Korean Journal of Medical Physics 2010; 21(3): 304-310
Published online September 25, 2010
Copyright © Korean Society of Medical Physics.
Sei-Kwon Kang, Soah Park, Taejin Hwang, Kwang-Ho Cheong,
Me Yeon Lee, Kyoung-Ju Kim, Do Hoon Oh, Hoonsik Bae
강세권ㆍ박소아ㆍ황태진ㆍ정광호ㆍ이미연ㆍ김경주ㆍ오도훈ㆍ배훈식
Less execution of the electron arc treatment could in large part be attributed to the lack of an adequate planning system. Unlike most linear accelerators providing the electron arc mode, no commercial planning systems for the electron arc plan are available at this time. In this work, with the expectation that an easily accessible planning system could promote electron arc therapy, a commercial planning system was commissioned and evaluated for the electron arc plan. For the electron arc plan with use of a Varian 21-EX, Pinnacle3 (ver. 7.4f), with an electron pencil beam algorithm, was commissioned in which the arc consisted of multiple static fields with a fixed beam opening. Film dosimetry and point measurements were executed for the evaluation of the computation. Beam modeling was not satisfactory with the calculation of lateral profiles. Contrary to good agreement within 1% of the calculated and measured depth profiles, the calculated lateral profiles showed underestimation compared with measurements, such that the distance-to-agreement (DTA) was 5.1 mm at a 50% dose level for 6 MeV and 6.7 mm for 12 MeV with similar results for the measured depths. Point and film measurements for the humanoid phantom revealed that the delivered dose was more than the calculation by approximately 10%. The electron arc plan, based on the pencil beam algorithm, provides qualitative information for the dose distribution. Dose verification before the treatment should be mandatory.
KeywordsElectron arc plan, Pencil beam algorithm, Plan system commissioning
pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829
Frequency: Quarterly