Progress in Medical Physics 2022; 33(1): 1-9
Published online March 31, 2022
https://doi.org/10.14316/pmp.2022.33.1.1
Copyright © Korean Society of Medical Physics.
Chul-Young Yi1 , Yun Ho Kim1 , Don Yeong Jeong2
Correspondence to:Chul-Young Yi
(cyyi@kriss.re.kr)
Tel: 82-42-868-5370
Fax: 82-42-868-5671
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Purpose: To evaluate the effective volume of the Korea Research Institute of Standards and Science free air chamber (KRISS FAC) L1 used for the primary standard device of the low-energy X-ray air kerma.
Methods: The mechanical dimensions were measured using a 3-dimensional coordinate measuring machine (3-d CMM, Model UMM 500, Carl Zeiss). The diameter of the diaphragm was measured by a ring gauge calibrator (Model KRISS-DM1, KRISS). The elongation of the collector length due to electric field distortion was determined from the capacitance measurement of the KRISS FAC considering the result of the finite element method (FEM) analysis using the code QuickField v6.4.
Results: The measured length of the collector was 15.8003±0.0014 mm with a 68% confidence level (k =1). The aperture diameter of the diaphragm was 10.0021±0.0002 mm (k =1). The mechanical measurement volume of the KRISS FAC L1 was 1.2415±0.0006 cm3 (k =1). The elongated length of the collector due to the electric field distortion was 0.170±0.021 mm. Considering the elongated length, the effective measurement volume of the KRISS FAC L1 was 1.2548±0.0019 cm3 (k =1).
Conclusions: The effective volume of the KRISS FAC L1 was determined from the mechanically measured value by adding the elongated volume due to the electric field distortion in the FAC. The effective volume will replace the existing mechanically determined volume in establishing and maintaining the primary standard of the low-energy X-ray.
KeywordsPrimary standard, Air kerma, Low-energy X-ray, Free air chamber, Effective volume
The air kerma is the basic physical quantity for measuring the X-ray dose for radiation protection purposes in both medical and the industrial fields. The air kerma standard of the X-ray beams is realized by the free air chamber (FAC), known as the “primary standard device” if it is used to establish the air kerma standard in the national metrology institute (NMI). Most NMIs have two FACs: one for measuring the low-energy X-ray and the other for measuring the medium-energy X-ray. According to the NMI of Korea, we used two FACs to realize the primary standards of air kermas for low- and medium-energy X-ray beams, including the mammography X-ray beams specified in the International Organization for Standardization (ISO) [1] and International Electrotechnical Commission (IEC) reference documents [2]. We manufactured the FACs with our own designs, mainly because they were not commercially available. The structural information of the FAC is critical in realizing the unit of air kerma, and is obtained with sufficient accuracy by measuring it during the manufacturing and assembly stages of the FAC.
The air kerma from the X-ray beams in the reference plane is measured by the FAC using the following equation:
where
where ρair is the dry air density,
Given that the air mass is usually determined by multiplying the dry air density by the measurement volume, as given in Eq. (2), the measurement volume of the FAC is an essential physical factor in conducting the absolute measurement of the air kerma using the FAC. As the air kerma is defined as the mean sum of the initial kinetic energies of all charged particles liberated in air by the incident photons divided by the mass of air [3], the volume of the FAC should be carefully evaluated when wishing to develop the primary standard air kerma ensuring traceability to the International System of Units.
We have recently found that the existing mechanically determined volume of the Korea Research Institute of Standards and Science (KRISS) FAC L1 for the low-energy X-ray measurement is different from the value deduced from the capacitance measurement. In this study, considering the result obtained from the finite element method (FEM), the change in the collector length due to the electric field distortion was evaluated from the capacitance measurement and, as a result, the effective volume of the KRISS FAC L1 was determined.
