Ex) Article Title, Author, Keywords
Ex) Article Title, Author, Keywords
Progress in Medical Physics 2017; 28(3): 111-121
Published online September 30, 2017
https://doi.org/10.14316/pmp.2017.28.3.111
Copyright © Korean Society of Medical Physics.
Talat Mahmood*, Mounir Ibrahim*, Muhammad Aqeel†
Correspondence to:Talat Mahmood
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Verification of dose distribution is an essential part of ensuring the treatment planning system’s (TPS) calculated dose will achieve the desired outcome in radiation therapy. Each measurement have uncertainty associated with it. It is desirable to reduce the measurement uncertainty. A best approach is to reduce the uncertainty associated with each step of the process to keep the total uncertainty under acceptable limits. Point dose patient specific quality assurance (QA) is recommended by American Association of Medical Physicists (AAPM) and European Society for Radiotherapy and Oncology (ESTRO) for all the complex radiation therapy treatment techniques. Relative and absolute point dose measurement methods are used to verify the TPS computed dose. Relative and absolute point dose measurement techniques have a number of steps to measure the point dose which includes chamber cross calibration, electrometer reading, chamber calibration coefficient, beam quality correction factor, reference conditions, influences quantities, machine stability, nominal calibration factor (for relative method) and absolute dose calibration of machine. Keeping these parameters in mind, the estimated relative percentage uncertainty associated with the absolute point dose measurement is 2.1% (
KeywordsUncertainty, EBRT, IMRT, Point dose, QA
Uncertainty associated with each quality assurance (QA) procedure in external beam radiation therapy (EBRT) should individually be evaluated to minimise the overall uncertainty. The rationale is to minimise the uncertainty in each QA procedure is to get a desirable clinical outcome, because normal tissue complication probability (NTCP) and tumour control probability (TCP) are directly related to the actual doses received by each organ at risk (OAR) and targets.1) The uncertainty associated with the actual delivered dose to the OARs and target volumes is the reason for an increase or decrease in the variation of TCP and NTCP parameters depending on the slop of the dose response curve. The therapeutic ratio is based on the slop of the TCP and NTCP curve.2,6)
Farmer type ionisation chambers are used as the local secondary standard for absolute dosimetry in many of clinical settings. Recently, a new FC23-C (volume 0.2cc) ionisation chamber (IBA Dosimetry Germany) became available to verify the point dose measurement for IMRT treatment plans. The chamber manufacturing company provides a secondary standard calibration laboratory (SSDL) chamber calibration coefficient in terms of absorbed dose in water (
The 3D complexity of the dose in Intensity modulation radiation therapy (IMRT) treatment plan, along with the beam geometry and the resulting dose distribution, means that the QA of IMRT dose distributions needs to concentrate more on the cumulative delivered dose rather than on a the QA of individual segments contributing to the overall dose delivered. Ezzell et al.8) recommends performing a point dose measurement for IMRT treatment plans to verify the TPS computed point dose prior to patient treatment because of the complexity of the IMRT treatment plan. For point dose measurements, Low et al.9) recommended, chamber should be made of tissue equivalent material.
All the components linked to measure the absolute or relative absorbed dose for patient specific QA needs to be analysed individually and estimate the standard uncertainty associated with each component. In this study, a rectangular distribution was assumed to estimate the type B uncertainties. A rectangular distribution to estimate the type B standard uncertainties is presented in
Where
In this study, 13 H&N patients were selected. These patients were planned with the step and shoot IMRT treatment technique. Relative point dose measurement for these patients were already been measured in the CIRS H&N phantom using semiflex ionisation chamber (Serial #: 1976, Volume: 0.125 cc) with electrometer (PTW UnidosE, Serial # 090753). A sample calculation for point dose measurement using relative method for H&N IMRT treatment plans is presented in
For absolute point dose measurements, a chamber calibration factor is required in term of the absorbed dose to water. The International Atomic Energy Agency (IAEA) technical report series (TRS) 398 dosimetry protocols was used to cross calibrate the field ionisation chamber.10)
The field ionisation chamber (IBA FC23-C, Serial number 2408) was cross calibrated with the local secondary standard ionisation chamber (NE 2571, Serial number 3036) using TRS 398. The reference chamber was calibrated (20 May 2016) at the APRANSA PSSDL. A solid water phantom (Gammax, Middleton, WI 53562, USA) was used to cross calibrate the field ionisation chamber because the reference chamber is not water proof.
