Ex) Article Title, Author, Keywords
Ex) Article Title, Author, Keywords
Korean Journal of Medical Physics 2004; 15(4): 197-201
Published online December 25, 2004
Copyright © Korean Society of Medical Physics.
Dong Joon Lee*, and Do Heui Lee†
이동준*·이도희†
Stereotactic brain biopsy using stereotactic head frame such as CRW (Radionics, USA) has demonstrated a precise lesion localizing accuracy. In this study, we developed the target point calculation program for brain lesion biopsy using CRW stereotactic head frame and designed a phantom for verify the new developed program. The phantom was designed to have capability to simulate clinical stereotactic brain biopsy. The phantom has 10 vertical rods whose diameters are 6mm and tip of each rods are 2mm. Each rod has different length, 150 mm×4 ea, 130 mm×4 ea, 110 mm×2 ea. CT images were acquired with Simens CT scanner as continuous transverse slice, 1 mm thickness in a 25 cm field of view and stored in a dicom file as a 256×256 matrix. As a result, the developed new target localization program will be useful for planning and training in complicated 3 dimensional stereotactic brain biopsy.
KeywordsStereotactic brain biopsy, Target localization program
pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829
Frequency: Quarterly