Original Article

Progress in Medical Physics 2021; 32(3): 70-81

Published online September 30, 2021 https://doi.org/10.14316/pmp.2021.32.3.70

Copyright © Korean Society of Medical Physics.

Dosimetric Evaluation of Low-Dose Spillage Volumes for Head and Neck Cancer Using Intensity-Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Treatment Techniques

Gourav Kumar1 , Manindra Bhushan1 , Lalit Kumar1 , Vimal Kishore2 , Kothanda Raman1 , Pawan Kumar1 , Soumitra Barik1 , Sandeep Purohit1

1Medical Physics Division & Department of Radiation Oncology, Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, 2Department of Applied Science & Humanities, Bundelkhand Institute of Engineering & Technology, Jhansi, India

Correspondence to:Lalit Kumar
(lalitk48@gmail.com)
Tel: +91-9599329464
Fax: +91-11-27051037

Received: June 5, 2021; Revised: August 5, 2021; Accepted: August 10, 2021

Purpose: This study was designed to investigate the dosimetric difference between intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) in head and neck cancer (HNC). The study primarily focuses on low-dose spillage evaluation between these two techniques.
Methods: This retrospective study involved 45 patients with HNC. The treatment plans were generated using the IMRT and VMAT techniques for all patients. Dosimetric comparisons were performed in terms of target coverage, organ-at-risk (OAR) sparing, and various parameters, including conformity index, uniformity index, homogeneity index, conformation number, low-dose volumes, and normal tissue integral dose (NTID).
Results: No significant (P>0.05) difference in planning target volume coverage (D95%) was observed between IMRT and VMAT plans for supraglottic larynx, hard palate, and tongue cancers. A decrease in dose volumes ranging from 1 Gy to 30 Gy was observed for VMAT plans compared with those for IMRT plans, except for V1Gy and V30Gy for supraglottic larynx cancer and V1Gy for tongue cancer. Moreover, decreases (P<0.05) in NTID were observed for VMAT plans compared with that for IMRT plans in supraglottic larynx (4.50%), hard palate (12.80%), and tongue (7.76%) cancers. In contrast, a slight increase in monitor units for VMAT compared with those for IMRT in supraglottic larynx (0.46%), hard palate (2.54%), and tongue (7.56%) cancers.
Conclusions: For advanced-stage HNC, both IMRT and VMAT offer satisfactory clinical plans. VMAT offers a conformal and homogeneous dose distribution with comparable OAR sparing and higher dose falloff outside the target volume than IMRT, which provides an edge to reduce the risk of secondary malignancies for HNC over IMRT.

KeywordsIntensity-modulated radiation therapy, Volumetric modulated arc therapy, Head and neck cancer, Low-dose spillage, Radiotherapy

Head and neck cancers (HNCs) are the most common type of malignancy spreading worldwide. In India, among all cancer cases reported every year, 35% are HNC cases [1,2]. Most HNC malignancies include tumors of the tongue, nasopharynx, oropharynx, hypopharynx, and oral cavity [3,4]. Among all HNCs, squamous cell carcinoma (SCC) constitutes more than 90% of the cases. The widespread use of tobacco and smoking are considered the most important factors for developing these malignancies. Due to the lack of appropriate knowledge and education, most Indian patients present with the advanced stage of the disease, therefore, making radiotherapy challenging. According to the 2018 Global Cancer Incidence, Mortality and Prevalence report, there were 834,860 new HNC cases reported every year, resulting in 431,131 deaths per year [5]. Limited options are available for treatment, such as surgery, chemotherapy, and radiotherapy for early and advanced stages of HNC.

The primary aim of radiotherapy is to deliver therapeutic doses to the target while sparing the surrounding organs-at-risk (OARs) as much as possible. The introduction of intensity-modulated radiation therapy (IMRT) in HNC replaced conventional three-dimensional conformal radiotherapy (3D-CRT), resulting in better target coverage, more conformity, and better sparing of OARs [6]. Radiation-induced toxicities are low in IMRT compared with the 3D-CRT technique [7]. IMRT is primarily useful for irregularly shaped target volumes to achieve a high degree of conformal dose distribution using large numbers of smaller radiation beamlets of varying intensities [8].

With the advancement in the IMRT technique, volumetric modulated arc therapy (VMAT), a rotational treatment technology, came into existence in 2007 and now widely accepted worldwide [9]. VMAT is the evolution of IMRT incorporating simultaneous variation of three parameters, namely, the gantry speed, multileaf collimator (MLC) speed, and dose rate. Because of steeper dose gradients, both IMRT and VMAT require accurate target and normal tissue delineation. Various studies in the literature have shown that VMAT delivers a more conformal dose to planning target volumes (PTVs), results in a lower dose spillage, and has a better normal tissue-sparing ability than the IMRT technique [10-14].

With the evolution of technology, dose escalation and conformity to the target has improved. However, normal tissues outside the target get irradiated unnecessarily. It has been reported in the literature that radiation exposure to normal tissues causes secondary malignancies. The risk of secondary malignancy depends on the location and amount of exposure received by the surrounding normal tissues. Diallo et al. [15] have reported that 66% of secondary malignancies occurred within the first 5 cm from the field boundary, 22% occurred after the first 5 cm from the field boundary, and 12% occurred within the target volume.

The dose distribution generated for HNC cases using IMRT and VMAT has been enormously detailed in the literature [11,16,17]. However, distinction in low-dose distribution generated using IMRT and VMAT takes a dip for HNC cases and should be evaluated. Low-dose exposure may contribute to secondary malignancies. Therefore, this study was designed to quantify the aforementioned difference in dose distribution generated for HNC cases.

This study focused on the selection of the most appropriate treatment technique between IMRT and VMAT for patients with HNC regarding the conformity index (CI), homogeneity index (HI), conformation number (CN), uniformity index (UI), and other dosimetric parameters. Few studies have reported the effects of low doses in radiotherapy planning. Therefore, this study was conducted to determine the dosimetric difference between IMRT and VMAT in terms of low-dose spillage.

1. Patient selection, contouring, and dose prescription

This retrospective study involved 45 patients with stage II and IIIB HNC with a mean age of ≥50 years. These patients presented with SCC of the supraglottic larynx and hard palate and postoperative cancer of the tongue. All patients received radiotherapy between 2014 and 2019 as part of their standard care at our institution using the simultaneous integrated boost (SIB) technique. Computed tomography (CT) scans with 3-mm slice thickness for each patient were performed in head-first-supine position along with a standard HN thermoplast orfit mask on Somaton Sensation CT scanner (Siemens medical system, Malvern, PA, USA). After that, CT image sets were exported to the Monaco (Elekta Medical system, Crawley, UK) treatment planning system (TPS) in the Digital Imaging and Communication in Medicine format.

CT images were registered and fused with positron emission tomography and magnetic resonance imaging images to contour gross target volume. The clinical target volume and PTV were created using a margin of 12 mm and 5 mm, respectively, according to the Radiation Therapy Oncology Group recommendations. Moreover, critical OARs were delineated on the CT images, namely, parotid glands, spinal cord, brainstem, larynx, mandible, and optic apparatus.

The dose prescriptions were 70 Gy (high-risk PTV), 63 Gy (intermediate-risk PTV), and 56 Gy (low-risk PTV) in 35 fractions for the supraglottic larynx; 70 Gy (high-risk PTV) and 56 Gy (low-risk PTV) in 35 fractions for the hard palate; and 60 Gy (tumor bed) and 54 Gy (nodal volumes) in 30 fractions for tongue cancer. According to the International Commission on Radiation Units and Measurements 83 [18], treatment plans were normalized in such a way that at least 95% of the PTV should cover 100% of the prescription dose. The dose objective for OARs were followed as prescribed by the Quantitative Analyses of Normal Tissue Effects in the Clinic group. Table 1 summarizes the planning dose objectives for the PTV and OARs. Treatment plans were generated using a 6-megavolt (MV) photon beam energized using a Synergy (Elekta Medical System) linear accelerator (linac). The Synergy linac is equipped with 40 pairs of MLC leaves with a width of 1 cm at the isocenter, which could deliver a maximum dose rate of 600 cGy/min at the dose maximum (Dmax).

Table 1 Planning dose objectives for the planning target volume (PTV) and organs-at-risk

StructureParameterPlan objective
PTVD95%
D110%
D93%
=100% of prescribed dose
<10% of PTV
<1% of PTV
Spinal cordDmax<45 Gy
BrainstemDmax<54 Gy
LipsDmean<35 Gy
Contralateral parotidDmean<26 Gy
LarynxDmean<50 Gy
MandibleDmax<75 Gy

2. Intensity-modulated radiation therapy planning

Clinically acceptable dynamic MLC plans were generated on the Monaco TPS (version 5.11.02; Elekta Medical System). The Monaco TPS uses the Monte Carlo algorithm for optimization and final dose calculation, which is the gold standard presently. In Monaco, IMRT incorporates cost function-based segment shape optimization. Seven coplanar fields of 6-MV photon beam with couch and collimator angles at 0° were used to design the respective plans. All gantry angles were equidistant from each other, that is, 0°, 51°, 102°, 153°, 204°, 255°, and 306°, respectively. Additionally, a spill structure with a margin of 5 mm from the target was drawn to control low-dose spillage outside PTVs. The recommended grid spacing of 3 mm and statistical uncertainty of 2% were employed for all treatment plans. The minimum segment size of 5 mm and maximum number of control point per field of 30 were chosen for optimization and dose calculation [19].

3. Volumetric modulated arc therapy planning

VMAT plans were created in the similar fashion as IMRT plans. All cost functions and dose objectives (target and OARs) were the same as those used in IMRT optimization. In VMAT planning, we used dual-arc ranges from 181° to 179.9° in clockwise and counterclockwise directions with an arc increment angle of 20°. The maximum number of control points per arc of 100 was chosen.

4. Dosimetric evaluation

Intercomparison of both the IMRT and VMAT plans was performed using analysis of dose-volume histogram (DVH) curves. Various dosimetric parameters were evaluated for the PTV and critical OARs using DVH data. The following dose indices were evaluated:

PTV: D95% (dose to the 95% volume of the PTV), D98%, D50%, D2%, Dmean, Dmin, and Dmax.

OARs: Dmax for serial organs, i.e., spinal cord, brainstem.

Mean dose for parallel organs, that is, parotid glands, lips, and larynx.

Low-dose volume analysis: V1Gy (volume of 1 Gy outside the target), V2Gy, V3Gy, V4Gy, V5Gy, V10Gy, V15Gy, V20Gy, V25Gy, V30Gy, D1% (dose to the 1% volume outside the target), D2%, D5%.

Additionally, a ring structure was delineated to be 5 cm beyond the PTV to estimate the mean dose to the normal tissues lying in that region.

Moreover, the UI, CI, HI, conformity number (CN), and normal tissue integral dose (NTID) were computed using the following formulas:

UI=D5GyD95Gy
CI=TV95% ccTV
HI=D2Gy   D98Gy D50Gy
CN=TVRITV×TVRIVRI
NTID= mean dose×volume of normal tissues outside the PTV

where TV stands for the target volume in cm3, TVRI stands for the volume covered by the reference isodose, and VRI stands for the volume of the prescription isodose outside the target volume (inside body).

