Progress in Medical Physics 2019; 30(1): 32-38
Published online March 31, 2019
https://doi.org/10.14316/pmp.2019.30.1.32
Copyright © Korean Society of Medical Physics.
Hyun Joon An1,2, Myeong Soo Kim1,2, Jiseong Kim1,2, Jaeman Son1,2, Chang Heon Choi1,2,3, Jong Min Park1,2,3,4, Jung-in Kim1,2,3
Correspondence to:Jung-in Kim, (madangin@gmail.com)
Tel: 82-2-2072-3573, Fax: 82-2-3410-2619
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
The objective of this study is to evaluate the geometrical accuracy of a patient-specific bolus based on a three-dimensional (3D) printed mold and casting method. Three breast cancer patients undergoing treatment for a superficial region were scanned using computed tomography (CT) and a designed bolus structure through a treatment planning system (TPS). For the fabrication of patient-specific bolus, we cast harmless certified silicone into 3D printed molds. The produced bolus was also imaged using CT under the same conditions as the patient CT to acquire its geometrical shape. We compared the shapes of the produced bolus with the planned bolus structure from the TPS by measuring the average distance between two structures after a surface registration. The result of the average difference in distance was within 1 mm and, as the worst case, the absolute difference did not exceed ±2 mm. The result of the geometric difference in the cross-section profile of each bolus were approximately 1 mm, which is a similar property of the average difference in distance. This discrepancy was negligible in affecting the dose reduction. The proposed fabrication of patient-specific bolus is useful for radiation therapy in the treatment of superficial regions, particularly those with an irregular shape.Purpose
Materials and Methods
Results
Conclusions
KeywordsBolus, Patient specific bolus, 3D printing, Dose build up, Geometric analysis
In radiation therapy, the primary goal is to deliver a sufficient amount of radiation to the target tumor while alleviating the effects on the adjacent normal tissue. When the planning target volume (PTV) is located near the surface area, a bolus composed of a tissue-equivalent material is used to provide an adequate dose build-up in the skin. A variety of boluses including simple water, mixtures, wax, and even metal have been used for clinical application.1) In most cases, a gel-sheet type bolus such as Superflab (Radiation Product Design, Albertvile, MN, USA) is typically used to cover large areas of the skin without modification. The effects of a bolus on the surface sufficiently increase the dose for the tangential fields of conventional and intensity modulated radiation therapy (IMRT).2) However, under certain conditions, such as when the skin presents a particularly irregular shape, it is difficult to avoid air gaps between the bolus and the patient's skin. According to Butson et al.,3) the effects of small air gaps of less than 2 mm under a bolus material have not been shown to reduce the surface dose. As the air gap increases by 4 mm, the surface dose decreases, and by 10 mm, up to a 10% reduction is achieved. Therefore, an accurate fit of the bolus to the patient's surface is important to an accurate dose delivery in near-surface tumors.
With significant advances in three-dimensional (3D) printer technologies, several studies have applied the concept of 3D-printed boluses. Kim et al.4) designed and fabricated a customized 3D printed bolus for a RANDO phantom, and Su et al.5) investigated optimizing a 3D printing bolus design for electron radiation therapy. Park et al.6) showed a reduction of the air gap compared to a commercial bolus when using a 3D-printed bolus. Ricotti et al.7) described modulating the shift of the build-up region by tuning the infill percentage of the 3D printed bolus. Beyond the phantom study, some studies have demonstrated the usefulness of 3D-printed boluses in actual patient treatment.8,9) Each of these studies was conducted using a direct 3D printing thermoplastic filament such as acrylonitrile butadiene styrene or polylactic acid. However, these materials have a problem in that they are not fitted to the patient's surface owing to the hardness characteristics. In order to avoid the air gap, ultrasound gel can be applied to the interface between the 3D-printed bolus and the skin,10) but the gel can flow along the irregular surface or it is difficult to use it for patients with skin ulcer or bleeding. Although 3D printable flexible materials have recently been made commercially available, such as NinjaFlex, Cheeta (both from NinjaTek, Manheim, PA, USA), TangoPlus (Stratasys, Eden Prairie, MN, USA), they still have a solid property.
In this study, we propose a different way of fabricating a patient-specific bolus from a 3D printed mold and using a casting method. The bolus was fabricated using a more flexible and soft material and analyzed for the geometrical changes during the molding and casting process.