Beam quality factor
The CIRS H&N phantom was scanned on a Toshiba Aquilion Large Bore CT (16 Slices) to create a QA phantom in the Pinnacle 9.8 TPS (Philips Radiation Oncology System, Fitchburg, WI) with the FC23-C ionisation chamber. The chamber sensitive volume was contoured during the creation of the QA phantom in the TPS and named as “sensitive volume”.
Before mapping the original treatment plans on the CIRS H&N QA phantom, the MUs of each treatment plan were recorded in the worksheet. After the plan mapping a few modifications were applied as listed below:
Remove the CT simulator couch,
Density of the sensitive volume overridden to 1 g/cm3,
Change the prescription to fixed MUs,
Set grid size (2 mm).
Couch was removed during the treatment plan mapping from the CIRS H&N phantom because the phantom was placed on the H&N extension board during the measurement of the point dose. Same setup was also used for the relative point dose measurement. To get a uniform dose distribution across the chamber sensitive volume the phantom position was adjusted in the TPS. This adjustment continued until the TPS computed standard deviation of chamber sensitive volume was ≤ 0.8%. Dose computation was computed using the collapsed cone convolution (CCC) algorithm.
To avoid the beam penetrating through the couch a carbon fiber head and neck extension board (Type-S FixatorTM Shoulder Suppression System by CIVCO) was used to treat the H&N IMRT patients. The CIRS H&N phantom was placed in net area of the H&N extension board because it has negligible transmission factor. This area of H&N extension board was selected only for this experiment to reduce one variable which can effect on absolute point dose measurement due to attenuation of the radiation through the couch. A shift was applied to the phantom as recorded during the mapping of the treatment plan in TPS. Absolute dose of the LINAC was to correct for the day to day output variation of LINAC. Air density correction factors were calculated using measured temperature and air pressure as recommended in TRS 398. The phantom was exposed using the planned gantry and collimator angles. Absolute dose is calculated by using
For relative dose measurement methods, the same steps were followed to measured cumulative charge of all the radiation beams for each IMRT treatment plan. To convert this cumulative charge reading to dose, a nominal calibration factor (NCF) of the chamber was carried out. Detail of nominal calibration is given below.
Before measuring the nominal calibration factor (NCF), leakage measurement test was performed. To calculate the NCF, chamber was exposed three times in water phantom with 6 MV for 200 MUs at calibration set up of the LINAC at 400 MU/min dose rate.
Similarly, the chamber was also exposed in the CIRS H&N phantom at Isocenter with 6 MV for 200 MUs at 400 MU/min dose rate (Fig. 1b) and record the electrometer reading. The detail calculation of the chamber nominal calibration factor (NCF) is explained in
Several factors contribute to the uncertainty calculation in the cross calibration of the chamber using the TRS 398 dosimetry protocols. Each factor should be accounted for to calculate the total uncertainty associated with this process. The electrometer reading for reference and fields chambers are corrected for several factors like reading scale, polarity, air density, humidity, leakage, radiation background, distance and ionic recombination.
Calibration factor (
In the TRS 398 dosimetry protocol, electron stopping powers for monoenergetic photon beam data is used which is presented in International Commissioning of Radiation Units and Measurements (ICRU) report 37, with the density effect model.13) Beam quality factor (
Uncertainty associated with the air density correction factor depends on the thermometer and barometer resolution, calibration certificate and long term stability. The resolution of the thermometer and barometer used in this study are 0.1°C and 0.1 hPa respectively. Das et al.14) also reported the affect of the temperature on chamber volume and its response; it should be taken into account if the temperature difference is higher than the normal calibration temperature. An estimated type B percentage standard uncertainty associated with the air density correction factor was reported by Castro et al.15) 0.2%.