5. Statistical analysis

The dosimetric difference between IMRT and VMAT was analyzed using Statistical Package for the Social Sciences (version 20; IBM Corp., Armonk, NY, USA) in terms of the mean, standard deviation, and P-values. The independent paired t-test with a confidence interval limit of 95% was performed to assess the dosimetric endpoints for the target and OARs. P-values of less than 0.05 were used to denote statistical significance.

6. Quality assurance

Moreover, patient-specific quality assurance (PSQA) was executed using ArcCHECK (Sun Nuclear, Melbourne, FL, USA), which is a 3D cylindrical array of 1,386 equally spaced (10 mm apart) diode detectors. All treatment plans were calculated, and the same plans were delivered on ArcCHECK. SNC (version 6.2; Sun Nuclear) was used for gamma analysis. A criterion of 3-mm distance to agreement (DTA) and 3% dose difference (DD) with a threshold value of 10% was used for PSQA analysis. Additionally, 2-mm DTA and 2% DD criteria with a threshold value of 10% was examined for gamma analysis. A tolerance of ±5% was considered acceptable between the ion-chamber-measured and TPS-calculated doses [20].

Clinically acceptable treatment plans were created using the IMRT and VMAT techniques for all patients. Qualitative and quantitative analyses were performed on dose distribution created for IMRT and VMAT plans. The data were derived from cumulative DVH data for each treatment plan. Table 2 and 3 illustrate the dosimetric data on the PTV and OARs for cancers of the supraglottic larynx, hard palate, and tongue using the IMRT and VMAT techniques. Fig. 1 shows the dose distribution in axial and coronal planes using IMRT and VMAT techniques along their DVH.

Table 2 The dosimetric parameters for PTV in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

TargetParameterSupraglottic larynxHard palateTongueP-value




IMRTVMATIMRTVMATIMRTVMATSupraHPTongue
PTV70D95% (Gy)70.0070.0070.0070.00
D98% (Gy)69.20±0.1669.07±0.1569.17±0.0868.98±0.160.0110.006
D50% (Gy)72.66±0.3473.02±0.4272.48±0.2873.10±0.360.0090.001
D2% (Gy)75.38±0.5276.13±0.6175.14±0.5176.31±0.750.0190.001
Mean (Gy)72.04±1.0772.29±1.2371.99±0.3473.01±0.340.5480.045
Dmax (Gy)78.54±1.2479.73±0.9678.60±1.1780.77±0.930.0070.001
Dmin (Gy)63.31±2.1962.11±2.2562.77±1.7461.57±2.150.1310.081
UI1.07±0.011.08±0.011.06±0.011.08±0.01<0.001<0.001
HI0.072±0.010.071±0.010.081±0.010.078±0.01<0.001<0.001
CI0.99±0.030.99±0.030.99±0.011.00±0.010.7910.093
CN0.88±0.030.90±0.030.92±0.030.93±0.030.9720.667
PTV63D95% (Gy)64.01±0.9163.91±1.180.511
D98% (Gy)62.95±0.9862.67±1.450.197
D50% (Gy)68.65±1.5269.24±1.460.001
D2% (Gy)74.69±0.7975.26±0.790.006
Mean (Gy)66.06±1.0666.19±1.620.767
Dmax (Gy)74.65±1.2875.29±0.960.192
Dmin (Gy)51.02±3.5351.35±4.050.717
UI1.15±0.011.16±0.020.102
HI0.171±0.010.168±0.020.032
CI0.99±0.030.98±0.030.341
CN0.86±0.050.88±0.060.142
PTV56D95% (Gy)57.78±0.9257.82±1.0458.83±0.8359.70±1.110.7710.055
D98% (Gy)56.94±0.8256.79±1.0357.80±0.7258.25±0.940.4230.066
D50% (Gy)63.75±3.6464.28±3.7267.40±3.3368.29±2.720.0030.011
D2% (Gy)70.63±5.7171.34±5.6274.12±1.5375.13±1.770.0020.002
Mean (Gy)58.93±1.1759.04±1.4762.51±0.9164.11±1.090.7690.002
Dmax (Gy)71.35±2.5070.92±2.7375.61±1.1476.55±0.990.4070.008
Dmin (Gy)46.54±3.6045.82±2.8945.65±3.2846.42±1.790.3780.518
UI1.21±0.091.11±0.371.24±0.031.24±0.040.3910.576
HI0.218±0.080.214±0.070.232±0.020.229±0.030.0510.302
CI1.00±0.011.00±0.011.00±0.010.99±0.010.8920.271
CN0.72±0.210.75±0.210.90±0.110.92±0.12<0.0010.113
PTV60D95% (Gy)60.0060.00
D98% (Gy)58.93±0.1858.76±0.25<0.001
D50% (Gy)62.84±0.3463.51±0.45<0.001
D2% (Gy)65.44±0.5366.81±0.74<0.001
Mean (Gy)62.69±0.3163.35±0.41<0.001
Dmax (Gy)69.43±1.1371.67±1.19<0.001
Dmin (Gy)50.49±2.2749.72±3.020.871
UI1.08±0.011.1±0.01<0.001
HI0.126±0.010.123±0.01<0.001
CI0.99±0.0010.99±0.010.244
CN0.88±0.040.89±0.040.478
PTV54D95% (Gy)54.69±0.4354.48±0.530.072
D98% (Gy)53.89±0.5453.38±0.580.001
D50% (Gy)57.23±1.3157.91±1.380.001
D2% (Gy)61.50±2.7862.69±2.900.001
Mean (Gy)56.81±0.9057.08±0.590.228
Dmax (Gy)64.16±1.4264.76±0.990.046
Dmin (Gy)43.06±4.1243.93±2.890.115
UI1.10±0.051.13±0.05<0.001
HI0.161±0.050.159±0.05<0.001
CI1.00±0.0010.99±0.010.078
CN0.69±0.180.72±0.180.101

Values are presented in mean±standard deviation.

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume; UI, uniformity index; HI, homogeneity index; CI, conformity index; CN, conformation number.


Table 3 The dosimetric parameters for OARs in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

SiteOARsParameter (Gy)IMRTVMATP-value
Supraglottic larynxContra lateral parotidDmean22.76±1.2423.41±1.21<0.001
D50%18.22±2.2818.31±0.810.853
BrainstemDmax33.88±2.5736.21±2.53<0.001
Spinal cordDmax32.77±1.6333.92±1.580.009
LipsDmean32.71±3.7434.65±2.960.016
LarynxDmean49.68±1.9150.47±1.520.004
MandibleDmax75.83±2.6176.46±2.190.102
Ring (5 cm beyond PTV)Dmean35.89±3.3235.48±3.040.301
Hard palateContra lateral parotidDmean24.64±2.3725.43±2.690.059
D50%20.24±0.9621.75±0.750.001
BrainstemDmax34.69±3.6535.68±3.050.536
Spinal cordDmax34.54±2.3536.38±2.360.029
LipsDmean34.72±1.9135.86±1.570.093
LarynxDmean47.84±1.0449.13±0.820.011
MandibleDmax77.02±1.8477.32±1.990.394
Ring (5 cm beyond PTV)Dmean36.44±3.7935.72±4.520.062
TongueContra lateral parotidDmean23.61±1.7124.76±1.76<0.001
D50%19.09±1.7620.77±1.65<0.001
BrainstemDmax30.64±3.2930.23±3.870.671
Spinal cordDmax35.44±1.7635.65±2.250.656
LipsDmean34.59±2.4135.33±1.960.206
LarynxDmean48.46±1.6648.71±2.550.527
MandibleDmax67.41±0.9468.78±1.28<0.001
Ring (5 cm beyond PTV)Dmean32.10±1.5831.38±1.710.001

Values are presented as mean±standard deviation.

OARs, organs-at-risk; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume.


Figure 1.Comparison of dose distribution in axial and coronal planes using (a) the IMRT and (b) VMAT techniques along with their (c) DVH. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; DVH, dose-volume histograms; PTV, planning target volume.

No significant (P>0.05) difference in PTV coverage (D95%) was observed between IMRT and VMAT plans for cancers of the supraglottic larynx, hard palate, and tongue. A noticeable increase (P<0.05) in D50% was observed for VMAT plans compared with that for IMRT plans for cancers of the supraglottic larynx, hard palate, and tongue. Moreover, a slight increase in the target mean dose and Dmax was observed for VMAT compared with those for IMRT for cancers of the supraglottic larynx, hard palate, and tongue, except for VMAT PTV56 for supraglottic larynx cancer. A significant (P<0.05) increase in UI was observed for VMAT plans compared with that for IMRT plans, except for PTV63 and PTV56 for supraglottic larynx cancer. Additionally, significant (P<0.05) reductions in HI were observed for VMAT plans compared with that for IMRT plans, except for PTV56 (P>0.05) for supraglottic larynx cancer. However, no significant (P>0.05) difference in CI was found between VMAT and IMRT plans. Additionally, a significant (P>0.05) increase in the CN number was noticed for VMAT plans compared with that for IMRT for all cases.

A slight increase in OAR dose was noted for VMAT plans compared with that for IMRT plans for all cases, except for the brainstem in tongue cancer (Table 3). Table 4 shows the increase in monitor units (MUs) for VMAT plans compared with those for IMRT plans in cancers of the supraglottic larynx (P>0.05, 0.46%), hard palate (P>0.05, 2.54%), and tongue (P<0.05, 7.56%). Additionally, significant (P<0.05) reductions in the NTID were noted for VMAT plans compared with that for IMRT plans in cancers of the supraglottic larynx (4.50%), hard palate (12.80%), and tongue (7.76%). Furthermore, a reduction in the mean dose to the ring structure (i.e., 5 cm beyond the PTV) was observed using the VMAT technique compared with that using the IMRT technique for cancers of the supraglottic larynx (1.14%, P>0.05), hard palate (1.97%, P>0.05), and tongue (2.24%, P<0.05).

Table 4 The MUs and NTID in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

SiteMUsNTID


IMRTVMATP-valueIMRTVMATP-value
Supraglottic larynx871.59±66.70875.64±84.560.9713.77±1.083.60±1.080.003
Hard palate854.57±80.32876.35±88.650.3074.53±1.833.95±1.350.025
Tongue915.35±79.48984.64±123.01<0.0013.35±1.093.09±0.91<0.001

Values are presented as mean±standard deviation.

MU, monitor unit; NTID, integral dose to normal tissue; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.


In low-dose volume evaluation, the doses to the 1%, 2%, and 5% volumes of normal tissue were evaluated and recorded in Table 5. A slight reduction (P>0.05) in D5% was observed for VMAT plans compared with that for IMRT plans for all HNC cases. Additionally, a reduction in dose volumes ranging from 1 Gy to 30 Gy was observed for VMAT plans compared with IMRT plans, except for V1Gy, V2Gy, and V30Gy for supraglottic larynx cancer and V1Gy for tongue cancer.

