Technical Note

Progress in Medical Physics 2019; 30(4): 155-159

Published online December 31, 2019

https://doi.org/10.14316/pmp.2019.30.4.155

Copyright © Korean Society of Medical Physics.

Radiological Characteristics of Materials Used in 3-Dimensional Printing with Various Infill Densities

So-Yeon Park , Noorie Choi , Byeong Geol Choi , Dong Myung Lee , Na Young Jang

Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Korea

Correspondence to:Na Young Jang (mumuki79@gmail.com)
Tel: 82-2-2225-4647 Fax: 82-2-2225-4640

Received: December 13, 2019; Revised: December 18, 2019; Accepted: December 18, 2019

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Radiological properties of newly introduced and existing 3-dimensional (3D) printing materials were evaluated by measuring their Hounsfield units (HUs) at varying infill densities. The six materials for 3D printing which consisted of acrylonitrile butadiene styrene (ABS), a unique ABS plastic blend manufactured by Zortrax (ULTRAT), high impact polystyrene (HIPS), polyethylene terephthalate glycol (PETG), polylactic acid (PLA), and a thermoplastic polyester elastomer manufactured by Zortrax (FLEX) were used. We used computed tomography (CT) imaging to determine the HU values of each material, and thus assess its suitability for various applications in radiation oncology. We found that several material and infill density combinations resembled the HU values of fat, soft tissues, and lungs; however, none of the tested materials exhibited HU values similar to that of bone. These results will help researchers and clinicians develop more appropriate instruments for improving the quality of radiation therapy. Using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.

Keywords3D printing, Computed tomography, Hounsfield unit, Infill density, Radiation oncology

Three-dimensional (3D) printing technology is widely used in the field of radiation oncology to make computed tomography (CT)-based patient-specific tools, thereby improving the quality of radiation therapy.17) For example, patient-specific compensators for radiotherapy are fabricated for use in areas with severe curvatures, such as the nose, feet, and hands, thus delivering more radiation to a tumor in the area and reducing radiation to the surrounding normal organs.4) For total body irradiation, 3D-printed patient-specific metal compensators reflecting the overall curves of the patient’s body shape can deliver more uniform radiation doses to the whole body than conventional methods.6) 3D printing technology has also been used for high dose rate (HDR) brachytherapy to create patient-specific elastic HDR applicators. These deliver accurate doses of radiation to the skin’s surface; this technology has been proven to improve treatment delivery accuracy.2,4,8) In addition, many attempts have been made to perform quality assurance in radiation therapy by producing patient-specific phantoms using patient CTs and 3D printers.5,913)

In the 3D printing field, advances have been made not only in printing technology, but also in the printing materials used. Various synthetic 3D printing materials have been developed, including polyethylene terephthalate glycol (PETG) and a unique acrylonitrile butadiene styrene (ABS) plastic blend manufactured by Zortrax (ULTRAT; Zortrax SA, Olsztyn, Poland) which is in a type of thermoplastic polymer. Flexible materials have also been developed, like the thermoplastic polyester elastomer manufactured by Zortrax (FLEX) which had great interlayer adhesion and dimensional tolerance. These materials can be used for a wide range of applications in 3D printing for radiation therapy, in combination with existing 3D printing materials like ABS, polylactic acid (PLA), and high impact polystyrene (HIPS). However, no prior research has evaluated the radiological properties of the various 3D printing materials currently available, knowledge of which is crucial for these material’s proper application.

In this study, the radiological properties of newly introduced and existing 3D printing materials were evaluated by measuring their Hounsfield units (HUs). The HU values were evaluated by varying the infill density of each printing material, thus allowing us to assess their applicability to the field of radiation oncology.

