Progress in Medical Physics 2019; 30(4): 155-159
Published online December 31, 2019
https://doi.org/10.14316/pmp.2019.30.4.155
Copyright © Korean Society of Medical Physics.
So-Yeon Park , Noorie Choi , Byeong Geol Choi , Dong Myung Lee , Na Young Jang
Correspondence to:Na Young Jang (mumuki79@gmail.com)
Tel: 82-2-2225-4647 Fax: 82-2-2225-4640
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Radiological properties of newly introduced and existing 3-dimensional (3D) printing materials were evaluated by measuring their Hounsfield units (HUs) at varying infill densities. The six materials for 3D printing which consisted of acrylonitrile butadiene styrene (ABS), a unique ABS plastic blend manufactured by Zortrax (ULTRAT), high impact polystyrene (HIPS), polyethylene terephthalate glycol (PETG), polylactic acid (PLA), and a thermoplastic polyester elastomer manufactured by Zortrax (FLEX) were used. We used computed tomography (CT) imaging to determine the HU values of each material, and thus assess its suitability for various applications in radiation oncology. We found that several material and infill density combinations resembled the HU values of fat, soft tissues, and lungs; however, none of the tested materials exhibited HU values similar to that of bone. These results will help researchers and clinicians develop more appropriate instruments for improving the quality of radiation therapy. Using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.
Keywords3D printing, Computed tomography, Hounsfield unit, Infill density, Radiation oncology
Three-dimensional (3D) printing technology is widely used in the field of radiation oncology to make computed tomography (CT)-based patient-specific tools, thereby improving the quality of radiation therapy.1–7) For example, patient-specific compensators for radiotherapy are fabricated for use in areas with severe curvatures, such as the nose, feet, and hands, thus delivering more radiation to a tumor in the area and reducing radiation to the surrounding normal organs.4) For total body irradiation, 3D-printed patient-specific metal compensators reflecting the overall curves of the patient’s body shape can deliver more uniform radiation doses to the whole body than conventional methods.6) 3D printing technology has also been used for high dose rate (HDR) brachytherapy to create patient-specific elastic HDR applicators. These deliver accurate doses of radiation to the skin’s surface; this technology has been proven to improve treatment delivery accuracy.2,4,8) In addition, many attempts have been made to perform quality assurance in radiation therapy by producing patient-specific phantoms using patient CTs and 3D printers.5,9–13)
In the 3D printing field, advances have been made not only in printing technology, but also in the printing materials used. Various synthetic 3D printing materials have been developed, including polyethylene terephthalate glycol (PETG) and a unique acrylonitrile butadiene styrene (ABS) plastic blend manufactured by Zortrax (ULTRAT; Zortrax SA, Olsztyn, Poland) which is in a type of thermoplastic polymer. Flexible materials have also been developed, like the thermoplastic polyester elastomer manufactured by Zortrax (FLEX) which had great interlayer adhesion and dimensional tolerance. These materials can be used for a wide range of applications in 3D printing for radiation therapy, in combination with existing 3D printing materials like ABS, polylactic acid (PLA), and high impact polystyrene (HIPS). However, no prior research has evaluated the radiological properties of the various 3D printing materials currently available, knowledge of which is crucial for these material’s proper application.
In this study, the radiological properties of newly introduced and existing 3D printing materials were evaluated by measuring their Hounsfield units (HUs). The HU values were evaluated by varying the infill density of each printing material, thus allowing us to assess their applicability to the field of radiation oncology.
In this study, a total of six 3D printing materials were used: ABS, ULTRAT, HIPS, PETG, PLA, and FLEX. Table 1 shows the physical properties of the six materials; of these, PETG exhibited the highest tensile strength (the mechanical strength of the material), FLEX had the highest bending stress (the stress that occurs inside a material by bending moment), and the lowest shore hardness (and was thus the most flexible material). HIPS had the lowest specific density (1.14 g/cm3), and PLA had the highest specific density (1.43 g/cm3).
