Korean Journal of Medical Physics 2004; 15(3): 167-172
Published online September 25, 2004
Copyright © Korean Society of Medical Physics.
Byung-Chul Cho, Hee-Chul Park, and Hoonsik Bae
조병철·박희철·배훈식
A Siemens HD-270 MLCTM, a virtual-micro MLC, allows to the application of a smooth field edge method due to the finite leaf-width of MLC. This technique was implemented into a Pinnacle planning system in order to evaluate the dose distributions during the planning stage. The necessary dosimetric aspects, such as undulation and effective penumbra, were investigated with variations in the resolutions of a virtual-micro MLC and field edge angle. The positional accuracy of the couch movement was also assessed for clinical implementation. The overhead time for planning and treatment was confirmed as negligible.
KeywordsVirtual-micro MLC, Isodose undulation, Effective penumbra, Treatment planning