검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Split Viewer

Review Article

Progress in Medical Physics 2020; 31(3): 35-53

Published online September 30, 2020 https://doi.org/10.14316/pmp.2020.31.3.35

Copyright © Korean Society of Medical Physics.

Magnetic Resonance Imaging: Historical Overview, Technical Developments, and Clinical Applications

Geon-Ho Jahng1 , Soonchan Park1 , Chang-Woo Ryu1 , Zang-Hee Cho2

1Department of Radiology, Kyung Hee University Hospital at Gangdong, College of Medicine, Kyung Hee University, 2Neuroscience Convergence Center, Korea University, Seoul, Korea

Correspondence to:Geon-Ho Jahng
(ghjahng@gmail.com)
Tel: 82-2-440-6187
Fax: 82-2-440-6932

Received: May 29, 2020; Revised: August 1, 2020; Accepted: September 1, 2020

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

The authors congratulate the cerebrations for the 30 years of the Korean Society of Medical Physics (http://www.ksmp.or.kr/). The paper is published to recognize the anniversary. Geon-Ho Jahng invited Professor Z. H. Cho to join to submit this manuscript because he has been one of the leaders in the field of magnetic resonance imaging (MRI) during the last 40 years. In this review, we describe the development and clinical histories of MRI internationally and domestically. We also discuss diffusion and perfusion MRI, molecular imaging using MRI and MR spectroscopy (MRS), and the hybrid systems, such as positron emission tomography一MRI (PET一MRI), MR-guided focused ultrasound surgery (MRgFUS), and MRI-guided linear accelerators (MRI-LINACs). In each part, we discuss the historical evolution of the developments, technical developments, and clinical applications.

KeywordsMagnetic resonance imaging, History, Technical development, Clinical application, Review

1. International contributions

The first nuclear magnetic resonance (NMR) signals from a living animal were acquired from an anesthetized rat in 1968 [1]. The capability of NMR to differentiate tumors from normal tissue was reported by Damadian in 1971 [2]. Magnetic resonance imaging (MRI) was developed by Lauterbur [3] based on the encoding spatial information of NMR signals with magnetic field gradients. The first cross-sectional image of a living mouse was published in 1974 by Lauterbur [4]. The echo-planar imaging (EPI) technique was developed by Mansfield [5]. The first MRI body scan of a human being was performed by Damadian in 1977 [6]. For their efforts, Paul C. Lauterbur and Peter Mansfield were awarded the Nobel Prize in Physiology and Medicine in 2003 in justification of the fundamental importance and applicability of MRI in medicine. The k-space was patented by Richard S. Likes in U.S.A (#U.S. Patent 4,307,343). The first clinical MRI system was installed in the early 1980s. The first whole-body MRI scanner was developed in Korea in 1982 and was commercialized in 1984 by Goldstar (Seoul, Korea). At same time, another whole-body MRI system was built at the University of Aberdeen and was used at St. Bartholomew’s Hospital (London, UK) in 1983. Superconducting MRI was developed and it was clinically used in 1985 to 1988 at Siemens (Erlangen, Germany), General Electric (GE, Boston, MA, USA), and Goldstar. Fig. 1 shows the timeline of MRI developments and a summary of the major contributions achieved internationally and domestically.

Figure 1.Timeline of MRI developments and summary of the major contributions. MRI, magnetic resonance imaging; M, magnetic; R, resonance; I, imaging; F, functional; NMR, nuclear magnetic resonance; BOLD, blood oxygen level-dependent; NIH, National Institutes of Health; PET, positron emission tomography; MRI–LINAC, MRI-guided linear accelerators.

2. Domestic contributions

The first MRI system was developed by the Korean Advanced Institute of Science and Technology (Daejeon, Korea), and was installed in Shin Hwa Hospital (Shin Hwa Nursing Hospital, Seoul, Korea) in 1984. This MRI scanner used a 0.2 T permanent magnet. Therefore, its use for clinical studies was limited considerably at that time. The first commercial MRI (Spectro-20000; Goldstar) was installed in Seoul University Hospital (Seoul, Korea) in 1987. This MRI system was a 2.0 Tesla superconductivity magnet. During this period, GE, and Siemens also developed the 1.5 T superconductive MRI systems. In 1990, the first animal 4.7 T MRI system (Biospec, Bruker, Switzerland) was installed in Seoul’s Asan Joong Ang Hospital (Asan Medical Center, Seoul, Korea). Among the many incremental developments in MRI in Korea, one of the notable progresses made during the early 21st century was the development of ultrahigh field (UHF) 7.0 T MRI and its applications [7 -11]. Another notable worldwide research activity was the development of the hybrid positron emission tomography–MRI (PET–MRI) system. Since the introduction of the UHF (7.0 T) MRI system in Korea, an effort was initiated for the construction of a hybrid system with 7.0 T MRI and PET that led to the development of one of the world’s most advanced PET–MRI systems in 2007. With these developments, a number of neurotransmitter studies was initiated to study the serotonergic distribution in the brainstem in vivo [7,8]. Another new and interesting development with UHF 7.0 T MRI was the super-resolution tractography initiative with resolution down to 200 µm [12].

3. Development of basic imaging contrast

In 1950, spin echoes and free induction decay were detected by Hahn [13,14]. Therefore, the spin echo (SE) is commonly referred to as Hahn’s echo. MR angiography was developed by Charles L. Dumoulin and Howard R. Hart at the General Electric in 1986 (“Blood-flow checker”. Popular Science: 12. 1987). The fluid attenuation inversion recovery (FLAIR) pulse sequence that yields high-signal regions in normal white matter, was demonstrated by Hajnal et al. in 1992 [15]. Blood Oxygen Level Dependent signal was recognized by Ogawa [16] at AT&T Bell labs in 1990. Susceptibility-weighted imaging (SWI) was developed by Reichenbach et al. [17] at Washington University in 1997.

Fig. 2 shows patient cases that demonstrate the use of imaging contrast agents acquired from a 83-year-old female, 71-year-old male, and from a 26-year-old male with a 3 T MRI system.

Figure 2.Patient cases to show imaging contrasts acquired from (a) 83-year-old female, (b) 71-year-old male, and (c) 26-year-old male using a 3 T MRI system. F, female; M, male; T1W, T1-weighted; T2W, T2-weighted; FLAIR, fluid attenuated inversion recovery; GRE, gradient-echo; TOF, time of flight; CE MRA, contrast-enhanced magnetic resonance angiography; DWI b0, diffusion-weighted image with b=0 s/mm2; DWI b1000, diffusion-weighted image with b=1000 s/mm2; ADC, apparent diffusion coefficient; CE T1W, contrast-enhanced T1-weighted; APT, amide proton transfer; DSC rCBV, dynamic susceptibility contrast relative cerebral blood volume; DSC rCBF, dynamic susceptibility contrast relative cerebral blood flow; DSC MTT, dynamic susceptibility contrast mean transit time; SVS, single voxel spectroscopy; DTI b1000, diffusion tensor imaging with b=1000 s/mm2; DTI FA, diffusion tensor imaging fractional anisotropy; fMRI, functional MRI.

Speed is always important to clinical practice. Fast MRI techniques were introduced and were based on the use of multiple refocusing pulses, commonly referred to as turbo spin-echo (TSE) imaging or turbo gradient-echo imaging. Rapid Acquisition with Refocused Echoes (RARE) technique was commercially implemented after Hennig [18] described it in 1986. In this sequence, the echo train length (ETL), also known as the turbo factor, denotes the number of echoes acquired at a given repetition time (TR). Spiral sequences were introduced in 1986 by Ahn et al. [19].

1. Technical developments

Several pulse sequences for fast imaging were developed, such as TSE or turbo field echo, half-Fourier, single-shot turbo spin echo (HASTE), gradient and spin echo (GRASE), balanced steady-state free-precession (bSSFP), EPI, and spiral [20]. First, the principle of operation of TSE or RARE sequences was based on the filling of many k-space lines after every radiofrequency (RF) excitation that reduced the MR scan time by reducing the number of RF excitations required for each image [21]. Second, HASTE was a single-shot version of the TSE technique [22] wherein slightly more than half the total number of required phase encoding lines of the k-space were acquired after a single RF excitation. Third, turbo gradient spin echo or GRASE, is a turbo SE sequence with additional phase encoding gradient echoes between successive 180° refocusing pulses [23]. Fourth, bSSFP is a gradient-echo sequence that accomplishes fast acquisitions by eliminating idle times [24]. In the most common case, the data are mapped line-by-line from the top to bottom parts of k-space in a lexicographic manner, commonly referred to as Cartesian sampling. Fifth, EPI allows the collection of all the data required to reconstruct an image after a single RF excitation [25] and spiral imaging acquires data using two oscillating gradients that create a spiral trajectory in k-space [19]. The single-shot approach of EPI or spiral is the fastest with a subsecond acquisition time. Therefore, these techniques are usually used in functional MRI (fMRI). Furthermore, radial trajectories have also been introduced [26].

Several reconstruction methods were also developed to achieve higher signal-to-noise ratios (SNR) and reduced scan times. Parallel imaging techniques in conjunction with the use of phased array coils have been developed to reduce scan times by acquiring a reduced amount of k-space data with an array of receiver coils. These techniques include the simultaneous acquisition of spatial harmonics (SMASH) [27], sensitivity encoding (SENSE) [28], and generalized auto-calibrating partially parallel acquisitions (GRAPPA) [29]. SENSE is based on reconstruction in the imaging domain, but GRAPPA is based on k-space domain reconstruction. These parallel imaging techniques are being further developed nowadays [30]. Recently, multiband excitation imaging and finger printing imaging techniques have been developed to reduce the scan times even further.

The simultaneous multislice (SMS) pulse sequence [31] applies a multiband composite RF pulse with a slice-selective gradient to simultaneously excite multiple slice planes [32]. The multiband technique is simultaneous image refocused (SIR) EPI or simultaneous echo refocused (SER) EPI that does not use parallel imaging and coil sensitivity for image separation [33]. Another recently developed fast imaging technique is magnetic resonance fingerprinting (MRF) [34]. This technique allows simultaneous and efficient measurements of multiple tissue properties with one acquisition [35]. In MRF, the acquisition parameters, such as the RF excitation angle, phase, repetition time, and k-space sampling trajectory, are varied throughout the acquisition. When implemented properly, this acquisition could generate a unique signal time course for each tissue. For every MRF sequence, the dictionary of signal evolutions can be generated on a computer using mathematical algorithms to predict spin behavior and signal evolution during the acquisition.

2. Clinical applications

Ultrafast imaging is used to eliminate the effects of physiological motion, thus capturing dynamic events in real time or shortening the total scan time. Shortening the scan time helps improve patient comfort and compliance, thereby minimizing motion during scans [36]. At higher imaging speeds, it becomes feasible to examine a wide range of relevant physiological processes or to freeze induced motion that may otherwise lead to artifacts. Relevant physiological processes include, among others, respiration [37], cardiac rhythm [38], and hemodynamics from neuronal activation [39]. For example, ultrafast imaging is often used in cardiovascular imaging [40,41].

SENSE or GRAPPA are commonly used in the clinical practice nowadays. Multiband technique is used in fMRI. However, there is always an inherent trade-off between imaging speed and quality. MRF was initially evaluated in brain relaxometry [35], prostate [42], liver [43], cardiac [44], musculoskeletal imaging [43], arterial spin labeling (ASL) perfusion measurement [45], and microvascular properties [46].

A diffusion process is described by Fick’s law. Einstein described the relationship between the mean-squared displacement and diffusion coefficient in Brownian motion [47]. Diffusion is dependent on concentration, pressure, and temperature (https://en.wikipedia.org/wiki/Diffusion). Diffusion MRI is currently a well-established technique that is used in routine clinical practice to identify lesions and to characterize them. Diffusion MRI was developed by Le Bihan in 1989 [48]. Diffusion tensor imaging (DTI) and fiber tractography were developed by Peter Basser and Le Bihan in the early 1990s [49 -51]. The diffusion coefficient for water at 37°C is approximately D=3×10−9 m2/s.

1. Technical developments

Diffusion-weighted imaging (DWI) was developed to investigate microstructural properties by evaluating the proton diffusion process. The technique is used to characterize the microscopic behaviors of protons noninvasively. The diffusion coefficient or apparent diffusion coefficient (ADC) is measured on the basis of the use of diffusion gradients around the refocusing pulse in the SE sequence, or the use of bipolar gradients in the gradient-echo sequence within the echo time (TE) time period, and is fitted by the mono-exponential decay curve [50].

S=S0ebD

where b is the b-factor that can be calculated using the strength of the applied diffusion gradient, the duration of the diffusion gradient, and the duration of the period between two diffusion gradients.

A two-compartment model was applied to measure diffusion from intracellular and extracellular compartments, respectively referred to as slow (Dslow) and fast (Dfast) diffusion with the corresponding volume fractions Fslow and Ffast [52].

S=S0 [FfastebDfast+FslowebDslow]

The intravoxel incoherent motion (IVIM) method was introduced to separate diffusion from flow effects in a voxel [50]. In this technique, the flowing vascular volume fraction of incoherently flowing blood in the tissue (F), pseudo-diffusion coefficient (D*) associated to the IVIM effect, and true diffusion coefficient (D) can be measured by applying multiple b-values.

S=S0 [FflowebD*+(1Fflow)ebD]

Because in vivo proton movement is not isotropic, anisotropic diffusion was introduced by assuming ellipsoidal diffusion, known as DTI [51]. The diffusion coefficient in the DTI model is a 3×3 symmetric matrix. Eigenvalues and eigenvectors can be extracted from the tensor matrix to calculate the isotropic diffusion effects by using the mean diffusivity (MD=(Dxx+Dyy+Dzz)/3), and anisotropic diffusion effects by using fractional anisotropy (FA) index that represents the variance of the three eigenvalues divided by norm of the eigenvalues.

Tensor models can be applied by assuming Gaussian distributions of proton movements. However, in biological structures, water molecular diffusion is hindered in the extra-axonal space and restricted in the intra-axonal space [53]. To take this effect into account, other models were introduced, such as the ball-and-stick model [54], Q-ball imaging (QBI) [55], diffusion spectrum imaging (DSI) [56], spherical deconvolution model [57], and neurite orientation dispersion and density imaging (NODDI) [58]. Furthermore, non-Gaussian diffusion kurtosis imaging, was applied in clinical studies [59].

Finally, tractography techniques were introduced to detect white matter fiber bundles by measuring the eigenvectors along the fibers [60,61]. Improvements of fiber tracking were further achieved by using probabilistic models [62]. Tractography is currently used to investigate white matter connectivity in healthy brains and in several pathologic conditions [63].

2. Clinical applications

Diffusion MRI techniques, including DWI, DTI, and tractography, are currently extensively used in clinical settings. DWI is routinely applied in stroke and tumor patients. Diffusion indices of MD and FA are altered in the presence of ischemic injury [64] and in neoplasms [65]. In brain neoplasms, ADC values have been shown to be decreased in highly cellular tumors, such as central nervous system (CNS) lymphoma, medulloblastoma, and high-grade glioma [66]. Diffusion imaging techniques were applied in Alzheimer’s disease [67 -71]. Furthermore, malignant lesions have lower ADC values compared with surrounding normal tissue, edema, and benign tumors in the brain, head and neck malignancies, prostate, and liver cancer [72]. Diffusion-weighted whole-body imaging with background suppression (DWIBS) was performed with the short tau inversion recovery (STIR) EPI sequence with a high b value for background suppression to evaluate metastatic lesions in the body [73]. DTI was used to evaluate dislocation, disruption, infiltration, and edema. Tractography was used to evaluate corticospinal tract fibers, optic radiation fibers, and language fibers in patients to perform presurgical planning in neoplastic brain tumor cases.

Perfusion refers to the delivery of blood to a capillary bed in tissue. This is different compared with bulk flow motion (https://en.wikipedia.org/wiki/Perfusion). August Krogh first described the mechanism of regulation of capillaries in muscle, and was awarded the Nobel Prize in Physiology or Medicine [74]. Perfusion MRI is categorized according to the use of a contrast agent or not. Dynamic-susceptibility-contrast (DSC)-based perfusion MRI was developed by Villringer et al. in 1988 [75]. The ASL technique was developed by Koretsky in 2012 [76].

1. Technical developments

Details of perfusion MRI were summarized in a previous paper [77]. Three important techniques are currently used in clinical practice to obtain perfusion-related parameters. The first-pass DSC-enhanced MR perfusion is based on the susceptibility effects of gadolinium-based contrast agents on the signal echo. Therefore, T2- or T2*-weighted imaging sequence, usually a gradient-echo or an EPI sequence, is used to obtain the signal attenuation of the time-series images. Cerebral blood volume (CBV) and flow (CBF) values as well as time-related parameters, such as the mean transit time (MTT) and time-to-peak can be mapped in each pixel. Convolution theory is used to evaluate the measured concentrations of the hemodynamic changes of the contrast agent as follows [78],

C(t)=CBF0tCa(t')R(tt')dt'

where Ca(t) is the measured arterial input function (AIF) that describes the shape of the tracer that enters a voxel, and R(t-t’) is the unknown residue function that describes the probability of the tracer entering a voxel. Therefore,

MTT=0R(t)dtCBV=0CT (t) dt/0Ca (t) dt

, and CBF=CBV/MTT.

The dynamic contrast-enhanced (DCE) MR perfusion is based on the relaxivity effects of gadolinium-based contrast agents on the signal echo. Therefore, a transverse relaxation (T1)-weighted imaging sequence (usually a three-dimensional sequence), is used to obtain signal increments of time-series images. The area under-the-curve can be mapped. Furthermore, a pharmacokinetic model is used to map permeability-related parameters such as Ktrans and Kep and the corresponding volume fractions such as vp and ve. The general equation used to express the hemodynamic event after injecting the contrast agent is expressed with the extended Tofts model as follows [79],

Ctissue (t)=vp Cp (t)+Ktrans 0tCp (t')exp -Ktransve[t-t']dt'

where Cp(t) is the concentration of contrast agent in blood, Ktrans is the permeability–surface area constant from the vascular to the extracellular space, and ve is volume fraction in the extravascular and extracellular space.

The ASL MR perfusion is based on an endogenous contrast agent using magnetically labeled arterial blood water as a diffusible flow tracer. Therefore, the proton-density-weighted sequence is used to obtain signal changes with and without the use of magnetically labeled blood by either continuous or pulsed RF pulses. CBF value can be quantified. A general kinetic model for the evaluation of the difference of the longitudinal magnetization in the tissue owing to the labeled blood can be expressed to quantify the blood flow as follows [80],

ΔM(t)=2M0bCBF0t[Ca(t')R(t-t')m(t-t')]dt'

where M0b is the equilibrium magnetization in a blood filled arterial voxel, Ca(t) is the delivery function, such as AIF, R(t-t’) is the residue function that describes the washout of tagged spins from a voxel, and m(t-t’) includes the longitudinal magnetization relaxation effects. The pseudo-continuous ASL (pCASL) technique was introduced to improve the ASL signals [81], and is currently used in clinical practice.

2. Clinical applications

Perfusion MRI is a promising tool used to assess stroke, tumors, and neurodegenerative diseases. DSC perfusion MRI is the standard technique used to evaluate brain diseases, such as stroke [82] and tumors [83]. A combination of perfusion and DWI is used to evaluate a mismatch between the size of a perfusion defect and the diffusion abnormality that is referred to as the ischemic penumbra [84]. Tumor grade, recurrence, postoperative changes, or radiation effects can be established with DSC perfusion imaging [85].