The cross-sectional view of the KRISS FAC L1 for measuring the low-energy X-ray is presented in Fig. 1. The base plate, in which the collecting electrode is located, was in the same plane as the collecting electrode, and within 5 μm. The distance between the guard strips was maintained using four sapphire disks with a nominal thickness of 2 mm. The difference in thickness in a layer composed of four sapphires was <1 μm. After piling all the guard strips over the base plate, we checked the assembled heights with a 3-dimensional coordinate measuring machine (3-d CMM) on the upper high-voltage (HV) plate. The maximum difference in height on the upper surface of the HV plate measured from the base plate was <10 μm. The front wall was comprised of a nominal 10-mm-thick stainless steel, the face of which was in contact with the rear plane of the diaphragm and used for the reference plane of the measurement. The flatness of the reference plane was within 1 μm. The subtended angle of the front face from the base plate was 90.0057±0.0007 degrees, with a 68% confidence level (
The mechanical dimensions were measured using the 3-d CMM (Model UMM 500, Carl Zeiss, Oberkochen, Germany) with the typical measurement uncertainty at a 68% confidence level (
where
where
The high voltage was loaded onto the HV plate in the FAC (see Fig. 1) during the ionizing current measurement. We loaded −2000 V to the HV plate of the KRISS FAC L1. If the electric field is distorted, it influences the measurement volume of the FAC. To evaluate how the electric field is distorted, we analyzed the electric field formed in our FAC using the two-dimensional FEM using the code QuickField v6.4. Based on the measured dimensions by the 3-d CMM, the FAC was modeled realistically in the QuickField and the electric field line was drawn along the height direction starting from the air gap between the collector and base palate, with the execution script named electric_field_lines.hta, which was included in the QuickField v6.4.
The capacitance of our FAC was measured with the charge mode of the Keithley 6517 electrometer (Keithley, Cleveland, OH, USA). Before the capacitance measurement, we calibrated the charge mode using a standard air capacitor and the constant voltage source according to the KRISS calibration procedure of the electrometer [6]. The uncertainty of the capacitance measurement was within 0.02% (
where
If the area of the plate (in a rectangular shape) with the length
where ∆
The collector dimensions of the KRISS FAC L1 were measured using the 3-d CMM. The collector length, including the air gap between the collector and base plate, was 15.8003±0.0014 mm at a 68% confidence level (
The electric field lines are shown in Fig. 2. The electric field line in Fig. 2a is drawn starting from the center of the air gap in the collector width, while that in Fig. 2b is drawn from the air gap along the collector length.
We found that the distortion of the electric field line was mainly originated from the thickness of the guard strip. The distortion becomes larger as the distance from the center is relatively close to the guard strips. The field line starting from the air gap shown in Fig. 2 looks straight but it bends toward the guard strips. The field lines in Fig. 2 are enlarged in Fig. 3. The elongation peaked on the beam axis, the value of which was 0.08225 mm, with an average value of 0.0433 mm in the length of the collector. In terms of the width, the peak value was 0.16950 mm, with an average of 0.0892 mm. The total length increase in the collector electrode evaluated by the FEM was obtained by adding the peak elongations in both sides of the collector, i.e., 0.1645 mm. If the elongation is constant along the height, the average elongated length of the collector electrode is 0.0865 mm. The elongated length evaluated from the FEM analysis showed a maximum variation of approximately 0.01 mm with the number of nodes we selected.
The collector area determined by the capacitance measurement was 0.83% larger than the area determined by the 3-d CMM measurement and the ∆l was 0.0895 mm from Eq. (6) of (∆
Because the elongated volume was smaller than the collection volume in the cuboid shape, it was necessary to evaluate and correct the electron energy lost outside the elongated region. For evaluating the energy deposition in the elongated region by the Monte–Carlo method, we modeled 14 air slabs, each of 5-mm thickness. Their lengths and widths were the same as those of the collector electrode. The air slabs were piled and placed around the middle of the air box with the same dimension as the KRISS FAC. In the simulation model, a low-energy X-ray beam 10-mm in diameter was incident on the front face of the air box. The beam axis was at the center of the air box, which was 35 mm-high from the bottom of the air box. The centers of the slabs were 77-mm away from the front face of the air box. The air slabs were paired in plane symmetry with the beam axis plane parallel to the air slabs as the symmetry plane. The energy deposition portion of the paired air slabs was calculated using the code EGSnrc [9] and the statistical uncertainty of the calculation was <0.03% (
The calculated result is given in Table 1 and 2. As expected, the portions of energy deposition in the slabs placed outer from the beam center were minor. The portion of energy deposition was >90% in the 1st paired slabs adjacent to the beam center for all beam codes in the investigation, with the exception of the beam codes Bureau International des Poids et Mesures (BIPM) 50b and BIPM50a.
