검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Split Viewer

Original Article

Progress in Medical Physics 2024; 35(3): 65-72

Published online September 30, 2024

https://doi.org/10.14316/pmp.2024.35.3.65

Copyright © Korean Society of Medical Physics.

Impact of Smaller Gantry Arc Increments on Volumetric Modulated Arc Radiation Therapy in the Monaco Treatment Planning System

Seonghee Kang1,2,3 , Hyejo Ryu4 , Do Hoon Oh4 , Lee Yoo4 , Minsoo Chun2,4,5

1Department of Radiation Oncology, Seoul National University Hospital, Seoul, 2Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, 3Biomedical Research Institute, Seoul National University Hospital, Seoul, 4Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gwangmyeong, 5Clinical Research Center, Chung-Ang University Gwang Myeong Hospital, Gwangmyeong, Korea

Correspondence to:Minsoo Chun
(ms1236@caumc.or.kr)
Tel: 82-2-2222-1860
Fax: 82-2-2222-1864

Received: August 26, 2024; Revised: August 28, 2024; Accepted: September 9, 2024

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Purpose: This study aims to evaluate the impact of smaller gantry arc increment (GAI) values on the plan quality and deliverability of volumetric modulated arc therapy (VMAT) for head and neck (HN) and prostate cancer cases using the Monaco treatment planning system. The study investigates whether a smaller GAI can enhance organ at risk (OAR) sparing without compromising target coverage or significantly increasing plan complexity.
Methods: VMAT plans were created for 20 patients (10 HN and 10 prostate cancer) using GAI values of 15° and 30°. Dose-volumetric parameters, such as conformity number, homogeneity and gradient indices, were assessed alongside plan complexity metrics like the modulation complexity score for VMAT (MCSv) and monitor unit (MU). Statistical significance was determined using the Wilcoxon signed-rank test.
Results: For HN cases, a 15° increment significantly reduced the D0.03cc for the spinal cord and the Dmean for both parotid glands compared to a 30° increment, improving OAR sparing. However, no significant differences were observed in the OAR doses for prostate cases. The 15° increment resulted in higher plan complexity, reflected by a lower MCSv, but the MU difference was not significant.
Conclusions: Smaller GAI values, such as 15°, can significantly reduce OAR doses in HN VMAT plans, offering potential clinical benefits despite increased plan complexity. However, no substantial advantages were observed in prostate cases. These findings suggest that smaller GAI values may be particularly beneficial for cases requiring high modulation.

KeywordsVolumetric modulated arc radiation therapy, Treatment planning, Prostate radiotherapy, Head and neck radiotherapy, Gantry arc increment

Volumetric modulated arc therapy (VMAT) is widely used in clinical practice as it can deliver precise dose coverage to the target while minimizing radiation exposure to the surrounding healthy tissues [1-3]. The VMAT planning process relies on an optimization algorithm within a treatment planning system (TPS) that uses an inverse planning approach. Although the optimization automatically determines the optimal beam parameters based on dose-volume constraints for targets and organs at risk (OARs), it still requires user-defined inputs such as prescription doses, collimator and couch angles, and gantry rotation settings [4,5].

In the Monaco TPS (Elekta AB), users must specify gantry arc increment (GAI) values, which define the angle for each sector of the arc [5,6]. The number of sectors is calculated by dividing the total gantry angle by the GAI. Within each sector, the multi-leaf collimator (MLC) moves in a consistent direction, reversing direction in the subsequent sector [5-8]. Increasing the number of sectors can enhance plan quality by allowing for more precise MLC modulation [7,8]. Although GAI significantly affects plan quality and treatment time, few studies have investigated the optimal GAI.

Nithya et al. [7] observed similar plan qualities with GAI values of 15°, 20°, 30°, and 40° in VMAT for patients with esophageal cancer, concluding that 40° was optimal due to its higher dose homogeneity and improved monitor unit (MU) efficiency. Similarly, Chen et al. [8] demonstrated that GAI values of 30° and 40° enhanced the plan quality and comparable deliverability in VMAT for patients with cervical cancer. However, the optimal GAI for plans requiring relatively higher degrees of modulation is poorly understood. Moreover, the impact of GAIs on treatment sites requiring different modulation levels is unclear. In this study, we examined VMAT plans for two clinical treatment sites requiring varying modulation levels: head and neck (HN) and prostate cancer, which require high and low modulation, respectively.

1. Patient selection, computed tomography scanning, and OAR contouring

Twenty patients (10 for HN, 10 for prostate) were selected following institutional review board approval (IRB No. 2405-164-068). Informed consent was waived due to the retrospective nature of the study. All patients underwent computed tomography (CT) scans using a Big Bore RT CT Simulator (Philips Health Systems) with appropriate immobilization devices and a slice thickness of 2 mm. After the scans, OARs were automatically contoured using the Contour ProtégéAITM module in MIM® software (Version 7.2.10; MIM Software Inc.). A radiation oncologist reviewed and confirmed the OAR contours and manually contoured the clinical target volume (CTV) according to established guidelines [9-14]. The planning target volume (PTV) was generated by expanding the CTV with a 3 mm margin in all directions for the HN cases and a 5 mm margin in all directions except for a 1 mm margin posteriorly for the prostate cases.

2. Treatment planning

VMAT plans were created using the Monaco TPS (Version 6.1.2; Elekta AB). Users must specify the GAI, with options ranging from 5° to 45°. Elekta recommends a 30° GAI value for general VMAT plans, which can be reduced if more modulation is needed [6]. To evaluate the effects of a smaller GAI, we compared increments of 30° and 15°, the latter being the vendor-recommended value. For the planning of HN and prostrate cases, 6 MV and 10 MV photon beams were utilized, respectively. Collimator angles of 45° and 315° were selected for each field, as a 90° difference between these angles is known to optimize the sparing of OARs, with literature showing no significant differences with other angle combinations [15,16]. A Monte Carlo dose calculation algorithm was employed with a statistical uncertainty of 1.0% per calculation and a dose grid resolution of 2 mm [4,17].