For this study, we chose three breast cancer patients who required a bolus to treat a superficial region. Each patient was CT scanned using a Philips Brilliance CT Big Bore (Philips Healthcare, Cleveland, OH, USA). All reconstructed CT images had a slice thickness of 3 mm and 1.15 mm in the x- and y-axis resolution. The retrospective use of the scan data was approved by the Institutional Review Board of Seoul National University Hospital (no. 1812-163-999).
Based on the acquired CT images, the bolus was designed using an Eclipse (Varian Medical Systems, Palo Alto, CA, USA) treatment planning system (TPS). The external surface of the patient body was defined using automatic body contouring. A 10-mm thick bolus was added outside the external surface to include the (PTV) area. The thickness was manually changed in some areas to closely adhere to the patient's body, as shown in Fig. 1.
A summary of the overall bolus fabrication process is shown in Fig. 2. The bolus structure for each patient was the exported digital imaging and communications in medicine radiotherapy (DICOM-RT) structure format. The segmented bolus structure was converted into an stereolithography (STL) file using an open-source software, 3D Slicer (
To evaluate the geometrical accuracy, we compared the shapes of the produced bolus with respect to the planned bolus structure from the TPS. The produced bolus was imaged using CT under the same conditions as the patient. Then, the measured bolus image was converted into an STL file for comparison using the mesh information.
The registration process was conducted using Polyworks 2014 software (InnovMetric, Quebec, QC, Canada), which uses the IMAlign module through a cloud-based method for surface matching under the principle of the iterative closest point (ICP) algorithm.11) This method iteratively minimizes the distance between the point in one cloud set and its closest point in the other set by estimating the transformation parameters. It allows a comparison of the resulting geometry differences by measuring the average distance between points. We also compared a cross-section profile of the registered bolus geometry and calculated the basic geometric parameter such as the surface area and volume.
The results of surface-based registration between a planned and measured bolus structure is shown in Fig. 4. The modeling information for each structure is described in Table 2. After iteratively calculating the distance between two points in each cloud as the minimum, the registered meshes were superimposed, as shown in Fig. 4c, and the average distance was visualized as a color map, shown in Fig. 4d. In this calculation, the mean and standard deviation of the difference in distance for each patient are plotted in Fig. 5.
In each patient case, we compared the geometric difference along the cross-section profile. The geometric values were calculated after dividing five cross-section profiles with equal spacing according to the bolus size. Fig. 6 shows the difference value according to each cross- section profile as a normal vector. Table 3 summarizes the geometric difference value of five cross-sections of each patient.
We also compared the volume and surface area between the planned bolus structure from the TPS and measured the bolus structure from a 3D printed mold and casting bolus. The calculated volume, surface area, and percent of difference are shown in Table 4.
In the present study, we introduced a bolus using a 3D printed mold and casting method, and investigated the geometrical accuracy based on the surface registration and basic geometric parameters.
The results of the average difference in distance was within 1 mm, and as the worst case, the absolute difference did not exceed ±2 mm. This discrepancy was negligible in affecting the dose reduction.3) The results of the cross-section profile showed similar properties with the average difference in distance. According to the cross-section results, cross-section (CS) 3, which was the bolus center, had a smaller mean difference than CS 1 and CS 5, which were the boundary of the bolus. CS 3 shows the cross-section near the PTV and is the most important region, providing a dose build-up. The reason for the large difference at the boundary was due to the positional error in the process of CT imaging of the fabricated bolus. This tendency was the most pronounced in a large sized bolus such as the case of patient 3. In addition, one of the limitations of the ICP-based registration is the use of a rigid transformation, which does not consider a non-rigid transformation, increasing the uncertainty of the comparison.
The difference in area between the planned and fabricated bolus structure was less than 5.5%, although the difference in volume was almost 10%. This most likely decreases the volume as the air bubbles are removed during the curing process of the silicone. Therefore, more attention should be paid to the changes in volume during the production process.
Compared to direct 3D printing approach, 3D-printed mold and casting method has several advantages. It is possible to make a bolus using silicone such as Dragon Skin. In addition, various materials can be used to make a bolus suitable for the purpose. It is possible to fabricate a bolus that is more flexible, softer, and less uncomfortable to the patient. It is also easy to make a large-sized bolus through the post-processing of 3D-printed molding. However, there is a disadvantage that labor intensity is required due to mold removal and curing process. The time to make a bolus depends on the output volume, but in general, the direct 3D printing method takes more time for 3D printing to fill the filaments at 100%. The 3D printed mold has only 1-mm thickness and 0% fill density, saving about 30~40% of the time for the 3D printing process.
