검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Original Article

Korean Journal of Medical Physics 2004; 15(1): 39-44

Published online March 25, 2004

Copyright © Korean Society of Medical Physics.

The Calculation of Energy Distributions for Clinical Electron

Beams from Mono Energetic Depth dose Data

단일에너지 깊이선량률 자료에 의한 치료용 전자선의 에너지분포 계산

Jeong Ok Lee*, and Dong Hyeok Jeong

이정옥*·정동혁

*Department of Radiotechnology, Wonkwang Health Science College,Department of Radiation Oncology, Wonkwang University Hospital

*원광보건대학 방사선과, 원광대학병원 방사선종양학과

Abstract

The energy distributions for clinically used electron beams from measured and calculated mono energetic depth dose values were calculated. The energy distributions having the minimum difference between the measured and reduced values of depth dose are determined by iterations based on least square method. The nominal energies of 6, 9, 12, 15 MeV clinical electron beams were examined. The Monte Carlo depth dose calculations with determined energy distributions were performed to evaluate those distributions. In a comparison of the calculated and measured depth dose data, the standard errors are estimated within ±3% from surface to R80 depth and within ±4% from the surface to near the range for all electron beams. This can be practically applied to determine the energy distributions for clinically used electron beams.

KeywordsEnergy distribution, Electron beam, Monte Carlo

Korean Society of Medical Physics

Vol.35 No.2
June 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line