Polarity effect depends on beam quality but cylindrical chambers do not have significantly dependency on beam quality.16) The uncertainty associated with the polarity factor was estimated by accounting the machine reproducibility with the external monitor chamber. The standard deviation is 0.1% was calculated for both chambers with accounting the LINAC reproducibility. Total uncertainty was estimated 0.14% (associated with the user and standards laboratory).
Ion recombination was calculated using a two voltage method.10) The uncertainty associated with two voltage methods is based on the difference between expected value from the Boag theory and the value measured by two voltages method. The maximum deviation in polynomial fit error reported by Weinhous et al.17)60Co beam quality was used to measure the ion recombination for reference ionisation chamber in standard laboratory. Similarly, two voltages method was used to calculate the ion recombination for reference ionisation chamber. The maximum difference 0.16% was reported between two voltage method and Zankowski et al.18) model for cylindrical chambers. The combined estimated value of the ion recombination was 0.16% for field and reference chamber.
No correction for humidity was needed because the calibration certificate was referred a relative humidity of 50% and this condition was also satisfied during our measurements.12) If no correction is made for humidity effect, a maximum error of 0.3% is estimated in the range of 0% to 100% relative humidity.15) Thus the uncertainty associated due to humidity is estimated 0.17% assuming the rectangular distribution.
The reproducibility of reference conditions include the source to surface distance (SSD), depth of measurement (d), setting of the field size. The uncertainty associated with the reproducibility of the reference conditions was reported 0.4% in TRS 398.10)
The standard deviation of the charge reading of well-behaved therapy level chamber should not exceed 0.3%.19) The variation in the charge reading depends on the following factors, reproducibility, display resolution, electrometer linearity, correct nulling of the electrometer and long term stability of the chamber. The uncertainty in the charge reading for both chambers (reference and field chamber) is estimated to 0.3%.
In our case, the reference chamber and electrometer were calibrated as a system in a standard dosimetry laboratory. During the calibration, filed ionisation chamber and electrometer calibrated as a system. An electrometer calibration correction factor is applied only when the electrometer and chamber are calibrated separately.
Overall relative percentage uncertainty associated with reference and field chamber is estimated for each influenced quantity. Calculated combined total uncertainty in the chamber calibration coefficient is presented in Table 1.
Uncertainty associated with each point dose measurement method is estimated and presented in the Table 2. Uncertainty associated with the nominal calibration factor for the relative point dose measurement method is estimated 0.7% including the setup reproducibility, beam monitor system, temperature change, charge reading and long term stability of the electrometer and chamber.
Results are presented in Table 1 for all the influence quantities required for the cross calibration of the field chamber. Calibration factor of the reference ionisation chamber in terms of absorb dose in water at 60Co beam quality was obtained from the PSDL calibration certificate.12) The cross calibration coefficient of the field chamber was obtained by using TRS 398 dosimetry protocol.
The measured absolute point dose difference was within ±3% as compared to the TPS computed point dose except patient 3. Results of the measured point dose with the relative method are also presented in Table 3.
The average percentage difference between the TPS computed dose and measured absolute and relative point dose was 1.4% and 1% respectively for all H&N IMRT treatment plans Fig. 2.
Standard deviation of absolute and relative point dose measurement methods are 0.9% and 1.2% respectively. The standard deviation shows that the absolute point dose measurement method has relatively better reproducibility than the relative method.
Both methods were analysed statistically to see the difference. An independent samples t-test was conducted to compare the means of percentage point dose difference for absolute and relative methods. There was not a significant difference in the score of absolute (M=1.37, SD=0.90) and relative (M=0.95, SD=1.20) point dose measurement methods (t (24)=−0.91,
The estimation of the type B uncertainties is partly based on subjective consideration and published literature. The components associated with the cross calibration factor have been shown to be able to considerably change the uncertainty in the chamber calibration coefficient due to beam quality correction factor. In this study, beam quality correction factor is based on the theoretical calculation and it does not consider chamber to chamber disparity. The uncertainty in chamber calibration coefficient will increase if the ionisation chamber is cross calibrated with the primary standard for photon beam due to this additional step of cross calibration. According to TRS-398 dosimetry protocol, the uncertainty in dose determination increases by approximately 0.2% if the field ionisation chamber is used to determine the absolute dose.10)
The uncertainty associated with this cross calibration (FC23-C ionisation chamber) was estimated 1.7% (
The estimated extended uncertainty (
In order to demonstrate if there is a significant difference between the methods, the subtraction of the mean on one against the other can be performed and compared to the uncertainty involved. This uncertainty can be assumed to follow the summation of the individual errors in quatradure.