Table 1 Energy deposition portions in the paired air slabs for the low-energy X-ray beams of BIPM quality
Location of the paired air slabs | Energy deposition portion (%) | ||||||
---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |||
35–40 | 30–35 | 98.87 | 93.81 | 93.06 | 79.56 | 65.88 | |
40–45 | 25–30 | 1.07 | 6.01 | 6.85 | 17.01 | 23.91 | |
45–50 | 20–25 | 0.02 | 0.14 | 0.07 | 2.72 | 7.50 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.55 | 2.06 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.12 | 0.51 | |
60–65 | 5–10 | 0.01 | 0.01 | 0.00 | 0.03 | 0.11 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.01 | 0.03 |
BIPM, Bureau International des Poids et Mesures.
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as the symmetry plane. The low-energy X-ray beams of BIPM quality were incident on the front face of the air box (see Fig. 4) and the beam size was 10-mm in diameter. The lengths and widths of the slabs were the same as those of the collector electrode, 15.8 mm and 70.0 mm, respectively.
Table 2 Energy deposition portions in the paired air slabs for the mammography X-ray beams
Location of the paired air slabs | Energy deposition portion (%) | ||||||||
---|---|---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||||
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |||
35–40 | 30–35 | 94.39 | 93.34 | 92.78 | 92.11 | 91.70 | 90.93 | 90.35 | |
40–45 | 25–30 | 5.56 | 6.61 | 7.16 | 7.82 | 8.20 | 8.89 | 9.34 | |
45–50 | 20–25 | 0.02 | 0.03 | 0.03 | 0.05 | 0.07 | 0.15 | 0.25 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.02 | 0.04 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | |
60–65 | 5–10 | 0.00 | 0.00 | 0.01 | 0.00 | 0.00 | 0.00 | 0.00 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 |
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as a symmetry plane. The low-energy mammography X-ray beams were incident on the front face of the air box (see Fig. 4). The number appearing in the beam code indicated the high voltage loaded on the X-ray tube with a molybdenum anode. The added filtration was 0.03 mm molybdenum, and the inherent filtration was 1 mm beryllium.
Note that the shape of the elongated volume defined by the electric field distortion is in the plane symmetry with the beam axis plane as the symmetry plane. Due to the plane symmetry and the charged particle equilibrium in the axial direction of the incident X-ray beam in the elongated region, the energy deposition portion in the thinly sliced elongated region is proportional to the cross-sectional area of the thin slice. We cut the elongated volume into 14 sections, each of which had a thickness of 5 mm. They were also paired symmetrically with the beam axis plane as the symmetry plane, as depicted in Fig. 5. In the figure, the sum of the cross-sectional areas of the pared slices, where
To correct the electron energy loss conveyed outside the elongated region, we redefined the electron loss correction of the KRISS FAC L1 as follows:
where
From Fig. 4–6, we have the following relationship:
where
with
Table 3 Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||
---|---|---|---|---|---|
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |
0.9852 | 0.9810 | 0.9806 | 0.9630 | 0.9363 | |
1.0002 | 1.0002 | 1.0002 | 1.0004 | 1.0007 |
BIPM, Bureau International des Poids et Mesures.