1. Types of materials used in 3-dimensional printing

In this study, a total of six 3D printing materials were used: ABS, ULTRAT, HIPS, PETG, PLA, and FLEX. Table 1 shows the physical properties of the six materials; of these, PETG exhibited the highest tensile strength (the mechanical strength of the material), FLEX had the highest bending stress (the stress that occurs inside a material by bending moment), and the lowest shore hardness (and was thus the most flexible material). HIPS had the lowest specific density (1.14 g/cm3), and PLA had the highest specific density (1.43 g/cm3).

2. Three-dimensional printing with various infill densities

The 3D printer used in this study was a Zortrax M300 Plus model (Zortrax SA, Olsztyn, Poland). The specifications of this printer are shown in Table 2. The Z-suite program (Zortrax SA, Olsztyn, Poland) was used for printing; the program’s printer settings are shown in Fig. 1. This program allows printing with various infill densities. For the study experiments, rectangles with dimensions of 4×4×2 cm were fabricated using HIPS (with the lowest specific density) and PLA (with the highest specific density). The infill densities of these rectangles were varied as 10%, 20%, 40%, 60%, and 100%. Rectangles of the other four kinds of materials of the same size were also fabricated, each with a 60% infill density.

3. Computed tomography imaging

In order to measure the HU values of the rectangles, CT images of the six materials were acquired. The scan thickness was 1 mm and the scan conditions were 120 kV and 200 mA. Under the same conditions, CT images were obtained of HIPS and PLA with varying infill densities. Fig. 2 shows images of HIPS samples with infill densities from 10% to 100%, alongside their respective CT images.

4. Hounsfield unit measurement

HUs are a value constituting the grayscale level of a CT image, and can be calculated through the image’s linear attenuation coefficient. The equation for calculating HUs is defined below:

HU=μmaterial-μwaterμwater×1000,

where μmaterial is the linear attenuation coefficient for an arbitrary material and μwater is the linear attenuation coefficient for water. Air has an HU value of −1,000, because its μ value is 0, and water has an HU value of 0. In this study, a 1×1 cm region of interest was obtained through a coronal view CT image to evaluate the HU for each material and infill density; and average HU values were obtained for all six materials with 60% infill density. Average HUs were also obtained for HIPS and PLA with 10%–100% infill densities. The HU values for HIPS and PLA at different infill densities were subjected to curve fitting using the equation below:

a·Infilldensity2+b·Infilldensity+c=Hounsfieldunit

This study investigated the radiological properties of six 3D printing materials, to determine how to incorporate a wide range of 3D printing technologies into radiation oncology. The average HU values of the six printing materials were obtained, and the average HU values of HIPS and PLA were analyzed across a range of infill densities.

The time required to print each material at each infill density was also measured. When the infill density was 10%, the average printing time was 32 minutes; when the infill density was 100%, the average time was 12 hours and 45 minutes. Considering that the volume of each prepared cube was 32 m3, a printing time of 12 hours and 45 minutes could be considered quite wasteful. Using 100% infill density should therefore be carefully considered while manufacturing patient-specific human phantoms.1113)

Table 3 shows the HU values for the six printing materials, each at an infill density of 60%. ABS, ULTRAT, HIPS, PETG, PLA, and FLEX had HU values of −535±12, −557±10, −542±7, −508±20, −530±25, and −633±15, respectively. The HU values were similar regardless of the specific density of the materials.

Fig. 3 includes a graph showing HU changing with the infill density of HIPS and PLA. The HU value increased with infill density for each material. The HU values for HIPS and PLA were −131 and 61, respectively, at an infill density of 100%. By comparison, at an infill density of 10%, the HU values for HIPS and PLA were −896 and −904, respectively, In addition, the parameters for fitting the relationship between infill density and HU using Equation (2) are shown in Table 4. In the case of HIPS and PLA, the R2 values were 0.994 and 0.976, respectively, indicating that the fitting was successful. Thus, when manufacturing patient-specific phantoms using current materials, HU values for lungs, soft tissues, and fats can be simulated by calculating appropriate infill densities. Matching the HU of bone, however, proved to be difficult when using a material with a high HU value.14) Additional studies should be conducted to find a material that can reflect the HU of bone. In addition, there is a need for a method for overcoming the limitations of the printing material itself (i.e., they cannot produce a sufficiently high HU), including a method for manufacturing a casting mold from the printing material and casting a material with a high HU. An additional study should be conducted to develop such a method.