The 3D printer used in this study was a Zortrax M300 Plus model (Zortrax SA, Olsztyn, Poland). The specifications of this printer are shown in Table 2. The Z-suite program (Zortrax SA, Olsztyn, Poland) was used for printing; the program’s printer settings are shown in Fig. 1. This program allows printing with various infill densities. For the study experiments, rectangles with dimensions of 4×4×2 cm were fabricated using HIPS (with the lowest specific density) and PLA (with the highest specific density). The infill densities of these rectangles were varied as 10%, 20%, 40%, 60%, and 100%. Rectangles of the other four kinds of materials of the same size were also fabricated, each with a 60% infill density.
In order to measure the HU values of the rectangles, CT images of the six materials were acquired. The scan thickness was 1 mm and the scan conditions were 120 kV and 200 mA. Under the same conditions, CT images were obtained of HIPS and PLA with varying infill densities. Fig. 2 shows images of HIPS samples with infill densities from 10% to 100%, alongside their respective CT images.
HUs are a value constituting the grayscale level of a CT image, and can be calculated through the image’s linear attenuation coefficient. The equation for calculating HUs is defined below:
where μmaterial is the linear attenuation coefficient for an arbitrary material and μwater is the linear attenuation coefficient for water. Air has an HU value of −1,000, because its μ value is 0, and water has an HU value of 0. In this study, a 1×1 cm region of interest was obtained through a coronal view CT image to evaluate the HU for each material and infill density; and average HU values were obtained for all six materials with 60% infill density. Average HUs were also obtained for HIPS and PLA with 10%–100% infill densities. The HU values for HIPS and PLA at different infill densities were subjected to curve fitting using the equation below:
This study investigated the radiological properties of six 3D printing materials, to determine how to incorporate a wide range of 3D printing technologies into radiation oncology. The average HU values of the six printing materials were obtained, and the average HU values of HIPS and PLA were analyzed across a range of infill densities.
The time required to print each material at each infill density was also measured. When the infill density was 10%, the average printing time was 32 minutes; when the infill density was 100%, the average time was 12 hours and 45 minutes. Considering that the volume of each prepared cube was 32 m3, a printing time of 12 hours and 45 minutes could be considered quite wasteful. Using 100% infill density should therefore be carefully considered while manufacturing patient-specific human phantoms.11–13)
Table 3 shows the HU values for the six printing materials, each at an infill density of 60%. ABS, ULTRAT, HIPS, PETG, PLA, and FLEX had HU values of −535±12, −557±10, −542±7, −508±20, −530±25, and −633±15, respectively. The HU values were similar regardless of the specific density of the materials.
Fig. 3 includes a graph showing HU changing with the infill density of HIPS and PLA. The HU value increased with infill density for each material. The HU values for HIPS and PLA were −131 and 61, respectively, at an infill density of 100%. By comparison, at an infill density of 10%, the HU values for HIPS and PLA were −896 and −904, respectively, In addition, the parameters for fitting the relationship between infill density and HU using
In this study, HU values were obtained for various 3D printing materials at different infill densities; the radiological characteristics of each material were then analyzed. The results suggest that using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.
This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Korea (grant number: VHSMC 19021).
