Korean Journal of Medical Physics 2011; 22(2): 79-84

Published online June 25, 2011

Copyright © Korean Society of Medical Physics.

Study on the 6 MV Photon Beam Characteristics and Analysis Method from Medical Linear Accelerators Using Geant4 Medical Linac2 Example

GEANT4 Medical Linac2 예제를 이용한 6 MV 선형가속기 광자선속의 기초특성과 연구방법

Byung Yong Kim*, Hyung Dong Kim*, Sung Jin Kim*, Se An Oh*, Jung Gu Kang, Sung Kyu Kim

김병용*ㆍ김형동*ㆍ김성진*ㆍ오세안*ㆍ강정구ㆍ김성규

*Department of Physics, College of Science, Yeungnam University, Gyongsan,Department of Radiation Oncology, Presbyterian Medical Center, Jeonju,Department of Therapeutic Radiology & Oncology, College of Medicine, Yeungnam University, Daegu, Korea

*영남대학교 이과대학 물리학과, 전주예수병원 방사선종양학과, 영남대학교 의과대학 방사선종양학교실

Abstract

In this study, Geant4 based Monte Carlo simulations were carried out for medical linear accelerator. Modified Medical Linac2 toolkit was used for calculation. The energy spectrum, most probable energy and the photon mean energy compared with the published results using the EGS4 code. The results well agreed with published results. The calculated results of photon fluence, energy fluence and mean energy according to the radius from the centre of the beam were analyzed. Monte Carlo simulation using Medical Linac2 code is considered to be useful for analysis of medical linear accelerator. Because the calculated results varies depending on Physics List model for same head structure. It it important to choose the right model for research purpose. Monte Carlo simulation using GEANT4 Medical Linac2 is a valuable for any novice to adopt this code to the study related to 6 MV photon fluence from medical linear accelerator.

KeywordsGeant4, Monte Carlo, Photon, Spectrum