For both HN and prostate cases, the PTV was prescribed to receive 70 Gy in 35 fractions. The OAR constraints are summarized in Table 1. The plans were optimized to ensure that 95% of the PTV received 70 Gy, with the dose of 0.03 cc (D0.03cc) of the PTV not exceeding 110% of the prescribed 70 Gy. After the optimization, the plans were normalized to ensure that 70 Gy covered 95% of the PTV.

Table 1 Dose constraints used in this study

CaseOARDose-volumetric parameterConstraint
HNSpinal cordDmax<45 Gy
Brain stemDmax<54 Gy
Left parotid glandDmean<26 Gy
Right parotid glandDmean<26 Gy
ProstateBladderV50Gy<50%
V70Gy<30%
RectumV50Gy<50%
V70Gy<25%
Left femoral headV50Gy<5%
Right femoral headV50Gy<5%

3. Evaluations

The quality of the plans was assessed in terms of dose-volumetric parameters listed in Table 1. The Dmean for OARs in prostate cases was also evaluated. Further, conformity number (CN), homogeneity index (HI), and gradient index (GI) were calculated as follows [18-21]:

CN=VT,100%VT×VT,100%V100%
HI=D2%D98%D50%
GI=V50%V100%

where VT is the volume of the PTV, VT,100% is the volume covered by the prescription dose, Vx% is the volume receiving x% of the prescription dose, and Dy% is the dose received by y% of the PTV.

Plan deliverability was evaluated based on the total MUs and the modulation complexity score for VMAT (MCSv). MCSv, a plan complexity metric developed by McNiven et al. [22] for intensity-modulated radiation therapy and later adapted by Masi et al. [23] for VMAT, measures plan complexity by evaluating aperture area variability (AAV) and leaf sequence variability (LSV). First, the maximum leaf position changes for each control point (posmax(CP)) was calculated using the following formula:

posmaxCP=maxposnN minposnN leafbank

Next, LSVcp and AAVcp were calculated as follows:

LSVcp=n=1N1posmax posn posn+1 N1×posmaxleftbank×n=1N1posmax posn posn+1 N1×posmaxrightbank
AAVcp=a=1Aposa leftbankposa rightbanka=1Amaxposa leftbankarcmaxposa rightbankarc

After that, MCS for each arc (MCSarc) was calculated as:

MCSarc=i=1I1AAVcpi +AAVcpi+1 2×LSVcpi +LSVcpi+1 2×MUcpi,i+1MUarc

Finally, MCSv was calculated as:

MCSv=1Kk=1KMCSarck

A lower MCSv indicates higher plan complexity, while a higher MCSv reflects lower complexity. Therefore, MCSv is adopted as a surrogate metric for plan deliverability.

Given the relatively small sample size in this study, we employed the Wilcoxon signed-rank test for statistical analysis, which is particularly advantageous for paired samples and small sample sizes [24]. Differences with P-values <0.05 were considered statistically significant. Statistical analysis was performed using the Real Statistics Resource Pack software (Release 9.1.1) plug-in for Excel.

1. Plan quality comparisons

Dose-volumetric analysis results are summarized in Table 2. Figs. 1 and 2 show the dose distributions and dose-volume histograms for the representative cases. For HN cases, the D0.03cc, HI, and GI for the PTV were similar across the groups. However, plans using 15° GAI values demonstrated significantly better conformity than those using a 30° increment. While the D0.03cc for the brainstem was lower in the 15° plans than in the 30° plans, this difference was not statistically significant. However, the D0.03cc for the spinal cord and the Dmean for the left and right parotid glands were significantly lower in the 15° plans than those in the 30° plans.

Table 2 Dose-volumetric analysis results and statistical significances

CaseOrganDV parameter30° GAI15° GAIP-value
HNPTVD0.03cc (Gy)76.25±0.7576.48±1.030.097
CN0.85±0.040.86±0.040.019*
HI0.06±0.010.06±0.010.246
GI3.03±0.562.93±0.460.097
Spinal cordD0.03cc (Gy)23.86±6.0320.84±4.780.003*
BrainstemD0.03cc (Gy)16.18±16.3315.81±16.760.216
Left parotid glandDmean (Gy)26.79±11.2024.75±11.200.032*
Right parotid glandDmean (Gy)32.74±14.7430.27±15.180.032*
ProstatePTVD0.03cc (Gy)74.70±0.5574.86±0.380.188
CN0.84±0.020.84±0.030.278
HI0.05±0.010.05±0.000.097
GI4.75±0.544.69±0.560.246
BladderV70Gy (%)27.74±17.0627.41±17.030.053
V50Gy (%)47.62±18.9248.10±18.260.326
Dmean (Gy)46.99±9.7747.63±9.770.216
RectumV70Gy (%)2.92±1.382.55±0.960.097
V50Gy (%)19.86±5.3118.32±4.540.097
Dmean (Gy)33.51±3.7333.06±4.330.423
Left femur headV50Gy (%)0.13±0.400.25±0.790.250
Dmean (Gy)19.18±1.8018.95±2.300.385
Right femur headV50Gy (%)0.10±0.220.04±0.090.250
Dmean (Gy)17.36±2.4318.64±2.470.003*

Figure 1.Sample dose distributions for representative cases in head and neck cancer with GAIs of (a) 30° and (b) 15°, and in prostate cancer with GAIs of (c) 30° and (d) 15°. GAI, gantry arc increment; PTV, planning target volume.