It can also be concluded from the results that absolute point dose measurement method does not produce results different from the relative point dose measurement method for head and neck IMRT treatment plans.
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
The study was approved by the institutional review board (IRB approval number; ETHLR.17.044).
Percentage relative standard uncertainty of the each factor associated with the field and reference ionisation chamber.
Items | Percentage relative uncertainty | |
---|---|---|
Reference chamber | Field chamber | |
0.4 | NA | |
1 | 1 | |
0.2 | 0.2 | |
0.1 | 0.1 | |
0.16 | 0.16 | |
Humidity | 0.17 | 0.17 |
Reproducibility of reference condition | 0.4 | 0.4 |
0.3 | 0.3 | |
0.12 | 0.12 | |
Percentage relative standard uncertainty ( | 1.24 | 1.18 |
Total uncertainty ( | 1.7% | |
Extended uncertainty ( | 3.4% |
Percentage relative standard uncertainty of the each factor associated with the absolute and relative point dose measurement methods and total percentage relative standard uncertainty and extended uncertainty.
Items | % Relative uncertainty (point dose measurement methods) | |
---|---|---|
Relative | Absolute | |
0.7 | 1.7 | |
NA | 1 | |
0.4 | 0.4 | |
Setup reproducibility | 0.4 | 0.4 |
Long term dosimeter stability | 0.3 | 0.3 |
Electrometer charge reading | 0.3 | 0.3 |
Beam Monitor | 0.12 | 0.12 |
Percentage relative standard uncertainty ( | 1.00 | 2.07 |
Extended uncertainty ( | 2.0 | 4.2 |
Percentage dose difference between measured dose (relative and absolute method) and TPS computed dose for H&N IMRT treatment plans.
Patients | Percentage Dose difference between measured and TPS computed dose | |
---|---|---|
Relative method | Absolute method | |
1 | −1.1 | 1.19 |
2 | −0.1 | 0.40 |
3 | 1.38 | 3.45 |
4 | 1.2 | 1.08 |
5 | 2.2 | 0.06 |
6 | 2.5 | 1.93 |
7 | 1.35 | 1.36 |
8 | 0.0 | 1.42 |
9 | 1.5 | 2.19 |
10 | 1.40 | 1.40 |
11 | 1.40 | 1.56 |
12 | −1.3 | 0.21 |
13 | 2.0 | 1.60 |
SD | 1.2 | 0.9 |
Progress in Medical Physics 2017; 28(3): 111-121
Published online September 30, 2017 https://doi.org/10.14316/pmp.2017.28.3.111
Copyright © Korean Society of Medical Physics.