Table 4 Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||||
---|---|---|---|---|---|---|---|
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |
0.9817 | 0.9809 | 0.9804 | 0.9798 | 0.9795 | 0.9787 | 0.9780 | |
1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 |
The effective volume measurement of our FAC will be obtained by replacing
where
The FEM analysis on the electric field of KRISS FAC L1 demonstrated that there was elongation in the measurement volume of the FAC due to electric field distortion caused by the thickness of the guard strips surrounding the collector electrode. This electric field distortion cannot be avoided as long as guard strips are used, and they are necessary in the current design of the parallel plate FAC. The amount of elongation was evaluated from the measured capacitance of the FAC. The energy loss correction for the elongated volume was also estimated.
Most NMIs have their own parallel plate-type FACs as the primary standard devices, and, without exception, there are guard strips inside; thus, the effect of the electric field distortion on the measurement volume of the FACs should be evaluated. We expect that the proposed methodology will be useful in the evaluation of the elongated volume.
We evaluated the effective volume of the FAC based on the precision measurement for the collector length and the aperture diameter of the diaphragm of the KRISS FAC L1. During the evaluation, the elongated length of the collector electrode was determined from the capacitance measurement of the FAC considering the FEM analysis result. The effective volume of the KRISS FAC L1 was 1.2548±0.0019 cm2 with a 68% confidence level (
In this study, we evaluated an essential physical parameter, the effective volume of the KRISS FAC L1, for the low-energy X-ray air kerma measurement. The electron loss corrections due to the elongated volume were evaluated thoroughly, with the values presented in Table 3 and 4. The effective volume will replace the existing mechanically determined volume and will be used to establish and maintain the primary standard of the low-energy X-ray air kerma of KRISS. Therefore, our findings contribute to the successful implementation of the air kerma metrology.
This work was supported by the Korea Research Institute of Standards and Science (KRISS) under the project “Development of Measurement Standards for Medical Radiation (2022).” The authors express gratitude to Dr. D.T. Burns in the Bureau International des Poids et Mesures (BIPM) for suggesting the introduction of (∆
The authors have nothing to disclose.
The data that support the findings of this study are available on request from the corresponding author.
Conceptualization: Chul-Young Yi. Data curation: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong. Formal analysis: Chul-Young Yi. Funding acquisition: none. Investigation: Chul-Young Yi and Yun Ho Kim. Methodology: Chul-Young Yi. Project administration: Chul-Young Yi. Resources: Chul-Young Yi and Don Yeong Jeong. Software: Chul-Young Yi and Yun Ho Kim. Supervision: Chul-Young Yi. Validation: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong. Writing–original draft: Chul-Young Yi. Writing–review and editing: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong.
Progress in Medical Physics 2022; 33(1): 1-9
Published online March 31, 2022 https://doi.org/10.14316/pmp.2022.33.1.1
Copyright © Korean Society of Medical Physics.
Chul-Young Yi1 , Yun Ho Kim1 , Don Yeong Jeong2
1Ionizing Radiation Metrology Group, 2Length Group, Korea Research Institute of Standards and Science (KRISS), Daejeon, Korea
Correspondence to:Chul-Young Yi
(cyyi@kriss.re.kr)
Tel: 82-42-868-5370
Fax: 82-42-868-5671
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Purpose: To evaluate the effective volume of the Korea Research Institute of Standards and Science free air chamber (KRISS FAC) L1 used for the primary standard device of the low-energy X-ray air kerma.
Methods: The mechanical dimensions were measured using a 3-dimensional coordinate measuring machine (3-d CMM, Model UMM 500, Carl Zeiss). The diameter of the diaphragm was measured by a ring gauge calibrator (Model KRISS-DM1, KRISS). The elongation of the collector length due to electric field distortion was determined from the capacitance measurement of the KRISS FAC considering the result of the finite element method (FEM) analysis using the code QuickField v6.4.
Results: The measured length of the collector was 15.8003±0.0014 mm with a 68% confidence level (k =1). The aperture diameter of the diaphragm was 10.0021±0.0002 mm (k =1). The mechanical measurement volume of the KRISS FAC L1 was 1.2415±0.0006 cm3 (k =1). The elongated length of the collector due to the electric field distortion was 0.170±0.021 mm. Considering the elongated length, the effective measurement volume of the KRISS FAC L1 was 1.2548±0.0019 cm3 (k =1).