Table 5 The low-dose spillage analysis for supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

ParameterSupraglottic larynxHard palateTongue



IMRTVMATP-valueIMRTVMATP-valueIMRTVMATP-value
D1% (Gy)40.67±10.2440.39±11.580.80740.32±12.1441.30±12.260.31640.62±11.7539.42±12.620.005
D2% (Gy)27.34±5.0727.6±6.760.76028.67±6.4828.71±7.640.97326.33±7.6624.91±8.110.006
D5% (Gy)19.31±5.3817.68±5.200.06220.07±6.8417.35±6.830.01015.68±4.5713.40±3.990.000
V1Gy (cc)7,993.9±2,173.18,046.7±2,177.00.2126,015.3±3,032.35,979.8±2,958.50.3826,442.9±1,306.56,515.3±1,322.50.070
V2Gy (cc)5,553.7±1,792.15,559.5±1,763.00.8714,260.9±2,294.64,218.6±2,228.50.5154,369.8±1,065.04,349.3±1,058.70.515
V3Gy (cc)4,597.8±1,681.44,531.9±1,602.10.1663,552.1±2,016.33,453.6±1,907.70.1683,559.2±1,029.53,423.7±1,011.10.000
V4Gy (cc)4,013.3±1,615.03,925.9±1,514.70.0913,095.2±1,835.62,972.5±1,733.10.0603,023.7±1,020.12,875.77±994.10.000
V5Gy (cc)3,572.8±1,520.63,466.8±1,402.10.0692,761.0±1,673.72,629.2±1,622.10.0272,616.1±990.72,500.3±969.80.001
V10Gy (cc)2,309.1±1,194.32,049.5±1,081.00.0071,822.7±1,230.61,482.8±1,058.90.0031,609.47±791.11,320.3±595.20.000
V15Gy (cc)1,703.2±981.41,205.5±692.70.0001,326.5±904.4844.7±568.80.0101,082.2±591.2712.3±304.20.000
V20Gy (cc)975.6±511.3747.6±414.20.002830.7±547.2542.2±323.870.024578.58±298.1452.1±198.50.002
V25Gy (cc)518.8±234.0498.1±254.70.570450.4±262.3359.2±179.10.085348.2±148.8313.4±149.70.028
V30Gy (cc)294.1±90.1332.36±153.60.165252.28±134.5243.0±100.10.719245.0±104.2232.8±123.10.225

Values are presented as mean±standard deviation.

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.


The PSQA analysis showed comparable gamma passing rates for both techniques. Point dose analysis results were within ±5% for all cases using the IMRT and VMAT techniques. The gamma passing rate (with 3 mm/3%) in the relative dose mode was 97.11%±1.69% and 97.44%±1.80%, whereas, in the absolute dose mode, the gamma passing rate was 96.45%±0.84% and 96.59%±2.13% for the IMRT and VMAT techniques, respectively. The gamma passing rate (with 2 mm/2%) rate in the relative dose mode was 88.46%±1.24% and 90.29%±2.12%, whereas, in the absolute dose mode, the gamma passing rate was 85.57%±1.91% and 87.71%±2.79% for the IMRT and VMAT techniques, respectively. Fig. 2 shows the gamma analysis of a representative patient: ArcCHECK-measured plan, TPS-calculated plan, difference between ArcCHECK-measured and TPS-calculated plans, and measured points that matched the acceptance criteria.

Figure 2.Gamma analysis of a representative patient. (a) ArcCHECK-measured plan, (b) TPS-calculated plan, (c) difference between the ArcCHECK-measured and TPS-calculated plans, and (d) measured points that matched the acceptance criteria were shown in yellow color and points that don't match acceptance criteria were shown in red color. TPS, treatment planning system.

This study analyzed the dosimetric difference between seven-field IMRT and double-arc VMAT treatment plans in supraglottic larynx, hard palate, and tongue cancers. The results revealed no significant difference in target coverage between IMRT and VMAT plans. Additionally, IMRT offers a slightly better OAR-sparing ability than VMAT; however, IMRT had CI, UI, and CN comparable to those of VMAT, but its HI was inferior to that of VMAT. However, VMAT offers less NTID, mean dose to ring structure (i.e., 5 cm beyond the PTV), and dose volumes ranging from 1 Gy to 30 Gy to normal tissues than IMRT.

Radhakrishana et al. [21] have demonstrated the comparable target coverage for the IMRT and VMAT techniques using the Monaco TPS for nasopharyngeal carcinoma using the SIB technique. In their study, VMAT has shown a more conformal dose distribution with a better or comparable OAR-sparing ability.

This study revealed that VMAT required more MUs than IMRT in HNC. Moreover, Nithya et al. [22] have reported similar findings in a comparative study between VMAT and IMRT for tongue cancer using the Monaco TPS. Additionally, Nithya et al. [22] have highlighted a dip in the literature about comparative studies for HNC using the Monaco TPS compared with Eclipse (Varian Medical System, Palo Alto, CA, USA) and other TPS. Several studies have reported that VMAT requires lesser MUs than IMRT for HNC [21,22]. HNC planning requires a high degree of modulation for a clinically acceptable treatment plan. Additionally, HNC presents a unique challenge of concave dose distribution along with close proximity of OARs to the target volume. The number of MUs depends on the size of the target and the degree of modulation employed in the optimization process for OAR sparing [23].

This study revealed that VMAT offers lesser NTID, mean dose to ring structure (i.e., 5 cm beyond the PTV), and dose volumes (ranging from 1 Gy to 30 Gy) to normal tissues than IMRT. The dose outside the treatment field depends on the target size and increases with an increase in the target volume because an increase in the target volume leads to more irradiation of the tissues along the target length [24]. The scatter contribution from the collimator and patient body along with softer energy spectrum outside the treatment field (compared with within the treatment field) plays a vital role in the irradiation dose to normal tissues outside the target volume [25,26]. Moreover, Fig. 3 shows a gradual decrease in the mean volume (V) encompassed by a certain dose (D) on a logarithmic scale for an increase in the dose and reveals that VMAT plans have a higher dose falloff than IMRT plans for supraglottic larynx, hard palate, and tongue cancers. This can be because VMAT offers more freedom in the optimization process restricting radiation doses to nearby OARs and normal tissues outside the target.

Figure 3.The rate of dose falloff beyond the target by plotting ln (D) versus ln (1/V) for (a) supraglottic larynx, (b) hard palate, and (c) tongue cancers. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.

Hall [27] and Hall and Wuu [28] advocated that a low-dose volume may not cause acute or subacute clinical morbidity but could be carcinogenic and reported that IMRT probably has a 1% to 1.75% higher risk of secondary malignancies than conventional radiotherapy in patients surviving for 10 years. Followill et al. [29] have reported an increased risk of secondary malignancies by 0.4% to 1.0% with IMRT compared with conventional RT techniques. Hauri and Schneider [30] have evaluated the impact of treatment techniques on out-of-field low-dose volumes and recommended VMAT over IMRT in terms of radiological protection to patients with cancer. Because VMAT requires less beam-on time than IMRT leads, a reduction in the out-of-field dose is expected.

Kry et al. [31] have reported that secondary malignancy is more likely to occur in areas surrounding the target receiving doses greater than 3 Gy. However, Diallo et al. [15] have reported that areas receiving doses smaller than 2.5 Gy are the prominent areas for the occurrence of secondary malignancies. In external beam radiotherapy, a larger area around the target gets exposed to radiation. Targets are exposed to a high dose of radiation to maintain the therapeutic level, and surrounding areas should be considered in plan evaluation to minimize exposure to radiation. Therefore, low-dose volumes should be carefully evaluated for the risk of secondary malignancies and for plan optimization for better consideration of nearby tissues in planning.

For advanced-stage HNCs, both IMRT and VMAT offer satisfactory clinical plans. For similar target coverage, VMAT offers a conformal and homogeneous dose distribution with a comparable OAR-sparing ability and a higher dose falloff outside the target volume compared with IMRT, which provides an edge over IMRT in reducing the risk of secondary malignancies.

The authors thank the management of Rajiv Gandhi Cancer Institute & Research Centre in New Delhi, India, for their continued support and encouragement to complete this research work.


The data that support the findings of this study are available on request from the corresponding author.


Conceptualization: Gourav Kumar and Lalit Kumar. Data curation: Gourav Kumar. Formal analysis: Manindra Bhushan. Funding acquisition: None. Investigation: Pawan Kumar and Soumitra Barik. Methodology: Sandeep Purohit. Project administration: Vimal Kishore. Resources: Soumitra Barik. Software: Sandeep Purohit and Pawan Kumar. Supervision: Manindra Bhushan and Vimal Kishore. Validation: Pawan Kumar and Kothanda Raman. Visualization: Manindra Bhushan. Writing–original draft: Gourav Kumar. Writing–review & editing: Lalit Kumar.

