Korean Journal of Medical Physics 2010; 21(2): 232-237

Published online June 25, 2010

Copyright © Korean Society of Medical Physics.

Development of Lead Free Shielding Material for Diagnostic Radiation Beams

의료영상용 방사선방호를 위한 무납차폐체 개발

Tae Jin Choi*, Young Kee Oh, Jin Hee Kim, Ok Bae Kim

최태진*ㆍ오영기ㆍ김진희ㆍ김옥배

Departments of *Medical Bio-Physics Engineering, Radiation Oncology, School of Medicine, Keimyung University, Daegu, Korea

계명대학교 의과대학 *의공학교실, 방사선종양학교실

Abstract

The shielding materials designed for replacement of lead equivalent materials for lighter apron than that of lead in diagnostic photon beams. The absorption characteristics of elements were applied to investigate the lead free material for design the shielding materials through the 50 kVp to 110 kVp x-ray energy in interval of 20 kVp respectively. The idea focused to the effect of K-edge absorption of variable elements excluding the lead material for weight reduction. The designed shielding materials composited of Tin 34.1%, Antimon 33.8% and Iodine 26.8% and Polyisoprene 5.3% gram weight account for 84 percent of weight of lead equivalent of 0.5 mm thickness. The size of lead-free shielder was 200×200×1.5 mm3 and 3.2 g/cm3 of density which is equivalent to 0.42 mm of Pb. The lead equivalent of 0.5 mm thickness generally used for shielding apron of diagnostic X rays which is transmitted 0.1% for 50 kVp, 0.9% for 70 kVp and 3.2% for 90 kVp and 4.8% for 110 kVp in experimental measurements. The experiment of transmittance for lead-free shielder has showed 0.3% for 50 kVp, 0.6% for 70 kVp, 2.0% for 90 kVp and 4.2% for 110 kVp within

±
0.1%. respectively. Using the attenuation coefficient of experiments for 0.5 mm Pb equivalent of lead-free materials showed 0.1%. 0.3%, 1.0% and 2.4%, respectively. Furthermore, the transmittance of lead-free shielder for scatter rays has showed the 2.4% in operation energy of 50 kVp and 5.9% in energy of 110 kVp against 2.4% and 5.1% for standard lead thickness within
±
0.2% discrepancy, respectively. In this experiment shows the designed lead-free shielder is very effective for reduction the apron weight in diagnostic radiation fields.

KeywordsRadiation protection, Lead-free shielding material, Spectrum of x-ray beam, Apron