Conclusions: The effective volume of the KRISS FAC L1 was determined from the mechanically measured value by adding the elongated volume due to the electric field distortion in the FAC. The effective volume will replace the existing mechanically determined volume in establishing and maintaining the primary standard of the low-energy X-ray.
Keywords: Primary standard, Air kerma, Low-energy X-ray, Free air chamber, Effective volume
The air kerma is the basic physical quantity for measuring the X-ray dose for radiation protection purposes in both medical and the industrial fields. The air kerma standard of the X-ray beams is realized by the free air chamber (FAC), known as the “primary standard device” if it is used to establish the air kerma standard in the national metrology institute (NMI). Most NMIs have two FACs: one for measuring the low-energy X-ray and the other for measuring the medium-energy X-ray. According to the NMI of Korea, we used two FACs to realize the primary standards of air kermas for low- and medium-energy X-ray beams, including the mammography X-ray beams specified in the International Organization for Standardization (ISO) [1] and International Electrotechnical Commission (IEC) reference documents [2]. We manufactured the FACs with our own designs, mainly because they were not commercially available. The structural information of the FAC is critical in realizing the unit of air kerma, and is obtained with sufficient accuracy by measuring it during the manufacturing and assembly stages of the FAC.
The air kerma from the X-ray beams in the reference plane is measured by the FAC using the following equation:
where
where ρair is the dry air density,
Given that the air mass is usually determined by multiplying the dry air density by the measurement volume, as given in Eq. (2), the measurement volume of the FAC is an essential physical factor in conducting the absolute measurement of the air kerma using the FAC. As the air kerma is defined as the mean sum of the initial kinetic energies of all charged particles liberated in air by the incident photons divided by the mass of air [3], the volume of the FAC should be carefully evaluated when wishing to develop the primary standard air kerma ensuring traceability to the International System of Units.
We have recently found that the existing mechanically determined volume of the Korea Research Institute of Standards and Science (KRISS) FAC L1 for the low-energy X-ray measurement is different from the value deduced from the capacitance measurement. In this study, considering the result obtained from the finite element method (FEM), the change in the collector length due to the electric field distortion was evaluated from the capacitance measurement and, as a result, the effective volume of the KRISS FAC L1 was determined.
The cross-sectional view of the KRISS FAC L1 for measuring the low-energy X-ray is presented in Fig. 1. The base plate, in which the collecting electrode is located, was in the same plane as the collecting electrode, and within 5 μm. The distance between the guard strips was maintained using four sapphire disks with a nominal thickness of 2 mm. The difference in thickness in a layer composed of four sapphires was <1 μm. After piling all the guard strips over the base plate, we checked the assembled heights with a 3-dimensional coordinate measuring machine (3-d CMM) on the upper high-voltage (HV) plate. The maximum difference in height on the upper surface of the HV plate measured from the base plate was <10 μm. The front wall was comprised of a nominal 10-mm-thick stainless steel, the face of which was in contact with the rear plane of the diaphragm and used for the reference plane of the measurement. The flatness of the reference plane was within 1 μm. The subtended angle of the front face from the base plate was 90.0057±0.0007 degrees, with a 68% confidence level (
The mechanical dimensions were measured using the 3-d CMM (Model UMM 500, Carl Zeiss, Oberkochen, Germany) with the typical measurement uncertainty at a 68% confidence level (
where
where
The high voltage was loaded onto the HV plate in the FAC (see Fig. 1) during the ionizing current measurement. We loaded −2000 V to the HV plate of the KRISS FAC L1. If the electric field is distorted, it influences the measurement volume of the FAC. To evaluate how the electric field is distorted, we analyzed the electric field formed in our FAC using the two-dimensional FEM using the code QuickField v6.4. Based on the measured dimensions by the 3-d CMM, the FAC was modeled realistically in the QuickField and the electric field line was drawn along the height direction starting from the air gap between the collector and base palate, with the execution script named electric_field_lines.hta, which was included in the QuickField v6.4.