  1. Bhattacharjee A, Chakraborty A, Purkaystha P. Prevalence of head and neck cancers in the North East-an institutional study. Indian J Otolaryngol Head Neck Surg. 2006;58:15-19.
    Pubmed KoreaMed CrossRef
  2. Sharma JD, Baishya N, Kataki AC, Kalita CR, Das AK, Rahman T. Head and neck squamous cell carcinoma in young adults: a hospital-based study. Indian J Med Paediatr Oncol. 2019;40(S1):S18-S22.
  3. Grégoire V, Lefebvre JL, Licitra L; Felip E; EHNS-ESMO-ESTRO Guidelines Working Group. Squamous cell carcinoma of the head and neck: EHNS-ESMO-ESTRO Clinical Practice Guidelines for diagnosis, treatment and follow-up. Ann Oncol. 2010;21 Suppl 5:v184-v186.
    Pubmed CrossRef
  4. Vigneswaran N, Williams MD. Epidemiologic trends in head and neck cancer and aids in diagnosis. Oral Maxillofac Surg Clin North Am. 2014;26:123-141.
    Pubmed KoreaMed CrossRef
  5. Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68:394-424.
    Pubmed CrossRef
  6. Gupta T, Agarwal J, Jain S, Phurailatpam R, Kannan S, Ghosh-Laskar S, et al. Three-dimensional conformal radiotherapy (3D-CRT) versus intensity modulated radiation therapy (IMRT) in squamous cell carcinoma of the head and neck: a randomized controlled trial. Radiother Oncol. 2012;104:343-348.
    Pubmed CrossRef
  7. Ge X, Liao Z, Yuan J, Mao D, Li Y, Yu E, et al. Radiotherapy-related quality of life in patients with head and neck cancers: a meta-analysis. Support Care Cancer. 2020;28:2701-2712.
    Pubmed CrossRef
  8. DE Felice F, Pranno N, Papi P, Brugnoletti O, Tombolini V, Polimeni A. Xerostomia and clinical outcomes in definitive intensity modulated radiotherapy (IMRT) versus three-dimensional conformal radiotherapy (3D-CRT) for head and neck squamous cell carcinoma: a meta-analysis. In Vivo. 2020;34:623-629.
    Pubmed KoreaMed CrossRef
  9. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317.
    Pubmed CrossRef
  10. Holt A, Van Gestel D, Arends MP, Korevaar EW, Schuring D, Kunze-Busch MC, et al. Multi-institutional comparison of volumetric modulated arc therapy vs. intensity-modulated radiation therapy for head-and-neck cancer: a planning study. Radiat Oncol. 2013;8:26.
    Pubmed KoreaMed CrossRef
  11. Studenski MT, Bar-Ad V, Siglin J, Cognetti D, Curry J, Tuluc M, et al. Clinical experience transitioning from IMRT to VMAT for head and neck cancer. Med Dosim. 2013;38:171-175.
    Pubmed CrossRef
  12. Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S. Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys. 2009;74:252-259.
    Pubmed CrossRef
  13. Kumar L, Yadav G, Raman K, Bhushan M, Pal M. The dosimetric impact of different photon beam energy on RapidArc radiotherapy planning for cervix carcinoma. J Med Phys. 2015;40:207-213.
    Pubmed KoreaMed CrossRef
  14. Kumar L, Yadav G, Samuvel KR, Bhushan M, Kumar P, Suhail M, et al. Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma. Rep Pract Oncol Radiother. 2017;22:10-18.
    Pubmed KoreaMed CrossRef
  15. Diallo I, Haddy N, Adjadj E, Samand A, Quiniou E, Chavaudra J, et al. Frequency distribution of second solid cancer locations in relation to the irradiated volume among 115 patients treated for childhood cancer. Int J Radiat Oncol Biol Phys. 2009;74:876-883.
    Pubmed CrossRef
  16. Nagarajan M, Banu R, Sathya B, Sundaram T, Chellapandian TP. Dosimetric evaluation and comparison between volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) plan in head and neck cancers. Gulf J Oncolog. 2020;1:45-50.
    Pubmed
  17. Osborn J. Is VMAT beneficial for patients undergoing radiotherapy to the head and neck? Radiography (Lond). 2017;23:73-76.
    Pubmed CrossRef
  18. ICRU Report 83. Prescribing, recording, and reporting intensity-modulated photon-beam therapy (IMRT). ICRU Report. Bethesda: ICRU. 2010: 83.
  19. IMPAC Medical Systems, Inc. Monaco® training guide. Stockholm: Elekta AB; 2013.
  20. Kumar L, Yadav G, Kishore V, Bhushan M, Gairola M, Tripathi D. Validation of the RapidArc delivery system using a volumetric phantom as per task group report 119 of the American Association of Physicists in Medicine. J Med Phys. 2019;44:126-134.
    Pubmed KoreaMed CrossRef
  21. Radhakrishnan S, Chandrasekaran A, Sarma Y, Balakrishnan S, Nandigam J. Dosimetric comparison between single and dual arc-volumetric modulated arc radiotherapy and intensity modulated radiotherapy for nasopharyngeal carcinoma using a simultaneous integrated boost technique. Asian Pac J Cancer Prev. 2017;18:1395-1402.
    Pubmed KoreaMed CrossRef
  22. Nithya L, Raj NA, Kumar A, Rathinamuthu S, Pandey MB. Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer. J Med Phys. 2014;39:121-126.
    Pubmed KoreaMed CrossRef
  23. Kumar L, Bhushan M, Kishore V, Yadav G, Gurjar OP. Dosimetric validation of Acuros® XB algorithm for RapidArc™ treatment technique: a post software upgrade analysis. J Cancer Res Ther. 2021. doi: 10.4103/jcrt.JCRT_1154_19.
  24. Stovall M, Blackwell CR, Cundiff J, Novack DH, Palta JR, Wagner LK, et al. Fetal dose from radiotherapy with photon beams: report of AAPM Radiation Therapy Committee Task Group No. 36. Med Phys. 1995;22:63-82.
    Pubmed CrossRef
  25. Edwards CR, Mountford PJ. Near surface photon energy spectra outside a 6 MV field edge. Phys Med Biol. 2004;49:N293-N301.
    Pubmed CrossRef
  26. Gayen S, Kombathula SH, Manna S, Varshney S, Pareek P. Dosimetric comparison of coplanar and non-coplanar volumetric-modulated arc therapy in head and neck cancer treated with radiotherapy. Radiat Oncol J. 2020;38:138-147.
    Pubmed KoreaMed CrossRef
  27. Hall EJ. Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys. 2006;65:1-7.
    Pubmed CrossRef
  28. Hall EJ, Wuu CS. Radiation-induced second cancers: the impact of 3D-CRT and IMRT. Int J Radiat Oncol Biol Phys. 2003;56:83-88.
    Pubmed CrossRef
  29. Followill D, Geis P, Boyer A. Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy. Int J Radiat Oncol Biol Phys. 1997;38:667-672.
    Pubmed CrossRef
  30. Hauri P, Schneider U. Whole-body dose equivalent including neutrons is similar for 6 MV and 15 MV IMRT, VMAT, and 3D conformal radiotherapy. J Appl Clin Med Phys. 2019;20:56-70.
    Pubmed KoreaMed CrossRef
  31. Kry SF, Bednarz B, Howell RM, Dauer L, Followill D, Klein E, et al. AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy. Med Phys. 2017;44:e391-e429.
    Pubmed CrossRef

Article

Original Article

Progress in Medical Physics 2021; 32(3): 70-81

Published online September 30, 2021 https://doi.org/10.14316/pmp.2021.32.3.70

Copyright © Korean Society of Medical Physics.

Dosimetric Evaluation of Low-Dose Spillage Volumes for Head and Neck Cancer Using Intensity-Modulated Radiation Therapy and Volumetric Modulated Arc Therapy Treatment Techniques

Gourav Kumar1 , Manindra Bhushan1 , Lalit Kumar1 , Vimal Kishore2 , Kothanda Raman1 , Pawan Kumar1 , Soumitra Barik1 , Sandeep Purohit1

1Medical Physics Division & Department of Radiation Oncology, Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, 2Department of Applied Science & Humanities, Bundelkhand Institute of Engineering & Technology, Jhansi, India

Correspondence to:Lalit Kumar
(lalitk48@gmail.com)
Tel: +91-9599329464
Fax: +91-11-27051037

Received: June 5, 2021; Revised: August 5, 2021; Accepted: August 10, 2021

Abstract

Purpose: This study was designed to investigate the dosimetric difference between intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) in head and neck cancer (HNC). The study primarily focuses on low-dose spillage evaluation between these two techniques.
Methods: This retrospective study involved 45 patients with HNC. The treatment plans were generated using the IMRT and VMAT techniques for all patients. Dosimetric comparisons were performed in terms of target coverage, organ-at-risk (OAR) sparing, and various parameters, including conformity index, uniformity index, homogeneity index, conformation number, low-dose volumes, and normal tissue integral dose (NTID).
Results: No significant (P>0.05) difference in planning target volume coverage (D95%) was observed between IMRT and VMAT plans for supraglottic larynx, hard palate, and tongue cancers. A decrease in dose volumes ranging from 1 Gy to 30 Gy was observed for VMAT plans compared with those for IMRT plans, except for V1Gy and V30Gy for supraglottic larynx cancer and V1Gy for tongue cancer. Moreover, decreases (P<0.05) in NTID were observed for VMAT plans compared with that for IMRT plans in supraglottic larynx (4.50%), hard palate (12.80%), and tongue (7.76%) cancers. In contrast, a slight increase in monitor units for VMAT compared with those for IMRT in supraglottic larynx (0.46%), hard palate (2.54%), and tongue (7.56%) cancers.
Conclusions: For advanced-stage HNC, both IMRT and VMAT offer satisfactory clinical plans. VMAT offers a conformal and homogeneous dose distribution with comparable OAR sparing and higher dose falloff outside the target volume than IMRT, which provides an edge to reduce the risk of secondary malignancies for HNC over IMRT.

Keywords: Intensity-modulated radiation therapy, Volumetric modulated arc therapy, Head and neck cancer, Low-dose spillage, Radiotherapy

Introduction

Head and neck cancers (HNCs) are the most common type of malignancy spreading worldwide. In India, among all cancer cases reported every year, 35% are HNC cases [1,2]. Most HNC malignancies include tumors of the tongue, nasopharynx, oropharynx, hypopharynx, and oral cavity [3,4]. Among all HNCs, squamous cell carcinoma (SCC) constitutes more than 90% of the cases. The widespread use of tobacco and smoking are considered the most important factors for developing these malignancies. Due to the lack of appropriate knowledge and education, most Indian patients present with the advanced stage of the disease, therefore, making radiotherapy challenging. According to the 2018 Global Cancer Incidence, Mortality and Prevalence report, there were 834,860 new HNC cases reported every year, resulting in 431,131 deaths per year [5]. Limited options are available for treatment, such as surgery, chemotherapy, and radiotherapy for early and advanced stages of HNC.

The primary aim of radiotherapy is to deliver therapeutic doses to the target while sparing the surrounding organs-at-risk (OARs) as much as possible. The introduction of intensity-modulated radiation therapy (IMRT) in HNC replaced conventional three-dimensional conformal radiotherapy (3D-CRT), resulting in better target coverage, more conformity, and better sparing of OARs [6]. Radiation-induced toxicities are low in IMRT compared with the 3D-CRT technique [7]. IMRT is primarily useful for irregularly shaped target volumes to achieve a high degree of conformal dose distribution using large numbers of smaller radiation beamlets of varying intensities [8].

With the advancement in the IMRT technique, volumetric modulated arc therapy (VMAT), a rotational treatment technology, came into existence in 2007 and now widely accepted worldwide [9]. VMAT is the evolution of IMRT incorporating simultaneous variation of three parameters, namely, the gantry speed, multileaf collimator (MLC) speed, and dose rate. Because of steeper dose gradients, both IMRT and VMAT require accurate target and normal tissue delineation. Various studies in the literature have shown that VMAT delivers a more conformal dose to planning target volumes (PTVs), results in a lower dose spillage, and has a better normal tissue-sparing ability than the IMRT technique [10-14].

With the evolution of technology, dose escalation and conformity to the target has improved. However, normal tissues outside the target get irradiated unnecessarily. It has been reported in the literature that radiation exposure to normal tissues causes secondary malignancies. The risk of secondary malignancy depends on the location and amount of exposure received by the surrounding normal tissues. Diallo et al. [15] have reported that 66% of secondary malignancies occurred within the first 5 cm from the field boundary, 22% occurred after the first 5 cm from the field boundary, and 12% occurred within the target volume.

The dose distribution generated for HNC cases using IMRT and VMAT has been enormously detailed in the literature [11,16,17]. However, distinction in low-dose distribution generated using IMRT and VMAT takes a dip for HNC cases and should be evaluated. Low-dose exposure may contribute to secondary malignancies. Therefore, this study was designed to quantify the aforementioned difference in dose distribution generated for HNC cases.