This study focused on a geometric analysis of a new bolus fabrication. In the future, we will proceed with an investigation into its dosimetric characteristics and clinical application.
As the results show, we determined the feasibility of using a 3D printed mold and casting method for a patient-specific bolus. The geometrical analysis demonstrated that the fabrication of the bolus did not change the shape effect of the dose reduction.
This study was supported by a grant from the National R&D Program for Cancer Control, Ministry of Health & Welfare, Republic of Korea (No. 1631200), and by a National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP) (No.2017M2A2A7A02020641 and 2017M2A2A7A02020643).
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
The study was approved by the institutional review board (IRB approval number; 1812-163-999).
Physical property of Dragon SkinTM 10 MEDIUM and 3D printer settings
Dragon SkinTM 10 MEDIUM | Zotrax M300 3D printer setting | ||
---|---|---|---|
Physical property | Value | Physical property | Value |
Physical Density | 1.179 g/cc | Extruder Temperature | 245°C |
Electron Density ratio compared to water | 1.134 | Extrusion Width | 0.4 mm |
Tensile strength | 475 psi | Layer Height | 0.14 mm |
Tensile Modulus | 22 psi | Speed | 100 mm/s |
Elongation at Break | 1000 % | Fill Density | 0% |
Shore Hardness# | 10 A | Fill Type | Mesh |
Fill Angle | 30° | ||
Nozzle Diameter | 0.4 mm | ||
Filament Diameter | 1.75 mm | ||
Layer Thickness | 0.14 mm | ||
Print Quality | High |
Three-dimentional modeling information for each structure
Modeling parameter | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Planned | Fabricated | Planned | Fabricated | Planned | Fabricated | |
Vertex | 62,682 | 30,866 | 29,460 | 40,088 | 163,398 | 194,514 |
Facet | 125,376 | 61,728 | 58,914 | 80,218 | 326,792 | 389,009 |
Geometric difference of cross-section profile for each patient (mm)
Number | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Mean | SD | Mean | SD | Mean | SD | |
CS 1 | 0.82 | 1.279 | 0.436 | 2.209 | −0.325 | 2.237 |
CS 2 | 1.076 | 1.195 | 0.376 | 1.166 | −0.237 | 2.044 |
CS 3 | 0.721 | 1.202 | 0.261 | 1.572 | 0.661 | 1.718 |
CS 4 | 0.954 | 1.645 | 0.280 | 1.405 | 0.369 | 1.694 |
CS 5 | 1.047 | 1.504 | 0.743 | 1.073 | −0.055 | 2.107 |
Comparison of volume and surface between planned and fabricated bolus
Property | Patient 1 | Patient 2 | Patient 3 | ||||||
---|---|---|---|---|---|---|---|---|---|
Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | |
Volume (mm3) | 1026.75 | 938.73 | 9.0 | 125.58 | 114.27 | 9.4 | 2084.63 | 1971.92 | 5.6 |
Area (mm2) | 2048.63 | 1938.14 | 5.5 | 313.65 | 304.79 | 2.9 | 3000.77 | 2988.07 | 0.4 |
Progress in Medical Physics 2019; 30(1): 32-38
Published online March 31, 2019 https://doi.org/10.14316/pmp.2019.30.1.32
Copyright © Korean Society of Medical Physics.