The capacitance of our FAC was measured with the charge mode of the Keithley 6517 electrometer (Keithley, Cleveland, OH, USA). Before the capacitance measurement, we calibrated the charge mode using a standard air capacitor and the constant voltage source according to the KRISS calibration procedure of the electrometer [6]. The uncertainty of the capacitance measurement was within 0.02% (
where
If the area of the plate (in a rectangular shape) with the length
where ∆
The collector dimensions of the KRISS FAC L1 were measured using the 3-d CMM. The collector length, including the air gap between the collector and base plate, was 15.8003±0.0014 mm at a 68% confidence level (
The electric field lines are shown in Fig. 2. The electric field line in Fig. 2a is drawn starting from the center of the air gap in the collector width, while that in Fig. 2b is drawn from the air gap along the collector length.
We found that the distortion of the electric field line was mainly originated from the thickness of the guard strip. The distortion becomes larger as the distance from the center is relatively close to the guard strips. The field line starting from the air gap shown in Fig. 2 looks straight but it bends toward the guard strips. The field lines in Fig. 2 are enlarged in Fig. 3. The elongation peaked on the beam axis, the value of which was 0.08225 mm, with an average value of 0.0433 mm in the length of the collector. In terms of the width, the peak value was 0.16950 mm, with an average of 0.0892 mm. The total length increase in the collector electrode evaluated by the FEM was obtained by adding the peak elongations in both sides of the collector, i.e., 0.1645 mm. If the elongation is constant along the height, the average elongated length of the collector electrode is 0.0865 mm. The elongated length evaluated from the FEM analysis showed a maximum variation of approximately 0.01 mm with the number of nodes we selected.
The collector area determined by the capacitance measurement was 0.83% larger than the area determined by the 3-d CMM measurement and the ∆l was 0.0895 mm from Eq. (6) of (∆
Because the elongated volume was smaller than the collection volume in the cuboid shape, it was necessary to evaluate and correct the electron energy lost outside the elongated region. For evaluating the energy deposition in the elongated region by the Monte–Carlo method, we modeled 14 air slabs, each of 5-mm thickness. Their lengths and widths were the same as those of the collector electrode. The air slabs were piled and placed around the middle of the air box with the same dimension as the KRISS FAC. In the simulation model, a low-energy X-ray beam 10-mm in diameter was incident on the front face of the air box. The beam axis was at the center of the air box, which was 35 mm-high from the bottom of the air box. The centers of the slabs were 77-mm away from the front face of the air box. The air slabs were paired in plane symmetry with the beam axis plane parallel to the air slabs as the symmetry plane. The energy deposition portion of the paired air slabs was calculated using the code EGSnrc [9] and the statistical uncertainty of the calculation was <0.03% (
The calculated result is given in Table 1 and 2. As expected, the portions of energy deposition in the slabs placed outer from the beam center were minor. The portion of energy deposition was >90% in the 1st paired slabs adjacent to the beam center for all beam codes in the investigation, with the exception of the beam codes Bureau International des Poids et Mesures (BIPM) 50b and BIPM50a.
Table 1 . Energy deposition portions in the paired air slabs for the low-energy X-ray beams of BIPM quality.
Location of the paired air slabs | Energy deposition portion (%) | ||||||
---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |||
35–40 | 30–35 | 98.87 | 93.81 | 93.06 | 79.56 | 65.88 | |
40–45 | 25–30 | 1.07 | 6.01 | 6.85 | 17.01 | 23.91 | |
45–50 | 20–25 | 0.02 | 0.14 | 0.07 | 2.72 | 7.50 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.55 | 2.06 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.12 | 0.51 | |
60–65 | 5–10 | 0.01 | 0.01 | 0.00 | 0.03 | 0.11 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.01 | 0.03 |
BIPM, Bureau International des Poids et Mesures..
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as the symmetry plane. The low-energy X-ray beams of BIPM quality were incident on the front face of the air box (see Fig. 4) and the beam size was 10-mm in diameter. The lengths and widths of the slabs were the same as those of the collector electrode, 15.8 mm and 70.0 mm, respectively..
