검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Original Article

Korean Journal of Medical Physics 2004; 15(3): 140-148

Published online September 25, 2004

Copyright © Korean Society of Medical Physics.

Absorbed Dose Analysis in Water for Proton Beam using PTRAN Code System

PTRAN 코드를 이용한 양성자선에 대한 물 흡수선량의 해석

Dong Hyeok Jeong, and Jin Young Kim*

정동혁·김진영*

Department of Radiation Oncology, Wonkwang University Hospital, *Department of Physics, Kunsan National University

원광대학병원 방사선종양학과, *군산대학교 물리학과

Abstract

The absorbed dose for proton beam in water was calculated using the PTRAN code system. The proton interactions with water and the description on absorbed dose calculations are discussed, and the file structure and an execution example of the PTRAN codes are described. For 60, 100, 150, 200, and 250 MeV proton beams it is demonstrated that the absorbed dose is determined from the sum of Coulomb interactions and nuclear reactions, and that the Bragg peak feature depends on the energy straggling and multiple scattering. The PTRAN code was useful for studying the fundamental mechanism of the absorbed dose to water for clinical proton beams.

KeywordsProton beam, Absorbed dose, PTRAN code

Korean Society of Medical Physics

Vol.35 No.2
June 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line