Figure 2.Sample dose-volume histograms for representative cases in (a) head and neck and (b) prostate cancer. The dotted and solid lines represent plans using 30° and 15° gantry arc increments, respectively. PTV, planning target volume.

For prostate cases, the D0.03cc, CN, HI, and GI were comparable across groups with no significant differences. Additionally, none of the dose-volumetric parameters for OARs showed statistical significance.

2. MU and complexity comparisons

The statistics for MU and MCSv are summarized in Table 3. For both HN and prostate cases, the mean MU for plans with 30° and 15° increments showed no significant difference. However, the mean MCSv for plans with 30° and 15° increments exhibited a statistically significant difference (P <0.05) in both cases.

Table 3 Statistics for MU and MCSv

CaseMetric30° GAI15° GAIP-value
HNMU1,268.21±225.921,309.73±223.150.216
MCSv0.17±0.080.09±0.010.007*
ProstateMU1,320.52±382.781,367.72±338.260.138
MCSv0.29±0.130.13±0.060.003*

This study demonstrated dose-volumetric analysis and plan deliverability for HN and prostate VMAT plans using different GAIs. While no significant differences in doses to OARs and targets were observed in prostate cases, plans using a 15° increment in HN cases achieved notable dose reductions, specifically in the D0.03cc for the spinal cord and the Dmean for both parotid glands. In both cases, plans with a 15° increment had higher MUs and lower MCSv than those with a 30° increment, although only the difference in MCSv was statistically significant.

Previous studies, such as those by Nithya et al. [7], demonstrated that using a larger increment (40°) improved the plan quality with fewer MUs in VMAT for patients with esophageal cancer. Similarly, Chen et al. [8] found that while plans with a 30° increment in VMAT for patients with cervical cancer had advantages, such as shorter treatment times, dose reductions with different increments did not show significant differences. While the findings for prostate cases in this study were consistent with previous research, HN plans using smaller GAIs might significantly reduce doses to OARs, albeit with increased plan complexity. Plan complexity for VMAT in HN cases is generally greater than that in prostate cases due to the typically irregular and elongated shape of the target volume in HN cases and the proximity of multiple OARs [23]. The complexity of plans using a 15° GAI was more pronounced than those using a 30° increment in both cases. These findings indicate that VMAT plans with smaller GAIs may offer potential benefits in reducing OAR doses in cases requiring higher modulation.

This study has some limitations. First, the sample size is relatively small for each case. To address this, we used the Wilcoxon signed-rank test for statistical analysis, as it is well-suited for small sample sizes and non-normally distributed data. Due to the limited sample size, the impact of dose reduction in prostate cases may not be as apparent. However, despite this limitation, the smaller GAI still demonstrated a significant OAR-sparing effect in HN cases. Another limitation is that plan deliverability was assessed using complexity metrics rather than performing a gamma analysis based on actual delivery. Although studies have shown that various complexity metrics are highly correlated with gamma passing rates [23], directly evaluating plan deliverability through gamma analysis might benefit future research.

Despite several limitations, our results showed that OAR doses can be potentially reduced with smaller GAIs, albeit with a slight increase in complexity. While the quality of VMAT plans is often determined by user-defined dose-volumetric parameters, a thorough understanding of the physical properties of the TPS and the mechanical parameters of the linear accelerator can enhance plan quality and deliverability, offering significant clinical benefits.

VMAT plans with a smaller GAI (15°) can significantly reduce OAR doses compared to those with a 30° GAI in HN cases. However, no significant dose reduction was observed with a smaller GAI in prostate cases. Although plan complexity and total MUs may slightly increase, using a smaller GAI can potentially reduce OAR doses in the Monaco TPS.

This work was supported by Korea Institute of Energy Technology Evalutaion and Planning (KETEP) grant funded by the Korea government (MOTIE) (20227410100040, Development of patch-type flexible personal dosimeter and real-time remote monitoring system using high-performance inorganic perovskite).

Conceptualization: Minsoo Chun. Data curation: Hyejo Ryu. Formal analysis: Hyejo Ryu, Lee Yoo. Funding acquisition: Seonghee Kang. Investigation: Seonghee Kang, Do Hoon Oh, Minsoo Chun. Methodology: Minsoo Chun. Project administration: Seonghee Kang. Resources: Hyejo Ryu, Do Hoon Oh. Software: Lee Yoo. Supervision: Minsoo Chun. Validation: Hyejo Ryu, Do Hoon Oh. Visualization: Hyejo Ryu, Lee Yoo. Writing – original draft: Seonghee Kang, Minsoo Chun. Writing – review & editing: Seonghee Kang, Hyejo Ryu, Minsoo Chun.

The study was approved by the Institutional Review Board of Chung-Ang University Gwang Myeong Hospital (IRB approval number; 2405-164-068).

