thus, approximately 1,400 cases of the procedures were investigated. Information was collected on the DAP, fluoroscopy time, patient gender, age, size, and weight. In addition, the information on the level of Complexity of the procedures (CI) to determine its difficulty was also classified into three levels (low, medium, and high). However, the study determined that the number of samples collected was not sufficient, and thus, the complexity level was not used. In the investigation, the overall DAP distribution of interventional procedures was 0.6–1136 Gy·cm2. The fluoroscopy time distribution ranged from 0.3 to 80 minutes. The number of digital subtraction angiography (DSA) image distributions ranged from 0 to 893 images. The DRL was set as the third quartile of the DAP, fluoroscopy time, and number of DSA images. Table 3 summarizes the results of determining the interventional procedure DRL in Spain in 2008.

In 2016, the SNSC conducted a study to reestablish the DRL.25) In this study, senior interventionists and medical physicists at eight university hospitals were surveyed. Seven procedures, including biliary drainage and hepatic chemoembolization, were performed, and approximately 1,600 cases were surveyed over three years from 2010 to 2013. The DRL was reset to the third quartile of the surveyed DAP, fluoroscopy time, and DSA image number distribution. Table 4 lists the results of determining the interventional radiology DRL in Spain in 2016.

4. Germany

In Germany, research on medical exposure status and DRL determination has been conducted, with focus on the Federal Office for Radiation Protection (BfS). The BfS revised the radiation-related regulations in 2003 to recommend the establishment of DRLs in diagnostic radiographs. Since then, periodic investigations and DRL determination of various medical radiation have been performed. The DRL for interventional radiology was set based on numerous databases collected nationally from the “ärztliche Stellen” (the Quality Assurance Commission), German Institute of Interventional Radiology (DeGIR), and AQUA Institute. In Germany, the DRLs were determined for interventional radiology in 2003, 2016, and 2019 based on the collected data.2628) In 2003, two procedures of the percutaneous transluminal angioplasty (PTA) and percutaneous transluminal coronary angioplasty were set. In 2016 and 2019, 12 and 11 procedures were announced, including the percutaneous coronary intervention, transcatheter aortic valve implantation, transcatheter arterial chemoembolization, endovascular treatment of abdominal aortic aneurysms, and PTA. Table 5 and 6 summarize Germany’s interventional radiology DRL as of 2019.28)

5. Switzerland

Swiss medical radiation survey research is conducted by the Swiss Federal Office of Public Health. In the early 2000s, the Federal Office of Public Health, in collaboration with two university radiation physics laboratories formed a working group for the radiation protection optimization of X-ray examination. The purpose of this group was to generate awareness of the radiation risks to physicians and radiologists to optimize dose-intensive radiographs. In 2007, DRL determination was performed for interventional radiology and diagnostic fluoroscopy.29) In this study, approximately 1,000 procedures were investigated in the radiology and cardiac disease departments of nine national and private hospitals for two years. Twelve procedures were investigated in total, and among these, six types of interventional radiology were examined, including hepatic embolization, biliary drainage, and stent insertion. The radiation dose information was examined for the DAP, fluoroscopy time, and number of images. Table 7 presents the DRL for Switzerland in 2007.29)

6. Finland

Research on medical radiation in Finland is conducted by the Radiation and Nuclear Safety Authority (STUK). The STUK measures or evaluates the patient doses every three years to establish the interventional radiology DRL. Finland has presented a DRL for interventional radiology in 2016.30) In Finland, in interventional radiology and cardiology procedures, the difficulty of the procedure may have a larger impact on the radiation dose than on the procedure steps or procedure equipment. Thus, it is recommended to use the radiation doses of numerous patients to set the DRL and consider the difficulties in the procedure, if necessary. Table 8 presents the DRL for interventional radiology in Finland conducted in 2016.

7. United Nations Scientific Committee on the Effects of Atomic Radiation

The United Nations Scientific Committee on the Effects of Atomic Radiation (UNSCEAR) is an international organization established by the United Nations (UN) for the analysis of the effects and risks of radiation on the human body. UNSCEAR collects information on the medical radiation exposure doses in all the countries to understand the current status of medical radiation safety management. The information is gathered by collecting questionnaires from representatives of each country or by researching national research reports and literatures. The UNSCEAR recently published a report on the results of worldwide medical radiation exposure surveys obtained from countries and institutions during 1997–2007.1) In this study, although the interventional radiology DRL was not set, the distributions and differences in the DAPs for different countries at a particular procedure were analyzed by the UNSCEAR. The UNSCEAR collected the results of the DAP and effective dose evaluation for 30 procedures regarding interventional radiology according to five major categories of biopsy, biliary tract, and urinary system. In the 2010 report, the DAP collected from each country was presented in each literature. Table 9 lists the distribution of the DAP according to the interventional radiology reported by the UNSCEAR.