Talat Mahmood*, Mounir Ibrahim*, Muhammad Aqeel†
*Medical Physics & Radiation Engineering Department, Canberra Hospital & Health Services, ACT, Australia, †Radiation Oncology Department, North-West General Hospital & Research Centre, Peshawar, Pakistan
Correspondence to:Talat Mahmood
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by- nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Verification of dose distribution is an essential part of ensuring the treatment planning system’s (TPS) calculated dose will achieve the desired outcome in radiation therapy. Each measurement have uncertainty associated with it. It is desirable to reduce the measurement uncertainty. A best approach is to reduce the uncertainty associated with each step of the process to keep the total uncertainty under acceptable limits. Point dose patient specific quality assurance (QA) is recommended by American Association of Medical Physicists (AAPM) and European Society for Radiotherapy and Oncology (ESTRO) for all the complex radiation therapy treatment techniques. Relative and absolute point dose measurement methods are used to verify the TPS computed dose. Relative and absolute point dose measurement techniques have a number of steps to measure the point dose which includes chamber cross calibration, electrometer reading, chamber calibration coefficient, beam quality correction factor, reference conditions, influences quantities, machine stability, nominal calibration factor (for relative method) and absolute dose calibration of machine. Keeping these parameters in mind, the estimated relative percentage uncertainty associated with the absolute point dose measurement is 2.1% (
Keywords: Uncertainty, EBRT, IMRT, Point dose, QA
Uncertainty associated with each quality assurance (QA) procedure in external beam radiation therapy (EBRT) should individually be evaluated to minimise the overall uncertainty. The rationale is to minimise the uncertainty in each QA procedure is to get a desirable clinical outcome, because normal tissue complication probability (NTCP) and tumour control probability (TCP) are directly related to the actual doses received by each organ at risk (OAR) and targets.1) The uncertainty associated with the actual delivered dose to the OARs and target volumes is the reason for an increase or decrease in the variation of TCP and NTCP parameters depending on the slop of the dose response curve. The therapeutic ratio is based on the slop of the TCP and NTCP curve.2,6)
Farmer type ionisation chambers are used as the local secondary standard for absolute dosimetry in many of clinical settings. Recently, a new FC23-C (volume 0.2cc) ionisation chamber (IBA Dosimetry Germany) became available to verify the point dose measurement for IMRT treatment plans. The chamber manufacturing company provides a secondary standard calibration laboratory (SSDL) chamber calibration coefficient in terms of absorbed dose in water (
The 3D complexity of the dose in Intensity modulation radiation therapy (IMRT) treatment plan, along with the beam geometry and the resulting dose distribution, means that the QA of IMRT dose distributions needs to concentrate more on the cumulative delivered dose rather than on a the QA of individual segments contributing to the overall dose delivered. Ezzell et al.8) recommends performing a point dose measurement for IMRT treatment plans to verify the TPS computed point dose prior to patient treatment because of the complexity of the IMRT treatment plan. For point dose measurements, Low et al.9) recommended, chamber should be made of tissue equivalent material.
All the components linked to measure the absolute or relative absorbed dose for patient specific QA needs to be analysed individually and estimate the standard uncertainty associated with each component. In this study, a rectangular distribution was assumed to estimate the type B uncertainties. A rectangular distribution to estimate the type B standard uncertainties is presented in
Where
In this study, 13 H&N patients were selected. These patients were planned with the step and shoot IMRT treatment technique. Relative point dose measurement for these patients were already been measured in the CIRS H&N phantom using semiflex ionisation chamber (Serial #: 1976, Volume: 0.125 cc) with electrometer (PTW UnidosE, Serial # 090753). A sample calculation for point dose measurement using relative method for H&N IMRT treatment plans is presented in
For absolute point dose measurements, a chamber calibration factor is required in term of the absorbed dose to water. The International Atomic Energy Agency (IAEA) technical report series (TRS) 398 dosimetry protocols was used to cross calibrate the field ionisation chamber.10)
The field ionisation chamber (IBA FC23-C, Serial number 2408) was cross calibrated with the local secondary standard ionisation chamber (NE 2571, Serial number 3036) using TRS 398. The reference chamber was calibrated (20 May 2016) at the APRANSA PSSDL. A solid water phantom (Gammax, Middleton, WI 53562, USA) was used to cross calibrate the field ionisation chamber because the reference chamber is not water proof.
Beam quality factor
The CIRS H&N phantom was scanned on a Toshiba Aquilion Large Bore CT (16 Slices) to create a QA phantom in the Pinnacle 9.8 TPS (Philips Radiation Oncology System, Fitchburg, WI) with the FC23-C ionisation chamber. The chamber sensitive volume was contoured during the creation of the QA phantom in the TPS and named as “sensitive volume”.
Before mapping the original treatment plans on the CIRS H&N QA phantom, the MUs of each treatment plan were recorded in the worksheet. After the plan mapping a few modifications were applied as listed below:
Remove the CT simulator couch,
Density of the sensitive volume overridden to 1 g/cm3,
Change the prescription to fixed MUs,
Set grid size (2 mm).
