Brachytherapy plays an essential role in cervical cancer treatment, as it can deliver a high dose to the tumor without sparing the surrounding normal tissue. For locally advanced cervical cancer, the treatment with a combination of concurrent chemotherapy and external beam radiation therapy followed by intracavitary radiotherapy (ICR) has been used, and it improves the overall survival [1,2].
Traditionally, ICR based on two-dimensional (2D) X-ray images were standardized by International Commission on Radiological Units Report 38 [3]. However, this protocol does not represent the characteristics of individual patients because of the point-based dose prescriptions. Recently, the advantages of image-guided three-dimensional (3D) ICR have been reported as reducing the late complication rates for cervical cancer [4]. Guidelines have been published by the brachytherapy Group of the Groupe Européen de Curiethérapie and European Society for Therapeutic Radiology and Oncology (GEC ESTRO) [5,6]. For the 3D image-based plans, the target and organs at risk (OAR) are delineated, and the dose was prescribed to the outermost point that covered the target.
The 3D imaging usage in the high-dose-rate (HDR) workflow using magnetic resonance imaging (MRI) image provides excellent soft tissue contrast, specifically in gynecologic diseases [7,8]. As the advantage in MR images, the images facilitate the target and OAR contouring during brachytherapy procedures. However, the actual localization of the brachytherapy applicator and radiation source on MRI still remains a challenge because MRI does not directly delineate the source path [9,10]. Therefore, computed tomography (CT) images are also required to define the source position and reconstruct the applicator using a metallic marker during treatment planning. This hybrid approach that needs both MR and CT images for the plan has limitations in the efficiency of brachytherapy, which is conducted in the clinical workflow [11]. For this reason, there are efforts to translate the procedure into only MRI, which could decrease the number of scans and associated patient discomfort as well as reduce the planning-related costs. Also, the uncertainty due to image registration between CT and MR images was excluded. However, the main limitation of migrating toward the MR-only workflow is that the implanted catheters are not reliably visualized on the MR images [12].
Several studies have been developed on the MR line marker, which is visualized on MR images. Shaaer et al. [11] proposed an MR line marker filled with a contrast agent and evaluated the robustness of visualizing the marker on the T1- and T2-weighted imaging MRI. Kim et al. [13] developed rubber-based MR markers with silicone oil to enhance the signal intensity in low-magnetic-field MRI. However, the markers have a limitation in representing a lower intensity signal than distilled water.
Recently, another approach has been proposed to directly convert MR images into synthetic CT (sCT) based on deep learning techniques in radiotherapy [14-18]. Previous studies using the deep learning model for MRI-to-CT conversion employed a supervised learning scheme [15]. The most commonly used neural network is the U-Net architecture, which is an encoder–decoder network with skip connections trained discriminatively. However, there are few studies of the sCT generation for 3D ICR.
In this study, we proposed a deep learning architecture combining two tasks (image generation and segmentation) to generate sCT images from MR images acquired from the low-magnetic-field MR scanner for improving metallic marker visibility.
We retrospectively added patients treated with 3D ICR for cervical cancer in a single institution (Seoul National University Hospital, Seoul, Korea). This study was approved by the Institutional Review Board at our institution (1708-051-876), and the requirement for informed consent was waived. Planning CT images were obtained using Brilliance Big Bore CT scanner (Philips, Cleveland, OH, USA) with a 1-mm slice thickness, a 512×512 matrix, a 120-kVp tube voltage, and a YA convolution kernel with iDose level 3. MR images were obtained using a 0.35-T MRI scanner integrated with the radiation therapy system of the MRIdian MRgRT system (ViewRay, Oakwood, OH, USA) 15 minutes after the CT scan. The true fast imaging with steady-state-free precession (TRUFI) sequence was selected. The applicator with metallic and line markers for CT and MR scans, respectively, were placed in the same patient setup.