Hyun Joon An1,2, Myeong Soo Kim1,2, Jiseong Kim1,2, Jaeman Son1,2, Chang Heon Choi1,2,3, Jong Min Park1,2,3,4, Jung-in Kim1,2,3
1Department of Radiation Oncology, Seoul National University Hospital, 2Biomedical Research Institute, Seoul National University Hospital, 3Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, 4Robotics Research Laboratory for Extreme Environments, Advanced Institutes of Convergence Technology, Suwon, Korea
Correspondence to:Jung-in Kim, (madangin@gmail.com)
Tel: 82-2-2072-3573, Fax: 82-2-3410-2619
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
The objective of this study is to evaluate the geometrical accuracy of a patient-specific bolus based on a three-dimensional (3D) printed mold and casting method. Three breast cancer patients undergoing treatment for a superficial region were scanned using computed tomography (CT) and a designed bolus structure through a treatment planning system (TPS). For the fabrication of patient-specific bolus, we cast harmless certified silicone into 3D printed molds. The produced bolus was also imaged using CT under the same conditions as the patient CT to acquire its geometrical shape. We compared the shapes of the produced bolus with the planned bolus structure from the TPS by measuring the average distance between two structures after a surface registration. The result of the average difference in distance was within 1 mm and, as the worst case, the absolute difference did not exceed ±2 mm. The result of the geometric difference in the cross-section profile of each bolus were approximately 1 mm, which is a similar property of the average difference in distance. This discrepancy was negligible in affecting the dose reduction. The proposed fabrication of patient-specific bolus is useful for radiation therapy in the treatment of superficial regions, particularly those with an irregular shape.Purpose
Materials and Methods
Results
Conclusions
Keywords: Bolus, Patient specific bolus, 3D printing, Dose build up, Geometric analysis
In radiation therapy, the primary goal is to deliver a sufficient amount of radiation to the target tumor while alleviating the effects on the adjacent normal tissue. When the planning target volume (PTV) is located near the surface area, a bolus composed of a tissue-equivalent material is used to provide an adequate dose build-up in the skin. A variety of boluses including simple water, mixtures, wax, and even metal have been used for clinical application.1) In most cases, a gel-sheet type bolus such as Superflab (Radiation Product Design, Albertvile, MN, USA) is typically used to cover large areas of the skin without modification. The effects of a bolus on the surface sufficiently increase the dose for the tangential fields of conventional and intensity modulated radiation therapy (IMRT).2) However, under certain conditions, such as when the skin presents a particularly irregular shape, it is difficult to avoid air gaps between the bolus and the patient's skin. According to Butson et al.,3) the effects of small air gaps of less than 2 mm under a bolus material have not been shown to reduce the surface dose. As the air gap increases by 4 mm, the surface dose decreases, and by 10 mm, up to a 10% reduction is achieved. Therefore, an accurate fit of the bolus to the patient's surface is important to an accurate dose delivery in near-surface tumors.
With significant advances in three-dimensional (3D) printer technologies, several studies have applied the concept of 3D-printed boluses. Kim et al.4) designed and fabricated a customized 3D printed bolus for a RANDO phantom, and Su et al.5) investigated optimizing a 3D printing bolus design for electron radiation therapy. Park et al.6) showed a reduction of the air gap compared to a commercial bolus when using a 3D-printed bolus. Ricotti et al.7) described modulating the shift of the build-up region by tuning the infill percentage of the 3D printed bolus. Beyond the phantom study, some studies have demonstrated the usefulness of 3D-printed boluses in actual patient treatment.8,9) Each of these studies was conducted using a direct 3D printing thermoplastic filament such as acrylonitrile butadiene styrene or polylactic acid. However, these materials have a problem in that they are not fitted to the patient's surface owing to the hardness characteristics. In order to avoid the air gap, ultrasound gel can be applied to the interface between the 3D-printed bolus and the skin,10) but the gel can flow along the irregular surface or it is difficult to use it for patients with skin ulcer or bleeding. Although 3D printable flexible materials have recently been made commercially available, such as NinjaFlex, Cheeta (both from NinjaTek, Manheim, PA, USA), TangoPlus (Stratasys, Eden Prairie, MN, USA), they still have a solid property.
In this study, we propose a different way of fabricating a patient-specific bolus from a 3D printed mold and using a casting method. The bolus was fabricated using a more flexible and soft material and analyzed for the geometrical changes during the molding and casting process.
For this study, we chose three breast cancer patients who required a bolus to treat a superficial region. Each patient was CT scanned using a Philips Brilliance CT Big Bore (Philips Healthcare, Cleveland, OH, USA). All reconstructed CT images had a slice thickness of 3 mm and 1.15 mm in the x- and y-axis resolution. The retrospective use of the scan data was approved by the Institutional Review Board of Seoul National University Hospital (no. 1812-163-999).
Based on the acquired CT images, the bolus was designed using an Eclipse (Varian Medical Systems, Palo Alto, CA, USA) treatment planning system (TPS). The external surface of the patient body was defined using automatic body contouring. A 10-mm thick bolus was added outside the external surface to include the (PTV) area. The thickness was manually changed in some areas to closely adhere to the patient's body, as shown in Fig. 1.
A summary of the overall bolus fabrication process is shown in Fig. 2. The bolus structure for each patient was the exported digital imaging and communications in medicine radiotherapy (DICOM-RT) structure format. The segmented bolus structure was converted into an stereolithography (STL) file using an open-source software, 3D Slicer (
To evaluate the geometrical accuracy, we compared the shapes of the produced bolus with respect to the planned bolus structure from the TPS. The produced bolus was imaged using CT under the same conditions as the patient. Then, the measured bolus image was converted into an STL file for comparison using the mesh information.