  1. Palma DA, Verbakel WF, Otto K, Senan S. New developments in arc radiation therapy: a review. Cancer Treat Rev. 2010;36:393-399.
    Pubmed CrossRef
  2. Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84:967-996.
    Pubmed KoreaMed CrossRef
  3. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317.
    Pubmed CrossRef
  4. Elekta AB. Monaco® XVMC and Pencil Beam Dose Calculation Reference Manual (E014439 / 02). 2024.
  5. Elekta AB. Hyperion VMAT Sequencer® Technical Reference (LRMMON0006). 2016.
  6. Elekta AB. Monaco® Training Guide (LTGMON0530). 2017.
  7. Nithya L, Raj NA, Rathinamuthu S, Sharma K, Pandey MB. Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: a planning study. J Med Phys. 2014;39:231-237.
    Pubmed KoreaMed CrossRef
  8. Chen A, Li Z, Chen L, Lin M, Li B, Chen F. The influence of increment of gantry on VMAT plan quality for cervical cancer. J Radiat Res Appl Sci. 2019;12:447-454.
    CrossRef
  9. Jensen K, Friborg J, Hansen CR, Samsøe E, Johansen J, Andersen M, et al. The Danish Head and Neck Cancer Group (DAHANCA) 2020 radiotherapy guidelines. Radiother Oncol. 2020;151:149-151.
    Pubmed CrossRef
  10. Gillison ML, Trotti AM, Harris J, Eisbruch A, Harari PM, Adelstein DJ, et al. Radiotherapy plus cetuximab or cisplatin in human papillomavirus-positive oropharyngeal cancer (NRG Oncology RTOG 1016): a randomised, multicentre, non-inferiority trial. Lancet. 2019;393:40-50.
    Pubmed KoreaMed CrossRef
  11. Machtay M, Torres-Saavedra P, Thorstad WL, Nguyen-Tan PF, Siu LL, Holsinger FC, et al. Randomized phase III trial of postoperative radiotherapy with or without cetuximab for intermediate-risk Squamous Cell Carcinoma of the Head and Neck (SCCHN): NRG/RTOG 0920. Int J Radiat Oncol Biol Phys. 2023;117:e1.
    CrossRef
  12. Li G, Li Y, Wang J, Gao X, Zhong Q, He L, et al. Guidelines for radiotherapy of prostate cancer (2020 edition). Precis Radiat Oncol. 2021;5:160-182.
    CrossRef
  13. Schaake W, van der Schaaf A, van Dijk LV, Bongaerts AH, van den Bergh AC, Langendijk JA. Normal tissue complication probability (NTCP) models for late rectal bleeding, stool frequency and fecal incontinence after radiotherapy in prostate cancer patients. Radiother Oncol. 2016;119:381-387.
    Pubmed CrossRef
  14. Chorbińska J, Krajewski W, Zdrojowy R. Urological complications after radiation therapy-nothing ventured, nothing gained: a narrative review. Transl Cancer Res. 2021;10:1096-1118.
    Pubmed KoreaMed CrossRef
  15. Li X, Wu J, Palta M, Zhang Y, Sheng Y, Zhang J, et al. A collimator setting optimization algorithm for dual-arc volumetric modulated arc therapy in pancreas stereotactic body radiation therapy. Technol Cancer Res Treat. 2019;18:1533033819870767.
    Pubmed KoreaMed CrossRef
  16. Sun W, Chen K, Li Y, Xia W, Dong L, Shi Y, et al. Optimization of collimator angles in dual-arc volumetric modulated arc therapy planning for whole-brain radiotherapy with hippocampus and inner ear sparing. Sci Rep. 2021;11:19035.
    Pubmed KoreaMed CrossRef
  17. Goodall SK, Ebert MA. Recommended dose voxel size and statistical uncertainty parameters for precision of Monte Carlo dose calculation in stereotactic radiotherapy. J Appl Clin Med Phys. 2020;21:120-130.
    Pubmed KoreaMed CrossRef
  18. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg. 2000;93(Suppl 3):219-222.
    Pubmed CrossRef
  19. International Commission on Radiation Units Measurements. ICRU report 83, prescribing, recording, and reporting photon-beam intensity-modulated radiation therapy (IMRT). J ICRU. 2010;10.
    Pubmed CrossRef
  20. Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105(Suppl):194-201.
    Pubmed CrossRef
  21. International Commission on Radiation Units Measurements. ICRU report 91, prescribing, recording, and reporting of stereotactic treatments with small photon beams. J ICRU. 2014;14.
    CrossRef
  22. McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505-515.
    Pubmed CrossRef
  23. Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718.
    Pubmed CrossRef
  24. Wilcoxon F; Individual comparisons by ranking methods. Breakthroughs in statistics: methodology and distribution. New York: Springer; 1992:196-202.
    CrossRef

Article

Original Article

Progress in Medical Physics 2024; 35(3): 65-72

Published online September 30, 2024 https://doi.org/10.14316/pmp.2024.35.3.65

Copyright © Korean Society of Medical Physics.

Impact of Smaller Gantry Arc Increments on Volumetric Modulated Arc Radiation Therapy in the Monaco Treatment Planning System

Seonghee Kang1,2,3 , Hyejo Ryu4 , Do Hoon Oh4 , Lee Yoo4 , Minsoo Chun2,4,5

1Department of Radiation Oncology, Seoul National University Hospital, Seoul, 2Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, 3Biomedical Research Institute, Seoul National University Hospital, Seoul, 4Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gwangmyeong, 5Clinical Research Center, Chung-Ang University Gwang Myeong Hospital, Gwangmyeong, Korea

Correspondence to:Minsoo Chun
(ms1236@caumc.or.kr)
Tel: 82-2-2222-1860
Fax: 82-2-2222-1864

Received: August 26, 2024; Revised: August 28, 2024; Accepted: September 9, 2024

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose: This study aims to evaluate the impact of smaller gantry arc increment (GAI) values on the plan quality and deliverability of volumetric modulated arc therapy (VMAT) for head and neck (HN) and prostate cancer cases using the Monaco treatment planning system. The study investigates whether a smaller GAI can enhance organ at risk (OAR) sparing without compromising target coverage or significantly increasing plan complexity.
Methods: VMAT plans were created for 20 patients (10 HN and 10 prostate cancer) using GAI values of 15° and 30°. Dose-volumetric parameters, such as conformity number, homogeneity and gradient indices, were assessed alongside plan complexity metrics like the modulation complexity score for VMAT (MCSv) and monitor unit (MU). Statistical significance was determined using the Wilcoxon signed-rank test.
Results: For HN cases, a 15° increment significantly reduced the D0.03cc for the spinal cord and the Dmean for both parotid glands compared to a 30° increment, improving OAR sparing. However, no significant differences were observed in the OAR doses for prostate cases. The 15° increment resulted in higher plan complexity, reflected by a lower MCSv, but the MU difference was not significant.
Conclusions: Smaller GAI values, such as 15°, can significantly reduce OAR doses in HN VMAT plans, offering potential clinical benefits despite increased plan complexity. However, no substantial advantages were observed in prostate cases. These findings suggest that smaller GAI values may be particularly beneficial for cases requiring high modulation.

