The Effects of the CT Voltages on the Dose Calculated by a Commercial RTP System

CT 관전압이 상용 전산화치료계획장치의 선량계산에 미치는 영향

Sei-kwon Kang, Byung-chul Cho, Hee-chul Park, and Hoonsik Bae

강세권·조병철·박희철·배훈식

Department of Radiation Oncology, Hallym University College of Medicine

한림대학교 의과대학, 한림대학교 성심병원 방사선종양학과

Abstract

The relationship between the dose calculated with a radiotherapy treatment planning system (RTPS) and CT number verses the relative electron density curve was investigated for various CT voltages and beam qualities. We obtained the relationship between the CT numbers and electron densities of the tissue equivalent materials for various CT voltages and beam qualities. At lower CT voltages, the higher density materials, like cortical bone, showed larger CT numbers and the soft tissues showed no variations. We performed a phantom study in a RTPS, where a phantom consisted of lung and bone regions in water. We calculated the dose received behind the lung and bone regions for 6 MV photon beams, in which the regions below the lung, water and bone received higher doses in this listed order. The result was the same for 10 MV photon beams. For the clinical application, the doses were calculated for the lung and pelvis. No difference was observed when using different electron density conversion tables with various CT voltages from a same CT. A relative dose difference of 1.5% was obtained when the CT machine for the density conversion table was different from that for the CT image for planning.

KeywordsRadiotherapy treatment planning system (RTPS), Computed tomography (CT), Electron density phantom, CT voltage

Korean Society of Medical Physics

Vol.35 No.4
December 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line