The registration process was conducted using Polyworks 2014 software (InnovMetric, Quebec, QC, Canada), which uses the IMAlign module through a cloud-based method for surface matching under the principle of the iterative closest point (ICP) algorithm.11) This method iteratively minimizes the distance between the point in one cloud set and its closest point in the other set by estimating the transformation parameters. It allows a comparison of the resulting geometry differences by measuring the average distance between points. We also compared a cross-section profile of the registered bolus geometry and calculated the basic geometric parameter such as the surface area and volume.
The results of surface-based registration between a planned and measured bolus structure is shown in Fig. 4. The modeling information for each structure is described in Table 2. After iteratively calculating the distance between two points in each cloud as the minimum, the registered meshes were superimposed, as shown in Fig. 4c, and the average distance was visualized as a color map, shown in Fig. 4d. In this calculation, the mean and standard deviation of the difference in distance for each patient are plotted in Fig. 5.
In each patient case, we compared the geometric difference along the cross-section profile. The geometric values were calculated after dividing five cross-section profiles with equal spacing according to the bolus size. Fig. 6 shows the difference value according to each cross- section profile as a normal vector. Table 3 summarizes the geometric difference value of five cross-sections of each patient.
We also compared the volume and surface area between the planned bolus structure from the TPS and measured the bolus structure from a 3D printed mold and casting bolus. The calculated volume, surface area, and percent of difference are shown in Table 4.
In the present study, we introduced a bolus using a 3D printed mold and casting method, and investigated the geometrical accuracy based on the surface registration and basic geometric parameters.
The results of the average difference in distance was within 1 mm, and as the worst case, the absolute difference did not exceed ±2 mm. This discrepancy was negligible in affecting the dose reduction.3) The results of the cross-section profile showed similar properties with the average difference in distance. According to the cross-section results, cross-section (CS) 3, which was the bolus center, had a smaller mean difference than CS 1 and CS 5, which were the boundary of the bolus. CS 3 shows the cross-section near the PTV and is the most important region, providing a dose build-up. The reason for the large difference at the boundary was due to the positional error in the process of CT imaging of the fabricated bolus. This tendency was the most pronounced in a large sized bolus such as the case of patient 3. In addition, one of the limitations of the ICP-based registration is the use of a rigid transformation, which does not consider a non-rigid transformation, increasing the uncertainty of the comparison.
The difference in area between the planned and fabricated bolus structure was less than 5.5%, although the difference in volume was almost 10%. This most likely decreases the volume as the air bubbles are removed during the curing process of the silicone. Therefore, more attention should be paid to the changes in volume during the production process.
Compared to direct 3D printing approach, 3D-printed mold and casting method has several advantages. It is possible to make a bolus using silicone such as Dragon Skin. In addition, various materials can be used to make a bolus suitable for the purpose. It is possible to fabricate a bolus that is more flexible, softer, and less uncomfortable to the patient. It is also easy to make a large-sized bolus through the post-processing of 3D-printed molding. However, there is a disadvantage that labor intensity is required due to mold removal and curing process. The time to make a bolus depends on the output volume, but in general, the direct 3D printing method takes more time for 3D printing to fill the filaments at 100%. The 3D printed mold has only 1-mm thickness and 0% fill density, saving about 30~40% of the time for the 3D printing process.
This study focused on a geometric analysis of a new bolus fabrication. In the future, we will proceed with an investigation into its dosimetric characteristics and clinical application.
As the results show, we determined the feasibility of using a 3D printed mold and casting method for a patient-specific bolus. The geometrical analysis demonstrated that the fabrication of the bolus did not change the shape effect of the dose reduction.
This study was supported by a grant from the National R&D Program for Cancer Control, Ministry of Health & Welfare, Republic of Korea (No. 1631200), and by a National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIP) (No.2017M2A2A7A02020641 and 2017M2A2A7A02020643).
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
The study was approved by the institutional review board (IRB approval number; 1812-163-999).