Keywords: Volumetric modulated arc radiation therapy, Treatment planning, Prostate radiotherapy, Head and neck radiotherapy, Gantry arc increment

Introduction

Volumetric modulated arc therapy (VMAT) is widely used in clinical practice as it can deliver precise dose coverage to the target while minimizing radiation exposure to the surrounding healthy tissues [1-3]. The VMAT planning process relies on an optimization algorithm within a treatment planning system (TPS) that uses an inverse planning approach. Although the optimization automatically determines the optimal beam parameters based on dose-volume constraints for targets and organs at risk (OARs), it still requires user-defined inputs such as prescription doses, collimator and couch angles, and gantry rotation settings [4,5].

In the Monaco TPS (Elekta AB), users must specify gantry arc increment (GAI) values, which define the angle for each sector of the arc [5,6]. The number of sectors is calculated by dividing the total gantry angle by the GAI. Within each sector, the multi-leaf collimator (MLC) moves in a consistent direction, reversing direction in the subsequent sector [5-8]. Increasing the number of sectors can enhance plan quality by allowing for more precise MLC modulation [7,8]. Although GAI significantly affects plan quality and treatment time, few studies have investigated the optimal GAI.

Nithya et al. [7] observed similar plan qualities with GAI values of 15°, 20°, 30°, and 40° in VMAT for patients with esophageal cancer, concluding that 40° was optimal due to its higher dose homogeneity and improved monitor unit (MU) efficiency. Similarly, Chen et al. [8] demonstrated that GAI values of 30° and 40° enhanced the plan quality and comparable deliverability in VMAT for patients with cervical cancer. However, the optimal GAI for plans requiring relatively higher degrees of modulation is poorly understood. Moreover, the impact of GAIs on treatment sites requiring different modulation levels is unclear. In this study, we examined VMAT plans for two clinical treatment sites requiring varying modulation levels: head and neck (HN) and prostate cancer, which require high and low modulation, respectively.

Materials and Methods

1. Patient selection, computed tomography scanning, and OAR contouring

Twenty patients (10 for HN, 10 for prostate) were selected following institutional review board approval (IRB No. 2405-164-068). Informed consent was waived due to the retrospective nature of the study. All patients underwent computed tomography (CT) scans using a Big Bore RT CT Simulator (Philips Health Systems) with appropriate immobilization devices and a slice thickness of 2 mm. After the scans, OARs were automatically contoured using the Contour ProtégéAITM module in MIM® software (Version 7.2.10; MIM Software Inc.). A radiation oncologist reviewed and confirmed the OAR contours and manually contoured the clinical target volume (CTV) according to established guidelines [9-14]. The planning target volume (PTV) was generated by expanding the CTV with a 3 mm margin in all directions for the HN cases and a 5 mm margin in all directions except for a 1 mm margin posteriorly for the prostate cases.

2. Treatment planning

VMAT plans were created using the Monaco TPS (Version 6.1.2; Elekta AB). Users must specify the GAI, with options ranging from 5° to 45°. Elekta recommends a 30° GAI value for general VMAT plans, which can be reduced if more modulation is needed [6]. To evaluate the effects of a smaller GAI, we compared increments of 30° and 15°, the latter being the vendor-recommended value. For the planning of HN and prostrate cases, 6 MV and 10 MV photon beams were utilized, respectively. Collimator angles of 45° and 315° were selected for each field, as a 90° difference between these angles is known to optimize the sparing of OARs, with literature showing no significant differences with other angle combinations [15,16]. A Monte Carlo dose calculation algorithm was employed with a statistical uncertainty of 1.0% per calculation and a dose grid resolution of 2 mm [4,17].

For both HN and prostate cases, the PTV was prescribed to receive 70 Gy in 35 fractions. The OAR constraints are summarized in Table 1. The plans were optimized to ensure that 95% of the PTV received 70 Gy, with the dose of 0.03 cc (D0.03cc) of the PTV not exceeding 110% of the prescribed 70 Gy. After the optimization, the plans were normalized to ensure that 70 Gy covered 95% of the PTV.

Table 1 . Dose constraints used in this study.

CaseOARDose-volumetric parameterConstraint
HNSpinal cordDmax<45 Gy
Brain stemDmax<54 Gy
Left parotid glandDmean<26 Gy
Right parotid glandDmean<26 Gy
ProstateBladderV50Gy<50%
V70Gy<30%
RectumV50Gy<50%
V70Gy<25%
Left femoral headV50Gy<5%
Right femoral headV50Gy<5%


3. Evaluations

The quality of the plans was assessed in terms of dose-volumetric parameters listed in Table 1. The Dmean for OARs in prostate cases was also evaluated. Further, conformity number (CN), homogeneity index (HI), and gradient index (GI) were calculated as follows [18-21]:

CN=VT,100%VT×VT,100%V100%
HI=D2%D98%D50%
GI=V50%V100%

where VT is the volume of the PTV, VT,100% is the volume covered by the prescription dose, Vx% is the volume receiving x% of the prescription dose, and Dy% is the dose received by y% of the PTV.