8. Korea

Survey research on the actual conditions of patient dose in the interventional radiation procedures in Korea is actively conducted with the MFDS. The MFDS set a DRL based on the collected information of the radiation dose of the interventional radiology performed in 2007 and 2012. In the 2012 study, 11 procedures were selected from among the interventional radiology procedures performed in Korea, which were relatively numerous procedures or were of high importance in relation to the radiation protection, including the TACE, arteriovenous fistula (AVF), percutaneous transluminal angioplasty with stent in arteries of the lower extremity (LE PTA and stent). The study collected patient radiation dose data from the eleven interventional radiology procedures selected from 24 major hospitals nationwide. The data surveys were performed according to the DAP and fluoroscopy time. In addition, 8,415 domestic data were obtained for each procedure, and the DRL was presented as the third quartile of the survey results.10) In addition, the accuracy of the built-in DAP meter of the IR equipment installed in Korea was evaluated, and the actual effects of the dose reduction parameters were evaluated by phantom measurement experiments.

Table 10 lists the results of the 2012 Korean interventional radiology DRL study.10) The DAP distribution for each procedure is found to be 0.01–3081 Gy·cm2 overall, with 8.5–1679 Gy·cm2 for the TACE and 0.1–686 Gy·cm2 for the transfemoral cerebral angiography (TFCA) procedure reprehensively. The fluoroscopy time distribution is overall 2–16,878 seconds, with 32–5,775 seconds for the TACE and 16–6,636 seconds for the TFCA procedure reprehensively. Even for the same procedure, the difference is more than several hundred times for the performer, patient, and hospital. The DRL is set to the third quartile of the DAP and fluoroscopy time distributions for each procedure. Representatively, the DRLs for the TACE and TFCA are 238 and 189 Gy·cm2 for the DAP and 1,224 and 686 seconds for the fluoroscopy time, respectively.

Fig. 5 and 6 display the comparison of the DAPs and fluoroscopy times in domestic and international interventional procedures. The domestic DRL is generally low compared to the DRLs in other countries. Comparing the DRL for the domestic DAP with the average of that in other countries, the domestic DRL is found to be about two times lower for the percutaneous transhepatic biliary drainage (PTBD) and biliary stent procedures. Further, for the TACE and percutaneous nephrostomy (PCN) procedures, it is approximately 55% and 18% lower, respectively. For the LE PTA and stent, aneurysm coil embolization, and Hickman insertion procedures, it is approximately 30%, 8%, and 30% higher, respectively. Further, when comparing the DRL for the fluoroscopy time with the average of the other countries, the domestic DRL is lower, except for the LE PTA and stent procedure, which is at a similar level. It is about approximately thrice lower for the PTBD and PCN procedures and approximately 50% lower for the biliary stent and Hickman insertion procedures. Moreover, it is approximately 30% lower for the TACE and aneurysm coil embolization procedures, and is similar to that of the other countries for the LE PTA and stent procedures.

Conclusions

The use of interventional radiology is rapidly increasing owing to the increase in its demand arising from the average lifespan increase and technology development. It is known that interventional radiology takes longer than other radiographies and uses relatively high radiation doses. Considering the increased usage and high radiation dose of these interventional procedures, the radiation protection of a patient and patient dose management are important during interventional procedures. Under these circumstances, we investigated the current status of the DRL setting in domestic and foreign interventional radiology procedures. Further, they were compared and analyzed to understand the radiation dose exposure status and patient dose management in interventional radiology for the Korean population.