Couch was removed during the treatment plan mapping from the CIRS H&N phantom because the phantom was placed on the H&N extension board during the measurement of the point dose. Same setup was also used for the relative point dose measurement. To get a uniform dose distribution across the chamber sensitive volume the phantom position was adjusted in the TPS. This adjustment continued until the TPS computed standard deviation of chamber sensitive volume was ≤ 0.8%. Dose computation was computed using the collapsed cone convolution (CCC) algorithm.
To avoid the beam penetrating through the couch a carbon fiber head and neck extension board (Type-S FixatorTM Shoulder Suppression System by CIVCO) was used to treat the H&N IMRT patients. The CIRS H&N phantom was placed in net area of the H&N extension board because it has negligible transmission factor. This area of H&N extension board was selected only for this experiment to reduce one variable which can effect on absolute point dose measurement due to attenuation of the radiation through the couch. A shift was applied to the phantom as recorded during the mapping of the treatment plan in TPS. Absolute dose of the LINAC was to correct for the day to day output variation of LINAC. Air density correction factors were calculated using measured temperature and air pressure as recommended in TRS 398. The phantom was exposed using the planned gantry and collimator angles. Absolute dose is calculated by using
For relative dose measurement methods, the same steps were followed to measured cumulative charge of all the radiation beams for each IMRT treatment plan. To convert this cumulative charge reading to dose, a nominal calibration factor (NCF) of the chamber was carried out. Detail of nominal calibration is given below.
Before measuring the nominal calibration factor (NCF), leakage measurement test was performed. To calculate the NCF, chamber was exposed three times in water phantom with 6 MV for 200 MUs at calibration set up of the LINAC at 400 MU/min dose rate.
Similarly, the chamber was also exposed in the CIRS H&N phantom at Isocenter with 6 MV for 200 MUs at 400 MU/min dose rate (Fig. 1b) and record the electrometer reading. The detail calculation of the chamber nominal calibration factor (NCF) is explained in
Several factors contribute to the uncertainty calculation in the cross calibration of the chamber using the TRS 398 dosimetry protocols. Each factor should be accounted for to calculate the total uncertainty associated with this process. The electrometer reading for reference and fields chambers are corrected for several factors like reading scale, polarity, air density, humidity, leakage, radiation background, distance and ionic recombination.
Calibration factor (
In the TRS 398 dosimetry protocol, electron stopping powers for monoenergetic photon beam data is used which is presented in International Commissioning of Radiation Units and Measurements (ICRU) report 37, with the density effect model.13) Beam quality factor (
Uncertainty associated with the air density correction factor depends on the thermometer and barometer resolution, calibration certificate and long term stability. The resolution of the thermometer and barometer used in this study are 0.1°C and 0.1 hPa respectively. Das et al.14) also reported the affect of the temperature on chamber volume and its response; it should be taken into account if the temperature difference is higher than the normal calibration temperature. An estimated type B percentage standard uncertainty associated with the air density correction factor was reported by Castro et al.15) 0.2%.
Polarity effect depends on beam quality but cylindrical chambers do not have significantly dependency on beam quality.16) The uncertainty associated with the polarity factor was estimated by accounting the machine reproducibility with the external monitor chamber. The standard deviation is 0.1% was calculated for both chambers with accounting the LINAC reproducibility. Total uncertainty was estimated 0.14% (associated with the user and standards laboratory).
Ion recombination was calculated using a two voltage method.10) The uncertainty associated with two voltage methods is based on the difference between expected value from the Boag theory and the value measured by two voltages method. The maximum deviation in polynomial fit error reported by Weinhous et al.17)60Co beam quality was used to measure the ion recombination for reference ionisation chamber in standard laboratory. Similarly, two voltages method was used to calculate the ion recombination for reference ionisation chamber. The maximum difference 0.16% was reported between two voltage method and Zankowski et al.18) model for cylindrical chambers. The combined estimated value of the ion recombination was 0.16% for field and reference chamber.