Plan deliverability was evaluated based on the total MUs and the modulation complexity score for VMAT (MCSv). MCSv, a plan complexity metric developed by McNiven et al. [22] for intensity-modulated radiation therapy and later adapted by Masi et al. [23] for VMAT, measures plan complexity by evaluating aperture area variability (AAV) and leaf sequence variability (LSV). First, the maximum leaf position changes for each control point (posmax(CP)) was calculated using the following formula:

posmaxCP=maxposnN minposnN leafbank

Next, LSVcp and AAVcp were calculated as follows:

LSVcp=n=1N1posmax posn posn+1 N1×posmaxleftbank×n=1N1posmax posn posn+1 N1×posmaxrightbank
AAVcp=a=1Aposa leftbankposa rightbanka=1Amaxposa leftbankarcmaxposa rightbankarc

After that, MCS for each arc (MCSarc) was calculated as:

MCSarc=i=1I1AAVcpi +AAVcpi+1 2×LSVcpi +LSVcpi+1 2×MUcpi,i+1MUarc

Finally, MCSv was calculated as:

MCSv=1Kk=1KMCSarck

A lower MCSv indicates higher plan complexity, while a higher MCSv reflects lower complexity. Therefore, MCSv is adopted as a surrogate metric for plan deliverability.

Given the relatively small sample size in this study, we employed the Wilcoxon signed-rank test for statistical analysis, which is particularly advantageous for paired samples and small sample sizes [24]. Differences with P-values <0.05 were considered statistically significant. Statistical analysis was performed using the Real Statistics Resource Pack software (Release 9.1.1) plug-in for Excel.

Results

1. Plan quality comparisons

Dose-volumetric analysis results are summarized in Table 2. Figs. 1 and 2 show the dose distributions and dose-volume histograms for the representative cases. For HN cases, the D0.03cc, HI, and GI for the PTV were similar across the groups. However, plans using 15° GAI values demonstrated significantly better conformity than those using a 30° increment. While the D0.03cc for the brainstem was lower in the 15° plans than in the 30° plans, this difference was not statistically significant. However, the D0.03cc for the spinal cord and the Dmean for the left and right parotid glands were significantly lower in the 15° plans than those in the 30° plans.

Table 2 . Dose-volumetric analysis results and statistical significances.

CaseOrganDV parameter30° GAI15° GAIP-value
HNPTVD0.03cc (Gy)76.25±0.7576.48±1.030.097
CN0.85±0.040.86±0.040.019*
HI0.06±0.010.06±0.010.246
GI3.03±0.562.93±0.460.097
Spinal cordD0.03cc (Gy)23.86±6.0320.84±4.780.003*
BrainstemD0.03cc (Gy)16.18±16.3315.81±16.760.216
Left parotid glandDmean (Gy)26.79±11.2024.75±11.200.032*
Right parotid glandDmean (Gy)32.74±14.7430.27±15.180.032*
ProstatePTVD0.03cc (Gy)74.70±0.5574.86±0.380.188
CN0.84±0.020.84±0.030.278
HI0.05±0.010.05±0.000.097
GI4.75±0.544.69±0.560.246
BladderV70Gy (%)27.74±17.0627.41±17.030.053
V50Gy (%)47.62±18.9248.10±18.260.326
Dmean (Gy)46.99±9.7747.63±9.770.216
RectumV70Gy (%)2.92±1.382.55±0.960.097
V50Gy (%)19.86±5.3118.32±4.540.097
Dmean (Gy)33.51±3.7333.06±4.330.423
Left femur headV50Gy (%)0.13±0.400.25±0.790.250
Dmean (Gy)19.18±1.8018.95±2.300.385
Right femur headV50Gy (%)0.10±0.220.04±0.090.250
Dmean (Gy)17.36±2.4318.64±2.470.003*


Figure 1. Sample dose distributions for representative cases in head and neck cancer with GAIs of (a) 30° and (b) 15°, and in prostate cancer with GAIs of (c) 30° and (d) 15°. GAI, gantry arc increment; PTV, planning target volume.

Figure 2. Sample dose-volume histograms for representative cases in (a) head and neck and (b) prostate cancer. The dotted and solid lines represent plans using 30° and 15° gantry arc increments, respectively. PTV, planning target volume.

For prostate cases, the D0.03cc, CN, HI, and GI were comparable across groups with no significant differences. Additionally, none of the dose-volumetric parameters for OARs showed statistical significance.

2. MU and complexity comparisons

The statistics for MU and MCSv are summarized in Table 3. For both HN and prostate cases, the mean MU for plans with 30° and 15° increments showed no significant difference. However, the mean MCSv for plans with 30° and 15° increments exhibited a statistically significant difference (P <0.05) in both cases.

Table 3 . Statistics for MU and MCSv.

CaseMetric30° GAI15° GAIP-value
HNMU1,268.21±225.921,309.73±223.150.216
MCSv0.17±0.080.09±0.010.007*
ProstateMU1,320.52±382.781,367.72±338.260.138
MCSv0.29±0.130.13±0.060.003*

Discussion

This study demonstrated dose-volumetric analysis and plan deliverability for HN and prostate VMAT plans using different GAIs. While no significant differences in doses to OARs and targets were observed in prostate cases, plans using a 15° increment in HN cases achieved notable dose reductions, specifically in the D0.03cc for the spinal cord and the Dmean for both parotid glands. In both cases, plans with a 15° increment had higher MUs and lower MCSv than those with a 30° increment, although only the difference in MCSv was statistically significant.