In Korea, a study on setting the interventional radiology DRL was conducted in 2012 by the Ministry of Food and Drug Safety. In this study, DRLs were set for the DAPs and fluoroscopy times of eleven procedures, which relatively have numerous procedures or are of high importance in relation to radiation protection. They included the TACE, AVF, and LE PTA and stent, among interventional radiology performed in Korea. Representatively, the DRLs for the TACE and TFCA were 237.7 and 188.5 Gy·cm2, respectively, and for the fluoroscopy time, 1,224 and 686 seconds, respectively. When comparing the domestic interventional radiology DRL with those of other countries, the domestic DRL was found to be low, in general. When comparing the DRL for the domestic DAP with the average values of the other countries, for the PTBD, biliary stent, TACE, and PCN procedures, it was found to be as low as 20% to as high as nearly twice the average values of the other countries. Further, for the LE PTA and stent, aneurysm coil embolization, and Hickman insertion procedures, it was approximately 30%, 8%, and 30% higher, respectively. In addition, when comparing the DRL for the fluoroscopy time with the average of the other countries, the domestic DRL was lower, except for the LE PTA and stent procedures, which were at similar levels.

Interventional radiology, as medical radiation exposure, benefits a patient and is an important procedure that is directly related to the life of a patient; thus, it is not appropriate to perform extremely strict management of the radiation exposure that a patient receives without considering the overall circumstances. However, to optimize the procedure and reduce patient dose, it is necessary to understand the current status of domestic patient dose management by investigating the patient exposure level and DRL setting during the interventional procedure. In the future, if the nationwide survey is continuously conducted for medical radiation safety management and patient protection, it will ultimately contribute to maximizing the benefit to patients from interventional radiology while minimizing the radiation dose of the patient.

Acknowledgements

This work was supported by the Nuclear Safety Research Program through the Korea Foundation of Nuclear Safety (KoFONS) using the financial resource granted by the Nuclear Safety and Security Commission (NSSC) of the Republic of Korea (No. 1803013).

Tables

Dose area products (DAPs) and fluoroscopy times for interventional procedures in the UK (2012)

Radiation dose unitProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)Biliary intervention102227933611143
Facet joint injection20302,72040.2166
Hickman line insertion213782920.1103
Nephrostomy243146480.52413
Oesophageal stent15241991337613
Pacemaker (permanent)31785,06270.2347
PTCA 1 stent14395,80534128140
Fluoroscopy time (s)Biliary intervention9212766871471,116849
Facet joint injection17282,64161411884
Hickman line insertion183469675451087
Nephrostomy2025332399571,630404
Oesophageal stent122116524951696298
Pacemaker (permanent)24634,475307103670358
PTCA 1 stent10355,44451684882675

Diagnostic reference levels for interventional procedures in the US (2012)

ProcedureNumberDose (Gy)DAP (Gy·cm2)Fluoroscopy time (s)
Transjugular intrahepatic portosystemic shunt creation30035253,600
Biliary drainage201.41001,800
Nephrostomy for obstruction120.440900
Nephrostomy for stone access140.7601,500
Pulmonary angiography2150.5110600
Inferior vena cava filter placement400.2560240
Renal or visceral angioplasty without stent21022001,200
Renal or visceral angioplasty with stent2002.32501,800
Iliac angioplasty without stent3001.252501,200
Iliac angioplasty with stent3501.93001,500
Bronchial artery embolization45022403,000
Hepatic chemoembolization3001.94001,500
Uterine fibroid embolization4503.64502,160
Other tumor embolization3252.63902,100
Gastrointestinal hemorrhage localization and treatment4253.85202,100
Embolization in the head for AVM1,50065508,100
Embolization in the head for aneurysm1,3504.753605,400
Embolization in the head for tumor1,7006.255012,000
Vertebroplasty12021201,260
Pelvic artery embolization for trauma or tumor5502.55502,100
Embolization in the spine for AVM or tumor1,50089507,800

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2008)

Dose IndexProcedureNumberMaximumMinimumMeanDRL
DAP (Gy·cm2)Fistulography180830.6912
Lower limb arteriography6856082.45873
Renal arteriography553977.77589
Biliary drainage2054280.66180
Hepatic chemoembolization15183027.4216289
Iliac stent701,13614.19194
Uterine embolization4567726.6170236
Fluoroscopy time (s)Fistulography1803,12018150150
Lower limb arteriography6853,40830180198
Renal arteriography551,65018336516
Biliary drainage2054,764368701,188
Hepatic chemoembolization1514,8001681,1881,458
Iliac stent704,02678552654
Uterine embolization452,670181,4281,818

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2016)

ProcedureNumberMean DAP (Gy·cm2)3th Q DAP (Gy·cm2)3th Q fluoroscopy time (s)
Lower extremity arteriography784--240
Renal arteriography37--336
Transjugular hepatic biopsies3030.845792
Biliary drainage31430.8301,038
Uterine fibroid embolization56189.62141,860
Colon endoprostheses31120.21692,790
Hepatic chemoembolization269216.13031,578
Femoropopliteal revascularization6283.91191,800
Iliac stent66105.71701,284