DCE perfusion MRI is often applied in brain diseases [86] and in patients with breast, prostate, pelvic, and muscle diseases, and can be useful in differentiating between tumor recurrence and radiation necrosis. The DCE parameters reflect a more extensive BBB disruption and a higher tumor grade [87], evaluate the treatment prediction [88], and differentiate pseudo-progression from true progressive disease in GBM patients [89].

ASL perfusion MRI is mainly applied in brain diseases, such as neurodegenerative [90,91], renal, and cardiovascular diseases. ASL perfusion MRI has been used to evaluate pseudo-progression in brain tumor [92]. This method is particularly useful for patients with poor intravenous access, infants and children, and pregnant women [93].

MR spectroscopy (MRS) is used to determine the molecular structure of compounds, or to detect their presence. Proton MRS is based on the proton’s magnetic moment and its interaction with magnetic fields. MRS is thus sensitive to certain aspects of tissue metabolism. Proton MRS can detect N-acetyl aspartate (NAA), creatine/phosphocreatine (Cr/PCr), choline (Cho), glucose (Glu), myoinositol (ml or mlns), lactate (Lac), alanine (Ala), glutamate and glutamine (GIx), citrate, and ethanol in the human body. MRS can detect other nuclei in compounds of biological interest, such as phosphorus-31 (found in PCr) or carbon-13 (found in glycogen). However, proton MRS is more routinely performed in clinical practice compared with 31P or 13C MRS. Therefore, in this review, we only discuss proton MRS.

Chemical exchange saturation transfer (CEST) is a novel MR technique that enables molecular imaging to obtain certain compounds at concentrations that are too low to impact the contrast of standard MRI and too low to be directly detected in MRS at typical water imaging resolutions.

1. Technical developments of MRS

1) Pulse sequences

A single voxel spectroscopy (SVS) study is performed with short or long TE values. The SVS pulse sequences are the following: point resolved spectroscopy (PRESS) [94], stimulated echo acquisition mode (STEAM) [95], image selected in vivo spectroscopy (ISIS) [96], and depth-resolved spectroscopy (DRESS) [97]. Longer TE results in the signal decrease as a result of the transverse relaxation (T2) that leads to the alteration of the phase of multiplet signals because of J-coupling [98]. TE values in the range of 135–144 ms are typically used, as this leads to the production of a spectrum in which the doublet signal of Lac with a J-coupling constant of nearly 7 Hz is entirely reversed owing to the short TE and long TR [99].

A spectroscopic imaging is an area of interest. Chemical shift imaging (CSI) is used for multiple-voxel spectroscopic acquisitions. In spectroscopic imaging, phase encoding gradients can be applied in all three dimensions to sample k-space to select a volume that resembles methods used in MRI [100].

2) Water suppression methods

The water resonance must be suppressed to detect the millimolar concentrations of other molecules/moieties. The most common suppression method is based on the use of a chemical shift selected (CHESS) pulse sequence [101].

3) Spectral quantification

The typical postprocessing techniques used include Fourier Transform (FT), baseline correction, zero filling, and phasing. Quantification of the MRS signal amplitude can provide a means for estimating the tissue concentration of the signal generating molecules. While MRS signals are usually acquired in the time domain as free induction decays or echoes, they are usually viewed and analyzed in the frequency domain. The frequency domain representation is derived from the acquired time domain data by the FT. Signal averaging is used in virtually all MRS studies to increase the SNR by averaging the signals obtained in repeated measurements. To quantify the proton spectrum in most of the clinical studies, the internal reference signal is typically used that is either the Cr signal at 3.05 ppm or the water signal at 4.69 ppm. When Cr is used as a reference signal, it is more common to report results as signal amplitude ratios, such as NAA/Cr, or Cho/Cr. One weak aspect related to the use of the Cr signal as a reference is attributed to the fact that the Cr signal is not as uniform throughout the normal brain [102]. Another weak aspect pertains to the fact that the assumption of the Cr levels do not change with disease and other physiological characteristics may be erroneous. If the water signal is used as a reference signal, its amplitude must be measured by performing a separate measurement in the same brain region without using water suppression.

A linear least-squares optimization procedure has been established and used in spectral fitting techniques. The most popular spectral fitting software is currently the LCModel [103,104]. Furthermore, spectrum fitting software is available either in the time [105] or frequency domains [106,107].

2. Molecular imaging tools other than MRS

CEST can be used to apply molecular imaging [108]. This technique is more appropriate compared with CSI or MR spectroscopy imaging because it provides relatively high resolution. The principle of CEST is based on the use of the magnetization transfer effects from other molecules to water molecules. Therefore, the requirement of CEST is that the chemical species must have in their structures a 1H proton that is exchangeable with those of water. Known endogenous diamagnetic CEST agents are involved with exchangeable groups of amide proton (-NH), amine proton (-NH2), and hydroxyl proton (-OH), whose chemical shifts are ~3.5, ~1.8–3.0, and ~0.5–1.5 parts per million (ppm), respectively [109]. Amide CEST is usually referred to as amide proton transfer (APT). CEST techniques have been applied to map glutamate (amine proton), creatine (amine proton), glycosaminoglycan (Gag) (hydroxyl proton), myoinositol (MI) (hydroxyl proton), and glucose (hydroxyl proton). The detailed principle of CEST technique has been described in numerous previous papers [110 -116].

3. Clinical applications

The goal of clinical spectroscopy is to provide physicians with biochemical information that will assist in the differential diagnosis when standard clinical and radiologic tests fail or are too invasive. Proton spectroscopy has attained clinical value in that it can monitor the evolution of diseases and associated therapies. Disease can sometimes lead to large changes in metabolite levels.

MR spectroscopy has also been shown to be of diagnostic value for evaluating and monitoring the progression of certain brain diseases, such as stroke [117], epilepsy, multiple sclerosis [118], Alzheimer’s disease [119,120], brain tumors [121] and other tumors, such as prostate cancer [122] and breast cancer [123]. Lactate signal levels are elevated in ischemic brain tissue. Choline signal levels are elevated in some neoplastic or inflamed tissues.

The main applications for amide CEST or APT are the detection of cancer and ischemic stroke. In tumor regions, the concentration of proteins are elevated compared with surrounding tissues, and thus lead to increased APT levels [124]. This method was applied to classify tumor progression from radiation necrosis [125]. The CEST technique is applied in stroke because reduced pH in the ischemic region leads to lowered APT exchange rate, and results in decreased CEST values [126]. CEST was applied in other-than-brain pathologies, such as breast [127], prostate [128], and knee [129].

1. PET–MRI

PET–MRI is an imaging system that incorporates MRI and PET to gain from the benefits of soft tissue morphological imaging (MRI) and metabolic imaging (PET). This hybrid system is mainly used in the fields of oncology and neurology for clinical and preclinical studies. Some systems operate in totally separate rooms, but other systems do operate in the same room with separate machines. In these cases, a bed is shared to transfer the subjects from MRI to PET or to fully integrated systems. The first whole-body PET–MRI systems were produced by Philips (Amsterdam, The Netherlands) and were installed in the US (Mount Sinai Medical Center, New York, NY) and in Switzerland (Geneva University Hospital, Geneva) in 2010. The system featured a PET and MRI scanner separated by a revolving bed. The simultaneous PET–MR acquisition system (Siemens) was installed in 2010. The fully integrated whole-body systems were provided in 2011 by Siemens and in 2014 by GE. In Korea, the first PET–MRI system was installed at the Gachon University Hospital (Inchen, Korea) [7,8].

1) Technical developments

In this part, we only discussed the issues of the fully integrated PET–MRI system. Placing PET detectors in the MR magnet can alter the local magnetic field strength causing the protons to spin at wrong frequencies, thus leading to the formation of severe image distortions and artifacts, such as susceptibility artifacts. The presence of PET hardware within the gradient coil significantly alters the MR eddy current characteristics of the system possibly leading to degraded spatial linearity. The gradient imperfection would impact imaging spatial resolution and image homogeneity. The lutetium-based scintillation crystals have acceptable magnetic properties [130]. The avalanche photodiode for PET can be used in a 7 T MRI system without inducing major effects to magnetic fields [131].

Attenuation correction describes a method to account for the self-absorption of the emitted annihilation photons, and is a prerequisite for accurate quantification of the PET data [132]. It is not possible to directly derive the attenuation properties of tissues from MRI measurements. MRI-based attenuation correction methods have been introduced [133]: segmentation-based [134] that are usually used in T1-weighted images [135] or Dixon-sequence-based images [136], atlas-based [133], and reconstruction-based [137] methods. Attenuation correction of bone is calculated by using ultrashort echo time images [138].

The simultaneous acquisition of MRI and PET data in the fully integrated PET–MRI systems has major advantages compared with sequential acquisitions. Two-dimensional or three-dimensional (3D) navigator MRI is used to correct respiratory motion in PET images [139]. 3D Cine MRI has been used to correct cardiac motion in PET images [140]. A high resolution MRI has been used for the correction of the partial volume effect in PET images [141]. MR-guided PET reconstruction techniques have been developed by incorporating “a priori” anatomical information from the MRI [142]. Finally, MR anatomical images have been used for aligning functional information obtained from PET [143].

2) Clinical applications

While PET provides a) the high sensitivity required to detect minute amounts of radiotracers and b) the ability to quantify radiotracer activity throughout the body in absolute terms, MRI provides excellent soft-tissue contrast according to multiple contrast mechanisms at high-spatial resolution. Therefore, PET/MRI imaging combines the diagnostic breadth and information content of both PET and MRI. The PET–MRI system has been used for the study of patients with hepatobiliary cancer [144], neuroendocrine tumors [145], pancreatic adenocarcinoma [146], prostate cancer [147], primary brain tumors [148], dementia [149], epilepsy [150], musculoskeletal tumor [151], and coronary artery disease [152].

2. MR-guided focused ultrasound surgery

High-intensity focused ultrasound (HIFU) is a noninvasive therapeutic technique that uses nonionizing ultrasonic waves to heat tissue [153]. HIFU has been combined with MRI to enable guidance of the treatment and monitoring. It is referred to as MR-guided focused ultrasound surgery (MRgFUS). It is a 3D imaging technique that features high-soft-tissue contrast and provides information about temperature, thus allowing the monitoring of ablation. In 1992, reports were published that described MRgFUS on ex vivo muscle tissue [154], and the following year on in vivo tissue [155]. MRgFUS developed by Hynynen [155] was later transferred to InsighTec in Haifa, Israel in 1998. The InsighTec ExAblate 2000 was the first MRgFUS system used to obtain Food and Drug Administration market approval in the US in 2004.

1) Technical developments

In MRgFUS, MR is used for both target localization and in vivo real time monitoring of temperature based on a technique referred to as MR thermometry [155], and for verifying tissue destruction using a postprocessing tool. The thermometric technique of temperature-dependent phase changes in gradient-recalled echo pulse sequences are commonly used to determine the temperature change [156]. The change in temperature is represented as

ΔT=ΔφγcB0TE

where Δϕ is the phase change, γ is the gyromagnetic ratio, c is the proton-resonance frequency shift constant (–0.01 ppm/°C) [157], Bo is the main magnetic field strength, and TE is the echo time. For the bone tumors, the calcification issues can be resolved be using susceptibility-weighted MRI to identify calcifications rather than computer tomography. An ultrashort TE sequence can be used to improve thermometry in bone.

2) Clinical applications

Clinical applications of MRgFUS are still limited. MRgFUS has been used in clinical studies in patients with uterine fibroids [158], bone metastases [159], prostate cancer [160], breast cancer [161], and brain diseases, such as brain tumors [162], intractable essential tremor [163], Parkinson’s disease [164], obsessive-compulsive disorder [165], major depressive disorders [166], and neuropathic pain [167].

3. MRI-guided linear accelerator

MRI-guided linear accelerator (MRI–LINAC) is a recently developed and advanced radiation treatment system. As indicating the name, the MRI–LINAC is fully integrated with the MRI for imaging soft-tissue tumors together with LINAC for the radiotherapy to treat cancers throughout the body. The advantage of MRI-based imaging on a linear accelerator has superior high-definition image quality, especially for some soft-tissue cancers compared with CT-based imaging in the traditional linear accelerators to visualize the target area and adjacent anatomy for treatment setup and delivery. The first technical prototype MR–Linac was developed and installed in the University Medical Center Utrecht in Utrecht, The Netherlands. Similar types of measurements have been performed on a hybrid MRI Cobalt-60 device [168]. The first clinically active MRI-guided radiation therapy machine (ViewRay) was installed at the Alvin J. Siteman Cancer Center at Barnes-Jewish Hospital at the Washington University School of Medicine (St. Louis, MO, USA). The treatment of the first patients was announced in February 2014.

1) Technical developments

LINAC is affected by MRI. Two main configurations of MRI–LINAC that are being pursued with the radiotherapy beam are either parallel or perpendicular to the main magnetic field. This configuration is affected by the interference between the delivery of the radiation beam of LINAC and MRI. The operation of the multileaf collimator in the strong magnetic field can be a problem in the shaping of the X-ray beam [169]. Both configurations have this problem, and vendors lowered magnetic field to minimize this issue. Another issue in the MRI–LINAC is that the accelerated electrons used to produce the X-ray beam can be deviated or defocused, thus causing a loss of the beam current. Previous studies showed that the perpendicular configuration is dominant to the total beam loss compared with the parallel one [170,171]. Skin dose can be increased by secondary electrons owing to the influence of the magnetic field [172]. In this case, the perpendicular configuration should be advantageous, although the electron return effect can still appear [173]. Receiver coils can cause attenuation of the primary beam and can increase the skin dose. Detailed explanation can be found in another paper [174].

MRI quality is also affected by the LINAC. Any RF noise generated by the LINAC can cause artifacts or noise in images. In addition, LINAC components, such as the accelerator or multileaf collimators cause inhomogeneity of the main magnetic field, thus worsening the imaging quality [175]. Finally, the radiation beam can affect conductors or electronics in the coil, causing imaging artifacts [176,177].

2) Clinical applications

The MRI-LINAC can adapt the radiation treatment plan based on movement of the organs or tumor, and also track the motion of the tumor. This system reduces complications after radiation treatments. The MRI–LINAC can be used to improve the personalization of the radiation therapy using existing contrast imaging mechanisms, such as diffusion, perfusion, functional, and metabolic, to evaluate treatment effects. The hybrid system has been focused on daily plan changes based on geometric changes in the organs-at-risk [178,179]. Furthermore, MRI has been used to evaluate radiation treatment effects [180]. Currently, this hybrid system is used to treat patients with prostate cancer [181], pelvic lymph nodes [182], and the esophagus [183].

The research was supported by the National Research Foundation of Korea grant funded by Ministry of Science and ICT (No. 2020R1A2C1004749, GHJ), Republic of Korea.

Conceptualization: Geon-Ho Jahng. Data curation: Geon-Ho Jahng, Soonchan Park, Chang-Woo Ryu, and Zang-Hee Cho. Writing–original drafting: Geon-Ho Jahng. Writing–review & editing: Geon-Ho Jahng, Soonchan Park, Chang-Woo Ryu, and Zang-Hee Cho.

