검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Original Article

Korean Journal of Medical Physics 2013; 24(4): 259-264

Published online December 30, 2013

Copyright © Korean Society of Medical Physics.

Development of DICOM Convert Program for the Geant4 Monte Carlo Simulation of the Radiotherapy

방사선치료의 Geant4 전산모사를 위한 DICOM 변환 프로그램 개발

Jeongku Kang*, Dong Joon Lee

강정구*ㆍ이동준

*Department of Radiation Oncology, Presbyterian Medical Center, Jeonju Department of Neurosurgery, Ilsan Paik Hospital, College of Medicine, Inje University, Goyang, Korea

*전주 예수병원 방사선종양학과, 인제대학교 의과대학 일산백병원 신경외과학교실

Received: November 26, 2013; Accepted: December 9, 2013

Abstract

The DICOM converter program of the Geant4 Monte Carlo simulation code for the application of radiotherapy was developed. We analysis the header part of the DICOM file and find various parameters, such as matrix size, pixel size, stored data bits, high bit, and padding values. Especially we evaluate every pixel value of the DICOM files. To conform the exact convert of the pixel values, we developed the verify program. As a result, the DICOM formats generated from difference CT vendors can be converted and verified for Genat4 calculations.

KeywordsMonte Carlo, Dicom, Radiotherapy, Geant4