Dose area products (DAPs) for interventional procedures in Germany (2019)

ProcedureDAP (Gy·cm2)Deff (mSv)

25th50th75thDRL
Thrombus aspiration after stroke (recanalization of cerebral arteries)519115818011
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)7412119225016
PCI203449489
Combined CA and PCI2840555510
TAVI2549828015
EVAR
 For thoracic aorta4711420323028
 For infrarenal abdominal aorta5510820332
 For suprarenal abdominal aorta479521836
TACE62121224230392
PTA
 For pelvis2244879023
 For thigh and knee815354010
 For lower leg and food61020256

Fluoroscopy times for interventional procedures in Germany (2019)

ProcedureFluoroscopy time (s)

25th50th75th
Thrombus aspiration after stroke (recanalization of cerebral arteries)7201,2602,100
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)1,2602,0403,240
PCI312558780
Combined CA and PCI354570780
TAVI4747201,080
EVAR
 For thoracic aorta4207201,140
 For infrarenal abdominal aorta8401,2601,980
 For suprarenal abdominal aorta8401,5603,120
TACE6601,0201,500
PTA
 For pelvis4206001,020
 For thigh and knee4206601,080
 For lower leg and food5401,0201,860

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Switzerland (2007)

CategoryProcedureFluoroscopy time (min)Frame (n)DAP (Gy·cm2)
DiagnosticBarium meal74060
Lower limb and iliac angiography8150210
Cerebral angiography154801,215
Barium enema1025150
Coronary angiography71,40080
Electrophysiology148010
InterventionalHepatic embolization30160620
Biliary drainage and stent insertion2530240
Cerebral embolization50800440
Iliac dilatation and stent insertion25200460
Percutaneous coronary intervention201,500110
Cardiac thermo-ablation20800140
CombinedCoronary angiography+percutaneous coronary intervention202,800260
Electrophysiology+cardiac thermo-ablation300280

Diagnostic reference levels (DRLs) for interventional procedures in Finland (2016)

ProcedureDAP (Gy·cm2)Fluoroscopy time (min)
Coronary artery examination using a contrast medium304
Percutaneous coronary intervention*7515
Pacemaker installation (excluding CRT pacemaker installation)3.55
Transcatheter aortic valve implantation9019
Electrophysiological treatment of atrial fibrillation2512
CRT pacemaker installation22-
Electrophysiological treatment of atrial flutter16-
Electrophysiological treatment of atrioventricular nodal reentrant tachycardia6-

DAP, dose area products; CRT, cardiac resynchronization therapy.

*Including potential angiography (ad hoc percutaneous coronary intervention).

Dose distributions for interventional procedures in the United Nations Scientific Committee on the Effects of Atomic Radiation (2010)

CategoryProcedureDAP (Gy·cm2)Effective dose (mSv)
BiopsyPathological specimen-1.6
Biopsy61.6
Small bowel biopsy10.26
Venous sampling-0.4
Biliary and urinary systemsBile duct drainage38–1509.9–38
Bile duct dilation/Stenting43–5411–14
Biliary intervention5414
Biliary duct stone extraction277
Lithotripsy51.3
Nephtostomy8–562.1–15
Ureteric stenting184.7
Kidney stent insertion4912.7
CardiovascularEmboliztion75–11420–30
Management of caricocele50.8–1316.4–38
Neuroembolization81–3202.3–11
Thrombolysis13.53.5
TIPS77–52420–87
Valvuloplasty16229
Vascular stenting40–425.8–10
Pelvic vein embolization-60
Insertion of caval filters4813
Removal of foreings bodies-7
Coronary angiography12.7–1473.1–16
PTCA11.8–2055.4–41
Stent procedures41–1667–13
Pacemaker insertion8.46–19-
Uterine fibroid embolization GastroinstinalUterine fibroid embolization30.6–2988–78
Feeding tube133.4
Gastrostomy133.4
Diliation/stenting oesophagus151.5
Dilation pyloric stenosis277
Colonic stent-7
Nerve injection1.70.2

Dose area products (DAPs) and fluoroscopy times for interventional procedures in Korea (2012)