  1. Jackson JA, Langham WH. Whole-body NMR spectrometer. Rev Sci Instrum. 1968;39:510-513.
    Pubmed CrossRef
  2. Damadian R. Tumor detection by nuclear magnetic resonance. Science. 1971;171:1151-1153.
    Pubmed CrossRef
  3. Lauterbur PC. Image formation by induced local interactions. Examples employing nuclear magnetic resonance. 1973. Clin Orthop Relat Res. 1989;244:3-6.
    Pubmed
  4. Lauterbur PC. Magnetic resonance zeugmatography. Pure Appl Chem. 1974;40:149-157.
    CrossRef
  5. Mansfield P, Grannell P. "Diffraction" and microscopy in solids and liquids by NMR. Phys Rev B. 1975;12:3618.
    CrossRef
  6. Damadian R, Minkoff L, Goldsmith M, Stanford M, Koutcher J. Field focusing nuclear magnetic resonance (FONAR): visualization of a tumor in a live animal. Science. 1976;194:1430-1432.
    Pubmed CrossRef
  7. Cho ZH, Kim YB, Han JY, Kim NB, Hwang SI, Kim SJ, et al. Altered T2* relaxation time of the hippocampus in major depressive disorder: implications of ultra-high field magnetic resonance imaging. J Psychiatr Res. 2010;44:881-886.
    Pubmed CrossRef
  8. Cho ZH, Lee YB, Kang CK, Yang JW, Jung IH, Park CA, et al. Microvascular imaging of asymptomatic MCA steno-occlusive patients using ultra-high-field 7T MRI. J Neurol. 2013;260:144-150.
    Pubmed CrossRef
  9. Cho ZH, Son YD, Kim HK, Kim KN, Oh SH, Han JY, et al. A fusion PET-MRI system with a high-resolution research tomograph-PET and ultra-high field 7.0 T-MRI for the molecular-genetic imaging of the brain. Proteomics. 2008;8:1302-1323.
    Pubmed CrossRef
  10. Kim M, Kim KE, Jeong SW, Hwang SW, Jo H, Lee J, et al. Effects of the ultra-high-frequency electrical field radiofrequency device on mouse skin: a histologic and molecular study. Plast Reconstr Surg. 2016;138:248e-255e.
    Pubmed CrossRef
  11. Oh SH, Chung JY, In MH, Zaitsev M, Kim YB, Speck O, et al. Distortion correction in EPI at ultra-high-field MRI using PSF mapping with optimal combination of shift detection dimension. Magn Reson Med. 2012;68:1239-1246.
    Pubmed CrossRef
  12. Richards K, Calamante F, Tournier JD, Kurniawan ND, Sadeghian F, Retchford AR, et al. Mapping somatosensory connectivity in adult mice using diffusion MRI tractography and super-resolution track density imaging. Neuroimage. 2014;102 Pt 2:381-392.
    Pubmed CrossRef
  13. Hahn EL. Spin echoes. Phys Rev. 1950;80:580.
    CrossRef
  14. Hahn EL. Nuclear induction due to free larmor precession. Phys Rev. 1950;77:297.
    CrossRef
  15. Hajnal JV, De Coene B, Lewis PD, Baudouin CJ, Cowan FM, Pennock JM, et al. High signal regions in normal white matter shown by heavily T2-weighted CSF nulled IR sequences. J Comput Assist Tomogr. 1992;16:506-513.
    Pubmed CrossRef
  16. Ogawa S, Lee TM, Kay AR, Tank DW. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci U S A. 1990;87:9868-9872.
    Pubmed CrossRef
  17. Reichenbach JR, Venkatesan R, Schillinger DJ, Kido DK, Haacke EM. Small vessels in the human brain: MR venography with deoxyhemoglobin as an intrinsic contrast agent. Radiology. 1997;204:272-277.
    Pubmed CrossRef
  18. Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast imaging method for clinical MR. Magn Reson Med. 1986;3:823-833.
    Pubmed CrossRef
  19. Ahn CB, Kim JH, Cho ZH. High-speed spiral-scan echo planar NMR imaging-I. IEEE Trans Med Imaging. 1986;5:2-7.
    Pubmed CrossRef
  20. Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn Reson Med. 1992;28:202-213.
    Pubmed CrossRef
  21. Hennig J, Friedburg H. Clinical applications and methodological developments of the RARE technique. Magn Reson Imaging. 1988;6:391-395.
    Pubmed CrossRef
  22. Kiefer B, Grassner J, I-lausman K. Image acquisition in a second with half fourier acquisition single shot turbo spin echo. J Magn Reson Imaging. 1994;4(P):86.
  23. Oshio K, Feinberg DA. GRASE (Gradient- and spin-echo) imaging: a novel fast MRI technique. Magn Reson Med. 1991;20:344-349.
    Pubmed CrossRef
  24. Nayak KS, Hargreaves BA, Hu BS, Nishimura DG, Pauly JM, Meyer CH. Spiral balanced steady-state free precession cardiac imaging. Magn Reson Med. 2005;53:1468-1473.
    Pubmed CrossRef
  25. Mansfield P. Multi-planar image formation using NMR spin echoes. J Phys C. 1977;10:L55.
    CrossRef
  26. Du J, Lu A, Block WF, Thornton FJ, Grist TM, Mistretta CA. Time-resolved undersampled projection reconstruction magnetic resonance imaging of the peripheral vessels using multi-echo acquisition. Magn Reson Med. 2005;53:730-734.
    Pubmed CrossRef
  27. Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38:591-603.
    Pubmed CrossRef
  28. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952-962.
    CrossRef
  29. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202-1210.
    Pubmed CrossRef
  30. Hamilton J, Franson D, Seiberlich N. Recent advances in parallel imaging for MRI. Prog Nucl Magn Reson Spectrosc. 2017;101:71-95.
    Pubmed CrossRef
  31. Feinberg DA, Setsompop K. Ultra-fast MRI of the human brain with simultaneous multi-slice imaging. J Magn Reson. 2013;229:90-100.
    Pubmed KoreaMed CrossRef
  32. Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging. 2001;13:313-317.
    Pubmed CrossRef
  33. Feinberg DA, Reese TG, Wedeen VJ. Simultaneous echo refocusing in EPI. Magn Reson Med. 2002;48:1-5.
    Pubmed CrossRef
  34. Panda A, Mehta BB, Coppo S, Jiang Y, Ma D, Seiberlich N, et al. Magnetic resonance fingerprinting-an overview. Curr Opin Biomed Eng. 2017;3:56-66.
    Pubmed KoreaMed CrossRef
  35. Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187-192.
    Pubmed KoreaMed CrossRef
  36. Tsao J. Ultrafast imaging: principles, pitfalls, solutions, and applications. J Magn Reson Imaging. 2010;32:252-266.
    Pubmed CrossRef
  37. Saini S, Stark DD, Rzedzian RR, Pykett IL, Rummeny E, Hahn PF, et al. Forty-millisecond MR imaging of the abdomen at 2.0 T. Radiology. 1989;173:111-116.
    Pubmed CrossRef
  38. Doyle M, Rzedzian R, Mansfield P, Coupland RE. Dynamic NMR cardiac imaging in a piglet. Br J Radiol. 1983;56:925-930.
    Pubmed CrossRef
  39. Glover GH, Lee AT. Motion artifacts in fMRI: comparison of 2DFT with PR and spiral scan methods. Magn Reson Med. 1995;33:624-635.
    Pubmed CrossRef
  40. Muthurangu V, Lurz P, Critchely JD, Deanfield JE, Taylor AM, Hansen MS. Real-time assessment of right and left ventricular volumes and function in patients with congenital heart disease by using high spatiotemporal resolution radial k-t SENSE. Radiology. 2008;248:782-791.
    Pubmed CrossRef
  41. Nayak KS, Cunningham CH, Santos JM, Pauly JM. Real-time cardiac MRI at 3 tesla. Magn Reson Med. 2004;51:655-660.
    Pubmed CrossRef
  42. Yu AC, Badve C, Ponsky LE, Pahwa S, Dastmalchian S, Rogers M, et al. Development of a combined MR fingerprinting and diffusion examination for prostate cancer. Radiology. 2017;283:729-738.
    Pubmed KoreaMed CrossRef
  43. Cloos MA, Knoll F, Zhao T, Block KT, Bruno M, Wiggins GC, et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016;7:12445.
    Pubmed KoreaMed CrossRef
  44. Hamilton JI, Jiang Y, Chen Y, Ma D, Lo WC, Griswold M, et al. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med. 2017;77:1446-1458.
    Pubmed CrossRef
  45. Su P, Mao D, Liu P, Li Y, Pinho MC, Welch BG, et al. Multiparametric estimation of brain hemodynamics with MR fingerprinting ASL. Magn Reson Med. 2017;78:1812-1823.
    Pubmed KoreaMed CrossRef
  46. Christen T, Pannetier NA, Ni WW, Qiu D, Moseley ME, Schuff N, et al. MR vascular fingerprinting: a new approach to compute cerebral blood volume, mean vessel radius, and oxygenation maps in the human brain. Neuroimage. 2014;89:262-270.
    Pubmed KoreaMed CrossRef
  47. Einstein A. Investigations on the theory of the Brownian movement. Chelmsford: Courier Corporation; 1956.
  48. Le Bihan D, Breton E, inventor; General Electric CGR SA, assignee. Method to measure the molecular diffusion and/or perfusion parameters of live tissue. United States patent US4809701A. 1989 Mar 7.
  49. Howe FA, Filler AG, Bell BA, Griffiths JR. Magnetic resonance neurography. Magn Reson Med. 1992;28:328-338.
    Pubmed CrossRef
  50. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology. 1986;161:401-407.
    Pubmed CrossRef
  51. Basser PJ, Mattiello J, LeBihan D. Estimation of the effective self-diffusion tensor from the NMR spin echo. J Magn Reson B. 1994;103:247-254.
    Pubmed CrossRef
  52. Niendorf T, Dijkhuizen RM, Norris DG, van Lookeren Campagne M, Nicolay K. Biexponential diffusion attenuation in various states of brain tissue: implications for diffusion-weighted imaging. Magn Reson Med. 1996;36:847-857.
    Pubmed CrossRef
  53. Assaf Y, Freidlin RZ, Rohde GK, Basser PJ. New modeling and experimental framework to characterize hindered and restricted water diffusion in brain white matter. Magn Reson Med. 2004;52:965-978.
    Pubmed CrossRef
  54. Behrens TE, Woolrich MW, Jenkinson M, Johansen-Berg H, Nunes RG, Clare S, et al. Characterization and propagation of uncertainty in diffusion-weighted MR imaging. Magn Reson Med. 2003;50:1077-1088.
    Pubmed CrossRef
  55. Tuch DS. Q-ball imaging. Magn Reson Med. 2004;52:1358-1372.
    Pubmed KoreaMed CrossRef
  56. Wedeen VJ, Wang RP, Schmahmann JD, Benner T, Tseng WY, Dai G, et al. Diffusion spectrum magnetic resonance imaging (DSI) tractography of crossing fibers. Neuroimage. 2008;41:1267-1277.
    Pubmed CrossRef
  57. Tournier JD, Calamante F, Gadian DG, Connelly A. Direct estimation of the fiber orientation density function from diffusion-weighted MRI data using spherical deconvolution. Neuroimage. 2004;23:1176-1185.
    Pubmed CrossRef
  58. Zhang H, Schneider T, Wheeler-Kingshott CA, Alexander DC. NODDI: practical in vivo neurite orientation dispersion and density imaging of the human brain. Neuroimage. 2012;61:1000-1016.
    Pubmed CrossRef
  59. Jensen JH, Helpern JA, Ramani A, Lu H, Kaczynski K. Diffusional kurtosis imaging: the quantification of non-gaussian water diffusion by means of magnetic resonance imaging. Magn Reson Med. 2005;53:1432-1440.
    Pubmed CrossRef
  60. Basser PJ, Pajevic S, Pierpaoli C, Duda J, Aldroubi A. In vivo fiber tractography using DT-MRI data. Magn Reson Med. 2000;44:625-632.
    CrossRef
  61. Mori S, Crain BJ, Chacko VP, van Zijl PC. Three-dimensional tracking of axonal projections in the brain by magnetic resonance imaging. Ann Neurol. 1999;45:265-269.
    Pubmed CrossRef
  62. Jones DK. Tractography gone wild: probabilistic fibre tracking using the wild bootstrap with diffusion tensor MRI. IEEE Trans Med Imaging. 2008;27:1268-1274.
    Pubmed CrossRef
  63. Bullmore E, Sporns O. Complex brain networks: graph theoretical analysis of structural and functional systems. Nat Rev Neurosci. 2009;10:186-198.
    Pubmed CrossRef
  64. Sotak CH. The role of diffusion tensor imaging in the evaluation of ischemic brain injury - a review. NMR Biomed. 2002;15:561-569.
    Pubmed CrossRef
  65. Alexander AL, Hurley SA, Samsonov AA, Adluru N, Hosseinbor AP, Mossahebi P, et al. Characterization of cerebral white matter properties using quantitative magnetic resonance imaging stains. Brain Connect. 2011;1:423-446.
    Pubmed KoreaMed CrossRef
  66. Baliyan V, Das CJ, Sharma R, Gupta AK. Diffusion weighted imaging: technique and applications. World J Radiol. 2016;8:785-798.
    Pubmed CrossRef
  67. Zhang Y, Schuff N, Jahng GH, Bayne W, Mori S, Schad L, et al. Diffusion tensor imaging of cingulum fibers in mild cognitive impairment and Alzheimer disease. Neurology. 2007;68:13-19.
    Pubmed KoreaMed CrossRef
  68. Zhang Y, Du AT, Hayasaka S, Jahng GH, Hlavin J, Zhan W, et al. Patterns of age-related water diffusion changes in human brain by concordance and discordance analysis. Neurobiol Aging. 2010;31:1991-2001.
    Pubmed KoreaMed CrossRef
  69. Jahng GH, Xu S, Weiner MW, Meyerhoff DJ, Park S, Schuff N. DTI studies in patients with Alzheimer's disease, mild cognitive impairment, or normal cognition with evaluation of the intrinsic background gradients. Neuroradiology. 2011;53:749-762.
    Pubmed CrossRef
  70. Jahng GH, Xu S. Local susceptibility causes diffusion alterations in patients with Alzheimer's disease and mild cognitive impairment. Brain Imaging Behav. 2012;6:426-436.
    Pubmed CrossRef
  71. Jahng GH, Xu S, Kim MJ. Mapping of distributions of a local b-matrix cross-term strength using diffusion tensor MRI in patients with Alzheimer's disease. Med Phys. 2012;39:6324-6331.
    Pubmed CrossRef
  72. Charles-Edwards EM, deSouza NM. Diffusion-weighted magnetic resonance imaging and its application to cancer. Cancer Imaging. 2006;6:135-143.
    Pubmed CrossRef
  73. Takahara T, Imai Y, Yamashita T, Yasuda S, Nasu S, Van Cauteren M. Diffusion weighted whole body imaging with background body signal suppression (DWIBS): technical improvement using free breathing, STIR and high resolution 3D display. Radiat Med. 2004;22:275-282.
    Pubmed
  74. Larsen EH. [August Krogh (1874-1949): 1920 nobel prize]. Ugeskr Laeger. 2007;169:2878. Danish.
  75. Villringer A, Rosen BR, Belliveau JW, Ackerman JL, Lauffer RB, Buxton RB, et al. Dynamic imaging with lanthanide chelates in normal brain: contrast due to magnetic susceptibility effects. Magn Reson Med. 1988;6:164-174.
    Pubmed CrossRef
  76. Koretsky AP. Early development of arterial spin labeling to measure regional brain blood flow by MRI. Neuroimage. 2012;62:602-607.
    Pubmed KoreaMed CrossRef
  77. Jahng GH, Li KL, Ostergaard L, Calamante F. Perfusion magnetic resonance imaging: a comprehensive update on principles and techniques. Korean J Radiol. 2014;15:554-577.
    Pubmed KoreaMed CrossRef
  78. Ostergaard L, Weisskoff RM, Chesler DA, Gyldensted C, Rosen BR. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysis. Magn Reson Med. 1996;36:715-725.
    Pubmed CrossRef
  79. Tofts PS. Modeling tracer kinetics in dynamic Gd-DTPA MR imaging. J Magn Reson Imaging. 1997;7:91-101.
    Pubmed CrossRef
  80. Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR. A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med. 1998;40:383-396.
    Pubmed CrossRef
  81. Dai W, Garcia D, de Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields. Magn Reson Med. 2008;60:1488-1497.
    Pubmed KoreaMed CrossRef
  82. Keir SL, Wardlaw JM. Systematic review of diffusion and perfusion imaging in acute ischemic stroke. Stroke. 2000;31:2723-2731.
    Pubmed CrossRef
  83. Cha S, Knopp EA, Johnson G, Wetzel SG, Litt AW, Zagzag D. Intracranial mass lesions: dynamic contrast-enhanced susceptibility-weighted echo-planar perfusion MR imaging. Radiology. 2002;223:11-29.
    Pubmed CrossRef
  84. Ostergaard L, Sorensen AG, Chesler DA, Weisskoff RM, Koroshetz WJ, Wu O, et al. Combined diffusion-weighted and perfusion-weighted flow heterogeneity magnetic resonance imaging in acute stroke. Stroke. 2000;31:1097-1103.
    Pubmed CrossRef
  85. Lee SJ, Kim JH, Kim YM, Lee GK, Lee EJ, Park IS, et al. Perfusion MR imaging in gliomas: comparison with histologic tumor grade. Korean J Radiol. 2001;2:1-7.
    Pubmed KoreaMed CrossRef
  86. Brix G, Kiessling F, Lucht R, Darai S, Wasser K, Delorme S, et al. Microcirculation and microvasculature in breast tumors: pharmacokinetic analysis of dynamic MR image series. Magn Reson Med. 2004;52:420-429.
    Pubmed CrossRef
  87. Choi HS, Kim AH, Ahn SS, Shin NY, Kim J, Lee SK. Glioma grading capability: comparisons among parameters from dynamic contrast-enhanced MRI and ADC value on DWI. Korean J Radiol. 2013;14:487-492.
    Pubmed KoreaMed CrossRef
  88. Ahn SS, Shin NY, Chang JH, Kim SH, Kim EH, Kim DW, et al. Prediction of methylguanine methyltransferase promoter methylation in glioblastoma using dynamic contrast-enhanced magnetic resonance and diffusion tensor imaging. J Neurosurg. 2014;121:367-373.
    Pubmed CrossRef
  89. Suh CH, Kim HS, Choi YJ, Kim N, Kim SJ. Prediction of pseudoprogression in patients with glioblastomas using the initial and final area under the curves ratio derived from dynamic contrast-enhanced T1-weighted perfusion MR imaging. AJNR Am J Neuroradiol. 2013;34:2278-2286.
    Pubmed CrossRef
  90. Kim SM, Kim MJ, Rhee HY, Ryu CW, Kim EJ, Petersen ET, et al. Regional cerebral perfusion in patients with Alzheimer's disease and mild cognitive impairment: effect of APOE epsilon4 allele. Neuroradiology. 2013;55:25-34.
    Pubmed CrossRef
  91. Johnson NA, Jahng GH, Weiner MW, Miller BL, Chui HC, Jagust WJ, et al. Pattern of cerebral hypoperfusion in Alzheimer disease and mild cognitive impairment measured with arterial spin-labeling MR imaging: initial experience. Radiology. 2005;234:851-859.
    Pubmed KoreaMed CrossRef
  92. Choi YJ, Kim HS, Jahng GH, Kim SJ, Suh DC. Pseudoprogression in patients with glioblastoma: added value of arterial spin labeling to dynamic susceptibility contrast perfusion MR imaging. Acta Radiol. 2013;54:448-454.
    Pubmed CrossRef
  93. Wang J, Licht DJ, Jahng GH, Liu CS, Rubin JT, Haselgrove J, et al. Pediatric perfusion imaging using pulsed arterial spin labeling. J Magn Reson Imaging. 2003;18:404-413.
    Pubmed CrossRef
  94. Bottomley PA. Spatial localization in NMR spectroscopy in vivo. Ann N Y Acad Sci. 1987;508:333-348.
    Pubmed CrossRef
  95. Frahm J, Bruhn H, Gyngell ML, Merboldt KD, Hänicke W, Sauter R. Localized high-resolution proton NMR spectroscopy using stimulated echoes: initial applications to human brain in vivo. Magn Reson Med. 1989;9:79-93.
    Pubmed CrossRef
  96. Ordidge RJ, Connelly A, Lohman JA. Image-selected in vivo spectroscopy (ISIS). A new technique for spatially selective NMR spectroscopy. J Magn Reson (1969). 1986;66:283-294.
    CrossRef
  97. Bottomley PA, Foster TB, Darrow RD. Depth-resolved surface-coil spectroscopy (DRESS) for in vivo 1H, 31P, and 13C NMR. J Magn Reson (1969). 1984;59:338-342.
    CrossRef
  98. Thrippleton MJ, Edden RA, Keeler J. Suppression of strong coupling artefacts in J-spectra. J Magn Reson. 2005;174:97-109.
    Pubmed CrossRef
  99. Gruber S, Pinker K, Riederer F, Chmelík M, Stadlbauer A, Bittsanský M, et al. Metabolic changes in the normal ageing brain: consistent findings from short and long echo time proton spectroscopy. Eur J Radiol. 2008;68:320-327.
    Pubmed CrossRef
  100. Skoch A, Jiru F, Bunke J. Spectroscopic imaging: basic principles. Eur J Radiol. 2008;67:230-239.
    Pubmed CrossRef
  101. Haase A, Frahm J, Hänicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30:341-344.
    Pubmed CrossRef
  102. Maudsley AA, Domenig C, Govind V, Darkazanli A, Studholme C, Arheart K, et al. Mapping of brain metabolite distributions by volumetric proton MR spectroscopic imaging (MRSI). Magn Reson Med. 2009;61:548-559.
    Pubmed KoreaMed CrossRef
  103. Provencher SW. Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med. 1993;30:672-679.
    Pubmed CrossRef
  104. Provencher SW. Automatic quantitation of localized in vivo 1H spectra with LCModel. NMR Biomed. 2001;14:260-264.
    Pubmed CrossRef
  105. Reynolds G, Wilson M, Peet A, Arvanitis TN. An algorithm for the automated quantitation of metabolites in in vitro NMR signals. Magn Reson Med. 2006;56:1211-1219.
    Pubmed CrossRef
  106. Naressi A, Couturier C, Castang I, de Beer R, Graveron-Demilly D. Java-based graphical user interface for MRUI, a software package for quantitation of in vivo/medical magnetic resonance spectroscopy signals. Comput Biol Med. 2001;31:269-286.
    Pubmed CrossRef
  107. Naressi A, Couturier C, Devos JM, Janssen M, Mangeat C, de Beer R, et al. Java-based graphical user interface for the MRUI quantitation package. MAGMA. 2001;12:141-152.
    Pubmed CrossRef
  108. van Zijl PC, Yadav NN. Chemical exchange saturation transfer (CEST): what is in a name and what isn't? Magn Reson Med. 2011;65:927-948.
    Pubmed CrossRef
  109. Vinogradov E, Sherry AD, Lenkinski RE. CEST: from basic principles to applications, challenges and opportunities. J Magn Reson. 2013;229:155-172.
    Pubmed KoreaMed CrossRef
  110. Park JE, Jahng GH, Jeong HK. Amide proton transfer imaging in clinics: basic concepts and current and future use in brain tumors and stroke. J Korean Soc Radiol. 2016;75:419-433.
    CrossRef
  111. Jahng GH, Oh JH. Physical modeling of chemical exchange saturation transfer imaging. Prog Med Phys. 2017;28:135-143.
    CrossRef
  112. Oh JH, Kim HG, Woo DC, Jeong HK, Lee SY, Jahng GH. Chemical-exchange-saturation-transfer magnetic resonance imaging to map gamma-aminobutyric acid, glutamate, myoinositol, glycine, and asparagine: phantom experiments. J Korean Phys Soc. 2017;70:545-553.
    CrossRef
  113. Jahng GH, Choi W, Chung JJ, Kim ST, Rhee HY. Mapping exchangeable protons to monitor protein alterations in the brain of an Alzheimer's disease mouse model by using MRI. Curr Alzheimer Res. 2018;15:1343-1353.
    Pubmed CrossRef
  114. Yoo CH, Oh J, Park S, Ryu CW, Kwon YK, Jahng GH. Comparative evaluation of the polynomial and spline fitting methods for the B0 correction of CEST MRI data acquired from human brains. Int J Imaging Syst Technol. 2019;29:272-282.
    CrossRef
  115. Oh JH, Kim HG, Woo DC, Rhee SJ, Lee SY, Jahng GH. Preliminary phantom experiments to map amino acids and neurotransmitters using MRI. Prog Med Phys. 2018;29:29-41.
    CrossRef
  116. Park S, Jang J, Oh JH, Ryu CW, Jahng GH. Assessment of the cerebrospinal fluid effect on the chemical exchange saturation transfer map obtained from the full Z-spectrum in the elderly human brain. Prog Med Phys. 2019;30:139-149.
    CrossRef
  117. Bruhn H, Frahm J, Gyngell ML, Merboldt KD, Hänicke W, Sauter R. Cerebral metabolism in man after acute stroke: new observations using localized proton NMR spectroscopy. Magn Reson Med. 1989;9:126-131.
    Pubmed CrossRef
  118. Arnold DL, Matthews PM, Francis G, Antel J. Proton magnetic resonance spectroscopy of human brain in vivo in the evaluation of multiple sclerosis: assessment of the load of disease. Magn Reson Med. 1990;14:154-159.
    Pubmed CrossRef
  119. Miller BL, Moats RA, Shonk T, Ernst T, Woolley S, Ross BD. Alzheimer disease: depiction of increased cerebral myo-inositol with proton MR spectroscopy. Radiology. 1993;187:433-437.
    Pubmed CrossRef
  120. Jahng GH, Oh J, Lee DW, Kim HG, Rhee HY, Shin W, et al. Glutamine and glutamate complex, as measured by functional magnetic resonance spectroscopy, alters during face-name association task in patients with mild cognitive impairment and Alzheimer's disease. J Alzheimers Dis. 2016;52:145-159.
    Pubmed CrossRef
  121. Bruhn H, Frahm J, Gyngell ML, Merboldt KD, Hänicke W, Sauter R, et al. Noninvasive differentiation of tumors with use of localized H-1 MR spectroscopy in vivo: initial experience in patients with cerebral tumors. Radiology. 1989;172:541-548.
    Pubmed CrossRef
  122. Sciarra A, Panebianco V, Ciccariello M, Salciccia S, Cattarino S, Lisi D, et al. Value of magnetic resonance spectroscopy imaging and dynamic contrast-enhanced imaging for detecting prostate cancer foci in men with prior negative biopsy. Clin Cancer Res. 2010;16:1875-1883.
    Pubmed CrossRef
  123. Haddadin IS, McIntosh A, Meisamy S, Corum C, Styczynski Snyder AL, Powell NJ, et al. Metabolite quantification and high-field MRS in breast cancer. NMR Biomed. 2009;22:65-76.
    Pubmed KoreaMed CrossRef
  124. Zhao X, Wen Z, Zhang G, Huang F, Lu S, Wang X, et al. Three-dimensional turbo-spin-echo amide proton transfer MR imaging at 3-Tesla and its application to high-grade human brain tumors. Mol Imaging Biol. 2013;15:114-122.
    Pubmed KoreaMed CrossRef
  125. Zhou J, Tryggestad E, Wen Z, Lal B, Zhou T, Grossman R, et al. Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides. Nat Med. 2011;17:130-134.
    Pubmed KoreaMed CrossRef
  126. Wang M, Hong X, Chang CF, Li Q, Ma B, Zhang H, et al. Simultaneous detection and separation of hyperacute intracerebral hemorrhage and cerebral ischemia using amide proton transfer MRI. Magn Reson Med. 2015;74:42-50.
    Pubmed KoreaMed CrossRef
  127. Dula AN, Arlinghaus LR, Dortch RD, Dewey BE, Whisenant JG, Ayers GD, et al. Amide proton transfer imaging of the breast at 3 T: establishing reproducibility and possible feasibility assessing chemotherapy response. Magn Reson Med. 2013;70:216-224.
    Pubmed CrossRef
  128. Jia G, Abaza R, Williams JD, Zynger DL, Zhou J, Shah ZK, et al. Amide proton transfer MR imaging of prostate cancer: a preliminary study. J Magn Reson Imaging. 2011;33:647-654.
    Pubmed KoreaMed CrossRef
  129. Ling W, Regatte RR, Navon G, Jerschow A. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc Natl Acad Sci U S A. 2008;105:2266-2270.
    Pubmed KoreaMed CrossRef
  130. Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci. 2003;50:1683-1685.
    CrossRef
  131. Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med. 2006;47:639-647.
    Pubmed
  132. Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys. 1998;25:2046-2053.
    Pubmed CrossRef
  133. Wagenknecht G, Kaiser HJ, Mottaghy FM, Herzog H. MRI for attenuation correction in PET: methods and challenges. MAGMA. 2013;26:99-113.
    Pubmed CrossRef
  134. Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Börnert P, et al. Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging. 2011;38:138-152.
    Pubmed CrossRef
  135. Beyer T, Lassen ML, Boellaard R, Delso G, Yaqub M, Sattler B, et al. Investigating the state-of-the-art in whole-body MR-based attenuation correction: an intra-individual, inter-system, inventory study on three clinical PET/MR systems. MAGMA. 2016;29:75-87.
    Pubmed CrossRef
  136. Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd'hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50:520-526.
    Pubmed CrossRef
  137. Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging. 1999;18:393-403.
    Pubmed CrossRef
  138. Johansson A, Karlsson M, Nyholm T. CT substitute derived from MRI sequences with ultrashort echo time. Med Phys. 2011;38:2708-2714.
    Pubmed CrossRef
  139. King AP, Buerger C, Tsoumpas C, Marsden PK, Schaeffter T. Thoracic respiratory motion estimation from MRI using a statistical model and a 2-D image navigator. Med Image Anal. 2012;16:252-264.
    Pubmed CrossRef
  140. Uribe S, Muthurangu V, Boubertakh R, Schaeffter T, Razavi R, Hill DL, et al. Whole-heart cine MRI using real-time respiratory self-gating. Magn Reson Med. 2007;57:606-613.
    Pubmed CrossRef
  141. Erlandsson K, Dickson J, Arridge S, Atkinson D, Ourselin S, Hutton BF. MR Imaging-guided partial volume correction of PET data in PET/MR imaging. PET Clin. 2016;11:161-177.
    Pubmed CrossRef
  142. Bai B, Li Q, Leahy RM. Magnetic resonance-guided positron emission tomography image reconstruction. Semin Nucl Med. 2013;43:30-44.
    Pubmed KoreaMed CrossRef
  143. Wahl RL, Quint LE, Cieslak RD, Aisen AM, Koeppe RA, Meyer CR. "Anatometabolic" tumor imaging: fusion of FDG PET with CT or MRI to localize foci of increased activity. J Nucl Med. 1993;34:1190-1197.
    Pubmed
  144. Kirchner J, Sawicki LM, Deuschl C, Grüneisen J, Beiderwellen K, Lauenstein TC, et al. 18 F-FDG PET/MR imaging in patients with suspected liver lesions: value of liver-specific contrast agent Gadobenate dimeglumine. PLoS One. 2017;12:e0180349.
    Pubmed KoreaMed CrossRef
  145. Baumann T, Rottenburger C, Nicolas G, Wild D. Gastroenteropancreatic neuroendocrine tumours (GEP-NET) - Imaging and staging. Best Pract Res Clin Endocrinol Metab. 2016;30:45-57.
    Pubmed CrossRef
  146. Chen BB, Tien YW, Chang MC, Cheng MF, Chang YT, Wu CH, et al. PET/MRI in pancreatic and periampullary cancer: correlating diffusion-weighted imaging, MR spectroscopy and glucose metabolic activity with clinical stage and prognosis. Eur J Nucl Med Mol Imaging. 2016;43:1753-1764.
    Pubmed CrossRef
  147. Eiber M, Weirich G, Holzapfel K, Souvatzoglou M, Haller B, Rauscher I, et al. Simultaneous 68Ga-PSMA HBED-CC PET/MRI improves the localization of primary prostate cancer. Eur Urol. 2016;70:829-836.
    Pubmed CrossRef
  148. Miller-Thomas MM, Benzinger TL. Neurologic applications of PET/MR imaging. Magn Reson Imaging Clin N Am. 2017;25:297-313.
    Pubmed KoreaMed CrossRef
  149. Jack CR Jr, Albert MS, Knopman DS, McKhann GM, Sperling RA, Carrillo MC, et al. Introduction to the recommendations from the National Institute on Aging-Alzheimer's Association workgroups on diagnostic guidelines for Alzheimer's disease. Alzheimers Dement. 2011;7:257-262.
    Pubmed CrossRef
  150. Lee KK, Salamon N. [18F] fluorodeoxyglucose-positron-emission tomography and MR imaging coregistration for presurgical evaluation of medically refractory epilepsy. AJNR Am J Neuroradiol. 2009;30:1811-1816.
    Pubmed KoreaMed CrossRef
  151. Kransdorf MJ, Bridges MD. Current developments and recent advances in musculoskeletal tumor imaging. Semin Musculoskelet Radiol. 2013;17:145-155.
    Pubmed CrossRef
  152. Abgral R, Dweck MR, Trivieri MG, Robson PM, Karakatsanis N, Mani V, et al. Clinical utility of combined FDG-PET/MR to assess myocardial disease. JACC Cardiovasc Imaging. 2017;10:594-597.
    Pubmed KoreaMed CrossRef
  153. Dubinsky TJ, Cuevas C, Dighe MK, Kolokythas O, Hwang JH. High-intensity focused ultrasound: current potential and oncologic applications. AJR Am J Roentgenol. 2008;190:191-199.
    Pubmed CrossRef
  154. Cline HE, Schenck JF, Hynynen K, Watkins RD, Souza SP, Jolesz FA. MR-guided focused ultrasound surgery. J Comput Assist Tomogr. 1992;16:956-965.
    Pubmed KoreaMed CrossRef
  155. Hynynen K, Darkazanli A, Unger E, Schenck JF. MRI-guided noninvasive ultrasound surgery. Med Phys. 1993;20:107-115.
    Pubmed CrossRef
  156. Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, et al. A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med. 1995;34:814-823.
    Pubmed CrossRef
  157. Rieke V, Butts Pauly K. MR thermometry. J Magn Reson Imaging. 2008;27:376-390.
    Pubmed CrossRef
  158. Lénárd ZM, McDannold NJ, Fennessy FM, Stewart EA, Jolesz FA, Hynynen K, et al. Uterine leiomyomas: MR imaging-guided focused ultrasound surgery--imaging predictors of success. Radiology. 2008;249:187-194.
    Pubmed KoreaMed CrossRef
  159. Gianfelice D, Gupta C, Kucharczyk W, Bret P, Havill D, Clemons M. Palliative treatment of painful bone metastases with MR imaging--guided focused ultrasound. Radiology. 2008;249:355-363.
    Pubmed CrossRef
  160. Benedict SH, De Meerleer G, Orton CG, Stancanello J. Point/counterpoint. High intensity focused ultrasound may be superior to radiation therapy for the treatment of early stage prostate cancer. Med Phys. 2011;38:3909-3912.
    Pubmed CrossRef
  161. Hynynen K, Pomeroy O, Smith DN, Huber PE, McDannold NJ, Kettenbach J, et al. MR imaging-guided focused ultrasound surgery of fibroadenomas in the breast: a feasibility study. Radiology. 2001;219:176-185.
    Pubmed CrossRef
  162. Park J, Aryal M, Vykhodtseva N, Zhang YZ, McDannold N. Evaluation of permeability, doxorubicin delivery, and drug retention in a rat brain tumor model after ultrasound-induced blood-tumor barrier disruption. J Control Release. 2017;250:77-85.
    Pubmed KoreaMed CrossRef
  163. Lipsman N, Schwartz ML, Huang Y, Lee L, Sankar T, Chapman M, et al. MR-guided focused ultrasound thalamotomy for essential tremor: a proof-of-concept study. Lancet Neurol. 2013;12:462-468.
    Pubmed CrossRef
  164. Magara A, Bühler R, Moser D, Kowalski M, Pourtehrani P, Jeanmonod D. First experience with MR-guided focused ultrasound in the treatment of Parkinson's disease. J Ther Ultrasound. 2014;2:11.
    Pubmed CrossRef
  165. Jung HH, Kim SJ, Roh D, Chang JG, Chang WS, Kweon EJ, et al. Bilateral thermal capsulotomy with MR-guided focused ultrasound for patients with treatment-refractory obsessive-compulsive disorder: a proof-of-concept study. Mol Psychiatry. 2015;20:1205-1211.
    Pubmed CrossRef
  166. Kim M, Kim CH, Jung HH, Kim SJ, Chang JW. Treatment of major depressive disorder via magnetic resonance-guided focused ultrasound surgery. Biol Psychiatry. 2018;83:e17-e18.
    Pubmed CrossRef
  167. Jeanmonod D, Werner B, Morel A, Michels L, Zadicario E, Schiff G, et al. Transcranial magnetic resonance imaging-guided focused ultrasound: noninvasive central lateral thalamotomy for chronic neuropathic pain. Neurosurg Focus. 2012;32:E1.
    Pubmed CrossRef
  168. Saenz DL, Yan Y, Christensen N, Henzler MA, Forrest LJ, Bayouth JE, et al. Characterization of a 0.35T MR system for phantom image quality stability and in vivo assessment of motion quantification. J Appl Clin Med Phys. 2015;16:30-40.
    Pubmed KoreaMed CrossRef
  169. Yun J, St Aubin J, Rathee S, Fallone BG. Brushed permanent magnet DC MLC motor operation in an external magnetic field. Med Phys. 2010;37:2131-2134.
    Pubmed CrossRef
  170. St Aubin J, Steciw S, Fallone BG. Effect of transverse magnetic fields on a simulated in-line 6 MV linac. Phys Med Biol. 2010;55:4861-4869.
    Pubmed CrossRef
  171. St Aubin J, Santos DM, Steciw S, Fallone BG. Effect of longitudinal magnetic fields on a simulated in-line 6 MV linac. Med Phys. 2010;37:4916-4923.
    Pubmed CrossRef
  172. Bielajew AF. The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beams. Med Phys. 1993;20:1171-1179.
    Pubmed CrossRef
  173. Raaijmakers AJ, Raaymakers BW, Lagendijk JJ. Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons. Phys Med Biol. 2005;50:1363-1376.
    Pubmed CrossRef
  174. Liney GP, Whelan B, Oborn B, Barton M, Keall P. MRI-linear accelerator radiotherapy systems. Clin Oncol (R Coll Radiol). 2018;30:686-691.
    Pubmed CrossRef
  175. Emmerich J, Laun FB, Pfaffenberger A, Schilling R, Denoix M, Maier F, et al. Technical note: on the size of susceptibility-induced MR image distortions in prostate and cervix in the context of MR-guided radiation therapy. Med Phys. 2018;45:1586-1593.
    Pubmed CrossRef
  176. Liney GP, Dong B, Begg J, Vial P, Zhang K, Lee F, et al. Technical note: experimental results from a prototype high-field inline MRI-linac. Med Phys. 2016;43:5188.
    Pubmed CrossRef
  177. Burke B, Wachowicz K, Fallone BG, Rathee S. Effect of radiation induced current on the quality of MR images in an integrated linac-MR system. Med Phys. 2012;39:6139-6147.
    Pubmed CrossRef
  178. Palacios MA, Bohoudi O, Bruynzeel AME, van Sörsen de Koste JR, Cobussen P, Slotman BJ, et al. Role of daily plan adaptation in MR-guided stereotactic ablative radiation therapy for adrenal metastases. Int J Radiat Oncol Biol Phys. 2018;102:426-433.
    Pubmed CrossRef
  179. Kontaxis C, Bol GH, Lagendijk JJ, Raaymakers BW. A new methodology for inter- and intrafraction plan adaptation for the MR-linac. Phys Med Biol. 2015;60:7485-7497.
    Pubmed CrossRef
  180. Rudra S, Jiang N, Rosenberg SA, Olsen JR, Roach MC, Wan L, et al. Using adaptive magnetic resonance image-guided radiation therapy for treatment of inoperable pancreatic cancer. Cancer Med. 2019;8:2123-2132.
    Pubmed KoreaMed CrossRef
  181. Murray J, Tree AC. Prostate cancer- advantages and disadvantages of MR-guided RT. Clin Transl Radiat Oncol. 2019;18:68-73.
    Pubmed CrossRef
  182. Werensteijn-Honingh AM, Kroon PS, Winkel D, Aalbers EM, van Asselen B, Bol GH, et al. Feasibility of stereotactic radiotherapy using a 1.5 T MR-linac: multi-fraction treatment of pelvic lymph node oligometastases. Radiother Oncol. 2019;134:50-54.
    Pubmed CrossRef
  183. Winkel D, Bol GH, Kroon PS, van Asselen B, Hackett SS, Werensteijn-Honingh AM, et al. Adaptive radiotherapy: The Elekta Unity MR-linac concept. Clin Transl Radiat Oncol. 2019;18:54-59.
    Pubmed KoreaMed CrossRef