Table 2 . Energy deposition portions in the paired air slabs for the mammography X-ray beams.
Location of the paired air slabs | Energy deposition portion (%) | ||||||||
---|---|---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||||
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |||
35–40 | 30–35 | 94.39 | 93.34 | 92.78 | 92.11 | 91.70 | 90.93 | 90.35 | |
40–45 | 25–30 | 5.56 | 6.61 | 7.16 | 7.82 | 8.20 | 8.89 | 9.34 | |
45–50 | 20–25 | 0.02 | 0.03 | 0.03 | 0.05 | 0.07 | 0.15 | 0.25 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.02 | 0.04 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | |
60–65 | 5–10 | 0.00 | 0.00 | 0.01 | 0.00 | 0.00 | 0.00 | 0.00 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 |
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as a symmetry plane. The low-energy mammography X-ray beams were incident on the front face of the air box (see Fig. 4). The number appearing in the beam code indicated the high voltage loaded on the X-ray tube with a molybdenum anode. The added filtration was 0.03 mm molybdenum, and the inherent filtration was 1 mm beryllium..
Note that the shape of the elongated volume defined by the electric field distortion is in the plane symmetry with the beam axis plane as the symmetry plane. Due to the plane symmetry and the charged particle equilibrium in the axial direction of the incident X-ray beam in the elongated region, the energy deposition portion in the thinly sliced elongated region is proportional to the cross-sectional area of the thin slice. We cut the elongated volume into 14 sections, each of which had a thickness of 5 mm. They were also paired symmetrically with the beam axis plane as the symmetry plane, as depicted in Fig. 5. In the figure, the sum of the cross-sectional areas of the pared slices, where
To correct the electron energy loss conveyed outside the elongated region, we redefined the electron loss correction of the KRISS FAC L1 as follows:
where
From Fig. 4–6, we have the following relationship:
where
with
Table 3 . Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||
---|---|---|---|---|---|
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |
0.9852 | 0.9810 | 0.9806 | 0.9630 | 0.9363 | |
1.0002 | 1.0002 | 1.0002 | 1.0004 | 1.0007 |
BIPM, Bureau International des Poids et Mesures..
Table 4 . Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||||
---|---|---|---|---|---|---|---|
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |
0.9817 | 0.9809 | 0.9804 | 0.9798 | 0.9795 | 0.9787 | 0.9780 | |
1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 |
The effective volume measurement of our FAC will be obtained by replacing
where
The FEM analysis on the electric field of KRISS FAC L1 demonstrated that there was elongation in the measurement volume of the FAC due to electric field distortion caused by the thickness of the guard strips surrounding the collector electrode. This electric field distortion cannot be avoided as long as guard strips are used, and they are necessary in the current design of the parallel plate FAC. The amount of elongation was evaluated from the measured capacitance of the FAC. The energy loss correction for the elongated volume was also estimated.
Most NMIs have their own parallel plate-type FACs as the primary standard devices, and, without exception, there are guard strips inside; thus, the effect of the electric field distortion on the measurement volume of the FACs should be evaluated. We expect that the proposed methodology will be useful in the evaluation of the elongated volume.
We evaluated the effective volume of the FAC based on the precision measurement for the collector length and the aperture diameter of the diaphragm of the KRISS FAC L1. During the evaluation, the elongated length of the collector electrode was determined from the capacitance measurement of the FAC considering the FEM analysis result. The effective volume of the KRISS FAC L1 was 1.2548±0.0019 cm2 with a 68% confidence level (
In this study, we evaluated an essential physical parameter, the effective volume of the KRISS FAC L1, for the low-energy X-ray air kerma measurement. The electron loss corrections due to the elongated volume were evaluated thoroughly, with the values presented in Table 3 and 4. The effective volume will replace the existing mechanically determined volume and will be used to establish and maintain the primary standard of the low-energy X-ray air kerma of KRISS. Therefore, our findings contribute to the successful implementation of the air kerma metrology.
