Korean Journal of Medical Physics 2004; 15(3): 156-160
Published online September 25, 2004
Copyright © Korean Society of Medical Physics.
Su Jin Lee*, Re Na Lee*, Byong Yong Yi†, Sang Wook Lim†, and Jin Ho Choi‡
이수진*·이레나*·이병용†·임상욱†·최진호‡
An essential quality assurance (QA) procedure in high dose rate (HDR) remote after-loading brachytherapy is that of the verification of the Ir-192 HDR source positioning accuracy. A number of methods using mechanical rulers or autoradiograph and video cameras have been reported to check the positional error of the Ir-192 source. In this study, the feasibility of a CMOS (Complementary Metal Oxide Semiconductor) PC camera, with a fluorescent screen, was investigated. The agreement between the planned and measured dwell position was better than 1 mm and dwell times better than 0.4 sec. Our results indicate that the CMOS PC camera system could be used as a QA tool for the on-line determination of the source position and dwell time.
KeywordsHigh dose rate brachytherapy, QA, Source positioning accuracy