Radiation dose unitsProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)TACE22353,139191.18.51,678.8237.7
AVF91423414.10.2113.117.3
LE PTA & Stent131627885.41.3571.0114.1
TFCA18292,083149.60.1685.5188.5
Aneurysm coil1518320316.935.83,080.5383.5
PTBD152556636.10.31,290.837.5
Biliary Stent213611847.40.7289.464.6
PCN1315219190.5189.122.4
Hickman152051940.0375.34.3
Chemoport20244792.10.0131.72.8
Perm-cath17174603.70.0237.44.4
Fluoroscopy time (s)TACE22353,139935.632.05,775.01,224.0
AVF914234649.830.03,715.0786.1
LE PTA & Stent13162781,090.842.05,897.01,460.4
TFCA18292,083599.916.06,636.0686.0
Aneurysm Coil15183202,325.5250.016,878.02,975.8
PTBD1525566379.316.06,777.0427.5
Biliary Stent2136118606.530.02,493.0759.8
PCN1315219222.98.01,643.0236.0
Hickman152051951.35.0987.054.0
Chemoport202447934.15.0428.443.0
Perm-cath171746056.32.0507.067.8

Fig 1.

Figure 1.Number of interventional procedures per year in Korea.
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Fig 2.

Figure 2.Dose distributions of diagnostic radiology and diagnostic reference level (DRL).
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Fig 3.

Figure 3.Concept of the dose area product (DAP) measurement.
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Fig 4.

Figure 4.Concept of the reference air kerma (RAK) measurement. DAP, dose area product.
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Fig 5.

Figure 5.Dose area products (DAPs) for interventional procedures by country. TACE, transcatheter arterial chemoembolization; PTBD, percutaneous transhepatic biliary drainage; PCN, percutaneous nephrostomy.
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Fig 6.

Figure 6.Fluoroscopy times for interventional procedures by country. TACE, transcatheter arterial chemoembolization; PTBD, percutaneous transhepatic biliary drainage; PCN, percutaneous nephrostomy.
Progress in Medical Physics 2019; 30: 75-88https://doi.org/10.14316/pmp.2019.30.4.75

Table 1 Dose area products (DAPs) and fluoroscopy times for interventional procedures in the UK (2012)

Radiation dose unitProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)Biliary intervention102227933611143
Facet joint injection20302,72040.2166
Hickman line insertion213782920.1103
Nephrostomy243146480.52413
Oesophageal stent15241991337613
Pacemaker (permanent)31785,06270.2347
PTCA 1 stent14395,80534128140
Fluoroscopy time (s)Biliary intervention9212766871471,116849
Facet joint injection17282,64161411884
Hickman line insertion183469675451087
Nephrostomy2025332399571,630404
Oesophageal stent122116524951696298
Pacemaker (permanent)24634,475307103670358
PTCA 1 stent10355,44451684882675

DRL, diagnostic reference level; PTCA, percutaneous transluminal coronary angioplasty.


Table 2 Diagnostic reference levels for interventional procedures in the US (2012)

ProcedureNumberDose (Gy)DAP (Gy·cm2)Fluoroscopy time (s)
Transjugular intrahepatic portosystemic shunt creation30035253,600
Biliary drainage201.41001,800
Nephrostomy for obstruction120.440900
Nephrostomy for stone access140.7601,500
Pulmonary angiography2150.5110600
Inferior vena cava filter placement400.2560240
Renal or visceral angioplasty without stent21022001,200
Renal or visceral angioplasty with stent2002.32501,800
Iliac angioplasty without stent3001.252501,200
Iliac angioplasty with stent3501.93001,500
Bronchial artery embolization45022403,000
Hepatic chemoembolization3001.94001,500
Uterine fibroid embolization4503.64502,160
Other tumor embolization3252.63902,100
Gastrointestinal hemorrhage localization and treatment4253.85202,100
Embolization in the head for AVM1,50065508,100
Embolization in the head for aneurysm1,3504.753605,400
Embolization in the head for tumor1,7006.255012,000
Vertebroplasty12021201,260
Pelvic artery embolization for trauma or tumor5502.55502,100
Embolization in the spine for AVM or tumor1,50089507,800

DAP, dose area products; AVM, arteriovenous malformation.