Physical property of Dragon SkinTM 10 MEDIUM and 3D printer settings
Dragon SkinTM 10 MEDIUM | Zotrax M300 3D printer setting | ||
---|---|---|---|
Physical property | Value | Physical property | Value |
Physical Density | 1.179 g/cc | Extruder Temperature | 245°C |
Electron Density ratio compared to water | 1.134 | Extrusion Width | 0.4 mm |
Tensile strength | 475 psi | Layer Height | 0.14 mm |
Tensile Modulus | 22 psi | Speed | 100 mm/s |
Elongation at Break | 1000 % | Fill Density | 0% |
Shore Hardness# | 10 A | Fill Type | Mesh |
Fill Angle | 30° | ||
Nozzle Diameter | 0.4 mm | ||
Filament Diameter | 1.75 mm | ||
Layer Thickness | 0.14 mm | ||
Print Quality | High |
Three-dimentional modeling information for each structure
Modeling parameter | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Planned | Fabricated | Planned | Fabricated | Planned | Fabricated | |
Vertex | 62,682 | 30,866 | 29,460 | 40,088 | 163,398 | 194,514 |
Facet | 125,376 | 61,728 | 58,914 | 80,218 | 326,792 | 389,009 |
Geometric difference of cross-section profile for each patient (mm)
Number | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Mean | SD | Mean | SD | Mean | SD | |
CS 1 | 0.82 | 1.279 | 0.436 | 2.209 | −0.325 | 2.237 |
CS 2 | 1.076 | 1.195 | 0.376 | 1.166 | −0.237 | 2.044 |
CS 3 | 0.721 | 1.202 | 0.261 | 1.572 | 0.661 | 1.718 |
CS 4 | 0.954 | 1.645 | 0.280 | 1.405 | 0.369 | 1.694 |
CS 5 | 1.047 | 1.504 | 0.743 | 1.073 | −0.055 | 2.107 |
Comparison of volume and surface between planned and fabricated bolus
Property | Patient 1 | Patient 2 | Patient 3 | ||||||
---|---|---|---|---|---|---|---|---|---|
Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | |
Volume (mm3) | 1026.75 | 938.73 | 9.0 | 125.58 | 114.27 | 9.4 | 2084.63 | 1971.92 | 5.6 |
Area (mm2) | 2048.63 | 1938.14 | 5.5 | 313.65 | 304.79 | 2.9 | 3000.77 | 2988.07 | 0.4 |
Table 1 Physical property of Dragon SkinTM 10 MEDIUM and 3D printer settings
Dragon SkinTM 10 MEDIUM | Zotrax M300 3D printer setting | ||
---|---|---|---|
Physical property | Value | Physical property | Value |
Physical Density | 1.179 g/cc | Extruder Temperature | 245°C |
Electron Density ratio compared to water | 1.134 | Extrusion Width | 0.4 mm |
Tensile strength | 475 psi | Layer Height | 0.14 mm |
Tensile Modulus | 22 psi | Speed | 100 mm/s |
Elongation at Break | 1000 % | Fill Density | 0% |
Shore Hardness# | 10 A | Fill Type | Mesh |
Fill Angle | 30° | ||
Nozzle Diameter | 0.4 mm | ||
Filament Diameter | 1.75 mm | ||
Layer Thickness | 0.14 mm | ||
Print Quality | High |
Table 2 Three-dimentional modeling information for each structure
Modeling parameter | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Planned | Fabricated | Planned | Fabricated | Planned | Fabricated | |
Vertex | 62,682 | 30,866 | 29,460 | 40,088 | 163,398 | 194,514 |
Facet | 125,376 | 61,728 | 58,914 | 80,218 | 326,792 | 389,009 |
Table 3 Geometric difference of cross-section profile for each patient (mm)
Number | Patient 1 | Patient 2 | Patient 3 | |||
---|---|---|---|---|---|---|
Mean | SD | Mean | SD | Mean | SD | |
CS 1 | 0.82 | 1.279 | 0.436 | 2.209 | −0.325 | 2.237 |
CS 2 | 1.076 | 1.195 | 0.376 | 1.166 | −0.237 | 2.044 |
CS 3 | 0.721 | 1.202 | 0.261 | 1.572 | 0.661 | 1.718 |
CS 4 | 0.954 | 1.645 | 0.280 | 1.405 | 0.369 | 1.694 |
CS 5 | 1.047 | 1.504 | 0.743 | 1.073 | −0.055 | 2.107 |
SD, standard deviation; CS, cross-section.
Table 4 Comparison of volume and surface between planned and fabricated bolus
Property | Patient 1 | Patient 2 | Patient 3 | ||||||
---|---|---|---|---|---|---|---|---|---|
Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | Planned | Fabricated | diff (%) | |
Volume (mm3) | 1026.75 | 938.73 | 9.0 | 125.58 | 114.27 | 9.4 | 2084.63 | 1971.92 | 5.6 |
Area (mm2) | 2048.63 | 1938.14 | 5.5 | 313.65 | 304.79 | 2.9 | 3000.77 | 2988.07 | 0.4 |