Article

Review Article

Progress in Medical Physics 2020; 31(3): 35-53

Published online September 30, 2020 https://doi.org/10.14316/pmp.2020.31.3.35

Copyright © Korean Society of Medical Physics.

Magnetic Resonance Imaging: Historical Overview, Technical Developments, and Clinical Applications

Geon-Ho Jahng1 , Soonchan Park1 , Chang-Woo Ryu1 , Zang-Hee Cho2

1Department of Radiology, Kyung Hee University Hospital at Gangdong, College of Medicine, Kyung Hee University, 2Neuroscience Convergence Center, Korea University, Seoul, Korea

Correspondence to:Geon-Ho Jahng
(ghjahng@gmail.com)
Tel: 82-2-440-6187
Fax: 82-2-440-6932

Received: May 29, 2020; Revised: August 1, 2020; Accepted: September 1, 2020

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

The authors congratulate the cerebrations for the 30 years of the Korean Society of Medical Physics (http://www.ksmp.or.kr/). The paper is published to recognize the anniversary. Geon-Ho Jahng invited Professor Z. H. Cho to join to submit this manuscript because he has been one of the leaders in the field of magnetic resonance imaging (MRI) during the last 40 years. In this review, we describe the development and clinical histories of MRI internationally and domestically. We also discuss diffusion and perfusion MRI, molecular imaging using MRI and MR spectroscopy (MRS), and the hybrid systems, such as positron emission tomography一MRI (PET一MRI), MR-guided focused ultrasound surgery (MRgFUS), and MRI-guided linear accelerators (MRI-LINACs). In each part, we discuss the historical evolution of the developments, technical developments, and clinical applications.