Table 3 Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2008)

Dose IndexProcedureNumberMaximumMinimumMeanDRL
DAP (Gy·cm2)Fistulography180830.6912
Lower limb arteriography6856082.45873
Renal arteriography553977.77589
Biliary drainage2054280.66180
Hepatic chemoembolization15183027.4216289
Iliac stent701,13614.19194
Uterine embolization4567726.6170236
Fluoroscopy time (s)Fistulography1803,12018150150
Lower limb arteriography6853,40830180198
Renal arteriography551,65018336516
Biliary drainage2054,764368701,188
Hepatic chemoembolization1514,8001681,1881,458
Iliac stent704,02678552654
Uterine embolization452,670181,4281,818

DRL, diagnostic reference level.


Table 4 Dose area products (DAPs) and fluoroscopy times for interventional procedures in Spain (2016)

ProcedureNumberMean DAP (Gy·cm2)3th Q DAP (Gy·cm2)3th Q fluoroscopy time (s)
Lower extremity arteriography784--240
Renal arteriography37--336
Transjugular hepatic biopsies3030.845792
Biliary drainage31430.8301,038
Uterine fibroid embolization56189.62141,860
Colon endoprostheses31120.21692,790
Hepatic chemoembolization269216.13031,578
Femoropopliteal revascularization6283.91191,800
Iliac stent66105.71701,284

Table 5 Dose area products (DAPs) for interventional procedures in Germany (2019)

ProcedureDAP (Gy·cm2)Deff (mSv)

25th50th75thDRL
Thrombus aspiration after stroke (recanalization of cerebral arteries)519115818011
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)7412119225016
PCI203449489
Combined CA and PCI2840555510
TAVI2549828015
EVAR
 For thoracic aorta4711420323028
 For infrarenal abdominal aorta5510820332
 For suprarenal abdominal aorta479521836
TACE62121224230392
PTA
 For pelvis2244879023
 For thigh and knee815354010
 For lower leg and food61020256

DRL, diagnostic reference level; Deff, effective dose; EVAR, endovascular treatment of abdominal aortic aneurysms; PCI, percutaneous coronary intervention; CA, coronary angiography; TAVI, transcatheter aortic valve implantation; TACE, transcatheter arterial chemoembolization; PTA, percutaneous transluminal angioplasty.


Table 6 Fluoroscopy times for interventional procedures in Germany (2019)

ProcedureFluoroscopy time (s)

25th50th75th
Thrombus aspiration after stroke (recanalization of cerebral arteries)7201,2602,100
Coiling of a cerebral aneurysm (EVAR of the cerebral artery)1,2602,0403,240
PCI312558780
Combined CA and PCI354570780
TAVI4747201,080
EVAR
 For thoracic aorta4207201,140
 For infrarenal abdominal aorta8401,2601,980
 For suprarenal abdominal aorta8401,5603,120
TACE6601,0201,500
PTA
 For pelvis4206001,020
 For thigh and knee4206601,080
 For lower leg and food5401,0201,860

EVAR, endovascular treatment of abdominal aortic aneurysms; PCI, percutaneous coronary intervention; CA, coronary angiography; TAVI, transcatheter aortic valve implantation; TACE, transcatheter arterial chemoembolization; PTA, percutaneous transluminal angioplasty.


Table 7 Dose area products (DAPs) and fluoroscopy times for interventional procedures in Switzerland (2007)

CategoryProcedureFluoroscopy time (min)Frame (n)DAP (Gy·cm2)
DiagnosticBarium meal74060
Lower limb and iliac angiography8150210
Cerebral angiography154801,215
Barium enema1025150
Coronary angiography71,40080
Electrophysiology148010
InterventionalHepatic embolization30160620
Biliary drainage and stent insertion2530240
Cerebral embolization50800440
Iliac dilatation and stent insertion25200460
Percutaneous coronary intervention201,500110
Cardiac thermo-ablation20800140
CombinedCoronary angiography+percutaneous coronary intervention202,800260
Electrophysiology+cardiac thermo-ablation300280

Table 8 Diagnostic reference levels (DRLs) for interventional procedures in Finland (2016)

ProcedureDAP (Gy·cm2)Fluoroscopy time (min)
Coronary artery examination using a contrast medium304
Percutaneous coronary intervention*7515
Pacemaker installation (excluding CRT pacemaker installation)3.55
Transcatheter aortic valve implantation9019
Electrophysiological treatment of atrial fibrillation2512
CRT pacemaker installation22-
Electrophysiological treatment of atrial flutter16-
Electrophysiological treatment of atrioventricular nodal reentrant tachycardia6-

DAP, dose area products; CRT, cardiac resynchronization therapy.