In this study, HU values were obtained for various 3D printing materials at different infill densities; the radiological characteristics of each material were then analyzed. The results suggest that using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.

This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Korea (grant number: VHSMC 19021).

Physical characteristics of materials used in 3-dimensional printing

VariableABSULTRATHIPSPETGPLAFLEX
Tensile strength (MPa)30.532.616.940.228.89.9
Bending stress (MPa)46.354.029.355.358.6174.3
Specific density (g/cm3)1.201.181.141.301.431.19
Shore hardness (D)69.273.473.271.477.631.0

Specifications for 3-dimensional printer

SpecificationSpecification value
Build volume (mm3)300×300×300
Layer resolution (μm)90–290
Minimal wall thickness (μm)400
Material diameter (nm)1.75
Nozzle diameter (mm)1.4
Maximum printing temperature (°F)554
Supported input file types.stl, .obj, .dxf, .3mf

Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%

ABSULTRATHIPSPETGPLAFLEX
HU values−535±12−557±10−542±7−508±20−530±25−633±15

Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA

MaterialParameterR-squared value

abc
HIPS0.0335.125−962.6070.994
PLA0.0791.286−892.8790.976
  1. Jones EL, Tonino Baldion A, Thomas C, Burrows T, Byrne N, and Newton V, et al. Introduction of novel 3D-printed superficial applicators for high-dose-rate skin brachytherapy. Brachytherapy 2017;16:409-414.
    Pubmed CrossRef
  2. Ricotti R, Vavassori A, Bazani A, Ciardo D, Pansini F, and Spoto R, et al. 3D-printed applicators for high dose rate brachytherapy: dosimetric assessment at different infill percentage. Phys Med 2016;32:1698-1706.
    Pubmed CrossRef
  3. Zhao Y, Moran K, Yewondwossen M, Allan J, Clarke S, and Rajaraman M, et al. Clinical applications of 3-dimensional printing in radiation therapy. Med Dosim 2017;42:150-155.
    Pubmed CrossRef
  4. Park SY, Kang S, Park JM, An HJ, Oh DH, and Kim JI. Development and dosimetric assessment of a patient-specific elastic skin applicator for high-dose-rate brachytherapy. Brachytherapy 2019;18:224-232.
    Pubmed CrossRef
  5. Madamesila J, McGeachy P, Villarreal Barajas JE, and Khan R. Characterizing 3D printing in the fabrication of variable density phantoms for quality assurance of radiotherapy. Phys Med 2016;32:242-247.
    Pubmed CrossRef
  6. Park SY, Kim JI, Joo YH, Lee JC, and Park JM. Total body irradiation with a compensator fabricated using a 3D optical scanner and a 3D printer. Phys Med Biol 2017;62:3735-3756.
    Pubmed CrossRef
  7. Park SY, Choi CH, Park JM, Chun M, Han JH, and Kim JI. A patient-specific polylactic acid bolus made by a 3D printer for breast cancer radiation therapy. PLoS One 2016;11:e0168063.
    Pubmed KoreaMed CrossRef
  8. Choi CH, Kim JI, and Park JM. A 3D-printed patient-specific applicator guide for use in high-dose-rate interstitial brachytherapy for tongue cancer: a phantom study. Phys Med Biol 2019;64 135002.
    Pubmed CrossRef
  9. Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos AA, Cai T, and Kumamaru KK, et al. Medical 3D printing for the radiologist. Radiographics 2015;35:1965-1988.
    Pubmed KoreaMed CrossRef
  10. Kadoya N, Abe K, Nemoto H, Sato K, Ieko Y, and Ito K, et al. Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy. Radiol Phys Technol 2019;12:351-356.
    Pubmed CrossRef
  11. Ehler ED, Barney BM, Higgins PD, and Dusenbery KE. Patient specific 3D printed phantom for IMRT quality assurance. Phys Med Biol 2014;59:5763-5773.
    Pubmed CrossRef
  12. Yea JW, Park JW, Kim SK, Kim DY, Kim JG, and Seo CY, et al. Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy. PLoS One 2017;12:e0181560.
    Pubmed KoreaMed CrossRef
  13. Lee CL, Dietrich MC, Desai UG, Das A, Yu S, and Xiang HF, et al. A 3D-printed patient-specific phantom for external beam radiation therapy of prostate cancer. J Eng Sci Med Diagn Ther 2018;1 041004.
    Pubmed KoreaMed CrossRef
  14. Oh SA, Kim MJ, Kang JS, Hwang HS, Kim YJ, and Kim SH, et al. Feasibility of fabricating variable density phantoms using 3D printing for quality assurance (QA) in radiotherapy. Prog Med Phys 2017;28:106-110.
    CrossRef