Keywords: Magnetic resonance imaging, History, Technical development, Clinical application, Review

History of Magnetic Resonance Imaging

1. International contributions

The first nuclear magnetic resonance (NMR) signals from a living animal were acquired from an anesthetized rat in 1968 [1]. The capability of NMR to differentiate tumors from normal tissue was reported by Damadian in 1971 [2]. Magnetic resonance imaging (MRI) was developed by Lauterbur [3] based on the encoding spatial information of NMR signals with magnetic field gradients. The first cross-sectional image of a living mouse was published in 1974 by Lauterbur [4]. The echo-planar imaging (EPI) technique was developed by Mansfield [5]. The first MRI body scan of a human being was performed by Damadian in 1977 [6]. For their efforts, Paul C. Lauterbur and Peter Mansfield were awarded the Nobel Prize in Physiology and Medicine in 2003 in justification of the fundamental importance and applicability of MRI in medicine. The k-space was patented by Richard S. Likes in U.S.A (#U.S. Patent 4,307,343). The first clinical MRI system was installed in the early 1980s. The first whole-body MRI scanner was developed in Korea in 1982 and was commercialized in 1984 by Goldstar (Seoul, Korea). At same time, another whole-body MRI system was built at the University of Aberdeen and was used at St. Bartholomew’s Hospital (London, UK) in 1983. Superconducting MRI was developed and it was clinically used in 1985 to 1988 at Siemens (Erlangen, Germany), General Electric (GE, Boston, MA, USA), and Goldstar. Fig. 1 shows the timeline of MRI developments and a summary of the major contributions achieved internationally and domestically.

Figure 1. Timeline of MRI developments and summary of the major contributions. MRI, magnetic resonance imaging; M, magnetic; R, resonance; I, imaging; F, functional; NMR, nuclear magnetic resonance; BOLD, blood oxygen level-dependent; NIH, National Institutes of Health; PET, positron emission tomography; MRI–LINAC, MRI-guided linear accelerators.

2. Domestic contributions

The first MRI system was developed by the Korean Advanced Institute of Science and Technology (Daejeon, Korea), and was installed in Shin Hwa Hospital (Shin Hwa Nursing Hospital, Seoul, Korea) in 1984. This MRI scanner used a 0.2 T permanent magnet. Therefore, its use for clinical studies was limited considerably at that time. The first commercial MRI (Spectro-20000; Goldstar) was installed in Seoul University Hospital (Seoul, Korea) in 1987. This MRI system was a 2.0 Tesla superconductivity magnet. During this period, GE, and Siemens also developed the 1.5 T superconductive MRI systems. In 1990, the first animal 4.7 T MRI system (Biospec, Bruker, Switzerland) was installed in Seoul’s Asan Joong Ang Hospital (Asan Medical Center, Seoul, Korea). Among the many incremental developments in MRI in Korea, one of the notable progresses made during the early 21st century was the development of ultrahigh field (UHF) 7.0 T MRI and its applications [7 -11]. Another notable worldwide research activity was the development of the hybrid positron emission tomography–MRI (PET–MRI) system. Since the introduction of the UHF (7.0 T) MRI system in Korea, an effort was initiated for the construction of a hybrid system with 7.0 T MRI and PET that led to the development of one of the world’s most advanced PET–MRI systems in 2007. With these developments, a number of neurotransmitter studies was initiated to study the serotonergic distribution in the brainstem in vivo [7,8]. Another new and interesting development with UHF 7.0 T MRI was the super-resolution tractography initiative with resolution down to 200 µm [12].

3. Development of basic imaging contrast

In 1950, spin echoes and free induction decay were detected by Hahn [13,14]. Therefore, the spin echo (SE) is commonly referred to as Hahn’s echo. MR angiography was developed by Charles L. Dumoulin and Howard R. Hart at the General Electric in 1986 (“Blood-flow checker”. Popular Science: 12. 1987). The fluid attenuation inversion recovery (FLAIR) pulse sequence that yields high-signal regions in normal white matter, was demonstrated by Hajnal et al. in 1992 [15]. Blood Oxygen Level Dependent signal was recognized by Ogawa [16] at AT&T Bell labs in 1990. Susceptibility-weighted imaging (SWI) was developed by Reichenbach et al. [17] at Washington University in 1997.

Fig. 2 shows patient cases that demonstrate the use of imaging contrast agents acquired from a 83-year-old female, 71-year-old male, and from a 26-year-old male with a 3 T MRI system.

Figure 2. Patient cases to show imaging contrasts acquired from (a) 83-year-old female, (b) 71-year-old male, and (c) 26-year-old male using a 3 T MRI system. F, female; M, male; T1W, T1-weighted; T2W, T2-weighted; FLAIR, fluid attenuated inversion recovery; GRE, gradient-echo; TOF, time of flight; CE MRA, contrast-enhanced magnetic resonance angiography; DWI b0, diffusion-weighted image with b=0 s/mm2; DWI b1000, diffusion-weighted image with b=1000 s/mm2; ADC, apparent diffusion coefficient; CE T1W, contrast-enhanced T1-weighted; APT, amide proton transfer; DSC rCBV, dynamic susceptibility contrast relative cerebral blood volume; DSC rCBF, dynamic susceptibility contrast relative cerebral blood flow; DSC MTT, dynamic susceptibility contrast mean transit time; SVS, single voxel spectroscopy; DTI b1000, diffusion tensor imaging with b=1000 s/mm2; DTI FA, diffusion tensor imaging fractional anisotropy; fMRI, functional MRI.

Fast Imaging Techniques

Speed is always important to clinical practice. Fast MRI techniques were introduced and were based on the use of multiple refocusing pulses, commonly referred to as turbo spin-echo (TSE) imaging or turbo gradient-echo imaging. Rapid Acquisition with Refocused Echoes (RARE) technique was commercially implemented after Hennig [18] described it in 1986. In this sequence, the echo train length (ETL), also known as the turbo factor, denotes the number of echoes acquired at a given repetition time (TR). Spiral sequences were introduced in 1986 by Ahn et al. [19].

1. Technical developments

Several pulse sequences for fast imaging were developed, such as TSE or turbo field echo, half-Fourier, single-shot turbo spin echo (HASTE), gradient and spin echo (GRASE), balanced steady-state free-precession (bSSFP), EPI, and spiral [20]. First, the principle of operation of TSE or RARE sequences was based on the filling of many k-space lines after every radiofrequency (RF) excitation that reduced the MR scan time by reducing the number of RF excitations required for each image [21]. Second, HASTE was a single-shot version of the TSE technique [22] wherein slightly more than half the total number of required phase encoding lines of the k-space were acquired after a single RF excitation. Third, turbo gradient spin echo or GRASE, is a turbo SE sequence with additional phase encoding gradient echoes between successive 180° refocusing pulses [23]. Fourth, bSSFP is a gradient-echo sequence that accomplishes fast acquisitions by eliminating idle times [24]. In the most common case, the data are mapped line-by-line from the top to bottom parts of k-space in a lexicographic manner, commonly referred to as Cartesian sampling. Fifth, EPI allows the collection of all the data required to reconstruct an image after a single RF excitation [25] and spiral imaging acquires data using two oscillating gradients that create a spiral trajectory in k-space [19]. The single-shot approach of EPI or spiral is the fastest with a subsecond acquisition time. Therefore, these techniques are usually used in functional MRI (fMRI). Furthermore, radial trajectories have also been introduced [26].

Several reconstruction methods were also developed to achieve higher signal-to-noise ratios (SNR) and reduced scan times. Parallel imaging techniques in conjunction with the use of phased array coils have been developed to reduce scan times by acquiring a reduced amount of k-space data with an array of receiver coils. These techniques include the simultaneous acquisition of spatial harmonics (SMASH) [27], sensitivity encoding (SENSE) [28], and generalized auto-calibrating partially parallel acquisitions (GRAPPA) [29]. SENSE is based on reconstruction in the imaging domain, but GRAPPA is based on k-space domain reconstruction. These parallel imaging techniques are being further developed nowadays [30]. Recently, multiband excitation imaging and finger printing imaging techniques have been developed to reduce the scan times even further.

The simultaneous multislice (SMS) pulse sequence [31] applies a multiband composite RF pulse with a slice-selective gradient to simultaneously excite multiple slice planes [32]. The multiband technique is simultaneous image refocused (SIR) EPI or simultaneous echo refocused (SER) EPI that does not use parallel imaging and coil sensitivity for image separation [33]. Another recently developed fast imaging technique is magnetic resonance fingerprinting (MRF) [34]. This technique allows simultaneous and efficient measurements of multiple tissue properties with one acquisition [35]. In MRF, the acquisition parameters, such as the RF excitation angle, phase, repetition time, and k-space sampling trajectory, are varied throughout the acquisition. When implemented properly, this acquisition could generate a unique signal time course for each tissue. For every MRF sequence, the dictionary of signal evolutions can be generated on a computer using mathematical algorithms to predict spin behavior and signal evolution during the acquisition.

2. Clinical applications

Ultrafast imaging is used to eliminate the effects of physiological motion, thus capturing dynamic events in real time or shortening the total scan time. Shortening the scan time helps improve patient comfort and compliance, thereby minimizing motion during scans [36]. At higher imaging speeds, it becomes feasible to examine a wide range of relevant physiological processes or to freeze induced motion that may otherwise lead to artifacts. Relevant physiological processes include, among others, respiration [37], cardiac rhythm [38], and hemodynamics from neuronal activation [39]. For example, ultrafast imaging is often used in cardiovascular imaging [40,41].

SENSE or GRAPPA are commonly used in the clinical practice nowadays. Multiband technique is used in fMRI. However, there is always an inherent trade-off between imaging speed and quality. MRF was initially evaluated in brain relaxometry [35], prostate [42], liver [43], cardiac [44], musculoskeletal imaging [43], arterial spin labeling (ASL) perfusion measurement [45], and microvascular properties [46].

Diffusion MRI

A diffusion process is described by Fick’s law. Einstein described the relationship between the mean-squared displacement and diffusion coefficient in Brownian motion [47]. Diffusion is dependent on concentration, pressure, and temperature (https://en.wikipedia.org/wiki/Diffusion). Diffusion MRI is currently a well-established technique that is used in routine clinical practice to identify lesions and to characterize them. Diffusion MRI was developed by Le Bihan in 1989 [48]. Diffusion tensor imaging (DTI) and fiber tractography were developed by Peter Basser and Le Bihan in the early 1990s [49 -51]. The diffusion coefficient for water at 37°C is approximately D=3×10−9 m2/s.

1. Technical developments

Diffusion-weighted imaging (DWI) was developed to investigate microstructural properties by evaluating the proton diffusion process. The technique is used to characterize the microscopic behaviors of protons noninvasively. The diffusion coefficient or apparent diffusion coefficient (ADC) is measured on the basis of the use of diffusion gradients around the refocusing pulse in the SE sequence, or the use of bipolar gradients in the gradient-echo sequence within the echo time (TE) time period, and is fitted by the mono-exponential decay curve [50].

S=S0ebD

where b is the b-factor that can be calculated using the strength of the applied diffusion gradient, the duration of the diffusion gradient, and the duration of the period between two diffusion gradients.

A two-compartment model was applied to measure diffusion from intracellular and extracellular compartments, respectively referred to as slow (Dslow) and fast (Dfast) diffusion with the corresponding volume fractions Fslow and Ffast [52].

S=S0 [FfastebDfast+FslowebDslow]

The intravoxel incoherent motion (IVIM) method was introduced to separate diffusion from flow effects in a voxel [50]. In this technique, the flowing vascular volume fraction of incoherently flowing blood in the tissue (F), pseudo-diffusion coefficient (D*) associated to the IVIM effect, and true diffusion coefficient (D) can be measured by applying multiple b-values.

S=S0 [FflowebD*+(1Fflow)ebD]

Because in vivo proton movement is not isotropic, anisotropic diffusion was introduced by assuming ellipsoidal diffusion, known as DTI [51]. The diffusion coefficient in the DTI model is a 3×3 symmetric matrix. Eigenvalues and eigenvectors can be extracted from the tensor matrix to calculate the isotropic diffusion effects by using the mean diffusivity (MD=(Dxx+Dyy+Dzz)/3), and anisotropic diffusion effects by using fractional anisotropy (FA) index that represents the variance of the three eigenvalues divided by norm of the eigenvalues.

Tensor models can be applied by assuming Gaussian distributions of proton movements. However, in biological structures, water molecular diffusion is hindered in the extra-axonal space and restricted in the intra-axonal space [53]. To take this effect into account, other models were introduced, such as the ball-and-stick model [54], Q-ball imaging (QBI) [55], diffusion spectrum imaging (DSI) [56], spherical deconvolution model [57], and neurite orientation dispersion and density imaging (NODDI) [58]. Furthermore, non-Gaussian diffusion kurtosis imaging, was applied in clinical studies [59].

Finally, tractography techniques were introduced to detect white matter fiber bundles by measuring the eigenvectors along the fibers [60,61]. Improvements of fiber tracking were further achieved by using probabilistic models [62]. Tractography is currently used to investigate white matter connectivity in healthy brains and in several pathologic conditions [63].

2. Clinical applications

Diffusion MRI techniques, including DWI, DTI, and tractography, are currently extensively used in clinical settings. DWI is routinely applied in stroke and tumor patients. Diffusion indices of MD and FA are altered in the presence of ischemic injury [64] and in neoplasms [65]. In brain neoplasms, ADC values have been shown to be decreased in highly cellular tumors, such as central nervous system (CNS) lymphoma, medulloblastoma, and high-grade glioma [66]. Diffusion imaging techniques were applied in Alzheimer’s disease [67 -71]. Furthermore, malignant lesions have lower ADC values compared with surrounding normal tissue, edema, and benign tumors in the brain, head and neck malignancies, prostate, and liver cancer [72]. Diffusion-weighted whole-body imaging with background suppression (DWIBS) was performed with the short tau inversion recovery (STIR) EPI sequence with a high b value for background suppression to evaluate metastatic lesions in the body [73]. DTI was used to evaluate dislocation, disruption, infiltration, and edema. Tractography was used to evaluate corticospinal tract fibers, optic radiation fibers, and language fibers in patients to perform presurgical planning in neoplastic brain tumor cases.

Perfusion MRI

Perfusion refers to the delivery of blood to a capillary bed in tissue. This is different compared with bulk flow motion (https://en.wikipedia.org/wiki/Perfusion). August Krogh first described the mechanism of regulation of capillaries in muscle, and was awarded the Nobel Prize in Physiology or Medicine [74]. Perfusion MRI is categorized according to the use of a contrast agent or not. Dynamic-susceptibility-contrast (DSC)-based perfusion MRI was developed by Villringer et al. in 1988 [75]. The ASL technique was developed by Koretsky in 2012 [76].

1. Technical developments

Details of perfusion MRI were summarized in a previous paper [77]. Three important techniques are currently used in clinical practice to obtain perfusion-related parameters. The first-pass DSC-enhanced MR perfusion is based on the susceptibility effects of gadolinium-based contrast agents on the signal echo. Therefore, T2- or T2*-weighted imaging sequence, usually a gradient-echo or an EPI sequence, is used to obtain the signal attenuation of the time-series images. Cerebral blood volume (CBV) and flow (CBF) values as well as time-related parameters, such as the mean transit time (MTT) and time-to-peak can be mapped in each pixel. Convolution theory is used to evaluate the measured concentrations of the hemodynamic changes of the contrast agent as follows [78],

C(t)=CBF0tCa(t')R(tt')dt'

where Ca(t) is the measured arterial input function (AIF) that describes the shape of the tracer that enters a voxel, and R(t-t’) is the unknown residue function that describes the probability of the tracer entering a voxel. Therefore,

MTT=0R(t)dtCBV=0CT (t) dt/0Ca (t) dt

, and CBF=CBV/MTT.

The dynamic contrast-enhanced (DCE) MR perfusion is based on the relaxivity effects of gadolinium-based contrast agents on the signal echo. Therefore, a transverse relaxation (T1)-weighted imaging sequence (usually a three-dimensional sequence), is used to obtain signal increments of time-series images. The area under-the-curve can be mapped. Furthermore, a pharmacokinetic model is used to map permeability-related parameters such as Ktrans and Kep and the corresponding volume fractions such as vp and ve. The general equation used to express the hemodynamic event after injecting the contrast agent is expressed with the extended Tofts model as follows [79],

Ctissue (t)=vp Cp (t)+Ktrans 0tCp (t')exp -Ktransve[t-t']dt'

where Cp(t) is the concentration of contrast agent in blood, Ktrans is the permeability–surface area constant from the vascular to the extracellular space, and ve is volume fraction in the extravascular and extracellular space.

The ASL MR perfusion is based on an endogenous contrast agent using magnetically labeled arterial blood water as a diffusible flow tracer. Therefore, the proton-density-weighted sequence is used to obtain signal changes with and without the use of magnetically labeled blood by either continuous or pulsed RF pulses. CBF value can be quantified. A general kinetic model for the evaluation of the difference of the longitudinal magnetization in the tissue owing to the labeled blood can be expressed to quantify the blood flow as follows [80],

ΔM(t)=2M0bCBF0t[Ca(t')R(t-t')m(t-t')]dt'

where M0b is the equilibrium magnetization in a blood filled arterial voxel, Ca(t) is the delivery function, such as AIF, R(t-t’) is the residue function that describes the washout of tagged spins from a voxel, and m(t-t’) includes the longitudinal magnetization relaxation effects. The pseudo-continuous ASL (pCASL) technique was introduced to improve the ASL signals [81], and is currently used in clinical practice.

2. Clinical applications

Perfusion MRI is a promising tool used to assess stroke, tumors, and neurodegenerative diseases. DSC perfusion MRI is the standard technique used to evaluate brain diseases, such as stroke [82] and tumors [83]. A combination of perfusion and DWI is used to evaluate a mismatch between the size of a perfusion defect and the diffusion abnormality that is referred to as the ischemic penumbra [84]. Tumor grade, recurrence, postoperative changes, or radiation effects can be established with DSC perfusion imaging [85].

DCE perfusion MRI is often applied in brain diseases [86] and in patients with breast, prostate, pelvic, and muscle diseases, and can be useful in differentiating between tumor recurrence and radiation necrosis. The DCE parameters reflect a more extensive BBB disruption and a higher tumor grade [87], evaluate the treatment prediction [88], and differentiate pseudo-progression from true progressive disease in GBM patients [89].

ASL perfusion MRI is mainly applied in brain diseases, such as neurodegenerative [90,91], renal, and cardiovascular diseases. ASL perfusion MRI has been used to evaluate pseudo-progression in brain tumor [92]. This method is particularly useful for patients with poor intravenous access, infants and children, and pregnant women [93].

Molecular MRI/MR Spectroscopy

MR spectroscopy (MRS) is used to determine the molecular structure of compounds, or to detect their presence. Proton MRS is based on the proton’s magnetic moment and its interaction with magnetic fields. MRS is thus sensitive to certain aspects of tissue metabolism. Proton MRS can detect N-acetyl aspartate (NAA), creatine/phosphocreatine (Cr/PCr), choline (Cho), glucose (Glu), myoinositol (ml or mlns), lactate (Lac), alanine (Ala), glutamate and glutamine (GIx), citrate, and ethanol in the human body. MRS can detect other nuclei in compounds of biological interest, such as phosphorus-31 (found in PCr) or carbon-13 (found in glycogen). However, proton MRS is more routinely performed in clinical practice compared with 31P or 13C MRS. Therefore, in this review, we only discuss proton MRS.

