Korean Journal of Medical Physics 2006; 17(2): 67-76
Published online June 25, 2006
Copyright © Korean Society of Medical Physics.
Kwang-Ho Cheong*†, Tae-Suk Suh*, Byung-Chul Cho†, Sei-Kwon Kang†, Kyoung-Joo Kim†, Hoon-Sik Bae†
정광호*†·서태석*·조병철†·강세권†·김경주†·배훈식†
We evaluated the accuracy of a patient setup error correction due to reference image quality for a 2D-2D matching process. Digitally reconstructed radiographs (DRRs) generated by use of the Pinnacle3 and the Eclipse for various regions of a humanoid phantom and a patient for different CT slice thickness were employed as a reference images and kV X-ray images from the On-Board Imager were registered to the reference DRRs. In comparison of the DRRs and profiles, DRR image quality was getting worse with an increase of CT image slice thickness. However there were only slight differences of setup errors evaluation between matching results for good and poor reference DRRs. Although DRR image quality did not strongly affect to the 2D-2D matching accuracy, there are still potential errors for matching procedure, therefore we recommend that DRR images are needed to be generated with less than 3mm slice thickness for 2D-2D matching.
KeywordsComputed tomography (CT), Digitally reconstructed radiographs (DRR), 2D-2D matching, On-board imager (OBI)