Commissioning and Validation of a Dedicated Scanning Nozzle at Samsung Proton Therapy Center

Kwangzoo Chung*, Younyih Han*, Sung Hwan Ahn*, Jin Sung Kim, Hideki Nonaka

*Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, Seoul, Korea, Division of Industrial Equipment, Sumitomo Heavy Industries Ltd., Niihama, Japan

Received: December 16, 2016; Revised: December 23, 2016; Accepted: December 24, 2016

Abstract

In this study, we present the commissioning and validation results of a dedicated scanning nozzle. The dedicated scanning nozzle is installed in one of the two gantry treatment rooms at Samsung Proton Therapy Center. Following a successful completion of the acceptance test, the commissioning process including the beam data measurement for treatment planning system has been conducted. Extended measurements have been conducted as a validation of the clinical performance of the nozzle and various quality assurance protocols have been prepared.

KeywordsPencil beam scanning, Commissioning, Proton therapy