*Including potential angiography (ad hoc percutaneous coronary intervention).


Table 9 Dose distributions for interventional procedures in the United Nations Scientific Committee on the Effects of Atomic Radiation (2010)

CategoryProcedureDAP (Gy·cm2)Effective dose (mSv)
BiopsyPathological specimen-1.6
Biopsy61.6
Small bowel biopsy10.26
Venous sampling-0.4
Biliary and urinary systemsBile duct drainage38–1509.9–38
Bile duct dilation/Stenting43–5411–14
Biliary intervention5414
Biliary duct stone extraction277
Lithotripsy51.3
Nephtostomy8–562.1–15
Ureteric stenting184.7
Kidney stent insertion4912.7
CardiovascularEmboliztion75–11420–30
Management of caricocele50.8–1316.4–38
Neuroembolization81–3202.3–11
Thrombolysis13.53.5
TIPS77–52420–87
Valvuloplasty16229
Vascular stenting40–425.8–10
Pelvic vein embolization-60
Insertion of caval filters4813
Removal of foreings bodies-7
Coronary angiography12.7–1473.1–16
PTCA11.8–2055.4–41
Stent procedures41–1667–13
Pacemaker insertion8.46–19-
Uterine fibroid embolization GastroinstinalUterine fibroid embolization30.6–2988–78
Feeding tube133.4
Gastrostomy133.4
Diliation/stenting oesophagus151.5
Dilation pyloric stenosis277
Colonic stent-7
Nerve injection1.70.2

DAP, dose area products; TIPS, transjugular intrahepatic portosystemic shunt; PTCA, percutaneous transluminal coronary angioplasty.


Table 10 Dose area products (DAPs) and fluoroscopy times for interventional procedures in Korea (2012)

Radiation dose unitsProcedureMeasurement (n)MeanMinimumMaximumDRL

HospitalRoomPatient
DAP (Gy·cm2)TACE22353,139191.18.51,678.8237.7
AVF91423414.10.2113.117.3
LE PTA & Stent131627885.41.3571.0114.1
TFCA18292,083149.60.1685.5188.5
Aneurysm coil1518320316.935.83,080.5383.5
PTBD152556636.10.31,290.837.5
Biliary Stent213611847.40.7289.464.6
PCN1315219190.5189.122.4
Hickman152051940.0375.34.3
Chemoport20244792.10.0131.72.8
Perm-cath17174603.70.0237.44.4
Fluoroscopy time (s)TACE22353,139935.632.05,775.01,224.0
AVF914234649.830.03,715.0786.1
LE PTA & Stent13162781,090.842.05,897.01,460.4
TFCA18292,083599.916.06,636.0686.0
Aneurysm Coil15183202,325.5250.016,878.02,975.8
PTBD1525566379.316.06,777.0427.5
Biliary Stent2136118606.530.02,493.0759.8
PCN1315219222.98.01,643.0236.0
Hickman152051951.35.0987.054.0
Chemoport202447934.15.0428.443.0
Perm-cath171746056.32.0507.067.8

DRL, diagnostic reference level; TACE, transcatheter arterial chemoembolization; AVF, arteriovenous fistula; TFCA, transfemoral cerebral angiography; PTBD, percutaneous transhepatic biliary drainage; PCN, percutaneous nephrostomy.