Physical characteristics of materials used in 3-dimensional printing
Variable | ABS | ULTRAT | HIPS | PETG | PLA | FLEX |
---|---|---|---|---|---|---|
Tensile strength (MPa) | 30.5 | 32.6 | 16.9 | 40.2 | 28.8 | 9.9 |
Bending stress (MPa) | 46.3 | 54.0 | 29.3 | 55.3 | 58.6 | 174.3 |
Specific density (g/cm3) | 1.20 | 1.18 | 1.14 | 1.30 | 1.43 | 1.19 |
Shore hardness (D) | 69.2 | 73.4 | 73.2 | 71.4 | 77.6 | 31.0 |
Specifications for 3-dimensional printer
Specification | Specification value |
---|---|
Build volume (mm3) | 300×300×300 |
Layer resolution (μm) | 90–290 |
Minimal wall thickness (μm) | 400 |
Material diameter (nm) | 1.75 |
Nozzle diameter (mm) | 1.4 |
Maximum printing temperature (°F) | 554 |
Supported input file types | .stl, .obj, .dxf, .3mf |
Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%
ABS | ULTRAT | HIPS | PETG | PLA | FLEX | |
---|---|---|---|---|---|---|
HU values | −535±12 | −557±10 | −542±7 | −508±20 | −530±25 | −633±15 |
Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA
Material | Parameter | R-squared value | ||
---|---|---|---|---|
a | b | c | ||
HIPS | 0.033 | 5.125 | −962.607 | 0.994 |
PLA | 0.079 | 1.286 | −892.879 | 0.976 |
Progress in Medical Physics 2019; 30(4): 155-159
Published online December 31, 2019 https://doi.org/10.14316/pmp.2019.30.4.155
Copyright © Korean Society of Medical Physics.
So-Yeon Park , Noorie Choi , Byeong Geol Choi , Dong Myung Lee , Na Young Jang
Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Korea
Correspondence to:Na Young Jang (mumuki79@gmail.com)
Tel: 82-2-2225-4647 Fax: 82-2-2225-4640
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Radiological properties of newly introduced and existing 3-dimensional (3D) printing materials were evaluated by measuring their Hounsfield units (HUs) at varying infill densities. The six materials for 3D printing which consisted of acrylonitrile butadiene styrene (ABS), a unique ABS plastic blend manufactured by Zortrax (ULTRAT), high impact polystyrene (HIPS), polyethylene terephthalate glycol (PETG), polylactic acid (PLA), and a thermoplastic polyester elastomer manufactured by Zortrax (FLEX) were used. We used computed tomography (CT) imaging to determine the HU values of each material, and thus assess its suitability for various applications in radiation oncology. We found that several material and infill density combinations resembled the HU values of fat, soft tissues, and lungs; however, none of the tested materials exhibited HU values similar to that of bone. These results will help researchers and clinicians develop more appropriate instruments for improving the quality of radiation therapy. Using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.
Keywords: 3D printing, Computed tomography, Hounsfield unit, Infill density, Radiation oncology
Three-dimensional (3D) printing technology is widely used in the field of radiation oncology to make computed tomography (CT)-based patient-specific tools, thereby improving the quality of radiation therapy.1–7) For example, patient-specific compensators for radiotherapy are fabricated for use in areas with severe curvatures, such as the nose, feet, and hands, thus delivering more radiation to a tumor in the area and reducing radiation to the surrounding normal organs.4) For total body irradiation, 3D-printed patient-specific metal compensators reflecting the overall curves of the patient’s body shape can deliver more uniform radiation doses to the whole body than conventional methods.6) 3D printing technology has also been used for high dose rate (HDR) brachytherapy to create patient-specific elastic HDR applicators. These deliver accurate doses of radiation to the skin’s surface; this technology has been proven to improve treatment delivery accuracy.2,4,8) In addition, many attempts have been made to perform quality assurance in radiation therapy by producing patient-specific phantoms using patient CTs and 3D printers.5,9–13)
In the 3D printing field, advances have been made not only in printing technology, but also in the printing materials used. Various synthetic 3D printing materials have been developed, including polyethylene terephthalate glycol (PETG) and a unique acrylonitrile butadiene styrene (ABS) plastic blend manufactured by Zortrax (ULTRAT; Zortrax SA, Olsztyn, Poland) which is in a type of thermoplastic polymer. Flexible materials have also been developed, like the thermoplastic polyester elastomer manufactured by Zortrax (FLEX) which had great interlayer adhesion and dimensional tolerance. These materials can be used for a wide range of applications in 3D printing for radiation therapy, in combination with existing 3D printing materials like ABS, polylactic acid (PLA), and high impact polystyrene (HIPS). However, no prior research has evaluated the radiological properties of the various 3D printing materials currently available, knowledge of which is crucial for these material’s proper application.
