Korean Journal of Medical Physics 2004; 15(4): 192-196
Published online December 25, 2004
Copyright © Korean Society of Medical Physics.
Hyun Do Huh*†, Sung Yong Park‡, Rena J Lee§, Dong Oh Shin∥, Soo Il Kwon†, John J K Loh*, and Jinho Choi¶
허현도*†·박성용‡·이레나§·신동오∥·권수일†·노준규*·최진호¶
The activity of Ir-192 sources for high dose rate (HDR) Brachytherapy in Korea were measured by using the well-type chamber and using the calibration Jig with the Farmer-type ionization chamber to compare the manufacturer certificated source strength which is supplied with each new Ir-192 source. The activity of two different source models used in six hospitals were measured. The range of measured activities to the manufacturer's suggested ones was -2.40% to +3.31% for the calibration Jig and -3.12% to 0.00% for the well-type chamber system. The source strength values given by the manufacturer for the 6 sources were within ±5% for the two different measuring equipment. Our results demonstrate that well-type chamber as wall as Farmer-type chamber system are appropriate system for the routine source calibration procedures in HDR brachytherapy. Whenever a new source is installed to use in clinics, a source calibration should be carried out.
KeywordsStrength of source, Ir-192 source, Zig phantom system, Well-type chamber