Previous studies, such as those by Nithya et al. [7], demonstrated that using a larger increment (40°) improved the plan quality with fewer MUs in VMAT for patients with esophageal cancer. Similarly, Chen et al. [8] found that while plans with a 30° increment in VMAT for patients with cervical cancer had advantages, such as shorter treatment times, dose reductions with different increments did not show significant differences. While the findings for prostate cases in this study were consistent with previous research, HN plans using smaller GAIs might significantly reduce doses to OARs, albeit with increased plan complexity. Plan complexity for VMAT in HN cases is generally greater than that in prostate cases due to the typically irregular and elongated shape of the target volume in HN cases and the proximity of multiple OARs [23]. The complexity of plans using a 15° GAI was more pronounced than those using a 30° increment in both cases. These findings indicate that VMAT plans with smaller GAIs may offer potential benefits in reducing OAR doses in cases requiring higher modulation.

This study has some limitations. First, the sample size is relatively small for each case. To address this, we used the Wilcoxon signed-rank test for statistical analysis, as it is well-suited for small sample sizes and non-normally distributed data. Due to the limited sample size, the impact of dose reduction in prostate cases may not be as apparent. However, despite this limitation, the smaller GAI still demonstrated a significant OAR-sparing effect in HN cases. Another limitation is that plan deliverability was assessed using complexity metrics rather than performing a gamma analysis based on actual delivery. Although studies have shown that various complexity metrics are highly correlated with gamma passing rates [23], directly evaluating plan deliverability through gamma analysis might benefit future research.

Despite several limitations, our results showed that OAR doses can be potentially reduced with smaller GAIs, albeit with a slight increase in complexity. While the quality of VMAT plans is often determined by user-defined dose-volumetric parameters, a thorough understanding of the physical properties of the TPS and the mechanical parameters of the linear accelerator can enhance plan quality and deliverability, offering significant clinical benefits.

Conclusions

VMAT plans with a smaller GAI (15°) can significantly reduce OAR doses compared to those with a 30° GAI in HN cases. However, no significant dose reduction was observed with a smaller GAI in prostate cases. Although plan complexity and total MUs may slightly increase, using a smaller GAI can potentially reduce OAR doses in the Monaco TPS.

Funding

This work was supported by Korea Institute of Energy Technology Evalutaion and Planning (KETEP) grant funded by the Korea government (MOTIE) (20227410100040, Development of patch-type flexible personal dosimeter and real-time remote monitoring system using high-performance inorganic perovskite).

Conflicts of Interest

The authors have nothing to disclose.

Availability of Data and Materials

All relevant data are within the paper and its Supporting Information files.

Author Contributions

Conceptualization: Minsoo Chun. Data curation: Hyejo Ryu. Formal analysis: Hyejo Ryu, Lee Yoo. Funding acquisition: Seonghee Kang. Investigation: Seonghee Kang, Do Hoon Oh, Minsoo Chun. Methodology: Minsoo Chun. Project administration: Seonghee Kang. Resources: Hyejo Ryu, Do Hoon Oh. Software: Lee Yoo. Supervision: Minsoo Chun. Validation: Hyejo Ryu, Do Hoon Oh. Visualization: Hyejo Ryu, Lee Yoo. Writing – original draft: Seonghee Kang, Minsoo Chun. Writing – review & editing: Seonghee Kang, Hyejo Ryu, Minsoo Chun.

Ethics Approval and Consent to Participate

The study was approved by the Institutional Review Board of Chung-Ang University Gwang Myeong Hospital (IRB approval number; 2405-164-068).

Fig 1.

Figure 1.Sample dose distributions for representative cases in head and neck cancer with GAIs of (a) 30° and (b) 15°, and in prostate cancer with GAIs of (c) 30° and (d) 15°. GAI, gantry arc increment; PTV, planning target volume.
Progress in Medical Physics 2024; 35: 65-72https://doi.org/10.14316/pmp.2024.35.3.65

Fig 2.

Figure 2.Sample dose-volume histograms for representative cases in (a) head and neck and (b) prostate cancer. The dotted and solid lines represent plans using 30° and 15° gantry arc increments, respectively. PTV, planning target volume.
Progress in Medical Physics 2024; 35: 65-72https://doi.org/10.14316/pmp.2024.35.3.65

Table 1 Dose constraints used in this study

CaseOARDose-volumetric parameterConstraint
HNSpinal cordDmax<45 Gy
Brain stemDmax<54 Gy
Left parotid glandDmean<26 Gy
Right parotid glandDmean<26 Gy
ProstateBladderV50Gy<50%
V70Gy<30%
RectumV50Gy<50%
V70Gy<25%
Left femoral headV50Gy<5%
Right femoral headV50Gy<5%

Table 2 Dose-volumetric analysis results and statistical significances

CaseOrganDV parameter30° GAI15° GAIP-value
HNPTVD0.03cc (Gy)76.25±0.7576.48±1.030.097
CN0.85±0.040.86±0.040.019*
HI0.06±0.010.06±0.010.246
GI3.03±0.562.93±0.460.097
Spinal cordD0.03cc (Gy)23.86±6.0320.84±4.780.003*
BrainstemD0.03cc (Gy)16.18±16.3315.81±16.760.216
Left parotid glandDmean (Gy)26.79±11.2024.75±11.200.032*
Right parotid glandDmean (Gy)32.74±14.7430.27±15.180.032*
ProstatePTVD0.03cc (Gy)74.70±0.5574.86±0.380.188
CN0.84±0.020.84±0.030.278
HI0.05±0.010.05±0.000.097
GI4.75±0.544.69±0.560.246
BladderV70Gy (%)27.74±17.0627.41±17.030.053
V50Gy (%)47.62±18.9248.10±18.260.326
Dmean (Gy)46.99±9.7747.63±9.770.216
RectumV70Gy (%)2.92±1.382.55±0.960.097
V50Gy (%)19.86±5.3118.32±4.540.097
Dmean (Gy)33.51±3.7333.06±4.330.423
Left femur headV50Gy (%)0.13±0.400.25±0.790.250
Dmean (Gy)19.18±1.8018.95±2.300.385
Right femur headV50Gy (%)0.10±0.220.04±0.090.250
Dmean (Gy)17.36±2.4318.64±2.470.003*