Article

Technical Note

Progress in Medical Physics 2019; 30(4): 155-159

Published online December 31, 2019 https://doi.org/10.14316/pmp.2019.30.4.155

Copyright © Korean Society of Medical Physics.

Radiological Characteristics of Materials Used in 3-Dimensional Printing with Various Infill Densities

So-Yeon Park , Noorie Choi , Byeong Geol Choi , Dong Myung Lee , Na Young Jang

Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Korea

Correspondence to:Na Young Jang (mumuki79@gmail.com)
Tel: 82-2-2225-4647 Fax: 82-2-2225-4640

Received: December 13, 2019; Revised: December 18, 2019; Accepted: December 18, 2019

This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Radiological properties of newly introduced and existing 3-dimensional (3D) printing materials were evaluated by measuring their Hounsfield units (HUs) at varying infill densities. The six materials for 3D printing which consisted of acrylonitrile butadiene styrene (ABS), a unique ABS plastic blend manufactured by Zortrax (ULTRAT), high impact polystyrene (HIPS), polyethylene terephthalate glycol (PETG), polylactic acid (PLA), and a thermoplastic polyester elastomer manufactured by Zortrax (FLEX) were used. We used computed tomography (CT) imaging to determine the HU values of each material, and thus assess its suitability for various applications in radiation oncology. We found that several material and infill density combinations resembled the HU values of fat, soft tissues, and lungs; however, none of the tested materials exhibited HU values similar to that of bone. These results will help researchers and clinicians develop more appropriate instruments for improving the quality of radiation therapy. Using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.

Keywords: 3D printing, Computed tomography, Hounsfield unit, Infill density, Radiation oncology

Introduction

Three-dimensional (3D) printing technology is widely used in the field of radiation oncology to make computed tomography (CT)-based patient-specific tools, thereby improving the quality of radiation therapy.17) For example, patient-specific compensators for radiotherapy are fabricated for use in areas with severe curvatures, such as the nose, feet, and hands, thus delivering more radiation to a tumor in the area and reducing radiation to the surrounding normal organs.4) For total body irradiation, 3D-printed patient-specific metal compensators reflecting the overall curves of the patient’s body shape can deliver more uniform radiation doses to the whole body than conventional methods.6) 3D printing technology has also been used for high dose rate (HDR) brachytherapy to create patient-specific elastic HDR applicators. These deliver accurate doses of radiation to the skin’s surface; this technology has been proven to improve treatment delivery accuracy.2,4,8) In addition, many attempts have been made to perform quality assurance in radiation therapy by producing patient-specific phantoms using patient CTs and 3D printers.5,913)