The overall procedure of the study is presented in Fig. 1. We conducted image registration from the CT image set to the MR image set using a deformable registration algorithm implemented in ViewRay treatment planning system. The CT images are resampled to the same resolution as the MR image dimension of 334×300×288 pixels with voxel size at ≈1.5×1.5×1.5 mm3. For the CT image set, the connected-component labeling method was applied to remove the CT couch that was not presented in the MR images. The intensity of the CT image was normalized from [−1024 HU, 2048 HU] to [0, 1]. The MR images were corrected for low-frequency intensities and non-uniformities using the N4 bias field correction method. Because the MR images have different intensity ranges for different patients, we conducted normalized imaging using the 0% to 99% percentile value of the intensity distribution of each image to stabilize network training. X-ray dummy source markers in CT images were segmented by an intensity-based threshold method followed by a manual modification to provide marker position information to the deep learning model.
The overall architecture of the proposed models is illustrated in Fig. 2. The architecture comprises two fully convolutional 3D convolutional neural networks. The first one (Generator) performs as a sCT generation from MR image, whereas the second one (Segmenter) was used as the segmentation network. The Generator was based on the 3D U-Net with residual blocks. Owing to the graphical processing unit (GPU) memory constraints, we adjusted the number of filters at each level. The filter numbers were 8, 16, 32, 64, and 128 in the encoder for reducing the spatial dimensions by a factor of two by each layer and 128, 64, 32, 16, and 8 in the decoder for upsampling. The Segmenter uses the same 3D U-Net architecture as the Generator. This network performs voxel-wise classification using the softmax function. The output of the network is the segmentation map of the input patch. We defined the loss function (lossgen) of the Generator as follows:
where lossL1 is the mean absolute error (MAE) between the sCT and deformed CT (dCT) and lossDice is computed on the basis of the Dice coefficient between the predicted segmentation result from the sCT and the ground truth annotation. For the Segmenter, we chose the loss function as Dice loss between the dCT and the ground truth segmentation map. The weights and bias in the Generator layers as well as the Segmenter layers were jointly trained using adaptive stochastic gradient descent optimizers. The networks were trained for 2,000 epochs using 3,040 paired MRI–CT axial slices from 19 patients, defining the training cohort using a GPU NVIDIA GTX 1080Ti (11 GB). The remaining patients were considered the testing cohort.
The performance of the sCT generation models was evaluated quantitatively. We utilized three metrics to assess the accuracy of the image translation for the test set: the MAE, root mean square error (RMSE), and structural similarity (SSIM) between the sCT generated from the preprocessed MR images using the two deep neural networks and the dCT. The MAE and RMSE consider comparison of the pixel-wise difference between two images. The MAE and RMSE may be calculated respectively as:
where i is a voxel within the body and N is the total number of voxels.
The SSIM represents a comparison of the luminance, contrast, and structure by computing a similarity score between two images. This metric may be calculated as:
where µx is the mean intensity of image x:
σx is an estimate of the signal contrast:
σxy is a correlation coefficient between image x and y:
and c1 and c2 are constants.
The proposed method was applied to the low-tesla MR images, and their sCT images were produced. The representative case of a patient with cervical cancer was selected in the test set shown in Fig. 3. Fig. 3 shows the original CT images and the generated sCT images using the 3D U-Net and the proposed model. The metallic marker in the image of sCT generated using 3D U-Net was almost invisible, whereas it was visible in the sCT produced using the proposed model. Additionally, sCTs show soft tissue and dense structure, such as the pelvic bones and spine. These sCTs present smoother boundaries than the original CT images.
The MAE, RMSE, and SSIM between the dCT and sCT images calculated from the test set are summarized in Table 1. Generally, the sCT of our model showed smaller errors (MAE and RMSE) Generally, the sCT of our model showed smaller errors (MAE and RMSE) than the sCT of 3D U-Net. The MAE values were 7.94±2.31 HU and 7.49±2.10 HU for 3D U-Net and our model, respectively. However, the similarity score difference was insignificant.
Errors (MAE and RMSE) and similarity (SSIM) relative between synthetic CT and deformed CT in the test group
Metrics | 3D U-Net | Proposed model |
---|---|---|
MAE (HU) | 7.94±2.31 | 7.49±2.10 |
RMSE (HU) | 44.72±8.99 | 42.85±7.88 |
SSIM | 0.92±0.02 | 0.92±0.02 |
Values are presented as mean±standard deviation.