No correction for humidity was needed because the calibration certificate was referred a relative humidity of 50% and this condition was also satisfied during our measurements.12) If no correction is made for humidity effect, a maximum error of 0.3% is estimated in the range of 0% to 100% relative humidity.15) Thus the uncertainty associated due to humidity is estimated 0.17% assuming the rectangular distribution.
The reproducibility of reference conditions include the source to surface distance (SSD), depth of measurement (d), setting of the field size. The uncertainty associated with the reproducibility of the reference conditions was reported 0.4% in TRS 398.10)
The standard deviation of the charge reading of well-behaved therapy level chamber should not exceed 0.3%.19) The variation in the charge reading depends on the following factors, reproducibility, display resolution, electrometer linearity, correct nulling of the electrometer and long term stability of the chamber. The uncertainty in the charge reading for both chambers (reference and field chamber) is estimated to 0.3%.
In our case, the reference chamber and electrometer were calibrated as a system in a standard dosimetry laboratory. During the calibration, filed ionisation chamber and electrometer calibrated as a system. An electrometer calibration correction factor is applied only when the electrometer and chamber are calibrated separately.
Overall relative percentage uncertainty associated with reference and field chamber is estimated for each influenced quantity. Calculated combined total uncertainty in the chamber calibration coefficient is presented in Table 1.
Uncertainty associated with each point dose measurement method is estimated and presented in the Table 2. Uncertainty associated with the nominal calibration factor for the relative point dose measurement method is estimated 0.7% including the setup reproducibility, beam monitor system, temperature change, charge reading and long term stability of the electrometer and chamber.
Results are presented in Table 1 for all the influence quantities required for the cross calibration of the field chamber. Calibration factor of the reference ionisation chamber in terms of absorb dose in water at 60Co beam quality was obtained from the PSDL calibration certificate.12) The cross calibration coefficient of the field chamber was obtained by using TRS 398 dosimetry protocol.
The measured absolute point dose difference was within ±3% as compared to the TPS computed point dose except patient 3. Results of the measured point dose with the relative method are also presented in Table 3.
The average percentage difference between the TPS computed dose and measured absolute and relative point dose was 1.4% and 1% respectively for all H&N IMRT treatment plans Fig. 2.
Standard deviation of absolute and relative point dose measurement methods are 0.9% and 1.2% respectively. The standard deviation shows that the absolute point dose measurement method has relatively better reproducibility than the relative method.
Both methods were analysed statistically to see the difference. An independent samples t-test was conducted to compare the means of percentage point dose difference for absolute and relative methods. There was not a significant difference in the score of absolute (M=1.37, SD=0.90) and relative (M=0.95, SD=1.20) point dose measurement methods (t (24)=−0.91,
The estimation of the type B uncertainties is partly based on subjective consideration and published literature. The components associated with the cross calibration factor have been shown to be able to considerably change the uncertainty in the chamber calibration coefficient due to beam quality correction factor. In this study, beam quality correction factor is based on the theoretical calculation and it does not consider chamber to chamber disparity. The uncertainty in chamber calibration coefficient will increase if the ionisation chamber is cross calibrated with the primary standard for photon beam due to this additional step of cross calibration. According to TRS-398 dosimetry protocol, the uncertainty in dose determination increases by approximately 0.2% if the field ionisation chamber is used to determine the absolute dose.10)
The uncertainty associated with this cross calibration (FC23-C ionisation chamber) was estimated 1.7% (
The estimated extended uncertainty (
In order to demonstrate if there is a significant difference between the methods, the subtraction of the mean on one against the other can be performed and compared to the uncertainty involved. This uncertainty can be assumed to follow the summation of the individual errors in quatradure.
It can also be concluded from the results that absolute point dose measurement method does not produce results different from the relative point dose measurement method for head and neck IMRT treatment plans.
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
The study was approved by the institutional review board (IRB approval number; ETHLR.17.044).
Percentage relative standard uncertainty of the each factor associated with the field and reference ionisation chamber.