In this study, the radiological properties of newly introduced and existing 3D printing materials were evaluated by measuring their Hounsfield units (HUs). The HU values were evaluated by varying the infill density of each printing material, thus allowing us to assess their applicability to the field of radiation oncology.
In this study, a total of six 3D printing materials were used: ABS, ULTRAT, HIPS, PETG, PLA, and FLEX. Table 1 shows the physical properties of the six materials; of these, PETG exhibited the highest tensile strength (the mechanical strength of the material), FLEX had the highest bending stress (the stress that occurs inside a material by bending moment), and the lowest shore hardness (and was thus the most flexible material). HIPS had the lowest specific density (1.14 g/cm3), and PLA had the highest specific density (1.43 g/cm3).
The 3D printer used in this study was a Zortrax M300 Plus model (Zortrax SA, Olsztyn, Poland). The specifications of this printer are shown in Table 2. The Z-suite program (Zortrax SA, Olsztyn, Poland) was used for printing; the program’s printer settings are shown in Fig. 1. This program allows printing with various infill densities. For the study experiments, rectangles with dimensions of 4×4×2 cm were fabricated using HIPS (with the lowest specific density) and PLA (with the highest specific density). The infill densities of these rectangles were varied as 10%, 20%, 40%, 60%, and 100%. Rectangles of the other four kinds of materials of the same size were also fabricated, each with a 60% infill density.
In order to measure the HU values of the rectangles, CT images of the six materials were acquired. The scan thickness was 1 mm and the scan conditions were 120 kV and 200 mA. Under the same conditions, CT images were obtained of HIPS and PLA with varying infill densities. Fig. 2 shows images of HIPS samples with infill densities from 10% to 100%, alongside their respective CT images.
HUs are a value constituting the grayscale level of a CT image, and can be calculated through the image’s linear attenuation coefficient. The equation for calculating HUs is defined below:
where μmaterial is the linear attenuation coefficient for an arbitrary material and μwater is the linear attenuation coefficient for water. Air has an HU value of −1,000, because its μ value is 0, and water has an HU value of 0. In this study, a 1×1 cm region of interest was obtained through a coronal view CT image to evaluate the HU for each material and infill density; and average HU values were obtained for all six materials with 60% infill density. Average HUs were also obtained for HIPS and PLA with 10%–100% infill densities. The HU values for HIPS and PLA at different infill densities were subjected to curve fitting using the equation below:
This study investigated the radiological properties of six 3D printing materials, to determine how to incorporate a wide range of 3D printing technologies into radiation oncology. The average HU values of the six printing materials were obtained, and the average HU values of HIPS and PLA were analyzed across a range of infill densities.
The time required to print each material at each infill density was also measured. When the infill density was 10%, the average printing time was 32 minutes; when the infill density was 100%, the average time was 12 hours and 45 minutes. Considering that the volume of each prepared cube was 32 m3, a printing time of 12 hours and 45 minutes could be considered quite wasteful. Using 100% infill density should therefore be carefully considered while manufacturing patient-specific human phantoms.11–13)
Table 3 shows the HU values for the six printing materials, each at an infill density of 60%. ABS, ULTRAT, HIPS, PETG, PLA, and FLEX had HU values of −535±12, −557±10, −542±7, −508±20, −530±25, and −633±15, respectively. The HU values were similar regardless of the specific density of the materials.
Fig. 3 includes a graph showing HU changing with the infill density of HIPS and PLA. The HU value increased with infill density for each material. The HU values for HIPS and PLA were −131 and 61, respectively, at an infill density of 100%. By comparison, at an infill density of 10%, the HU values for HIPS and PLA were −896 and −904, respectively, In addition, the parameters for fitting the relationship between infill density and HU using
In this study, HU values were obtained for various 3D printing materials at different infill densities; the radiological characteristics of each material were then analyzed. The results suggest that using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.