Chemical exchange saturation transfer (CEST) is a novel MR technique that enables molecular imaging to obtain certain compounds at concentrations that are too low to impact the contrast of standard MRI and too low to be directly detected in MRS at typical water imaging resolutions.

1. Technical developments of MRS

1) Pulse sequences

A single voxel spectroscopy (SVS) study is performed with short or long TE values. The SVS pulse sequences are the following: point resolved spectroscopy (PRESS) [94], stimulated echo acquisition mode (STEAM) [95], image selected in vivo spectroscopy (ISIS) [96], and depth-resolved spectroscopy (DRESS) [97]. Longer TE results in the signal decrease as a result of the transverse relaxation (T2) that leads to the alteration of the phase of multiplet signals because of J-coupling [98]. TE values in the range of 135–144 ms are typically used, as this leads to the production of a spectrum in which the doublet signal of Lac with a J-coupling constant of nearly 7 Hz is entirely reversed owing to the short TE and long TR [99].

A spectroscopic imaging is an area of interest. Chemical shift imaging (CSI) is used for multiple-voxel spectroscopic acquisitions. In spectroscopic imaging, phase encoding gradients can be applied in all three dimensions to sample k-space to select a volume that resembles methods used in MRI [100].

2) Water suppression methods

The water resonance must be suppressed to detect the millimolar concentrations of other molecules/moieties. The most common suppression method is based on the use of a chemical shift selected (CHESS) pulse sequence [101].

3) Spectral quantification

The typical postprocessing techniques used include Fourier Transform (FT), baseline correction, zero filling, and phasing. Quantification of the MRS signal amplitude can provide a means for estimating the tissue concentration of the signal generating molecules. While MRS signals are usually acquired in the time domain as free induction decays or echoes, they are usually viewed and analyzed in the frequency domain. The frequency domain representation is derived from the acquired time domain data by the FT. Signal averaging is used in virtually all MRS studies to increase the SNR by averaging the signals obtained in repeated measurements. To quantify the proton spectrum in most of the clinical studies, the internal reference signal is typically used that is either the Cr signal at 3.05 ppm or the water signal at 4.69 ppm. When Cr is used as a reference signal, it is more common to report results as signal amplitude ratios, such as NAA/Cr, or Cho/Cr. One weak aspect related to the use of the Cr signal as a reference is attributed to the fact that the Cr signal is not as uniform throughout the normal brain [102]. Another weak aspect pertains to the fact that the assumption of the Cr levels do not change with disease and other physiological characteristics may be erroneous. If the water signal is used as a reference signal, its amplitude must be measured by performing a separate measurement in the same brain region without using water suppression.

A linear least-squares optimization procedure has been established and used in spectral fitting techniques. The most popular spectral fitting software is currently the LCModel [103,104]. Furthermore, spectrum fitting software is available either in the time [105] or frequency domains [106,107].

2. Molecular imaging tools other than MRS

CEST can be used to apply molecular imaging [108]. This technique is more appropriate compared with CSI or MR spectroscopy imaging because it provides relatively high resolution. The principle of CEST is based on the use of the magnetization transfer effects from other molecules to water molecules. Therefore, the requirement of CEST is that the chemical species must have in their structures a 1H proton that is exchangeable with those of water. Known endogenous diamagnetic CEST agents are involved with exchangeable groups of amide proton (-NH), amine proton (-NH2), and hydroxyl proton (-OH), whose chemical shifts are ~3.5, ~1.8–3.0, and ~0.5–1.5 parts per million (ppm), respectively [109]. Amide CEST is usually referred to as amide proton transfer (APT). CEST techniques have been applied to map glutamate (amine proton), creatine (amine proton), glycosaminoglycan (Gag) (hydroxyl proton), myoinositol (MI) (hydroxyl proton), and glucose (hydroxyl proton). The detailed principle of CEST technique has been described in numerous previous papers [110 -116].

3. Clinical applications

The goal of clinical spectroscopy is to provide physicians with biochemical information that will assist in the differential diagnosis when standard clinical and radiologic tests fail or are too invasive. Proton spectroscopy has attained clinical value in that it can monitor the evolution of diseases and associated therapies. Disease can sometimes lead to large changes in metabolite levels.

MR spectroscopy has also been shown to be of diagnostic value for evaluating and monitoring the progression of certain brain diseases, such as stroke [117], epilepsy, multiple sclerosis [118], Alzheimer’s disease [119,120], brain tumors [121] and other tumors, such as prostate cancer [122] and breast cancer [123]. Lactate signal levels are elevated in ischemic brain tissue. Choline signal levels are elevated in some neoplastic or inflamed tissues.

The main applications for amide CEST or APT are the detection of cancer and ischemic stroke. In tumor regions, the concentration of proteins are elevated compared with surrounding tissues, and thus lead to increased APT levels [124]. This method was applied to classify tumor progression from radiation necrosis [125]. The CEST technique is applied in stroke because reduced pH in the ischemic region leads to lowered APT exchange rate, and results in decreased CEST values [126]. CEST was applied in other-than-brain pathologies, such as breast [127], prostate [128], and knee [129].

Hybrid MRI Systems

1. PET–MRI

PET–MRI is an imaging system that incorporates MRI and PET to gain from the benefits of soft tissue morphological imaging (MRI) and metabolic imaging (PET). This hybrid system is mainly used in the fields of oncology and neurology for clinical and preclinical studies. Some systems operate in totally separate rooms, but other systems do operate in the same room with separate machines. In these cases, a bed is shared to transfer the subjects from MRI to PET or to fully integrated systems. The first whole-body PET–MRI systems were produced by Philips (Amsterdam, The Netherlands) and were installed in the US (Mount Sinai Medical Center, New York, NY) and in Switzerland (Geneva University Hospital, Geneva) in 2010. The system featured a PET and MRI scanner separated by a revolving bed. The simultaneous PET–MR acquisition system (Siemens) was installed in 2010. The fully integrated whole-body systems were provided in 2011 by Siemens and in 2014 by GE. In Korea, the first PET–MRI system was installed at the Gachon University Hospital (Inchen, Korea) [7,8].

1) Technical developments

In this part, we only discussed the issues of the fully integrated PET–MRI system. Placing PET detectors in the MR magnet can alter the local magnetic field strength causing the protons to spin at wrong frequencies, thus leading to the formation of severe image distortions and artifacts, such as susceptibility artifacts. The presence of PET hardware within the gradient coil significantly alters the MR eddy current characteristics of the system possibly leading to degraded spatial linearity. The gradient imperfection would impact imaging spatial resolution and image homogeneity. The lutetium-based scintillation crystals have acceptable magnetic properties [130]. The avalanche photodiode for PET can be used in a 7 T MRI system without inducing major effects to magnetic fields [131].

Attenuation correction describes a method to account for the self-absorption of the emitted annihilation photons, and is a prerequisite for accurate quantification of the PET data [132]. It is not possible to directly derive the attenuation properties of tissues from MRI measurements. MRI-based attenuation correction methods have been introduced [133]: segmentation-based [134] that are usually used in T1-weighted images [135] or Dixon-sequence-based images [136], atlas-based [133], and reconstruction-based [137] methods. Attenuation correction of bone is calculated by using ultrashort echo time images [138].

The simultaneous acquisition of MRI and PET data in the fully integrated PET–MRI systems has major advantages compared with sequential acquisitions. Two-dimensional or three-dimensional (3D) navigator MRI is used to correct respiratory motion in PET images [139]. 3D Cine MRI has been used to correct cardiac motion in PET images [140]. A high resolution MRI has been used for the correction of the partial volume effect in PET images [141]. MR-guided PET reconstruction techniques have been developed by incorporating “a priori” anatomical information from the MRI [142]. Finally, MR anatomical images have been used for aligning functional information obtained from PET [143].

2) Clinical applications

While PET provides a) the high sensitivity required to detect minute amounts of radiotracers and b) the ability to quantify radiotracer activity throughout the body in absolute terms, MRI provides excellent soft-tissue contrast according to multiple contrast mechanisms at high-spatial resolution. Therefore, PET/MRI imaging combines the diagnostic breadth and information content of both PET and MRI. The PET–MRI system has been used for the study of patients with hepatobiliary cancer [144], neuroendocrine tumors [145], pancreatic adenocarcinoma [146], prostate cancer [147], primary brain tumors [148], dementia [149], epilepsy [150], musculoskeletal tumor [151], and coronary artery disease [152].

2. MR-guided focused ultrasound surgery

High-intensity focused ultrasound (HIFU) is a noninvasive therapeutic technique that uses nonionizing ultrasonic waves to heat tissue [153]. HIFU has been combined with MRI to enable guidance of the treatment and monitoring. It is referred to as MR-guided focused ultrasound surgery (MRgFUS). It is a 3D imaging technique that features high-soft-tissue contrast and provides information about temperature, thus allowing the monitoring of ablation. In 1992, reports were published that described MRgFUS on ex vivo muscle tissue [154], and the following year on in vivo tissue [155]. MRgFUS developed by Hynynen [155] was later transferred to InsighTec in Haifa, Israel in 1998. The InsighTec ExAblate 2000 was the first MRgFUS system used to obtain Food and Drug Administration market approval in the US in 2004.

1) Technical developments

In MRgFUS, MR is used for both target localization and in vivo real time monitoring of temperature based on a technique referred to as MR thermometry [155], and for verifying tissue destruction using a postprocessing tool. The thermometric technique of temperature-dependent phase changes in gradient-recalled echo pulse sequences are commonly used to determine the temperature change [156]. The change in temperature is represented as

ΔT=ΔφγcB0TE

where Δϕ is the phase change, γ is the gyromagnetic ratio, c is the proton-resonance frequency shift constant (–0.01 ppm/°C) [157], Bo is the main magnetic field strength, and TE is the echo time. For the bone tumors, the calcification issues can be resolved be using susceptibility-weighted MRI to identify calcifications rather than computer tomography. An ultrashort TE sequence can be used to improve thermometry in bone.

2) Clinical applications

Clinical applications of MRgFUS are still limited. MRgFUS has been used in clinical studies in patients with uterine fibroids [158], bone metastases [159], prostate cancer [160], breast cancer [161], and brain diseases, such as brain tumors [162], intractable essential tremor [163], Parkinson’s disease [164], obsessive-compulsive disorder [165], major depressive disorders [166], and neuropathic pain [167].

3. MRI-guided linear accelerator

MRI-guided linear accelerator (MRI–LINAC) is a recently developed and advanced radiation treatment system. As indicating the name, the MRI–LINAC is fully integrated with the MRI for imaging soft-tissue tumors together with LINAC for the radiotherapy to treat cancers throughout the body. The advantage of MRI-based imaging on a linear accelerator has superior high-definition image quality, especially for some soft-tissue cancers compared with CT-based imaging in the traditional linear accelerators to visualize the target area and adjacent anatomy for treatment setup and delivery. The first technical prototype MR–Linac was developed and installed in the University Medical Center Utrecht in Utrecht, The Netherlands. Similar types of measurements have been performed on a hybrid MRI Cobalt-60 device [168]. The first clinically active MRI-guided radiation therapy machine (ViewRay) was installed at the Alvin J. Siteman Cancer Center at Barnes-Jewish Hospital at the Washington University School of Medicine (St. Louis, MO, USA). The treatment of the first patients was announced in February 2014.

1) Technical developments

LINAC is affected by MRI. Two main configurations of MRI–LINAC that are being pursued with the radiotherapy beam are either parallel or perpendicular to the main magnetic field. This configuration is affected by the interference between the delivery of the radiation beam of LINAC and MRI. The operation of the multileaf collimator in the strong magnetic field can be a problem in the shaping of the X-ray beam [169]. Both configurations have this problem, and vendors lowered magnetic field to minimize this issue. Another issue in the MRI–LINAC is that the accelerated electrons used to produce the X-ray beam can be deviated or defocused, thus causing a loss of the beam current. Previous studies showed that the perpendicular configuration is dominant to the total beam loss compared with the parallel one [170,171]. Skin dose can be increased by secondary electrons owing to the influence of the magnetic field [172]. In this case, the perpendicular configuration should be advantageous, although the electron return effect can still appear [173]. Receiver coils can cause attenuation of the primary beam and can increase the skin dose. Detailed explanation can be found in another paper [174].

MRI quality is also affected by the LINAC. Any RF noise generated by the LINAC can cause artifacts or noise in images. In addition, LINAC components, such as the accelerator or multileaf collimators cause inhomogeneity of the main magnetic field, thus worsening the imaging quality [175]. Finally, the radiation beam can affect conductors or electronics in the coil, causing imaging artifacts [176,177].

2) Clinical applications

The MRI-LINAC can adapt the radiation treatment plan based on movement of the organs or tumor, and also track the motion of the tumor. This system reduces complications after radiation treatments. The MRI–LINAC can be used to improve the personalization of the radiation therapy using existing contrast imaging mechanisms, such as diffusion, perfusion, functional, and metabolic, to evaluate treatment effects. The hybrid system has been focused on daily plan changes based on geometric changes in the organs-at-risk [178,179]. Furthermore, MRI has been used to evaluate radiation treatment effects [180]. Currently, this hybrid system is used to treat patients with prostate cancer [181], pelvic lymph nodes [182], and the esophagus [183].

Acknowledgements

The research was supported by the National Research Foundation of Korea grant funded by Ministry of Science and ICT (No. 2020R1A2C1004749, GHJ), Republic of Korea.

Conflicts of Interest

The authors have nothing to disclose.

Availability of Data|Materials

The authors confirm that the data supporting the findings of this study are available within the article.

Author Contributions

Conceptualization: Geon-Ho Jahng. Data curation: Geon-Ho Jahng, Soonchan Park, Chang-Woo Ryu, and Zang-Hee Cho. Writing–original drafting: Geon-Ho Jahng. Writing–review & editing: Geon-Ho Jahng, Soonchan Park, Chang-Woo Ryu, and Zang-Hee Cho.

Fig 1.

Figure 1.Timeline of MRI developments and summary of the major contributions. MRI, magnetic resonance imaging; M, magnetic; R, resonance; I, imaging; F, functional; NMR, nuclear magnetic resonance; BOLD, blood oxygen level-dependent; NIH, National Institutes of Health; PET, positron emission tomography; MRI–LINAC, MRI-guided linear accelerators.
Progress in Medical Physics 2020; 31: 35-53https://doi.org/10.14316/pmp.2020.31.3.35

Fig 2.

Figure 2.Patient cases to show imaging contrasts acquired from (a) 83-year-old female, (b) 71-year-old male, and (c) 26-year-old male using a 3 T MRI system. F, female; M, male; T1W, T1-weighted; T2W, T2-weighted; FLAIR, fluid attenuated inversion recovery; GRE, gradient-echo; TOF, time of flight; CE MRA, contrast-enhanced magnetic resonance angiography; DWI b0, diffusion-weighted image with b=0 s/mm2; DWI b1000, diffusion-weighted image with b=1000 s/mm2; ADC, apparent diffusion coefficient; CE T1W, contrast-enhanced T1-weighted; APT, amide proton transfer; DSC rCBV, dynamic susceptibility contrast relative cerebral blood volume; DSC rCBF, dynamic susceptibility contrast relative cerebral blood flow; DSC MTT, dynamic susceptibility contrast mean transit time; SVS, single voxel spectroscopy; DTI b1000, diffusion tensor imaging with b=1000 s/mm2; DTI FA, diffusion tensor imaging fractional anisotropy; fMRI, functional MRI.
Progress in Medical Physics 2020; 31: 35-53https://doi.org/10.14316/pmp.2020.31.3.35