This study focused on the selection of the most appropriate treatment technique between IMRT and VMAT for patients with HNC regarding the conformity index (CI), homogeneity index (HI), conformation number (CN), uniformity index (UI), and other dosimetric parameters. Few studies have reported the effects of low doses in radiotherapy planning. Therefore, this study was conducted to determine the dosimetric difference between IMRT and VMAT in terms of low-dose spillage.

Materials and Methods

1. Patient selection, contouring, and dose prescription

This retrospective study involved 45 patients with stage II and IIIB HNC with a mean age of ≥50 years. These patients presented with SCC of the supraglottic larynx and hard palate and postoperative cancer of the tongue. All patients received radiotherapy between 2014 and 2019 as part of their standard care at our institution using the simultaneous integrated boost (SIB) technique. Computed tomography (CT) scans with 3-mm slice thickness for each patient were performed in head-first-supine position along with a standard HN thermoplast orfit mask on Somaton Sensation CT scanner (Siemens medical system, Malvern, PA, USA). After that, CT image sets were exported to the Monaco (Elekta Medical system, Crawley, UK) treatment planning system (TPS) in the Digital Imaging and Communication in Medicine format.

CT images were registered and fused with positron emission tomography and magnetic resonance imaging images to contour gross target volume. The clinical target volume and PTV were created using a margin of 12 mm and 5 mm, respectively, according to the Radiation Therapy Oncology Group recommendations. Moreover, critical OARs were delineated on the CT images, namely, parotid glands, spinal cord, brainstem, larynx, mandible, and optic apparatus.

The dose prescriptions were 70 Gy (high-risk PTV), 63 Gy (intermediate-risk PTV), and 56 Gy (low-risk PTV) in 35 fractions for the supraglottic larynx; 70 Gy (high-risk PTV) and 56 Gy (low-risk PTV) in 35 fractions for the hard palate; and 60 Gy (tumor bed) and 54 Gy (nodal volumes) in 30 fractions for tongue cancer. According to the International Commission on Radiation Units and Measurements 83 [18], treatment plans were normalized in such a way that at least 95% of the PTV should cover 100% of the prescription dose. The dose objective for OARs were followed as prescribed by the Quantitative Analyses of Normal Tissue Effects in the Clinic group. Table 1 summarizes the planning dose objectives for the PTV and OARs. Treatment plans were generated using a 6-megavolt (MV) photon beam energized using a Synergy (Elekta Medical System) linear accelerator (linac). The Synergy linac is equipped with 40 pairs of MLC leaves with a width of 1 cm at the isocenter, which could deliver a maximum dose rate of 600 cGy/min at the dose maximum (Dmax).

Table 1 . Planning dose objectives for the planning target volume (PTV) and organs-at-risk.

StructureParameterPlan objective
PTVD95%
D110%
D93%
=100% of prescribed dose
<10% of PTV
<1% of PTV
Spinal cordDmax<45 Gy
BrainstemDmax<54 Gy
LipsDmean<35 Gy
Contralateral parotidDmean<26 Gy
LarynxDmean<50 Gy
MandibleDmax<75 Gy


2. Intensity-modulated radiation therapy planning

Clinically acceptable dynamic MLC plans were generated on the Monaco TPS (version 5.11.02; Elekta Medical System). The Monaco TPS uses the Monte Carlo algorithm for optimization and final dose calculation, which is the gold standard presently. In Monaco, IMRT incorporates cost function-based segment shape optimization. Seven coplanar fields of 6-MV photon beam with couch and collimator angles at 0° were used to design the respective plans. All gantry angles were equidistant from each other, that is, 0°, 51°, 102°, 153°, 204°, 255°, and 306°, respectively. Additionally, a spill structure with a margin of 5 mm from the target was drawn to control low-dose spillage outside PTVs. The recommended grid spacing of 3 mm and statistical uncertainty of 2% were employed for all treatment plans. The minimum segment size of 5 mm and maximum number of control point per field of 30 were chosen for optimization and dose calculation [19].

3. Volumetric modulated arc therapy planning

VMAT plans were created in the similar fashion as IMRT plans. All cost functions and dose objectives (target and OARs) were the same as those used in IMRT optimization. In VMAT planning, we used dual-arc ranges from 181° to 179.9° in clockwise and counterclockwise directions with an arc increment angle of 20°. The maximum number of control points per arc of 100 was chosen.

4. Dosimetric evaluation

Intercomparison of both the IMRT and VMAT plans was performed using analysis of dose-volume histogram (DVH) curves. Various dosimetric parameters were evaluated for the PTV and critical OARs using DVH data. The following dose indices were evaluated:

PTV: D95% (dose to the 95% volume of the PTV), D98%, D50%, D2%, Dmean, Dmin, and Dmax.

OARs: Dmax for serial organs, i.e., spinal cord, brainstem.

Mean dose for parallel organs, that is, parotid glands, lips, and larynx.

Low-dose volume analysis: V1Gy (volume of 1 Gy outside the target), V2Gy, V3Gy, V4Gy, V5Gy, V10Gy, V15Gy, V20Gy, V25Gy, V30Gy, D1% (dose to the 1% volume outside the target), D2%, D5%.

Additionally, a ring structure was delineated to be 5 cm beyond the PTV to estimate the mean dose to the normal tissues lying in that region.

Moreover, the UI, CI, HI, conformity number (CN), and normal tissue integral dose (NTID) were computed using the following formulas:

UI=D5GyD95Gy
CI=TV95% ccTV
HI=D2Gy   D98Gy D50Gy
CN=TVRITV×TVRIVRI
NTID= mean dose×volume of normal tissues outside the PTV

where TV stands for the target volume in cm3, TVRI stands for the volume covered by the reference isodose, and VRI stands for the volume of the prescription isodose outside the target volume (inside body).

5. Statistical analysis

The dosimetric difference between IMRT and VMAT was analyzed using Statistical Package for the Social Sciences (version 20; IBM Corp., Armonk, NY, USA) in terms of the mean, standard deviation, and P-values. The independent paired t-test with a confidence interval limit of 95% was performed to assess the dosimetric endpoints for the target and OARs. P-values of less than 0.05 were used to denote statistical significance.

6. Quality assurance

Moreover, patient-specific quality assurance (PSQA) was executed using ArcCHECK (Sun Nuclear, Melbourne, FL, USA), which is a 3D cylindrical array of 1,386 equally spaced (10 mm apart) diode detectors. All treatment plans were calculated, and the same plans were delivered on ArcCHECK. SNC (version 6.2; Sun Nuclear) was used for gamma analysis. A criterion of 3-mm distance to agreement (DTA) and 3% dose difference (DD) with a threshold value of 10% was used for PSQA analysis. Additionally, 2-mm DTA and 2% DD criteria with a threshold value of 10% was examined for gamma analysis. A tolerance of ±5% was considered acceptable between the ion-chamber-measured and TPS-calculated doses [20].

Results

Clinically acceptable treatment plans were created using the IMRT and VMAT techniques for all patients. Qualitative and quantitative analyses were performed on dose distribution created for IMRT and VMAT plans. The data were derived from cumulative DVH data for each treatment plan. Table 2 and 3 illustrate the dosimetric data on the PTV and OARs for cancers of the supraglottic larynx, hard palate, and tongue using the IMRT and VMAT techniques. Fig. 1 shows the dose distribution in axial and coronal planes using IMRT and VMAT techniques along their DVH.

Table 2 . The dosimetric parameters for PTV in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques.

TargetParameterSupraglottic larynxHard palateTongueP-value




IMRTVMATIMRTVMATIMRTVMATSupraHPTongue
PTV70D95% (Gy)70.0070.0070.0070.00
D98% (Gy)69.20±0.1669.07±0.1569.17±0.0868.98±0.160.0110.006
D50% (Gy)72.66±0.3473.02±0.4272.48±0.2873.10±0.360.0090.001
D2% (Gy)75.38±0.5276.13±0.6175.14±0.5176.31±0.750.0190.001
Mean (Gy)72.04±1.0772.29±1.2371.99±0.3473.01±0.340.5480.045
Dmax (Gy)78.54±1.2479.73±0.9678.60±1.1780.77±0.930.0070.001
Dmin (Gy)63.31±2.1962.11±2.2562.77±1.7461.57±2.150.1310.081
UI1.07±0.011.08±0.011.06±0.011.08±0.01<0.001<0.001
HI0.072±0.010.071±0.010.081±0.010.078±0.01<0.001<0.001
CI0.99±0.030.99±0.030.99±0.011.00±0.010.7910.093
CN0.88±0.030.90±0.030.92±0.030.93±0.030.9720.667
PTV63D95% (Gy)64.01±0.9163.91±1.180.511
D98% (Gy)62.95±0.9862.67±1.450.197
D50% (Gy)68.65±1.5269.24±1.460.001
D2% (Gy)74.69±0.7975.26±0.790.006
Mean (Gy)66.06±1.0666.19±1.620.767
Dmax (Gy)74.65±1.2875.29±0.960.192
Dmin (Gy)51.02±3.5351.35±4.050.717
UI1.15±0.011.16±0.020.102
HI0.171±0.010.168±0.020.032
CI0.99±0.030.98±0.030.341
CN0.86±0.050.88±0.060.142
PTV56D95% (Gy)57.78±0.9257.82±1.0458.83±0.8359.70±1.110.7710.055
D98% (Gy)56.94±0.8256.79±1.0357.80±0.7258.25±0.940.4230.066
D50% (Gy)63.75±3.6464.28±3.7267.40±3.3368.29±2.720.0030.011
D2% (Gy)70.63±5.7171.34±5.6274.12±1.5375.13±1.770.0020.002
Mean (Gy)58.93±1.1759.04±1.4762.51±0.9164.11±1.090.7690.002
Dmax (Gy)71.35±2.5070.92±2.7375.61±1.1476.55±0.990.4070.008
Dmin (Gy)46.54±3.6045.82±2.8945.65±3.2846.42±1.790.3780.518
UI1.21±0.091.11±0.371.24±0.031.24±0.040.3910.576
HI0.218±0.080.214±0.070.232±0.020.229±0.030.0510.302
CI1.00±0.011.00±0.011.00±0.010.99±0.010.8920.271
CN0.72±0.210.75±0.210.90±0.110.92±0.12<0.0010.113
PTV60D95% (Gy)60.0060.00
D98% (Gy)58.93±0.1858.76±0.25<0.001
D50% (Gy)62.84±0.3463.51±0.45<0.001
D2% (Gy)65.44±0.5366.81±0.74<0.001
Mean (Gy)62.69±0.3163.35±0.41<0.001
Dmax (Gy)69.43±1.1371.67±1.19<0.001
Dmin (Gy)50.49±2.2749.72±3.020.871
UI1.08±0.011.1±0.01<0.001
HI0.126±0.010.123±0.01<0.001
CI0.99±0.0010.99±0.010.244
CN0.88±0.040.89±0.040.478
PTV54D95% (Gy)54.69±0.4354.48±0.530.072
D98% (Gy)53.89±0.5453.38±0.580.001
D50% (Gy)57.23±1.3157.91±1.380.001
D2% (Gy)61.50±2.7862.69±2.900.001
Mean (Gy)56.81±0.9057.08±0.590.228
Dmax (Gy)64.16±1.4264.76±0.990.046
Dmin (Gy)43.06±4.1243.93±2.890.115
UI1.10±0.051.13±0.05<0.001
HI0.161±0.050.159±0.05<0.001
CI1.00±0.0010.99±0.010.078
CN0.69±0.180.72±0.180.101

Values are presented in mean±standard deviation..