Table 3 Statistics for MU and MCSv

CaseMetric30° GAI15° GAIP-value
HNMU1,268.21±225.921,309.73±223.150.216
MCSv0.17±0.080.09±0.010.007*
ProstateMU1,320.52±382.781,367.72±338.260.138
MCSv0.29±0.130.13±0.060.003*

References

  1. Palma DA, Verbakel WF, Otto K, Senan S. New developments in arc radiation therapy: a review. Cancer Treat Rev. 2010;36:393-399.
    Pubmed CrossRef
  2. Teoh M, Clark CH, Wood K, Whitaker S, Nisbet A. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol. 2011;84:967-996.
    Pubmed KoreaMed CrossRef
  3. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317.
    Pubmed CrossRef
  4. Elekta AB. Monaco® XVMC and Pencil Beam Dose Calculation Reference Manual (E014439 / 02). 2024.
  5. Elekta AB. Hyperion VMAT Sequencer® Technical Reference (LRMMON0006). 2016.
  6. Elekta AB. Monaco® Training Guide (LTGMON0530). 2017.
  7. Nithya L, Raj NA, Rathinamuthu S, Sharma K, Pandey MB. Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: a planning study. J Med Phys. 2014;39:231-237.
    Pubmed KoreaMed CrossRef
  8. Chen A, Li Z, Chen L, Lin M, Li B, Chen F. The influence of increment of gantry on VMAT plan quality for cervical cancer. J Radiat Res Appl Sci. 2019;12:447-454.
    CrossRef
  9. Jensen K, Friborg J, Hansen CR, Samsøe E, Johansen J, Andersen M, et al. The Danish Head and Neck Cancer Group (DAHANCA) 2020 radiotherapy guidelines. Radiother Oncol. 2020;151:149-151.
    Pubmed CrossRef
  10. Gillison ML, Trotti AM, Harris J, Eisbruch A, Harari PM, Adelstein DJ, et al. Radiotherapy plus cetuximab or cisplatin in human papillomavirus-positive oropharyngeal cancer (NRG Oncology RTOG 1016): a randomised, multicentre, non-inferiority trial. Lancet. 2019;393:40-50.
    Pubmed KoreaMed CrossRef
  11. Machtay M, Torres-Saavedra P, Thorstad WL, Nguyen-Tan PF, Siu LL, Holsinger FC, et al. Randomized phase III trial of postoperative radiotherapy with or without cetuximab for intermediate-risk Squamous Cell Carcinoma of the Head and Neck (SCCHN): NRG/RTOG 0920. Int J Radiat Oncol Biol Phys. 2023;117:e1.
    CrossRef
  12. Li G, Li Y, Wang J, Gao X, Zhong Q, He L, et al. Guidelines for radiotherapy of prostate cancer (2020 edition). Precis Radiat Oncol. 2021;5:160-182.
    CrossRef
  13. Schaake W, van der Schaaf A, van Dijk LV, Bongaerts AH, van den Bergh AC, Langendijk JA. Normal tissue complication probability (NTCP) models for late rectal bleeding, stool frequency and fecal incontinence after radiotherapy in prostate cancer patients. Radiother Oncol. 2016;119:381-387.
    Pubmed CrossRef
  14. Chorbińska J, Krajewski W, Zdrojowy R. Urological complications after radiation therapy-nothing ventured, nothing gained: a narrative review. Transl Cancer Res. 2021;10:1096-1118.
    Pubmed KoreaMed CrossRef
  15. Li X, Wu J, Palta M, Zhang Y, Sheng Y, Zhang J, et al. A collimator setting optimization algorithm for dual-arc volumetric modulated arc therapy in pancreas stereotactic body radiation therapy. Technol Cancer Res Treat. 2019;18:1533033819870767.
    Pubmed KoreaMed CrossRef
  16. Sun W, Chen K, Li Y, Xia W, Dong L, Shi Y, et al. Optimization of collimator angles in dual-arc volumetric modulated arc therapy planning for whole-brain radiotherapy with hippocampus and inner ear sparing. Sci Rep. 2021;11:19035.
    Pubmed KoreaMed CrossRef
  17. Goodall SK, Ebert MA. Recommended dose voxel size and statistical uncertainty parameters for precision of Monte Carlo dose calculation in stereotactic radiotherapy. J Appl Clin Med Phys. 2020;21:120-130.
    Pubmed KoreaMed CrossRef
  18. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg. 2000;93(Suppl 3):219-222.
    Pubmed CrossRef
  19. International Commission on Radiation Units Measurements. ICRU report 83, prescribing, recording, and reporting photon-beam intensity-modulated radiation therapy (IMRT). J ICRU. 2010;10.
    Pubmed CrossRef
  20. Paddick I, Lippitz B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg. 2006;105(Suppl):194-201.
    Pubmed CrossRef
  21. International Commission on Radiation Units Measurements. ICRU report 91, prescribing, recording, and reporting of stereotactic treatments with small photon beams. J ICRU. 2014;14.
    CrossRef
  22. McNiven AL, Sharpe MB, Purdie TG. A new metric for assessing IMRT modulation complexity and plan deliverability. Med Phys. 2010;37:505-515.
    Pubmed CrossRef
  23. Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718.
    Pubmed CrossRef
  24. Wilcoxon F; Individual comparisons by ranking methods. Breakthroughs in statistics: methodology and distribution. New York: Springer; 1992:196-202.
    CrossRef
Korean Society of Medical Physics

Vol.35 No.4
December 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line