Kyoung Won Jang*, Dong Hyun Cho*, Sang Hun Shin*, Wook Jae Yoo*, Jae Hun Jun*, Bongsoo Lee*, Joo Hyun Moon†, Byung Gi Park‡
장경원*ㆍ조동현*ㆍ신상훈*ㆍ유욱재*ㆍ전재훈*ㆍ이봉수*ㆍ문주현†ㆍ박병기‡
In this study, we have fabricated a one-dimensional fiber-optic dosimeter for electron beam therapy dosimetry. Each fiber-optic dosimeter has an organic scintillator with a plastic optical fiber and it is embedded and arrayed in the plastic phantom to measure one-dimensional high energy electron beam profile of clinical linear accelerator. The scintillating lights generated from each sensor probe are guided by plastic optical fibers to the multi-channel photodiode amplifier system. We have measured one-dimensional electron beam profiles in a PMMA phantom according to different field sizes and energies of electron beam. Also, the isodose and three-dimensional percent depth dose curves in a PMMA phantom are obtained using a one-dimensional fiber-optic dosimeter with different electron beam energies.
KeywordsDosimeter, Optical fiber, Organic scintillator, Electron beam