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume; UI, uniformity index; HI, homogeneity index; CI, conformity index; CN, conformation number..



Table 3 . The dosimetric parameters for OARs in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques.

SiteOARsParameter (Gy)IMRTVMATP-value
Supraglottic larynxContra lateral parotidDmean22.76±1.2423.41±1.21<0.001
D50%18.22±2.2818.31±0.810.853
BrainstemDmax33.88±2.5736.21±2.53<0.001
Spinal cordDmax32.77±1.6333.92±1.580.009
LipsDmean32.71±3.7434.65±2.960.016
LarynxDmean49.68±1.9150.47±1.520.004
MandibleDmax75.83±2.6176.46±2.190.102
Ring (5 cm beyond PTV)Dmean35.89±3.3235.48±3.040.301
Hard palateContra lateral parotidDmean24.64±2.3725.43±2.690.059
D50%20.24±0.9621.75±0.750.001
BrainstemDmax34.69±3.6535.68±3.050.536
Spinal cordDmax34.54±2.3536.38±2.360.029
LipsDmean34.72±1.9135.86±1.570.093
LarynxDmean47.84±1.0449.13±0.820.011
MandibleDmax77.02±1.8477.32±1.990.394
Ring (5 cm beyond PTV)Dmean36.44±3.7935.72±4.520.062
TongueContra lateral parotidDmean23.61±1.7124.76±1.76<0.001
D50%19.09±1.7620.77±1.65<0.001
BrainstemDmax30.64±3.2930.23±3.870.671
Spinal cordDmax35.44±1.7635.65±2.250.656
LipsDmean34.59±2.4135.33±1.960.206
LarynxDmean48.46±1.6648.71±2.550.527
MandibleDmax67.41±0.9468.78±1.28<0.001
Ring (5 cm beyond PTV)Dmean32.10±1.5831.38±1.710.001

Values are presented as mean±standard deviation..

OARs, organs-at-risk; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume..



Figure 1. Comparison of dose distribution in axial and coronal planes using (a) the IMRT and (b) VMAT techniques along with their (c) DVH. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; DVH, dose-volume histograms; PTV, planning target volume.

No significant (P>0.05) difference in PTV coverage (D95%) was observed between IMRT and VMAT plans for cancers of the supraglottic larynx, hard palate, and tongue. A noticeable increase (P<0.05) in D50% was observed for VMAT plans compared with that for IMRT plans for cancers of the supraglottic larynx, hard palate, and tongue. Moreover, a slight increase in the target mean dose and Dmax was observed for VMAT compared with those for IMRT for cancers of the supraglottic larynx, hard palate, and tongue, except for VMAT PTV56 for supraglottic larynx cancer. A significant (P<0.05) increase in UI was observed for VMAT plans compared with that for IMRT plans, except for PTV63 and PTV56 for supraglottic larynx cancer. Additionally, significant (P<0.05) reductions in HI were observed for VMAT plans compared with that for IMRT plans, except for PTV56 (P>0.05) for supraglottic larynx cancer. However, no significant (P>0.05) difference in CI was found between VMAT and IMRT plans. Additionally, a significant (P>0.05) increase in the CN number was noticed for VMAT plans compared with that for IMRT for all cases.

A slight increase in OAR dose was noted for VMAT plans compared with that for IMRT plans for all cases, except for the brainstem in tongue cancer (Table 3). Table 4 shows the increase in monitor units (MUs) for VMAT plans compared with those for IMRT plans in cancers of the supraglottic larynx (P>0.05, 0.46%), hard palate (P>0.05, 2.54%), and tongue (P<0.05, 7.56%). Additionally, significant (P<0.05) reductions in the NTID were noted for VMAT plans compared with that for IMRT plans in cancers of the supraglottic larynx (4.50%), hard palate (12.80%), and tongue (7.76%). Furthermore, a reduction in the mean dose to the ring structure (i.e., 5 cm beyond the PTV) was observed using the VMAT technique compared with that using the IMRT technique for cancers of the supraglottic larynx (1.14%, P>0.05), hard palate (1.97%, P>0.05), and tongue (2.24%, P<0.05).

Table 4 . The MUs and NTID in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques.

SiteMUsNTID


IMRTVMATP-valueIMRTVMATP-value
Supraglottic larynx871.59±66.70875.64±84.560.9713.77±1.083.60±1.080.003
Hard palate854.57±80.32876.35±88.650.3074.53±1.833.95±1.350.025
Tongue915.35±79.48984.64±123.01<0.0013.35±1.093.09±0.91<0.001

Values are presented as mean±standard deviation..

MU, monitor unit; NTID, integral dose to normal tissue; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy..



In low-dose volume evaluation, the doses to the 1%, 2%, and 5% volumes of normal tissue were evaluated and recorded in Table 5. A slight reduction (P>0.05) in D5% was observed for VMAT plans compared with that for IMRT plans for all HNC cases. Additionally, a reduction in dose volumes ranging from 1 Gy to 30 Gy was observed for VMAT plans compared with IMRT plans, except for V1Gy, V2Gy, and V30Gy for supraglottic larynx cancer and V1Gy for tongue cancer.

Table 5 . The low-dose spillage analysis for supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques.

ParameterSupraglottic larynxHard palateTongue



IMRTVMATP-valueIMRTVMATP-valueIMRTVMATP-value
D1% (Gy)40.67±10.2440.39±11.580.80740.32±12.1441.30±12.260.31640.62±11.7539.42±12.620.005
D2% (Gy)27.34±5.0727.6±6.760.76028.67±6.4828.71±7.640.97326.33±7.6624.91±8.110.006
D5% (Gy)19.31±5.3817.68±5.200.06220.07±6.8417.35±6.830.01015.68±4.5713.40±3.990.000
V1Gy (cc)7,993.9±2,173.18,046.7±2,177.00.2126,015.3±3,032.35,979.8±2,958.50.3826,442.9±1,306.56,515.3±1,322.50.070
V2Gy (cc)5,553.7±1,792.15,559.5±1,763.00.8714,260.9±2,294.64,218.6±2,228.50.5154,369.8±1,065.04,349.3±1,058.70.515
V3Gy (cc)4,597.8±1,681.44,531.9±1,602.10.1663,552.1±2,016.33,453.6±1,907.70.1683,559.2±1,029.53,423.7±1,011.10.000
V4Gy (cc)4,013.3±1,615.03,925.9±1,514.70.0913,095.2±1,835.62,972.5±1,733.10.0603,023.7±1,020.12,875.77±994.10.000
V5Gy (cc)3,572.8±1,520.63,466.8±1,402.10.0692,761.0±1,673.72,629.2±1,622.10.0272,616.1±990.72,500.3±969.80.001
V10Gy (cc)2,309.1±1,194.32,049.5±1,081.00.0071,822.7±1,230.61,482.8±1,058.90.0031,609.47±791.11,320.3±595.20.000
V15Gy (cc)1,703.2±981.41,205.5±692.70.0001,326.5±904.4844.7±568.80.0101,082.2±591.2712.3±304.20.000
V20Gy (cc)975.6±511.3747.6±414.20.002830.7±547.2542.2±323.870.024578.58±298.1452.1±198.50.002
V25Gy (cc)518.8±234.0498.1±254.70.570450.4±262.3359.2±179.10.085348.2±148.8313.4±149.70.028
V30Gy (cc)294.1±90.1332.36±153.60.165252.28±134.5243.0±100.10.719245.0±104.2232.8±123.10.225

Values are presented as mean±standard deviation..

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy..



The PSQA analysis showed comparable gamma passing rates for both techniques. Point dose analysis results were within ±5% for all cases using the IMRT and VMAT techniques. The gamma passing rate (with 3 mm/3%) in the relative dose mode was 97.11%±1.69% and 97.44%±1.80%, whereas, in the absolute dose mode, the gamma passing rate was 96.45%±0.84% and 96.59%±2.13% for the IMRT and VMAT techniques, respectively. The gamma passing rate (with 2 mm/2%) rate in the relative dose mode was 88.46%±1.24% and 90.29%±2.12%, whereas, in the absolute dose mode, the gamma passing rate was 85.57%±1.91% and 87.71%±2.79% for the IMRT and VMAT techniques, respectively. Fig. 2 shows the gamma analysis of a representative patient: ArcCHECK-measured plan, TPS-calculated plan, difference between ArcCHECK-measured and TPS-calculated plans, and measured points that matched the acceptance criteria.

Figure 2. Gamma analysis of a representative patient. (a) ArcCHECK-measured plan, (b) TPS-calculated plan, (c) difference between the ArcCHECK-measured and TPS-calculated plans, and (d) measured points that matched the acceptance criteria were shown in yellow color and points that don't match acceptance criteria were shown in red color. TPS, treatment planning system.

Discussion

This study analyzed the dosimetric difference between seven-field IMRT and double-arc VMAT treatment plans in supraglottic larynx, hard palate, and tongue cancers. The results revealed no significant difference in target coverage between IMRT and VMAT plans. Additionally, IMRT offers a slightly better OAR-sparing ability than VMAT; however, IMRT had CI, UI, and CN comparable to those of VMAT, but its HI was inferior to that of VMAT. However, VMAT offers less NTID, mean dose to ring structure (i.e., 5 cm beyond the PTV), and dose volumes ranging from 1 Gy to 30 Gy to normal tissues than IMRT.

Radhakrishana et al. [21] have demonstrated the comparable target coverage for the IMRT and VMAT techniques using the Monaco TPS for nasopharyngeal carcinoma using the SIB technique. In their study, VMAT has shown a more conformal dose distribution with a better or comparable OAR-sparing ability.

This study revealed that VMAT required more MUs than IMRT in HNC. Moreover, Nithya et al. [22] have reported similar findings in a comparative study between VMAT and IMRT for tongue cancer using the Monaco TPS. Additionally, Nithya et al. [22] have highlighted a dip in the literature about comparative studies for HNC using the Monaco TPS compared with Eclipse (Varian Medical System, Palo Alto, CA, USA) and other TPS. Several studies have reported that VMAT requires lesser MUs than IMRT for HNC [21,22]. HNC planning requires a high degree of modulation for a clinically acceptable treatment plan. Additionally, HNC presents a unique challenge of concave dose distribution along with close proximity of OARs to the target volume. The number of MUs depends on the size of the target and the degree of modulation employed in the optimization process for OAR sparing [23].