In the 3D printing field, advances have been made not only in printing technology, but also in the printing materials used. Various synthetic 3D printing materials have been developed, including polyethylene terephthalate glycol (PETG) and a unique acrylonitrile butadiene styrene (ABS) plastic blend manufactured by Zortrax (ULTRAT; Zortrax SA, Olsztyn, Poland) which is in a type of thermoplastic polymer. Flexible materials have also been developed, like the thermoplastic polyester elastomer manufactured by Zortrax (FLEX) which had great interlayer adhesion and dimensional tolerance. These materials can be used for a wide range of applications in 3D printing for radiation therapy, in combination with existing 3D printing materials like ABS, polylactic acid (PLA), and high impact polystyrene (HIPS). However, no prior research has evaluated the radiological properties of the various 3D printing materials currently available, knowledge of which is crucial for these material’s proper application.

In this study, the radiological properties of newly introduced and existing 3D printing materials were evaluated by measuring their Hounsfield units (HUs). The HU values were evaluated by varying the infill density of each printing material, thus allowing us to assess their applicability to the field of radiation oncology.

Materials and Methods

1. Types of materials used in 3-dimensional printing

In this study, a total of six 3D printing materials were used: ABS, ULTRAT, HIPS, PETG, PLA, and FLEX. Table 1 shows the physical properties of the six materials; of these, PETG exhibited the highest tensile strength (the mechanical strength of the material), FLEX had the highest bending stress (the stress that occurs inside a material by bending moment), and the lowest shore hardness (and was thus the most flexible material). HIPS had the lowest specific density (1.14 g/cm3), and PLA had the highest specific density (1.43 g/cm3).

2. Three-dimensional printing with various infill densities

The 3D printer used in this study was a Zortrax M300 Plus model (Zortrax SA, Olsztyn, Poland). The specifications of this printer are shown in Table 2. The Z-suite program (Zortrax SA, Olsztyn, Poland) was used for printing; the program’s printer settings are shown in Fig. 1. This program allows printing with various infill densities. For the study experiments, rectangles with dimensions of 4×4×2 cm were fabricated using HIPS (with the lowest specific density) and PLA (with the highest specific density). The infill densities of these rectangles were varied as 10%, 20%, 40%, 60%, and 100%. Rectangles of the other four kinds of materials of the same size were also fabricated, each with a 60% infill density.

3. Computed tomography imaging

In order to measure the HU values of the rectangles, CT images of the six materials were acquired. The scan thickness was 1 mm and the scan conditions were 120 kV and 200 mA. Under the same conditions, CT images were obtained of HIPS and PLA with varying infill densities. Fig. 2 shows images of HIPS samples with infill densities from 10% to 100%, alongside their respective CT images.

4. Hounsfield unit measurement

HUs are a value constituting the grayscale level of a CT image, and can be calculated through the image’s linear attenuation coefficient. The equation for calculating HUs is defined below:

HU=μmaterial-μwaterμwater×1000,

where μmaterial is the linear attenuation coefficient for an arbitrary material and μwater is the linear attenuation coefficient for water. Air has an HU value of −1,000, because its μ value is 0, and water has an HU value of 0. In this study, a 1×1 cm region of interest was obtained through a coronal view CT image to evaluate the HU for each material and infill density; and average HU values were obtained for all six materials with 60% infill density. Average HUs were also obtained for HIPS and PLA with 10%–100% infill densities. The HU values for HIPS and PLA at different infill densities were subjected to curve fitting using the equation below:

a·Infilldensity2+b·Infilldensity+c=Hounsfieldunit

Results and Discussion

This study investigated the radiological properties of six 3D printing materials, to determine how to incorporate a wide range of 3D printing technologies into radiation oncology. The average HU values of the six printing materials were obtained, and the average HU values of HIPS and PLA were analyzed across a range of infill densities.

