Korean Journal of Medical Physics 2013; 24(3): 171-175

Published online September 30, 2013

Copyright © Korean Society of Medical Physics.

Development of Phantom for Evaluate the Suitability of Ir-192 HDR Source with Brachytherapy Tools

근접치료용 하나로 생산 Ir-192 선원의 임상기기 적합성평가용 팬톰개발

Kyo Chul Shin*, Sang Gyu Choi*, Ki Hwan Kim, Kwang Jae Son, Dong Hyeok Jeong§, Jeung Kee Kim§

신교철*ㆍ최상규*ㆍ김기환ㆍ손광재ㆍ정동혁§ㆍ김정기§

Department of Radiation Oncology, *Dankook University Hospital, Cheonan,Chungnam National University Hospital, Korea Atomic Energy Research Institute, Daejeon, §Dongnam Institute of Radiological and Medical Sciences, Busan, Korea

*단국대학교병원, 충남대학교병원, 한국원자력연구원, §동남권원자력의학원 방사선종양학과

Abstract

Applicator of various kind of number ten kinds is used to raise from efficiency of brachytherapy to maximum. The compatibility of radiation source and applicator is very important subject for safety brachytherapy. Developed high dose rate brachytherapy source through Hanaro nuclear reactor in Korea Atomic Energy Research Institute and improve compatibility with using equipment in present. In this research, we wished to evaluate stability mechanical safety of radiation source and we developed phantom for evaluate several quality about Ir-192 sealed source that improve newly in Korea Atomic Energy Research Institute and is improved. The result for suitability of Ir-192 HDR source with brachytherapy tools that did normal operation in 2.2∼2.7 cm extent about change of equal curvature and consider change of sudden curvature that did normal operation in radius 1.5∼1.8 cm extent.

KeywordsIr-192 source, HDR, Brachytherapy, Mechanical safety