This study revealed that VMAT offers lesser NTID, mean dose to ring structure (i.e., 5 cm beyond the PTV), and dose volumes (ranging from 1 Gy to 30 Gy) to normal tissues than IMRT. The dose outside the treatment field depends on the target size and increases with an increase in the target volume because an increase in the target volume leads to more irradiation of the tissues along the target length [24]. The scatter contribution from the collimator and patient body along with softer energy spectrum outside the treatment field (compared with within the treatment field) plays a vital role in the irradiation dose to normal tissues outside the target volume [25,26]. Moreover, Fig. 3 shows a gradual decrease in the mean volume (V) encompassed by a certain dose (D) on a logarithmic scale for an increase in the dose and reveals that VMAT plans have a higher dose falloff than IMRT plans for supraglottic larynx, hard palate, and tongue cancers. This can be because VMAT offers more freedom in the optimization process restricting radiation doses to nearby OARs and normal tissues outside the target.

Figure 3. The rate of dose falloff beyond the target by plotting ln (D) versus ln (1/V) for (a) supraglottic larynx, (b) hard palate, and (c) tongue cancers. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.

Hall [27] and Hall and Wuu [28] advocated that a low-dose volume may not cause acute or subacute clinical morbidity but could be carcinogenic and reported that IMRT probably has a 1% to 1.75% higher risk of secondary malignancies than conventional radiotherapy in patients surviving for 10 years. Followill et al. [29] have reported an increased risk of secondary malignancies by 0.4% to 1.0% with IMRT compared with conventional RT techniques. Hauri and Schneider [30] have evaluated the impact of treatment techniques on out-of-field low-dose volumes and recommended VMAT over IMRT in terms of radiological protection to patients with cancer. Because VMAT requires less beam-on time than IMRT leads, a reduction in the out-of-field dose is expected.

Kry et al. [31] have reported that secondary malignancy is more likely to occur in areas surrounding the target receiving doses greater than 3 Gy. However, Diallo et al. [15] have reported that areas receiving doses smaller than 2.5 Gy are the prominent areas for the occurrence of secondary malignancies. In external beam radiotherapy, a larger area around the target gets exposed to radiation. Targets are exposed to a high dose of radiation to maintain the therapeutic level, and surrounding areas should be considered in plan evaluation to minimize exposure to radiation. Therefore, low-dose volumes should be carefully evaluated for the risk of secondary malignancies and for plan optimization for better consideration of nearby tissues in planning.

Conclusions

For advanced-stage HNCs, both IMRT and VMAT offer satisfactory clinical plans. For similar target coverage, VMAT offers a conformal and homogeneous dose distribution with a comparable OAR-sparing ability and a higher dose falloff outside the target volume compared with IMRT, which provides an edge over IMRT in reducing the risk of secondary malignancies.

Acknowledgements

The authors thank the management of Rajiv Gandhi Cancer Institute & Research Centre in New Delhi, India, for their continued support and encouragement to complete this research work.

Conflicts of Interest


The authors have nothing to disclose.

Availability of Data and Materials


The data that support the findings of this study are available on request from the corresponding author.

Author Contributions


Conceptualization: Gourav Kumar and Lalit Kumar. Data curation: Gourav Kumar. Formal analysis: Manindra Bhushan. Funding acquisition: None. Investigation: Pawan Kumar and Soumitra Barik. Methodology: Sandeep Purohit. Project administration: Vimal Kishore. Resources: Soumitra Barik. Software: Sandeep Purohit and Pawan Kumar. Supervision: Manindra Bhushan and Vimal Kishore. Validation: Pawan Kumar and Kothanda Raman. Visualization: Manindra Bhushan. Writing–original draft: Gourav Kumar. Writing–review & editing: Lalit Kumar.

Fig 1.

Figure 1.Comparison of dose distribution in axial and coronal planes using (a) the IMRT and (b) VMAT techniques along with their (c) DVH. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; DVH, dose-volume histograms; PTV, planning target volume.
Progress in Medical Physics 2021; 32: 70-81https://doi.org/10.14316/pmp.2021.32.3.70

Fig 2.

Figure 2.Gamma analysis of a representative patient. (a) ArcCHECK-measured plan, (b) TPS-calculated plan, (c) difference between the ArcCHECK-measured and TPS-calculated plans, and (d) measured points that matched the acceptance criteria were shown in yellow color and points that don't match acceptance criteria were shown in red color. TPS, treatment planning system.
Progress in Medical Physics 2021; 32: 70-81https://doi.org/10.14316/pmp.2021.32.3.70

Fig 3.

Figure 3.The rate of dose falloff beyond the target by plotting ln (D) versus ln (1/V) for (a) supraglottic larynx, (b) hard palate, and (c) tongue cancers. IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.
Progress in Medical Physics 2021; 32: 70-81https://doi.org/10.14316/pmp.2021.32.3.70

Table 1 Planning dose objectives for the planning target volume (PTV) and organs-at-risk

StructureParameterPlan objective
PTVD95%
D110%
D93%
=100% of prescribed dose
<10% of PTV
<1% of PTV
Spinal cordDmax<45 Gy
BrainstemDmax<54 Gy
LipsDmean<35 Gy
Contralateral parotidDmean<26 Gy
LarynxDmean<50 Gy
MandibleDmax<75 Gy

Table 2 The dosimetric parameters for PTV in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

TargetParameterSupraglottic larynxHard palateTongueP-value




IMRTVMATIMRTVMATIMRTVMATSupraHPTongue
PTV70D95% (Gy)70.0070.0070.0070.00
D98% (Gy)69.20±0.1669.07±0.1569.17±0.0868.98±0.160.0110.006
D50% (Gy)72.66±0.3473.02±0.4272.48±0.2873.10±0.360.0090.001
D2% (Gy)75.38±0.5276.13±0.6175.14±0.5176.31±0.750.0190.001
Mean (Gy)72.04±1.0772.29±1.2371.99±0.3473.01±0.340.5480.045
Dmax (Gy)78.54±1.2479.73±0.9678.60±1.1780.77±0.930.0070.001
Dmin (Gy)63.31±2.1962.11±2.2562.77±1.7461.57±2.150.1310.081
UI1.07±0.011.08±0.011.06±0.011.08±0.01<0.001<0.001
HI0.072±0.010.071±0.010.081±0.010.078±0.01<0.001<0.001
CI0.99±0.030.99±0.030.99±0.011.00±0.010.7910.093
CN0.88±0.030.90±0.030.92±0.030.93±0.030.9720.667
PTV63D95% (Gy)64.01±0.9163.91±1.180.511
D98% (Gy)62.95±0.9862.67±1.450.197
D50% (Gy)68.65±1.5269.24±1.460.001
D2% (Gy)74.69±0.7975.26±0.790.006
Mean (Gy)66.06±1.0666.19±1.620.767
Dmax (Gy)74.65±1.2875.29±0.960.192
Dmin (Gy)51.02±3.5351.35±4.050.717
UI1.15±0.011.16±0.020.102
HI0.171±0.010.168±0.020.032
CI0.99±0.030.98±0.030.341
CN0.86±0.050.88±0.060.142
PTV56D95% (Gy)57.78±0.9257.82±1.0458.83±0.8359.70±1.110.7710.055
D98% (Gy)56.94±0.8256.79±1.0357.80±0.7258.25±0.940.4230.066
D50% (Gy)63.75±3.6464.28±3.7267.40±3.3368.29±2.720.0030.011
D2% (Gy)70.63±5.7171.34±5.6274.12±1.5375.13±1.770.0020.002
Mean (Gy)58.93±1.1759.04±1.4762.51±0.9164.11±1.090.7690.002
Dmax (Gy)71.35±2.5070.92±2.7375.61±1.1476.55±0.990.4070.008
Dmin (Gy)46.54±3.6045.82±2.8945.65±3.2846.42±1.790.3780.518
UI1.21±0.091.11±0.371.24±0.031.24±0.040.3910.576
HI0.218±0.080.214±0.070.232±0.020.229±0.030.0510.302
CI1.00±0.011.00±0.011.00±0.010.99±0.010.8920.271
CN0.72±0.210.75±0.210.90±0.110.92±0.12<0.0010.113
PTV60D95% (Gy)60.0060.00
D98% (Gy)58.93±0.1858.76±0.25<0.001
D50% (Gy)62.84±0.3463.51±0.45<0.001
D2% (Gy)65.44±0.5366.81±0.74<0.001
Mean (Gy)62.69±0.3163.35±0.41<0.001
Dmax (Gy)69.43±1.1371.67±1.19<0.001
Dmin (Gy)50.49±2.2749.72±3.020.871
UI1.08±0.011.1±0.01<0.001
HI0.126±0.010.123±0.01<0.001
CI0.99±0.0010.99±0.010.244
CN0.88±0.040.89±0.040.478
PTV54D95% (Gy)54.69±0.4354.48±0.530.072
D98% (Gy)53.89±0.5453.38±0.580.001
D50% (Gy)57.23±1.3157.91±1.380.001
D2% (Gy)61.50±2.7862.69±2.900.001
Mean (Gy)56.81±0.9057.08±0.590.228
Dmax (Gy)64.16±1.4264.76±0.990.046
Dmin (Gy)43.06±4.1243.93±2.890.115
UI1.10±0.051.13±0.05<0.001
HI0.161±0.050.159±0.05<0.001
CI1.00±0.0010.99±0.010.078
CN0.69±0.180.72±0.180.101

Values are presented in mean±standard deviation.

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume; UI, uniformity index; HI, homogeneity index; CI, conformity index; CN, conformation number.


Table 3 The dosimetric parameters for OARs in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

SiteOARsParameter (Gy)IMRTVMATP-value
Supraglottic larynxContra lateral parotidDmean22.76±1.2423.41±1.21<0.001
D50%18.22±2.2818.31±0.810.853
BrainstemDmax33.88±2.5736.21±2.53<0.001
Spinal cordDmax32.77±1.6333.92±1.580.009
LipsDmean32.71±3.7434.65±2.960.016
LarynxDmean49.68±1.9150.47±1.520.004
MandibleDmax75.83±2.6176.46±2.190.102
Ring (5 cm beyond PTV)Dmean35.89±3.3235.48±3.040.301
Hard palateContra lateral parotidDmean24.64±2.3725.43±2.690.059
D50%20.24±0.9621.75±0.750.001
BrainstemDmax34.69±3.6535.68±3.050.536
Spinal cordDmax34.54±2.3536.38±2.360.029
LipsDmean34.72±1.9135.86±1.570.093
LarynxDmean47.84±1.0449.13±0.820.011
MandibleDmax77.02±1.8477.32±1.990.394
Ring (5 cm beyond PTV)Dmean36.44±3.7935.72±4.520.062
TongueContra lateral parotidDmean23.61±1.7124.76±1.76<0.001
D50%19.09±1.7620.77±1.65<0.001
BrainstemDmax30.64±3.2930.23±3.870.671
Spinal cordDmax35.44±1.7635.65±2.250.656
LipsDmean34.59±2.4135.33±1.960.206
LarynxDmean48.46±1.6648.71±2.550.527
MandibleDmax67.41±0.9468.78±1.28<0.001
Ring (5 cm beyond PTV)Dmean32.10±1.5831.38±1.710.001

Values are presented as mean±standard deviation.

