Korean Journal of Medical Physics 2004; 15(4): 173-178
Published online December 25, 2004
Copyright © Korean Society of Medical Physics.
Joo Young Kim, Sung Yong Park, Doo Hyun Lee, Seok Ho Lee, Tae Hyun Kim, and Kwan Ho Cho
김주영·박성용·이두현·이석호·김태현·조관호
Purpose: To compare desimetrically intensity-modulated radiotherapy treatment plans with commercially available multileaf collimators (MLCs) of different leaf width for intracranial lesions. Materials and Methods: Twelve patients with intracranial lesions were treated with BrainLAB's micro-MLCs (mMLCs) and performed with the BrainSCAN ver. 5.2 planning software. They were replanned using the Varian 120 and 80 MLCs. These collimators have minimum leaf width of 3 mm, 5 mm and 10 mm at isocenter, respectively. PTV was 3.3∼339.2 cm3 and the number of beams was 3∼7. These three plans were compared with respect to the uniformity and the conformity indices, doses to critical organ and normal tissue. Results: For the uniformity index of the planning target volume (PTV), there were no statistically significant differences between mMLCs and 120 MLCs (p=0.057) and between 120 MLCs and 80 MLCs (p=0.388). However, there was a difference between mMLCs and 80 MLCs (p<0.001). Maximum target dose to the PTV showed no dependency with respect to the leaf width. On the contrary, there were statistically significant differences in the conformity indices between mMLCs and 120 MLCs (p=0.003), between mMLCs and 80 MLCs (p=0.003) and between 120 MLCs and 80 MLCs (p=0.003). The volume of brainstem irradiated to ≥70% dose and to ≥50% dose was increased as the leaf width of MLCs increased. In particular, the volume of normal tissue irradiated is obviously changed for different leaf width. Volumetric increments for MLCs with leaf widths of 5 mm and 10 mm were 6.3% and 23.2% to the normal tissue irradiated to ≥50% dose, and 8.7% and 32.7% to the normal tissue irradiated to ≥70% dose, respectively, compared to the volume for MLCs with leaf width of 3 mm. Conclusions: The uniformity index and maximum target dose to the PTV showed no dependency with respect to leaf width of MLCs. However, the conformity index was improved as the leaf width decreased. For the sparing of normal brain tissue, treatment plans with MLCs of 3 mm leaf width is more effective, compared to ones with MLCs of 5 mm and 10 mm leaf widths.
KeywordsMultileaf collimators (MLCs), Uniformity index, Conformity index, Sparing of critical organ and normal brain tissue
pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829
Frequency: Quarterly