검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Original Article

Korean Journal of Medical Physics 2016; 27(3): 125-130

Published online September 30, 2016

Copyright © Korean Society of Medical Physics.

Optimization of Energy Modulation Filter for Dual Energy CBCT Using Geant4 Monte-Carlo Simulation

Eun Bin Ju*, So Hyun Ahn, Sang Gyu Choi, Rena Lee§

*Department of Medical Science, Ewha Womans University, Department of Radiation Oncology, School of Medicine, Yonsei University, Seoul, Department of Radiation Oncology, School of Medicine, Dankook University, Cheonan, §Department of Biomedical Engineering, School of Medicine, Ewha Womans University, Seoul, Korea

Received: September 6, 2016; Revised: September 22, 2016; Accepted: September 23, 2016

Abstract

Dual energy computed tomography (DECT) is used to classify two materials and quantify the mass density of each material in the human body. An energy modulation filter based DECT could acquire two images, which are generated by the low- and high-energy photon spectra, in one scan, with one tube and detector. In the case of DECT using the energy modulation filter, the filter should perform the optimization process for the type of materials and thicknesses for generating two photon spectra. In this study, Geant4 Monte-Carlo simulation toolkit was used to execute the optimization process for determining the property of the energy modulation filter. In the process, various materials used for the energy modulation filter are copper (Cu, 8.96 g/cm3), niobium (Nb, 8.57 g/cm3), stannum (Sn, 7.31 g/cm3), gold (Au, 19.32 g/cm3), and lead (Pb, 11.34 g/cm3). The thickness of the modulation filter varied from 0.1 mm to 1.0 mm. To evaluate the overlap region of the low- and high-energy spectrum, Geant4 Monte-Carlo simulation is used. The variation of the photon flux and the mean energy of photon spectrum that passes through the energy modulation filter are evaluated. In the primary photon spectrum of 80 kVp, the optimal modulation filter is a 0.1 mm lead filter that can acquire the same mean energy of 140 kVp photon spectrum. The lead filter of 0.1 mm based dual energy CBCT is required to increase the tube current 4.37 times than the original tube current owing to the 77.1% attenuation in the filter.

KeywordsDual energy CT, Modulation filter, Monte carlo simulation

Korean Society of Medical Physics

Vol.35 No.3
September 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line