References

  1. Jackson JA, Langham WH. Whole-body NMR spectrometer. Rev Sci Instrum. 1968;39:510-513.
    Pubmed CrossRef
  2. Damadian R. Tumor detection by nuclear magnetic resonance. Science. 1971;171:1151-1153.
    Pubmed CrossRef
  3. Lauterbur PC. Image formation by induced local interactions. Examples employing nuclear magnetic resonance. 1973. Clin Orthop Relat Res. 1989;244:3-6.
    Pubmed
  4. Lauterbur PC. Magnetic resonance zeugmatography. Pure Appl Chem. 1974;40:149-157.
    CrossRef
  5. Mansfield P, Grannell P. "Diffraction" and microscopy in solids and liquids by NMR. Phys Rev B. 1975;12:3618.
    CrossRef
  6. Damadian R, Minkoff L, Goldsmith M, Stanford M, Koutcher J. Field focusing nuclear magnetic resonance (FONAR): visualization of a tumor in a live animal. Science. 1976;194:1430-1432.
    Pubmed CrossRef
  7. Cho ZH, Kim YB, Han JY, Kim NB, Hwang SI, Kim SJ, et al. Altered T2* relaxation time of the hippocampus in major depressive disorder: implications of ultra-high field magnetic resonance imaging. J Psychiatr Res. 2010;44:881-886.
    Pubmed CrossRef
  8. Cho ZH, Lee YB, Kang CK, Yang JW, Jung IH, Park CA, et al. Microvascular imaging of asymptomatic MCA steno-occlusive patients using ultra-high-field 7T MRI. J Neurol. 2013;260:144-150.
    Pubmed CrossRef
  9. Cho ZH, Son YD, Kim HK, Kim KN, Oh SH, Han JY, et al. A fusion PET-MRI system with a high-resolution research tomograph-PET and ultra-high field 7.0 T-MRI for the molecular-genetic imaging of the brain. Proteomics. 2008;8:1302-1323.
    Pubmed CrossRef
  10. Kim M, Kim KE, Jeong SW, Hwang SW, Jo H, Lee J, et al. Effects of the ultra-high-frequency electrical field radiofrequency device on mouse skin: a histologic and molecular study. Plast Reconstr Surg. 2016;138:248e-255e.
    Pubmed CrossRef
  11. Oh SH, Chung JY, In MH, Zaitsev M, Kim YB, Speck O, et al. Distortion correction in EPI at ultra-high-field MRI using PSF mapping with optimal combination of shift detection dimension. Magn Reson Med. 2012;68:1239-1246.
    Pubmed CrossRef
  12. Richards K, Calamante F, Tournier JD, Kurniawan ND, Sadeghian F, Retchford AR, et al. Mapping somatosensory connectivity in adult mice using diffusion MRI tractography and super-resolution track density imaging. Neuroimage. 2014;102 Pt 2:381-392.
    Pubmed CrossRef
  13. Hahn EL. Spin echoes. Phys Rev. 1950;80:580.
    CrossRef
  14. Hahn EL. Nuclear induction due to free larmor precession. Phys Rev. 1950;77:297.
    CrossRef
  15. Hajnal JV, De Coene B, Lewis PD, Baudouin CJ, Cowan FM, Pennock JM, et al. High signal regions in normal white matter shown by heavily T2-weighted CSF nulled IR sequences. J Comput Assist Tomogr. 1992;16:506-513.
    Pubmed CrossRef
  16. Ogawa S, Lee TM, Kay AR, Tank DW. Brain magnetic resonance imaging with contrast dependent on blood oxygenation. Proc Natl Acad Sci U S A. 1990;87:9868-9872.
    Pubmed CrossRef
  17. Reichenbach JR, Venkatesan R, Schillinger DJ, Kido DK, Haacke EM. Small vessels in the human brain: MR venography with deoxyhemoglobin as an intrinsic contrast agent. Radiology. 1997;204:272-277.
    Pubmed CrossRef
  18. Hennig J, Nauerth A, Friedburg H. RARE imaging: a fast imaging method for clinical MR. Magn Reson Med. 1986;3:823-833.
    Pubmed CrossRef
  19. Ahn CB, Kim JH, Cho ZH. High-speed spiral-scan echo planar NMR imaging-I. IEEE Trans Med Imaging. 1986;5:2-7.
    Pubmed CrossRef
  20. Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Magn Reson Med. 1992;28:202-213.
    Pubmed CrossRef
  21. Hennig J, Friedburg H. Clinical applications and methodological developments of the RARE technique. Magn Reson Imaging. 1988;6:391-395.
    Pubmed CrossRef
  22. Kiefer B, Grassner J, I-lausman K. Image acquisition in a second with half fourier acquisition single shot turbo spin echo. J Magn Reson Imaging. 1994;4(P):86.
  23. Oshio K, Feinberg DA. GRASE (Gradient- and spin-echo) imaging: a novel fast MRI technique. Magn Reson Med. 1991;20:344-349.
    Pubmed CrossRef
  24. Nayak KS, Hargreaves BA, Hu BS, Nishimura DG, Pauly JM, Meyer CH. Spiral balanced steady-state free precession cardiac imaging. Magn Reson Med. 2005;53:1468-1473.
    Pubmed CrossRef
  25. Mansfield P. Multi-planar image formation using NMR spin echoes. J Phys C. 1977;10:L55.
    CrossRef
  26. Du J, Lu A, Block WF, Thornton FJ, Grist TM, Mistretta CA. Time-resolved undersampled projection reconstruction magnetic resonance imaging of the peripheral vessels using multi-echo acquisition. Magn Reson Med. 2005;53:730-734.
    Pubmed CrossRef
  27. Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38:591-603.
    Pubmed CrossRef
  28. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952-962.
    CrossRef
  29. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202-1210.
    Pubmed CrossRef
  30. Hamilton J, Franson D, Seiberlich N. Recent advances in parallel imaging for MRI. Prog Nucl Magn Reson Spectrosc. 2017;101:71-95.
    Pubmed CrossRef
  31. Feinberg DA, Setsompop K. Ultra-fast MRI of the human brain with simultaneous multi-slice imaging. J Magn Reson. 2013;229:90-100.
    Pubmed KoreaMed CrossRef
  32. Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging. 2001;13:313-317.
    Pubmed CrossRef
  33. Feinberg DA, Reese TG, Wedeen VJ. Simultaneous echo refocusing in EPI. Magn Reson Med. 2002;48:1-5.
    Pubmed CrossRef
  34. Panda A, Mehta BB, Coppo S, Jiang Y, Ma D, Seiberlich N, et al. Magnetic resonance fingerprinting-an overview. Curr Opin Biomed Eng. 2017;3:56-66.
    Pubmed KoreaMed CrossRef
  35. Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187-192.
    Pubmed KoreaMed CrossRef
  36. Tsao J. Ultrafast imaging: principles, pitfalls, solutions, and applications. J Magn Reson Imaging. 2010;32:252-266.
    Pubmed CrossRef
  37. Saini S, Stark DD, Rzedzian RR, Pykett IL, Rummeny E, Hahn PF, et al. Forty-millisecond MR imaging of the abdomen at 2.0 T. Radiology. 1989;173:111-116.
    Pubmed CrossRef
  38. Doyle M, Rzedzian R, Mansfield P, Coupland RE. Dynamic NMR cardiac imaging in a piglet. Br J Radiol. 1983;56:925-930.
    Pubmed CrossRef
  39. Glover GH, Lee AT. Motion artifacts in fMRI: comparison of 2DFT with PR and spiral scan methods. Magn Reson Med. 1995;33:624-635.
    Pubmed CrossRef
  40. Muthurangu V, Lurz P, Critchely JD, Deanfield JE, Taylor AM, Hansen MS. Real-time assessment of right and left ventricular volumes and function in patients with congenital heart disease by using high spatiotemporal resolution radial k-t SENSE. Radiology. 2008;248:782-791.
    Pubmed CrossRef
  41. Nayak KS, Cunningham CH, Santos JM, Pauly JM. Real-time cardiac MRI at 3 tesla. Magn Reson Med. 2004;51:655-660.
    Pubmed CrossRef
  42. Yu AC, Badve C, Ponsky LE, Pahwa S, Dastmalchian S, Rogers M, et al. Development of a combined MR fingerprinting and diffusion examination for prostate cancer. Radiology. 2017;283:729-738.
    Pubmed KoreaMed CrossRef
  43. Cloos MA, Knoll F, Zhao T, Block KT, Bruno M, Wiggins GC, et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016;7:12445.
    Pubmed KoreaMed CrossRef
  44. Hamilton JI, Jiang Y, Chen Y, Ma D, Lo WC, Griswold M, et al. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med. 2017;77:1446-1458.
    Pubmed CrossRef
  45. Su P, Mao D, Liu P, Li Y, Pinho MC, Welch BG, et al. Multiparametric estimation of brain hemodynamics with MR fingerprinting ASL. Magn Reson Med. 2017;78:1812-1823.
    Pubmed KoreaMed CrossRef
  46. Christen T, Pannetier NA, Ni WW, Qiu D, Moseley ME, Schuff N, et al. MR vascular fingerprinting: a new approach to compute cerebral blood volume, mean vessel radius, and oxygenation maps in the human brain. Neuroimage. 2014;89:262-270.
    Pubmed KoreaMed CrossRef
  47. Einstein A. Investigations on the theory of the Brownian movement. Chelmsford: Courier Corporation; 1956.
  48. Le Bihan D, Breton E, inventor; General Electric CGR SA, assignee. Method to measure the molecular diffusion and/or perfusion parameters of live tissue. United States patent US4809701A. 1989 Mar 7.
  49. Howe FA, Filler AG, Bell BA, Griffiths JR. Magnetic resonance neurography. Magn Reson Med. 1992;28:328-338.
    Pubmed CrossRef
  50. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis E, Laval-Jeantet M. MR imaging of intravoxel incoherent motions: application to diffusion and perfusion in neurologic disorders. Radiology. 1986;161:401-407.
    Pubmed CrossRef
  51. Basser PJ, Mattiello J, LeBihan D. Estimation of the effective self-diffusion tensor from the NMR spin echo. J Magn Reson B. 1994;103:247-254.
    Pubmed CrossRef
  52. Niendorf T, Dijkhuizen RM, Norris DG, van Lookeren Campagne M, Nicolay K. Biexponential diffusion attenuation in various states of brain tissue: implications for diffusion-weighted imaging. Magn Reson Med. 1996;36:847-857.
    Pubmed CrossRef
  53. Assaf Y, Freidlin RZ, Rohde GK, Basser PJ. New modeling and experimental framework to characterize hindered and restricted water diffusion in brain white matter. Magn Reson Med. 2004;52:965-978.
    Pubmed CrossRef
  54. Behrens TE, Woolrich MW, Jenkinson M, Johansen-Berg H, Nunes RG, Clare S, et al. Characterization and propagation of uncertainty in diffusion-weighted MR imaging. Magn Reson Med. 2003;50:1077-1088.
    Pubmed CrossRef
  55. Tuch DS. Q-ball imaging. Magn Reson Med. 2004;52:1358-1372.
    Pubmed KoreaMed CrossRef
  56. Wedeen VJ, Wang RP, Schmahmann JD, Benner T, Tseng WY, Dai G, et al. Diffusion spectrum magnetic resonance imaging (DSI) tractography of crossing fibers. Neuroimage. 2008;41:1267-1277.
    Pubmed CrossRef
  57. Tournier JD, Calamante F, Gadian DG, Connelly A. Direct estimation of the fiber orientation density function from diffusion-weighted MRI data using spherical deconvolution. Neuroimage. 2004;23:1176-1185.
    Pubmed CrossRef
  58. Zhang H, Schneider T, Wheeler-Kingshott CA, Alexander DC. NODDI: practical in vivo neurite orientation dispersion and density imaging of the human brain. Neuroimage. 2012;61:1000-1016.
    Pubmed CrossRef
  59. Jensen JH, Helpern JA, Ramani A, Lu H, Kaczynski K. Diffusional kurtosis imaging: the quantification of non-gaussian water diffusion by means of magnetic resonance imaging. Magn Reson Med. 2005;53:1432-1440.
    Pubmed CrossRef
  60. Basser PJ, Pajevic S, Pierpaoli C, Duda J, Aldroubi A. In vivo fiber tractography using DT-MRI data. Magn Reson Med. 2000;44:625-632.
    CrossRef
  61. Mori S, Crain BJ, Chacko VP, van Zijl PC. Three-dimensional tracking of axonal projections in the brain by magnetic resonance imaging. Ann Neurol. 1999;45:265-269.
    Pubmed CrossRef
  62. Jones DK. Tractography gone wild: probabilistic fibre tracking using the wild bootstrap with diffusion tensor MRI. IEEE Trans Med Imaging. 2008;27:1268-1274.
    Pubmed CrossRef
  63. Bullmore E, Sporns O. Complex brain networks: graph theoretical analysis of structural and functional systems. Nat Rev Neurosci. 2009;10:186-198.
    Pubmed CrossRef
  64. Sotak CH. The role of diffusion tensor imaging in the evaluation of ischemic brain injury - a review. NMR Biomed. 2002;15:561-569.
    Pubmed CrossRef
  65. Alexander AL, Hurley SA, Samsonov AA, Adluru N, Hosseinbor AP, Mossahebi P, et al. Characterization of cerebral white matter properties using quantitative magnetic resonance imaging stains. Brain Connect. 2011;1:423-446.
    Pubmed KoreaMed CrossRef
  66. Baliyan V, Das CJ, Sharma R, Gupta AK. Diffusion weighted imaging: technique and applications. World J Radiol. 2016;8:785-798.
    Pubmed CrossRef
  67. Zhang Y, Schuff N, Jahng GH, Bayne W, Mori S, Schad L, et al. Diffusion tensor imaging of cingulum fibers in mild cognitive impairment and Alzheimer disease. Neurology. 2007;68:13-19.
    Pubmed KoreaMed CrossRef
  68. Zhang Y, Du AT, Hayasaka S, Jahng GH, Hlavin J, Zhan W, et al. Patterns of age-related water diffusion changes in human brain by concordance and discordance analysis. Neurobiol Aging. 2010;31:1991-2001.
    Pubmed KoreaMed CrossRef
  69. Jahng GH, Xu S, Weiner MW, Meyerhoff DJ, Park S, Schuff N. DTI studies in patients with Alzheimer's disease, mild cognitive impairment, or normal cognition with evaluation of the intrinsic background gradients. Neuroradiology. 2011;53:749-762.
    Pubmed CrossRef
  70. Jahng GH, Xu S. Local susceptibility causes diffusion alterations in patients with Alzheimer's disease and mild cognitive impairment. Brain Imaging Behav. 2012;6:426-436.
    Pubmed CrossRef
  71. Jahng GH, Xu S, Kim MJ. Mapping of distributions of a local b-matrix cross-term strength using diffusion tensor MRI in patients with Alzheimer's disease. Med Phys. 2012;39:6324-6331.
    Pubmed CrossRef
  72. Charles-Edwards EM, deSouza NM. Diffusion-weighted magnetic resonance imaging and its application to cancer. Cancer Imaging. 2006;6:135-143.
    Pubmed CrossRef
  73. Takahara T, Imai Y, Yamashita T, Yasuda S, Nasu S, Van Cauteren M. Diffusion weighted whole body imaging with background body signal suppression (DWIBS): technical improvement using free breathing, STIR and high resolution 3D display. Radiat Med. 2004;22:275-282.
    Pubmed
  74. Larsen EH. [August Krogh (1874-1949): 1920 nobel prize]. Ugeskr Laeger. 2007;169:2878. Danish.
  75. Villringer A, Rosen BR, Belliveau JW, Ackerman JL, Lauffer RB, Buxton RB, et al. Dynamic imaging with lanthanide chelates in normal brain: contrast due to magnetic susceptibility effects. Magn Reson Med. 1988;6:164-174.
    Pubmed CrossRef
  76. Koretsky AP. Early development of arterial spin labeling to measure regional brain blood flow by MRI. Neuroimage. 2012;62:602-607.
    Pubmed KoreaMed CrossRef
  77. Jahng GH, Li KL, Ostergaard L, Calamante F. Perfusion magnetic resonance imaging: a comprehensive update on principles and techniques. Korean J Radiol. 2014;15:554-577.
    Pubmed KoreaMed CrossRef
  78. Ostergaard L, Weisskoff RM, Chesler DA, Gyldensted C, Rosen BR. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysis. Magn Reson Med. 1996;36:715-725.
    Pubmed CrossRef
  79. Tofts PS. Modeling tracer kinetics in dynamic Gd-DTPA MR imaging. J Magn Reson Imaging. 1997;7:91-101.
    Pubmed CrossRef
  80. Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR. A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med. 1998;40:383-396.
    Pubmed CrossRef
  81. Dai W, Garcia D, de Bazelaire C, Alsop DC. Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields. Magn Reson Med. 2008;60:1488-1497.
    Pubmed KoreaMed CrossRef
  82. Keir SL, Wardlaw JM. Systematic review of diffusion and perfusion imaging in acute ischemic stroke. Stroke. 2000;31:2723-2731.
    Pubmed CrossRef
  83. Cha S, Knopp EA, Johnson G, Wetzel SG, Litt AW, Zagzag D. Intracranial mass lesions: dynamic contrast-enhanced susceptibility-weighted echo-planar perfusion MR imaging. Radiology. 2002;223:11-29.
    Pubmed CrossRef
  84. Ostergaard L, Sorensen AG, Chesler DA, Weisskoff RM, Koroshetz WJ, Wu O, et al. Combined diffusion-weighted and perfusion-weighted flow heterogeneity magnetic resonance imaging in acute stroke. Stroke. 2000;31:1097-1103.
    Pubmed CrossRef
  85. Lee SJ, Kim JH, Kim YM, Lee GK, Lee EJ, Park IS, et al. Perfusion MR imaging in gliomas: comparison with histologic tumor grade. Korean J Radiol. 2001;2:1-7.
    Pubmed KoreaMed CrossRef
  86. Brix G, Kiessling F, Lucht R, Darai S, Wasser K, Delorme S, et al. Microcirculation and microvasculature in breast tumors: pharmacokinetic analysis of dynamic MR image series. Magn Reson Med. 2004;52:420-429.
    Pubmed CrossRef
  87. Choi HS, Kim AH, Ahn SS, Shin NY, Kim J, Lee SK. Glioma grading capability: comparisons among parameters from dynamic contrast-enhanced MRI and ADC value on DWI. Korean J Radiol. 2013;14:487-492.
    Pubmed KoreaMed CrossRef
  88. Ahn SS, Shin NY, Chang JH, Kim SH, Kim EH, Kim DW, et al. Prediction of methylguanine methyltransferase promoter methylation in glioblastoma using dynamic contrast-enhanced magnetic resonance and diffusion tensor imaging. J Neurosurg. 2014;121:367-373.
    Pubmed CrossRef
  89. Suh CH, Kim HS, Choi YJ, Kim N, Kim SJ. Prediction of pseudoprogression in patients with glioblastomas using the initial and final area under the curves ratio derived from dynamic contrast-enhanced T1-weighted perfusion MR imaging. AJNR Am J Neuroradiol. 2013;34:2278-2286.
    Pubmed CrossRef
  90. Kim SM, Kim MJ, Rhee HY, Ryu CW, Kim EJ, Petersen ET, et al. Regional cerebral perfusion in patients with Alzheimer's disease and mild cognitive impairment: effect of APOE epsilon4 allele. Neuroradiology. 2013;55:25-34.
    Pubmed CrossRef
  91. Johnson NA, Jahng GH, Weiner MW, Miller BL, Chui HC, Jagust WJ, et al. Pattern of cerebral hypoperfusion in Alzheimer disease and mild cognitive impairment measured with arterial spin-labeling MR imaging: initial experience. Radiology. 2005;234:851-859.
    Pubmed KoreaMed CrossRef
  92. Choi YJ, Kim HS, Jahng GH, Kim SJ, Suh DC. Pseudoprogression in patients with glioblastoma: added value of arterial spin labeling to dynamic susceptibility contrast perfusion MR imaging. Acta Radiol. 2013;54:448-454.
    Pubmed CrossRef
  93. Wang J, Licht DJ, Jahng GH, Liu CS, Rubin JT, Haselgrove J, et al. Pediatric perfusion imaging using pulsed arterial spin labeling. J Magn Reson Imaging. 2003;18:404-413.
    Pubmed CrossRef
  94. Bottomley PA. Spatial localization in NMR spectroscopy in vivo. Ann N Y Acad Sci. 1987;508:333-348.
    Pubmed CrossRef
  95. Frahm J, Bruhn H, Gyngell ML, Merboldt KD, Hänicke W, Sauter R. Localized high-resolution proton NMR spectroscopy using stimulated echoes: initial applications to human brain in vivo. Magn Reson Med. 1989;9:79-93.
    Pubmed CrossRef
  96. Ordidge RJ, Connelly A, Lohman JA. Image-selected in vivo spectroscopy (ISIS). A new technique for spatially selective NMR spectroscopy. J Magn Reson (1969). 1986;66:283-294.
    CrossRef
  97. Bottomley PA, Foster TB, Darrow RD. Depth-resolved surface-coil spectroscopy (DRESS) for in vivo 1H, 31P, and 13C NMR. J Magn Reson (1969). 1984;59:338-342.
    CrossRef