This work was supported by the Korea Research Institute of Standards and Science (KRISS) under the project “Development of Measurement Standards for Medical Radiation (2022).” The authors express gratitude to Dr. D.T. Burns in the Bureau International des Poids et Mesures (BIPM) for suggesting the introduction of (∆
The authors have nothing to disclose.
The data that support the findings of this study are available on request from the corresponding author.
Conceptualization: Chul-Young Yi. Data curation: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong. Formal analysis: Chul-Young Yi. Funding acquisition: none. Investigation: Chul-Young Yi and Yun Ho Kim. Methodology: Chul-Young Yi. Project administration: Chul-Young Yi. Resources: Chul-Young Yi and Don Yeong Jeong. Software: Chul-Young Yi and Yun Ho Kim. Supervision: Chul-Young Yi. Validation: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong. Writing–original draft: Chul-Young Yi. Writing–review and editing: Chul-Young Yi, Yun Ho Kim, and Don Yeong Jeong.
Table 1 Energy deposition portions in the paired air slabs for the low-energy X-ray beams of BIPM quality
Location of the paired air slabs | Energy deposition portion (%) | ||||||
---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |||
35–40 | 30–35 | 98.87 | 93.81 | 93.06 | 79.56 | 65.88 | |
40–45 | 25–30 | 1.07 | 6.01 | 6.85 | 17.01 | 23.91 | |
45–50 | 20–25 | 0.02 | 0.14 | 0.07 | 2.72 | 7.50 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.55 | 2.06 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.12 | 0.51 | |
60–65 | 5–10 | 0.01 | 0.01 | 0.00 | 0.03 | 0.11 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.01 | 0.03 |
BIPM, Bureau International des Poids et Mesures.
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as the symmetry plane. The low-energy X-ray beams of BIPM quality were incident on the front face of the air box (see Fig. 4) and the beam size was 10-mm in diameter. The lengths and widths of the slabs were the same as those of the collector electrode, 15.8 mm and 70.0 mm, respectively.
Table 2 Energy deposition portions in the paired air slabs for the mammography X-ray beams
Location of the paired air slabs | Energy deposition portion (%) | ||||||||
---|---|---|---|---|---|---|---|---|---|
Upper slab (mm) | Lower slab (mm) | Beam code | |||||||
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |||
35–40 | 30–35 | 94.39 | 93.34 | 92.78 | 92.11 | 91.70 | 90.93 | 90.35 | |
40–45 | 25–30 | 5.56 | 6.61 | 7.16 | 7.82 | 8.20 | 8.89 | 9.34 | |
45–50 | 20–25 | 0.02 | 0.03 | 0.03 | 0.05 | 0.07 | 0.15 | 0.25 | |
50–55 | 15–20 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.02 | 0.04 | |
55–60 | 10–15 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | 0.01 | |
60–65 | 5–10 | 0.00 | 0.00 | 0.01 | 0.00 | 0.00 | 0.00 | 0.00 | |
65–70 | 0–5 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 |
The air slabs were piled symmetrically with the beam axis plane placed 35-mm-high from the collector electrode as a symmetry plane. The low-energy mammography X-ray beams were incident on the front face of the air box (see Fig. 4). The number appearing in the beam code indicated the high voltage loaded on the X-ray tube with a molybdenum anode. The added filtration was 0.03 mm molybdenum, and the inherent filtration was 1 mm beryllium.
Table 3 Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||
---|---|---|---|---|---|
BIPM10 | BIPM30 | BIPM25 | BIPM50b | BIPM50a | |
0.9852 | 0.9810 | 0.9806 | 0.9630 | 0.9363 | |
1.0002 | 1.0002 | 1.0002 | 1.0004 | 1.0007 |
BIPM, Bureau International des Poids et Mesures.
Table 4 Energy deposition fraction in the elongated region and the correction factor
Quantity | Beam code | ||||||
---|---|---|---|---|---|---|---|
M20 | M23 | M25 | M28 | M30 | M35 | M40 | |
0.9817 | 0.9809 | 0.9804 | 0.9798 | 0.9795 | 0.9787 | 0.9780 | |
1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 | 1.0002 |