Korean Journal of Medical Physics 2011; 22(1): 12-17

Published online March 25, 2011

Copyright © Korean Society of Medical Physics.

A Correction Method of Dose to Attenuation Rate of Transmitting Photon Beam Through Couch Top for Radiosurgery Using Novalis

노발리스를 이용한 방사선 수술시 치료테이블을 투과하는 광자선의 감쇠율에 따른 선량 보정방법

Sung Joon Kim*, Hyun Soo Shin, Seung Young Ko, Hyeli Park, Ja Young Kim, Bo Mi Lee, Ji Woon Yea, Sung Kyu Kim

김성준*ㆍ신현수ㆍ고승영ㆍ박혜리ㆍ김자영ㆍ이보미ㆍ예지원ㆍ김성규

*Department of Physics, College of Science, Yeungnam University, Gyongsan, Department of Therapeutic Radiology & Oncology, CHA Bundang Medical Center, CHA University, Seongnam, Department of Therapeutic Radiology & Oncology, College of Medicine, Yeungnam University, Daegu, Korea

*영남대학교 이과대학 물리학과, 차의과학대학교 분당차병원 방사선종양학과, 영남대학교 의과대학 방사선종양학교실

Abstract

This study has its own goal to deliver the accurate dose on the target volume by calculating and modifying the attenuation rate of photon beam transmitting the couch top with geometric model. The experiment was that the transmission rate and attenuation rate of photon beam transmitting the couch top was predicted by the geometric model, then compared and analyzed with what was measured experimentally based on that. The result showed that the predicted value by the geometric model accorded closely with the experimental value. In addition, in order to judge whether the practical clinical application is available, the point dose, measured after modifying the attenuation rate modelinged according to the treatment plan of a patient of spine radiosurgery, was compared with the one done nothing. The result was that the former showed decreased error range with treatment planned one than the latter. This papers calculated the transmission and attenuation rate with the geometric model transmitting the couch top and verified it experimentally. This method is expected to be very useful in not only the radiosurgery using Novalis but also the general radiation therapy.

KeywordsNovalis, Couch top, Radiosurgery, Transmission rate, Attenuation rate