References

  1. UNSCEAR 2008 Report. Sources and effects of ionizing radiation. UNSCEAR Report. Vienna: UNSCEAR 2010:2008.
  2. Study on the radiation exposure of the workers of the interventional radiology. Seoul: Ministry of Food and Drug Safety 2013.
  3. ICRP Publication 60: 1990 Recommendations of the International Commission on Radiological Protection. Ottawa: International Commission on Radiological Protection 1991.
  4. ICRP Publication 73: Radiological protection and safety in medicine. Ottawa: International Commission on Radiological Protection 1991.
  5. ICRP Publication 105: Radiological protection in medicine. Ottawa: International Commission on Radiological Protection 2007.
  6. ICRP Publication 135: Diagnostic reference levels in medical imaging. Ottawa: International Commission on Radiological Protection 2017.
  7. ICRP Supporting Guidance 2: Radiation and your patient: a guide for medical practitioners. Ottawa: International Commission on Radiological Protection 2002.
  8. IAEA Safety Series No. 115: International basic safety standards for protection against ionizing radiation and for the safety of radiation sources. Vienna: International Commission on Radiological Protection 1996.
  9. Evaluation of patient dose in interventional radiology. Seoul: Ministry of Food and Drug Safety 2007.
  10. Evaluation of patient doses from interventional radiology procedures. Seoul: Ministry of Food and Drug Safety 2012.
  11. HPA-CRCE-034: Doses to patients from radiographic and fluoroscopic x-ray imaging procedures in the UK - 2010 review. London: Health Protection Agency 2012.
  12. ICRU Report 74: Patient dosimetry for X rays used in medical imaging. Bethesda: International Commission in Radiation Units and Measurement 2005.
  13. IAEA Technical Report Series No. 457: Dosimetry in diagnostic radiology: an international code of practice. Vienna: International Atomic Energy Agency 2007.
  14. IEC 60532: Radiation protection instrumentation - installed dose rate meters, warning assemblies and monitors - X and gamma radiation of energy between 50 keV and 7 MeV. Geneva: International Electrotechnical Commission 2010.
  15. ICRP Publication 120: Radiological protection in cardiology. Ottawa: International Commission on Radiological Protection 2013.
  16. Fletcher DW, Miller DL, Balter S, and Taylor MA. Comparison of four techniques to estimate radiation dose to skin during angiographic and interventional radiology procedures. J Vasc Interv Radiol 2002;13:391-397.
    Pubmed CrossRef
  17. Chida K, Saito H, Otani H, Kohzuki M, Takahashi S, and Yamada S, et al. Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. AJR Am J Roentgenol 2006;186:774-778.
    Pubmed CrossRef
  18. NCRP Report No. 168: Radiation dose management for fluoroscopically-guided interventional medical procedures. Bethesda: National Council on Radiation Protection & Measurements 2010.
  19. Chambers CE, Fetterly KA, Holzer R, Lin PJ, Blankenship JC, and Balter S, et al. Radiation safety program for the cardiac catheterization laboratory. Catheter Cardiovasc Interv 2011;77:546-556.
    Pubmed CrossRef
  20. HPA-RPD-029: Doses to patients from radiographic and fluoroscopic X-ray imaging procedures in the UK-2005 review. London: Health Protection Agency 2007.
  21. Miller DL, Kwon D, and Bonavia GH. Reference levels for patient radiation doses in interventional radiology: proposed initial values for U.S. practice. Radiology 2009;253:753-764.
    Pubmed KoreaMed CrossRef
  22. NCRP Report 172: Reference levels and achievable doses in medical and dental imaging: recommendations for the United States. Bethesda: National Council on Radiation Protection & Measurements 2012.
  23. 97/43/ EURATOM: The European directive on health protection of individuals against the dangers of ionising radiation in relation to medical exposures. Rome: EURopean ATOMic Energy Community 1997.
  24. Vano E, Sanchez R, Fernandez JM, Gallego JJ, Verdu JF, and de Garay MG, et al. Patient dose reference levels for interventional radiology: a national approach. Cardiovasc Intervent Radiol 2009;32:19-24.
    Pubmed CrossRef
  25. Ruiz-Cruces R, Vano E, Carrera-Magariño F, Moreno-Rodriguez F, Soler-Cantos MM, and Canis-Lopez M, et al. Diagnostic reference levels and complexity indices in interventional radiology: a national programme. Eur Radiol 2016;26:4268-4276.
    Pubmed CrossRef
  26. Bundesamt für Strahlenschutz: Bekanntmachung der aktualisierten diagnostischen Referenzwerte für nuklear-medizinische Untersuchunge. Salzgitter: Bundesamt für Strahlenschutz 2003.
  27. Bekanntmachung der aktualisierten diagnostischen Referenzwerte für diagnostische und interventionelle Röntgenuntersuchungen. Salzgitter: Bundesamt für Strahlenschutz 2016.
  28. Schegerer AA, Loose R, Heuser L, and Brix G. Diagnostic reference levels for diagnostic and interventional X-ray procedures in Germany: update and handling. Rofo 2019;191:739-751.
    Pubmed CrossRef
  29. Aroua A, Rickli H, Stauffer JC, Schnyder P, Trueb PR, and Valley JF, et al. How to set up and apply reference levels in fluoroscopy at a national level. Eur Radiol 2007;17:1621-1633.
    Pubmed CrossRef
  30. Reference levels for the patient’s radiation exposure in cardiology. Helsinki: Radiation and Nuclear Safety Authority 2016.
Korean Society of Medical Physics

Vol.35 No.2
June 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line