Korean Journal of Medical Physics 2004; 15(2): 77-83
Published online June 25, 2004
Copyright © Korean Society of Medical Physics.
Sei-kwon Kang, Byung-chul Cho, Hee-chul Park, and Hoonsik Bae
강세권·조병철·박희철·배훈식
The commissioning of a model-based treatment planning system requires many parameters to fit the measured depth doses and transverse profiles. For the commissioning of the Pinnacle3 system, through the Monte Carlo (MC) simulation, the necessary parameters, including the photon spectrum, contaminant electrons, off-axis softening and fluency of photons, were observed. Through the simulation the parameters contained valuable information, but the calculated results of the Pinnacle3 using the MC-derived parameters showed discrepancies with those measured for the off-axis softening and the fluency of photons. Even though the MC calculation produces reasonable values for the commissioning, the thorough physical basis of the Pinnacle3's commissioning process is needed in order to directly use the MC derived parameters.
KeywordsRTP commissioning, Monte Carlo, Convolution/superposition algorithm