OARs, organs-at-risk; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy; PTV, planning target volume.


Table 4 The MUs and NTID in supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

SiteMUsNTID


IMRTVMATP-valueIMRTVMATP-value
Supraglottic larynx871.59±66.70875.64±84.560.9713.77±1.083.60±1.080.003
Hard palate854.57±80.32876.35±88.650.3074.53±1.833.95±1.350.025
Tongue915.35±79.48984.64±123.01<0.0013.35±1.093.09±0.91<0.001

Values are presented as mean±standard deviation.

MU, monitor unit; NTID, integral dose to normal tissue; IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.


Table 5 The low-dose spillage analysis for supraglottic larynx, hard palate, and tongue cancers using the IMRT and VMAT techniques

ParameterSupraglottic larynxHard palateTongue



IMRTVMATP-valueIMRTVMATP-valueIMRTVMATP-value
D1% (Gy)40.67±10.2440.39±11.580.80740.32±12.1441.30±12.260.31640.62±11.7539.42±12.620.005
D2% (Gy)27.34±5.0727.6±6.760.76028.67±6.4828.71±7.640.97326.33±7.6624.91±8.110.006
D5% (Gy)19.31±5.3817.68±5.200.06220.07±6.8417.35±6.830.01015.68±4.5713.40±3.990.000
V1Gy (cc)7,993.9±2,173.18,046.7±2,177.00.2126,015.3±3,032.35,979.8±2,958.50.3826,442.9±1,306.56,515.3±1,322.50.070
V2Gy (cc)5,553.7±1,792.15,559.5±1,763.00.8714,260.9±2,294.64,218.6±2,228.50.5154,369.8±1,065.04,349.3±1,058.70.515
V3Gy (cc)4,597.8±1,681.44,531.9±1,602.10.1663,552.1±2,016.33,453.6±1,907.70.1683,559.2±1,029.53,423.7±1,011.10.000
V4Gy (cc)4,013.3±1,615.03,925.9±1,514.70.0913,095.2±1,835.62,972.5±1,733.10.0603,023.7±1,020.12,875.77±994.10.000
V5Gy (cc)3,572.8±1,520.63,466.8±1,402.10.0692,761.0±1,673.72,629.2±1,622.10.0272,616.1±990.72,500.3±969.80.001
V10Gy (cc)2,309.1±1,194.32,049.5±1,081.00.0071,822.7±1,230.61,482.8±1,058.90.0031,609.47±791.11,320.3±595.20.000
V15Gy (cc)1,703.2±981.41,205.5±692.70.0001,326.5±904.4844.7±568.80.0101,082.2±591.2712.3±304.20.000
V20Gy (cc)975.6±511.3747.6±414.20.002830.7±547.2542.2±323.870.024578.58±298.1452.1±198.50.002
V25Gy (cc)518.8±234.0498.1±254.70.570450.4±262.3359.2±179.10.085348.2±148.8313.4±149.70.028
V30Gy (cc)294.1±90.1332.36±153.60.165252.28±134.5243.0±100.10.719245.0±104.2232.8±123.10.225

Values are presented as mean±standard deviation.

IMRT, intensity-modulated radiation therapy; VMAT, volumetric modulated arc therapy.


References

  1. Bhattacharjee A, Chakraborty A, Purkaystha P. Prevalence of head and neck cancers in the North East-an institutional study. Indian J Otolaryngol Head Neck Surg. 2006;58:15-19.
    Pubmed KoreaMed CrossRef
  2. Sharma JD, Baishya N, Kataki AC, Kalita CR, Das AK, Rahman T. Head and neck squamous cell carcinoma in young adults: a hospital-based study. Indian J Med Paediatr Oncol. 2019;40(S1):S18-S22.
  3. Grégoire V, Lefebvre JL, Licitra L; Felip E; EHNS-ESMO-ESTRO Guidelines Working Group. Squamous cell carcinoma of the head and neck: EHNS-ESMO-ESTRO Clinical Practice Guidelines for diagnosis, treatment and follow-up. Ann Oncol. 2010;21 Suppl 5:v184-v186.
    Pubmed CrossRef
  4. Vigneswaran N, Williams MD. Epidemiologic trends in head and neck cancer and aids in diagnosis. Oral Maxillofac Surg Clin North Am. 2014;26:123-141.
    Pubmed KoreaMed CrossRef
  5. Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68:394-424.
    Pubmed CrossRef
  6. Gupta T, Agarwal J, Jain S, Phurailatpam R, Kannan S, Ghosh-Laskar S, et al. Three-dimensional conformal radiotherapy (3D-CRT) versus intensity modulated radiation therapy (IMRT) in squamous cell carcinoma of the head and neck: a randomized controlled trial. Radiother Oncol. 2012;104:343-348.
    Pubmed CrossRef
  7. Ge X, Liao Z, Yuan J, Mao D, Li Y, Yu E, et al. Radiotherapy-related quality of life in patients with head and neck cancers: a meta-analysis. Support Care Cancer. 2020;28:2701-2712.
    Pubmed CrossRef
  8. DE Felice F, Pranno N, Papi P, Brugnoletti O, Tombolini V, Polimeni A. Xerostomia and clinical outcomes in definitive intensity modulated radiotherapy (IMRT) versus three-dimensional conformal radiotherapy (3D-CRT) for head and neck squamous cell carcinoma: a meta-analysis. In Vivo. 2020;34:623-629.
    Pubmed KoreaMed CrossRef
  9. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310-317.
    Pubmed CrossRef
  10. Holt A, Van Gestel D, Arends MP, Korevaar EW, Schuring D, Kunze-Busch MC, et al. Multi-institutional comparison of volumetric modulated arc therapy vs. intensity-modulated radiation therapy for head-and-neck cancer: a planning study. Radiat Oncol. 2013;8:26.
    Pubmed KoreaMed CrossRef
  11. Studenski MT, Bar-Ad V, Siglin J, Cognetti D, Curry J, Tuluc M, et al. Clinical experience transitioning from IMRT to VMAT for head and neck cancer. Med Dosim. 2013;38:171-175.
    Pubmed CrossRef
  12. Verbakel WF, Cuijpers JP, Hoffmans D, Bieker M, Slotman BJ, Senan S. Volumetric intensity-modulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study. Int J Radiat Oncol Biol Phys. 2009;74:252-259.
    Pubmed CrossRef
  13. Kumar L, Yadav G, Raman K, Bhushan M, Pal M. The dosimetric impact of different photon beam energy on RapidArc radiotherapy planning for cervix carcinoma. J Med Phys. 2015;40:207-213.
    Pubmed KoreaMed CrossRef
  14. Kumar L, Yadav G, Samuvel KR, Bhushan M, Kumar P, Suhail M, et al. Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma. Rep Pract Oncol Radiother. 2017;22:10-18.
    Pubmed KoreaMed CrossRef
  15. Diallo I, Haddy N, Adjadj E, Samand A, Quiniou E, Chavaudra J, et al. Frequency distribution of second solid cancer locations in relation to the irradiated volume among 115 patients treated for childhood cancer. Int J Radiat Oncol Biol Phys. 2009;74:876-883.
    Pubmed CrossRef
  16. Nagarajan M, Banu R, Sathya B, Sundaram T, Chellapandian TP. Dosimetric evaluation and comparison between volumetric modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) plan in head and neck cancers. Gulf J Oncolog. 2020;1:45-50.
    Pubmed
  17. Osborn J. Is VMAT beneficial for patients undergoing radiotherapy to the head and neck? Radiography (Lond). 2017;23:73-76.
    Pubmed CrossRef
  18. ICRU Report 83. Prescribing, recording, and reporting intensity-modulated photon-beam therapy (IMRT). ICRU Report. Bethesda: ICRU. 2010: 83.
  19. IMPAC Medical Systems, Inc. Monaco® training guide. Stockholm: Elekta AB; 2013.
  20. Kumar L, Yadav G, Kishore V, Bhushan M, Gairola M, Tripathi D. Validation of the RapidArc delivery system using a volumetric phantom as per task group report 119 of the American Association of Physicists in Medicine. J Med Phys. 2019;44:126-134.
    Pubmed KoreaMed CrossRef
  21. Radhakrishnan S, Chandrasekaran A, Sarma Y, Balakrishnan S, Nandigam J. Dosimetric comparison between single and dual arc-volumetric modulated arc radiotherapy and intensity modulated radiotherapy for nasopharyngeal carcinoma using a simultaneous integrated boost technique. Asian Pac J Cancer Prev. 2017;18:1395-1402.
    Pubmed KoreaMed CrossRef
  22. Nithya L, Raj NA, Kumar A, Rathinamuthu S, Pandey MB. Comparative analysis of volumetric-modulated arc therapy and intensity-modulated radiotherapy for base of tongue cancer. J Med Phys. 2014;39:121-126.
    Pubmed KoreaMed CrossRef
  23. Kumar L, Bhushan M, Kishore V, Yadav G, Gurjar OP. Dosimetric validation of Acuros® XB algorithm for RapidArc™ treatment technique: a post software upgrade analysis. J Cancer Res Ther. 2021. doi: 10.4103/jcrt.JCRT_1154_19.
  24. Stovall M, Blackwell CR, Cundiff J, Novack DH, Palta JR, Wagner LK, et al. Fetal dose from radiotherapy with photon beams: report of AAPM Radiation Therapy Committee Task Group No. 36. Med Phys. 1995;22:63-82.
    Pubmed CrossRef
  25. Edwards CR, Mountford PJ. Near surface photon energy spectra outside a 6 MV field edge. Phys Med Biol. 2004;49:N293-N301.
    Pubmed CrossRef
  26. Gayen S, Kombathula SH, Manna S, Varshney S, Pareek P. Dosimetric comparison of coplanar and non-coplanar volumetric-modulated arc therapy in head and neck cancer treated with radiotherapy. Radiat Oncol J. 2020;38:138-147.
    Pubmed KoreaMed CrossRef
  27. Hall EJ. Intensity-modulated radiation therapy, protons, and the risk of second cancers. Int J Radiat Oncol Biol Phys. 2006;65:1-7.
    Pubmed CrossRef
  28. Hall EJ, Wuu CS. Radiation-induced second cancers: the impact of 3D-CRT and IMRT. Int J Radiat Oncol Biol Phys. 2003;56:83-88.
    Pubmed CrossRef
  29. Followill D, Geis P, Boyer A. Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy. Int J Radiat Oncol Biol Phys. 1997;38:667-672.
    Pubmed CrossRef
  30. Hauri P, Schneider U. Whole-body dose equivalent including neutrons is similar for 6 MV and 15 MV IMRT, VMAT, and 3D conformal radiotherapy. J Appl Clin Med Phys. 2019;20:56-70.
    Pubmed KoreaMed CrossRef
  31. Kry SF, Bednarz B, Howell RM, Dauer L, Followill D, Klein E, et al. AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy. Med Phys. 2017;44:e391-e429.
    Pubmed CrossRef
Korean Society of Medical Physics

Vol.35 No.2
June 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line