This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Korea (grant number: VHSMC 19021).
Physical characteristics of materials used in 3-dimensional printing
Variable | ABS | ULTRAT | HIPS | PETG | PLA | FLEX |
---|---|---|---|---|---|---|
Tensile strength (MPa) | 30.5 | 32.6 | 16.9 | 40.2 | 28.8 | 9.9 |
Bending stress (MPa) | 46.3 | 54.0 | 29.3 | 55.3 | 58.6 | 174.3 |
Specific density (g/cm3) | 1.20 | 1.18 | 1.14 | 1.30 | 1.43 | 1.19 |
Shore hardness (D) | 69.2 | 73.4 | 73.2 | 71.4 | 77.6 | 31.0 |
Specifications for 3-dimensional printer
Specification | Specification value |
---|---|
Build volume (mm3) | 300×300×300 |
Layer resolution (μm) | 90–290 |
Minimal wall thickness (μm) | 400 |
Material diameter (nm) | 1.75 |
Nozzle diameter (mm) | 1.4 |
Maximum printing temperature (°F) | 554 |
Supported input file types | .stl, .obj, .dxf, .3mf |
Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%
ABS | ULTRAT | HIPS | PETG | PLA | FLEX | |
---|---|---|---|---|---|---|
HU values | −535±12 | −557±10 | −542±7 | −508±20 | −530±25 | −633±15 |
Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA
Material | Parameter | R-squared value | ||
---|---|---|---|---|
a | b | c | ||
HIPS | 0.033 | 5.125 | −962.607 | 0.994 |
PLA | 0.079 | 1.286 | −892.879 | 0.976 |
Table 1 Physical characteristics of materials used in 3-dimensional printing
Variable | ABS | ULTRAT | HIPS | PETG | PLA | FLEX |
---|---|---|---|---|---|---|
Tensile strength (MPa) | 30.5 | 32.6 | 16.9 | 40.2 | 28.8 | 9.9 |
Bending stress (MPa) | 46.3 | 54.0 | 29.3 | 55.3 | 58.6 | 174.3 |
Specific density (g/cm3) | 1.20 | 1.18 | 1.14 | 1.30 | 1.43 | 1.19 |
Shore hardness (D) | 69.2 | 73.4 | 73.2 | 71.4 | 77.6 | 31.0 |
ABS, acrylonitrile butadiene styrene; ULTRAT, ABS plastic blend unique manufactured by Zortrax; HIPS, high impact polystyrene; PETG, polyethylene terephthalate glycol; PLA, polylactic acid; FLEX, thermoplastic polyester elastomer manufactured by Zortrax.
Table 2 Specifications for 3-dimensional printer
Specification | Specification value |
---|---|
Build volume (mm3) | 300×300×300 |
Layer resolution (μm) | 90–290 |
Minimal wall thickness (μm) | 400 |
Material diameter (nm) | 1.75 |
Nozzle diameter (mm) | 1.4 |
Maximum printing temperature (°F) | 554 |
Supported input file types | .stl, .obj, .dxf, .3mf |
Table 3 Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%
ABS | ULTRAT | HIPS | PETG | PLA | FLEX | |
---|---|---|---|---|---|---|
HU values | −535±12 | −557±10 | −542±7 | −508±20 | −530±25 | −633±15 |
ABS, acrylonitrile butadiene styrene; ULTRAT, ABS plastic blend unique manufactured by Zortrax; HIPS, high impact polystyrene; PETG, polyethylene terephthalate glycol; PLA, polylactic acid; FLEX, thermoplastic polyester elastomer manufactured by Zortrax.
Table 4 Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA
Material | Parameter | R-squared value | ||
---|---|---|---|---|
a | b | c | ||
HIPS | 0.033 | 5.125 | −962.607 | 0.994 |
PLA | 0.079 | 1.286 | −892.879 | 0.976 |
HIPS, high impact polystyrene; PLA; polylactic acid.