  98. Thrippleton MJ, Edden RA, Keeler J. Suppression of strong coupling artefacts in J-spectra. J Magn Reson. 2005;174:97-109.
    Pubmed CrossRef
  99. Gruber S, Pinker K, Riederer F, Chmelík M, Stadlbauer A, Bittsanský M, et al. Metabolic changes in the normal ageing brain: consistent findings from short and long echo time proton spectroscopy. Eur J Radiol. 2008;68:320-327.
    Pubmed CrossRef
  100. Skoch A, Jiru F, Bunke J. Spectroscopic imaging: basic principles. Eur J Radiol. 2008;67:230-239.
    Pubmed CrossRef
  101. Haase A, Frahm J, Hänicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30:341-344.
    Pubmed CrossRef
  102. Maudsley AA, Domenig C, Govind V, Darkazanli A, Studholme C, Arheart K, et al. Mapping of brain metabolite distributions by volumetric proton MR spectroscopic imaging (MRSI). Magn Reson Med. 2009;61:548-559.
    Pubmed KoreaMed CrossRef
  103. Provencher SW. Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med. 1993;30:672-679.
    Pubmed CrossRef
  104. Provencher SW. Automatic quantitation of localized in vivo 1H spectra with LCModel. NMR Biomed. 2001;14:260-264.
    Pubmed CrossRef
  105. Reynolds G, Wilson M, Peet A, Arvanitis TN. An algorithm for the automated quantitation of metabolites in in vitro NMR signals. Magn Reson Med. 2006;56:1211-1219.
    Pubmed CrossRef
  106. Naressi A, Couturier C, Castang I, de Beer R, Graveron-Demilly D. Java-based graphical user interface for MRUI, a software package for quantitation of in vivo/medical magnetic resonance spectroscopy signals. Comput Biol Med. 2001;31:269-286.
    Pubmed CrossRef
  107. Naressi A, Couturier C, Devos JM, Janssen M, Mangeat C, de Beer R, et al. Java-based graphical user interface for the MRUI quantitation package. MAGMA. 2001;12:141-152.
    Pubmed CrossRef
  108. van Zijl PC, Yadav NN. Chemical exchange saturation transfer (CEST): what is in a name and what isn't? Magn Reson Med. 2011;65:927-948.
    Pubmed CrossRef
  109. Vinogradov E, Sherry AD, Lenkinski RE. CEST: from basic principles to applications, challenges and opportunities. J Magn Reson. 2013;229:155-172.
    Pubmed KoreaMed CrossRef
  110. Park JE, Jahng GH, Jeong HK. Amide proton transfer imaging in clinics: basic concepts and current and future use in brain tumors and stroke. J Korean Soc Radiol. 2016;75:419-433.
    CrossRef
  111. Jahng GH, Oh JH. Physical modeling of chemical exchange saturation transfer imaging. Prog Med Phys. 2017;28:135-143.
    CrossRef
  112. Oh JH, Kim HG, Woo DC, Jeong HK, Lee SY, Jahng GH. Chemical-exchange-saturation-transfer magnetic resonance imaging to map gamma-aminobutyric acid, glutamate, myoinositol, glycine, and asparagine: phantom experiments. J Korean Phys Soc. 2017;70:545-553.
    CrossRef
  113. Jahng GH, Choi W, Chung JJ, Kim ST, Rhee HY. Mapping exchangeable protons to monitor protein alterations in the brain of an Alzheimer's disease mouse model by using MRI. Curr Alzheimer Res. 2018;15:1343-1353.
    Pubmed CrossRef
  114. Yoo CH, Oh J, Park S, Ryu CW, Kwon YK, Jahng GH. Comparative evaluation of the polynomial and spline fitting methods for the B0 correction of CEST MRI data acquired from human brains. Int J Imaging Syst Technol. 2019;29:272-282.
    CrossRef
  115. Oh JH, Kim HG, Woo DC, Rhee SJ, Lee SY, Jahng GH. Preliminary phantom experiments to map amino acids and neurotransmitters using MRI. Prog Med Phys. 2018;29:29-41.
    CrossRef
  116. Park S, Jang J, Oh JH, Ryu CW, Jahng GH. Assessment of the cerebrospinal fluid effect on the chemical exchange saturation transfer map obtained from the full Z-spectrum in the elderly human brain. Prog Med Phys. 2019;30:139-149.
    CrossRef
  117. Bruhn H, Frahm J, Gyngell ML, Merboldt KD, Hänicke W, Sauter R. Cerebral metabolism in man after acute stroke: new observations using localized proton NMR spectroscopy. Magn Reson Med. 1989;9:126-131.
    Pubmed CrossRef
  118. Arnold DL, Matthews PM, Francis G, Antel J. Proton magnetic resonance spectroscopy of human brain in vivo in the evaluation of multiple sclerosis: assessment of the load of disease. Magn Reson Med. 1990;14:154-159.
    Pubmed CrossRef
  119. Miller BL, Moats RA, Shonk T, Ernst T, Woolley S, Ross BD. Alzheimer disease: depiction of increased cerebral myo-inositol with proton MR spectroscopy. Radiology. 1993;187:433-437.
    Pubmed CrossRef
  120. Jahng GH, Oh J, Lee DW, Kim HG, Rhee HY, Shin W, et al. Glutamine and glutamate complex, as measured by functional magnetic resonance spectroscopy, alters during face-name association task in patients with mild cognitive impairment and Alzheimer's disease. J Alzheimers Dis. 2016;52:145-159.
    Pubmed CrossRef
  121. Bruhn H, Frahm J, Gyngell ML, Merboldt KD, Hänicke W, Sauter R, et al. Noninvasive differentiation of tumors with use of localized H-1 MR spectroscopy in vivo: initial experience in patients with cerebral tumors. Radiology. 1989;172:541-548.
    Pubmed CrossRef
  122. Sciarra A, Panebianco V, Ciccariello M, Salciccia S, Cattarino S, Lisi D, et al. Value of magnetic resonance spectroscopy imaging and dynamic contrast-enhanced imaging for detecting prostate cancer foci in men with prior negative biopsy. Clin Cancer Res. 2010;16:1875-1883.
    Pubmed CrossRef
  123. Haddadin IS, McIntosh A, Meisamy S, Corum C, Styczynski Snyder AL, Powell NJ, et al. Metabolite quantification and high-field MRS in breast cancer. NMR Biomed. 2009;22:65-76.
    Pubmed KoreaMed CrossRef
  124. Zhao X, Wen Z, Zhang G, Huang F, Lu S, Wang X, et al. Three-dimensional turbo-spin-echo amide proton transfer MR imaging at 3-Tesla and its application to high-grade human brain tumors. Mol Imaging Biol. 2013;15:114-122.
    Pubmed KoreaMed CrossRef
  125. Zhou J, Tryggestad E, Wen Z, Lal B, Zhou T, Grossman R, et al. Differentiation between glioma and radiation necrosis using molecular magnetic resonance imaging of endogenous proteins and peptides. Nat Med. 2011;17:130-134.
    Pubmed KoreaMed CrossRef
  126. Wang M, Hong X, Chang CF, Li Q, Ma B, Zhang H, et al. Simultaneous detection and separation of hyperacute intracerebral hemorrhage and cerebral ischemia using amide proton transfer MRI. Magn Reson Med. 2015;74:42-50.
    Pubmed KoreaMed CrossRef
  127. Dula AN, Arlinghaus LR, Dortch RD, Dewey BE, Whisenant JG, Ayers GD, et al. Amide proton transfer imaging of the breast at 3 T: establishing reproducibility and possible feasibility assessing chemotherapy response. Magn Reson Med. 2013;70:216-224.
    Pubmed CrossRef
  128. Jia G, Abaza R, Williams JD, Zynger DL, Zhou J, Shah ZK, et al. Amide proton transfer MR imaging of prostate cancer: a preliminary study. J Magn Reson Imaging. 2011;33:647-654.
    Pubmed KoreaMed CrossRef
  129. Ling W, Regatte RR, Navon G, Jerschow A. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc Natl Acad Sci U S A. 2008;105:2266-2270.
    Pubmed KoreaMed CrossRef
  130. Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci. 2003;50:1683-1685.
    CrossRef
  131. Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med. 2006;47:639-647.
    Pubmed
  132. Kinahan PE, Townsend DW, Beyer T, Sashin D. Attenuation correction for a combined 3D PET/CT scanner. Med Phys. 1998;25:2046-2053.
    Pubmed CrossRef
  133. Wagenknecht G, Kaiser HJ, Mottaghy FM, Herzog H. MRI for attenuation correction in PET: methods and challenges. MAGMA. 2013;26:99-113.
    Pubmed CrossRef
  134. Schulz V, Torres-Espallardo I, Renisch S, Hu Z, Ojha N, Börnert P, et al. Automatic, three-segment, MR-based attenuation correction for whole-body PET/MR data. Eur J Nucl Med Mol Imaging. 2011;38:138-152.
    Pubmed CrossRef
  135. Beyer T, Lassen ML, Boellaard R, Delso G, Yaqub M, Sattler B, et al. Investigating the state-of-the-art in whole-body MR-based attenuation correction: an intra-individual, inter-system, inventory study on three clinical PET/MR systems. MAGMA. 2016;29:75-87.
    Pubmed CrossRef
  136. Martinez-Möller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd'hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50:520-526.
    Pubmed CrossRef
  137. Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging. 1999;18:393-403.
    Pubmed CrossRef
  138. Johansson A, Karlsson M, Nyholm T. CT substitute derived from MRI sequences with ultrashort echo time. Med Phys. 2011;38:2708-2714.
    Pubmed CrossRef
  139. King AP, Buerger C, Tsoumpas C, Marsden PK, Schaeffter T. Thoracic respiratory motion estimation from MRI using a statistical model and a 2-D image navigator. Med Image Anal. 2012;16:252-264.
    Pubmed CrossRef
  140. Uribe S, Muthurangu V, Boubertakh R, Schaeffter T, Razavi R, Hill DL, et al. Whole-heart cine MRI using real-time respiratory self-gating. Magn Reson Med. 2007;57:606-613.
    Pubmed CrossRef
  141. Erlandsson K, Dickson J, Arridge S, Atkinson D, Ourselin S, Hutton BF. MR Imaging-guided partial volume correction of PET data in PET/MR imaging. PET Clin. 2016;11:161-177.
    Pubmed CrossRef
  142. Bai B, Li Q, Leahy RM. Magnetic resonance-guided positron emission tomography image reconstruction. Semin Nucl Med. 2013;43:30-44.
    Pubmed KoreaMed CrossRef
  143. Wahl RL, Quint LE, Cieslak RD, Aisen AM, Koeppe RA, Meyer CR. "Anatometabolic" tumor imaging: fusion of FDG PET with CT or MRI to localize foci of increased activity. J Nucl Med. 1993;34:1190-1197.
    Pubmed
  144. Kirchner J, Sawicki LM, Deuschl C, Grüneisen J, Beiderwellen K, Lauenstein TC, et al. 18 F-FDG PET/MR imaging in patients with suspected liver lesions: value of liver-specific contrast agent Gadobenate dimeglumine. PLoS One. 2017;12:e0180349.
    Pubmed KoreaMed CrossRef
  145. Baumann T, Rottenburger C, Nicolas G, Wild D. Gastroenteropancreatic neuroendocrine tumours (GEP-NET) - Imaging and staging. Best Pract Res Clin Endocrinol Metab. 2016;30:45-57.
    Pubmed CrossRef
  146. Chen BB, Tien YW, Chang MC, Cheng MF, Chang YT, Wu CH, et al. PET/MRI in pancreatic and periampullary cancer: correlating diffusion-weighted imaging, MR spectroscopy and glucose metabolic activity with clinical stage and prognosis. Eur J Nucl Med Mol Imaging. 2016;43:1753-1764.
    Pubmed CrossRef
  147. Eiber M, Weirich G, Holzapfel K, Souvatzoglou M, Haller B, Rauscher I, et al. Simultaneous 68Ga-PSMA HBED-CC PET/MRI improves the localization of primary prostate cancer. Eur Urol. 2016;70:829-836.
    Pubmed CrossRef
  148. Miller-Thomas MM, Benzinger TL. Neurologic applications of PET/MR imaging. Magn Reson Imaging Clin N Am. 2017;25:297-313.
    Pubmed KoreaMed CrossRef
  149. Jack CR Jr, Albert MS, Knopman DS, McKhann GM, Sperling RA, Carrillo MC, et al. Introduction to the recommendations from the National Institute on Aging-Alzheimer's Association workgroups on diagnostic guidelines for Alzheimer's disease. Alzheimers Dement. 2011;7:257-262.
    Pubmed CrossRef
  150. Lee KK, Salamon N. [18F] fluorodeoxyglucose-positron-emission tomography and MR imaging coregistration for presurgical evaluation of medically refractory epilepsy. AJNR Am J Neuroradiol. 2009;30:1811-1816.
    Pubmed KoreaMed CrossRef
  151. Kransdorf MJ, Bridges MD. Current developments and recent advances in musculoskeletal tumor imaging. Semin Musculoskelet Radiol. 2013;17:145-155.
    Pubmed CrossRef
  152. Abgral R, Dweck MR, Trivieri MG, Robson PM, Karakatsanis N, Mani V, et al. Clinical utility of combined FDG-PET/MR to assess myocardial disease. JACC Cardiovasc Imaging. 2017;10:594-597.
    Pubmed KoreaMed CrossRef
  153. Dubinsky TJ, Cuevas C, Dighe MK, Kolokythas O, Hwang JH. High-intensity focused ultrasound: current potential and oncologic applications. AJR Am J Roentgenol. 2008;190:191-199.
    Pubmed CrossRef
  154. Cline HE, Schenck JF, Hynynen K, Watkins RD, Souza SP, Jolesz FA. MR-guided focused ultrasound surgery. J Comput Assist Tomogr. 1992;16:956-965.
    Pubmed KoreaMed CrossRef
  155. Hynynen K, Darkazanli A, Unger E, Schenck JF. MRI-guided noninvasive ultrasound surgery. Med Phys. 1993;20:107-115.
    Pubmed CrossRef
  156. Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, et al. A precise and fast temperature mapping using water proton chemical shift. Magn Reson Med. 1995;34:814-823.
    Pubmed CrossRef
  157. Rieke V, Butts Pauly K. MR thermometry. J Magn Reson Imaging. 2008;27:376-390.
    Pubmed CrossRef
  158. Lénárd ZM, McDannold NJ, Fennessy FM, Stewart EA, Jolesz FA, Hynynen K, et al. Uterine leiomyomas: MR imaging-guided focused ultrasound surgery--imaging predictors of success. Radiology. 2008;249:187-194.
    Pubmed KoreaMed CrossRef
  159. Gianfelice D, Gupta C, Kucharczyk W, Bret P, Havill D, Clemons M. Palliative treatment of painful bone metastases with MR imaging--guided focused ultrasound. Radiology. 2008;249:355-363.
    Pubmed CrossRef
  160. Benedict SH, De Meerleer G, Orton CG, Stancanello J. Point/counterpoint. High intensity focused ultrasound may be superior to radiation therapy for the treatment of early stage prostate cancer. Med Phys. 2011;38:3909-3912.
    Pubmed CrossRef
  161. Hynynen K, Pomeroy O, Smith DN, Huber PE, McDannold NJ, Kettenbach J, et al. MR imaging-guided focused ultrasound surgery of fibroadenomas in the breast: a feasibility study. Radiology. 2001;219:176-185.
    Pubmed CrossRef
  162. Park J, Aryal M, Vykhodtseva N, Zhang YZ, McDannold N. Evaluation of permeability, doxorubicin delivery, and drug retention in a rat brain tumor model after ultrasound-induced blood-tumor barrier disruption. J Control Release. 2017;250:77-85.
    Pubmed KoreaMed CrossRef
  163. Lipsman N, Schwartz ML, Huang Y, Lee L, Sankar T, Chapman M, et al. MR-guided focused ultrasound thalamotomy for essential tremor: a proof-of-concept study. Lancet Neurol. 2013;12:462-468.
    Pubmed CrossRef
  164. Magara A, Bühler R, Moser D, Kowalski M, Pourtehrani P, Jeanmonod D. First experience with MR-guided focused ultrasound in the treatment of Parkinson's disease. J Ther Ultrasound. 2014;2:11.
    Pubmed CrossRef
  165. Jung HH, Kim SJ, Roh D, Chang JG, Chang WS, Kweon EJ, et al. Bilateral thermal capsulotomy with MR-guided focused ultrasound for patients with treatment-refractory obsessive-compulsive disorder: a proof-of-concept study. Mol Psychiatry. 2015;20:1205-1211.
    Pubmed CrossRef
  166. Kim M, Kim CH, Jung HH, Kim SJ, Chang JW. Treatment of major depressive disorder via magnetic resonance-guided focused ultrasound surgery. Biol Psychiatry. 2018;83:e17-e18.
    Pubmed CrossRef
  167. Jeanmonod D, Werner B, Morel A, Michels L, Zadicario E, Schiff G, et al. Transcranial magnetic resonance imaging-guided focused ultrasound: noninvasive central lateral thalamotomy for chronic neuropathic pain. Neurosurg Focus. 2012;32:E1.
    Pubmed CrossRef
  168. Saenz DL, Yan Y, Christensen N, Henzler MA, Forrest LJ, Bayouth JE, et al. Characterization of a 0.35T MR system for phantom image quality stability and in vivo assessment of motion quantification. J Appl Clin Med Phys. 2015;16:30-40.
    Pubmed KoreaMed CrossRef
  169. Yun J, St Aubin J, Rathee S, Fallone BG. Brushed permanent magnet DC MLC motor operation in an external magnetic field. Med Phys. 2010;37:2131-2134.
    Pubmed CrossRef
  170. St Aubin J, Steciw S, Fallone BG. Effect of transverse magnetic fields on a simulated in-line 6 MV linac. Phys Med Biol. 2010;55:4861-4869.
    Pubmed CrossRef
  171. St Aubin J, Santos DM, Steciw S, Fallone BG. Effect of longitudinal magnetic fields on a simulated in-line 6 MV linac. Med Phys. 2010;37:4916-4923.
    Pubmed CrossRef
  172. Bielajew AF. The effect of strong longitudinal magnetic fields on dose deposition from electron and photon beams. Med Phys. 1993;20:1171-1179.
    Pubmed CrossRef
  173. Raaijmakers AJ, Raaymakers BW, Lagendijk JJ. Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons. Phys Med Biol. 2005;50:1363-1376.
    Pubmed CrossRef
  174. Liney GP, Whelan B, Oborn B, Barton M, Keall P. MRI-linear accelerator radiotherapy systems. Clin Oncol (R Coll Radiol). 2018;30:686-691.
    Pubmed CrossRef
  175. Emmerich J, Laun FB, Pfaffenberger A, Schilling R, Denoix M, Maier F, et al. Technical note: on the size of susceptibility-induced MR image distortions in prostate and cervix in the context of MR-guided radiation therapy. Med Phys. 2018;45:1586-1593.
    Pubmed CrossRef
  176. Liney GP, Dong B, Begg J, Vial P, Zhang K, Lee F, et al. Technical note: experimental results from a prototype high-field inline MRI-linac. Med Phys. 2016;43:5188.
    Pubmed CrossRef
  177. Burke B, Wachowicz K, Fallone BG, Rathee S. Effect of radiation induced current on the quality of MR images in an integrated linac-MR system. Med Phys. 2012;39:6139-6147.
    Pubmed CrossRef
  178. Palacios MA, Bohoudi O, Bruynzeel AME, van Sörsen de Koste JR, Cobussen P, Slotman BJ, et al. Role of daily plan adaptation in MR-guided stereotactic ablative radiation therapy for adrenal metastases. Int J Radiat Oncol Biol Phys. 2018;102:426-433.
    Pubmed CrossRef
  179. Kontaxis C, Bol GH, Lagendijk JJ, Raaymakers BW. A new methodology for inter- and intrafraction plan adaptation for the MR-linac. Phys Med Biol. 2015;60:7485-7497.
    Pubmed CrossRef
  180. Rudra S, Jiang N, Rosenberg SA, Olsen JR, Roach MC, Wan L, et al. Using adaptive magnetic resonance image-guided radiation therapy for treatment of inoperable pancreatic cancer. Cancer Med. 2019;8:2123-2132.
    Pubmed KoreaMed CrossRef
  181. Murray J, Tree AC. Prostate cancer- advantages and disadvantages of MR-guided RT. Clin Transl Radiat Oncol. 2019;18:68-73.
    Pubmed CrossRef
  182. Werensteijn-Honingh AM, Kroon PS, Winkel D, Aalbers EM, van Asselen B, Bol GH, et al. Feasibility of stereotactic radiotherapy using a 1.5 T MR-linac: multi-fraction treatment of pelvic lymph node oligometastases. Radiother Oncol. 2019;134:50-54.
    Pubmed CrossRef
  183. Winkel D, Bol GH, Kroon PS, van Asselen B, Hackett SS, Werensteijn-Honingh AM, et al. Adaptive radiotherapy: The Elekta Unity MR-linac concept. Clin Transl Radiat Oncol. 2019;18:54-59.
    Pubmed KoreaMed CrossRef
Korean Society of Medical Physics

Vol.34 No.2
June 2023

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line