Items | Percentage relative uncertainty | |
---|---|---|
Reference chamber | Field chamber | |
0.4 | NA | |
1 | 1 | |
0.2 | 0.2 | |
0.1 | 0.1 | |
0.16 | 0.16 | |
Humidity | 0.17 | 0.17 |
Reproducibility of reference condition | 0.4 | 0.4 |
0.3 | 0.3 | |
0.12 | 0.12 | |
Percentage relative standard uncertainty ( | 1.24 | 1.18 |
Total uncertainty ( | 1.7% | |
Extended uncertainty ( | 3.4% |
Percentage relative standard uncertainty of the each factor associated with the absolute and relative point dose measurement methods and total percentage relative standard uncertainty and extended uncertainty.
Items | % Relative uncertainty (point dose measurement methods) | |
---|---|---|
Relative | Absolute | |
0.7 | 1.7 | |
NA | 1 | |
0.4 | 0.4 | |
Setup reproducibility | 0.4 | 0.4 |
Long term dosimeter stability | 0.3 | 0.3 |
Electrometer charge reading | 0.3 | 0.3 |
Beam Monitor | 0.12 | 0.12 |
Percentage relative standard uncertainty ( | 1.00 | 2.07 |
Extended uncertainty ( | 2.0 | 4.2 |
Percentage dose difference between measured dose (relative and absolute method) and TPS computed dose for H&N IMRT treatment plans.
Patients | Percentage Dose difference between measured and TPS computed dose | |
---|---|---|
Relative method | Absolute method | |
1 | −1.1 | 1.19 |
2 | −0.1 | 0.40 |
3 | 1.38 | 3.45 |
4 | 1.2 | 1.08 |
5 | 2.2 | 0.06 |
6 | 2.5 | 1.93 |
7 | 1.35 | 1.36 |
8 | 0.0 | 1.42 |
9 | 1.5 | 2.19 |
10 | 1.40 | 1.40 |
11 | 1.40 | 1.56 |
12 | −1.3 | 0.21 |
13 | 2.0 | 1.60 |
SD | 1.2 | 0.9 |
Table 1 Percentage relative standard uncertainty of the each factor associated with the field and reference ionisation chamber.
Items | Percentage relative uncertainty | |
---|---|---|
Reference chamber | Field chamber | |
0.4 | NA | |
1 | 1 | |
0.2 | 0.2 | |
0.1 | 0.1 | |
0.16 | 0.16 | |
Humidity | 0.17 | 0.17 |
Reproducibility of reference condition | 0.4 | 0.4 |
0.3 | 0.3 | |
0.12 | 0.12 | |
Percentage relative standard uncertainty ( | 1.24 | 1.18 |
Total uncertainty ( | 1.7% | |
Extended uncertainty ( | 3.4% |
Table 2 Percentage relative standard uncertainty of the each factor associated with the absolute and relative point dose measurement methods and total percentage relative standard uncertainty and extended uncertainty.
Items | % Relative uncertainty (point dose measurement methods) | |
---|---|---|
Relative | Absolute | |
0.7 | 1.7 | |
NA | 1 | |
0.4 | 0.4 | |
Setup reproducibility | 0.4 | 0.4 |
Long term dosimeter stability | 0.3 | 0.3 |
Electrometer charge reading | 0.3 | 0.3 |
Beam Monitor | 0.12 | 0.12 |
Percentage relative standard uncertainty ( | 1.00 | 2.07 |
Extended uncertainty ( | 2.0 | 4.2 |
Table 3 Percentage dose difference between measured dose (relative and absolute method) and TPS computed dose for H&N IMRT treatment plans.
Patients | Percentage Dose difference between measured and TPS computed dose | |
---|---|---|
Relative method | Absolute method | |
1 | −1.1 | 1.19 |
2 | −0.1 | 0.40 |
3 | 1.38 | 3.45 |
4 | 1.2 | 1.08 |
5 | 2.2 | 0.06 |
6 | 2.5 | 1.93 |
7 | 1.35 | 1.36 |
8 | 0.0 | 1.42 |
9 | 1.5 | 2.19 |
10 | 1.40 | 1.40 |
11 | 1.40 | 1.56 |
12 | −1.3 | 0.21 |
13 | 2.0 | 1.60 |
SD | 1.2 | 0.9 |
pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829
Frequency: Quarterly