MAE, mean absolute error; RMSE, root mean square error; SSIM, structural similarity; CT, computed tomography; 3D, three-dimensional.
This study presented the feasibility of the sCT generation from the low-tesla MR image that improved the visibility of small metallic objects by employing the two-task-driven deep neural networks. Our method leverages a joint learning strategy that involves two networks: a Generator that converts MR images to CT images while preserving the region of interest and a segmentation network that predicts the segmentation mask of metallic dummy markers from the sCT images. Our findings show that the traditional U-Net model with loss function of pixel-wise difference has limited performance for generating small objects, such as the marker. Even though the region with the metallic markers in the CT images causes a large error in loss, the error is only a small portion of the total loss. Without incorporating the segmentation task in the deep learning architecture, the small objects were not preserved in the synthetic images, as shown in Fig. 2. For the proposed architecture, we designated the generated image as an input of the segmentation model and calculated the segmentation loss of the metallic objects to provide additional information to the generation model. As a result, the metallic markers could be well determined in the synthetic images of the proposed model than those of the 3D U-Net model.
The hybrid procedure currently used in the clinic for intracavitary gynecologic brachytherapy involves inevitable setup errors. To align the applicator position, rigid image registration was conducted between the CT and MR images to minimize the geometric distortion of the applicator. In this case, there is a disparity between the two images, such as on the location of internal organs. This effect could not be revealed in the dose calculation following the TG-43 protocol [19], which assumes a homogeneous water geometry. However, the influence might be emphasized when the model-based dose calculation according to TG-186 [20] is used, which assigns the tissue densities based on contoured organs. The MR-only workflow with sCT image can minimize the geometric discrepancy while accurately assigning the tissue mass densities derived from the sCT image for dose calculation.
Our study still has several limitations. We generated the sCT images from only the 0.35-T MR images acquired with TruFISP sequence used in radiation oncology. The MR images obtained using a higher-tesla machine of radiology could also be used for the 3D ICR treatment planning. The different image characteristics of the higher tesla MR scanners or other MR sequences affect the learning-based sCT generation model. Therefore, applying the proposed model to the MR-only ICR using other diagnostic MR images would be a valuable research endeavor in the future. Moreover, we evaluated the sCT images using only intensity-based comparison with original CT images. For clinical implementation, the sCT generation algorithm should be validated by comparing the dosimetric parameters between the ICR plans from original and synthetic images. The evaluation of the accuracy of the applicator reconstruction using the marker of sCT image is also necessary. These dosimetric and geometric considerations remained in further studies. Recently, it has been reported that adversarial loss using a generative adversarial neural network could increase the sharpness of the image compared with that using only the L1 or L2 loss. Therefore, the image quality of the sCT image might have the potential to be improved using other complex deep learning architectures or appropriate objective functions.
Our study presents the two-task-based deep learning models for generating the sCT images using low-tesla MR images for 3D ICR. This approach will be useful to the MR-only workflow in HDR brachytherapy.
This study was supported by a grant from the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP) (No. 2019M2A2B4096540).
The authors have nothing to disclose.
The data that support the findings of this study are available on request from the corresponding author.
Conceptualization: Hyeongmin Jin and Chang Heon Choi. Data curation: Hyun-Cheol Kang. Formal analysis: Hyeongmin Jin. Funding acquisition: Chang Heon Choi. Investigation: Hyeongmin Jin and Seonghee Kang. Methodology: Hyeongmin Jin and Seonghee Kang. Project administration: Hyun-Cheol Kang and Chang Heon Choi. Resources: Seonghee Kang and Hyun-Cheol Kang. Software: Hyeongmin Jin. Supervision: Hyun-Cheol Kang. Validation: Hyeongmin Jin and Seonghee Kang. Visualization: Hyeongmin Jin. Writing–original draft: Hyeongmin Jin. Writing–review & editing: Seonghee Kang and Chang Heon Choi.