The time required to print each material at each infill density was also measured. When the infill density was 10%, the average printing time was 32 minutes; when the infill density was 100%, the average time was 12 hours and 45 minutes. Considering that the volume of each prepared cube was 32 m3, a printing time of 12 hours and 45 minutes could be considered quite wasteful. Using 100% infill density should therefore be carefully considered while manufacturing patient-specific human phantoms.1113)

Table 3 shows the HU values for the six printing materials, each at an infill density of 60%. ABS, ULTRAT, HIPS, PETG, PLA, and FLEX had HU values of −535±12, −557±10, −542±7, −508±20, −530±25, and −633±15, respectively. The HU values were similar regardless of the specific density of the materials.

Fig. 3 includes a graph showing HU changing with the infill density of HIPS and PLA. The HU value increased with infill density for each material. The HU values for HIPS and PLA were −131 and 61, respectively, at an infill density of 100%. By comparison, at an infill density of 10%, the HU values for HIPS and PLA were −896 and −904, respectively, In addition, the parameters for fitting the relationship between infill density and HU using Equation (2) are shown in Table 4. In the case of HIPS and PLA, the R2 values were 0.994 and 0.976, respectively, indicating that the fitting was successful. Thus, when manufacturing patient-specific phantoms using current materials, HU values for lungs, soft tissues, and fats can be simulated by calculating appropriate infill densities. Matching the HU of bone, however, proved to be difficult when using a material with a high HU value.14) Additional studies should be conducted to find a material that can reflect the HU of bone. In addition, there is a need for a method for overcoming the limitations of the printing material itself (i.e., they cannot produce a sufficiently high HU), including a method for manufacturing a casting mold from the printing material and casting a material with a high HU. An additional study should be conducted to develop such a method.

Conclusions

In this study, HU values were obtained for various 3D printing materials at different infill densities; the radiological characteristics of each material were then analyzed. The results suggest that using optimized infill densities will help improve the quality of radiation therapy by producing customized instruments for each field of radiation therapy.

Acknowledgements

This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Korea (grant number: VHSMC 19021).

Tables

Physical characteristics of materials used in 3-dimensional printing

VariableABSULTRATHIPSPETGPLAFLEX
Tensile strength (MPa)30.532.616.940.228.89.9
Bending stress (MPa)46.354.029.355.358.6174.3
Specific density (g/cm3)1.201.181.141.301.431.19
Shore hardness (D)69.273.473.271.477.631.0

Specifications for 3-dimensional printer

SpecificationSpecification value
Build volume (mm3)300×300×300
Layer resolution (μm)90–290
Minimal wall thickness (μm)400
Material diameter (nm)1.75
Nozzle diameter (mm)1.4
Maximum printing temperature (°F)554
Supported input file types.stl, .obj, .dxf, .3mf

Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%

ABSULTRATHIPSPETGPLAFLEX
HU values−535±12−557±10−542±7−508±20−530±25−633±15

Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA

MaterialParameterR-squared value

abc
HIPS0.0335.125−962.6070.994
PLA0.0791.286−892.8790.976

Fig 1.

Figure 1.Printing settings in Z-Suite program to select the infill density values.
Progress in Medical Physics 2019; 30: 155-159https://doi.org/10.14316/pmp.2019.30.4.155

Fig 2.

Figure 2.(a) High impact polystyrene material with infill density values from 10% to 100% and (b) computed tomography (CT) images in coronal view obtained by the Brilliance CT Big Bore.
Progress in Medical Physics 2019; 30: 155-159https://doi.org/10.14316/pmp.2019.30.4.155

Fig 3.

Figure 3.Correlations between Hounsfield unit (HUs) and infill density values for high impact polystyrene (HIPS) and polylactic acid (PLA) are plotted with dashed lines, respectively. Polynomial fitting curves for these correlations for HIPS and PLA are plotted with solid lines, respectively.
Progress in Medical Physics 2019; 30: 155-159https://doi.org/10.14316/pmp.2019.30.4.155

Table 1 Physical characteristics of materials used in 3-dimensional printing

VariableABSULTRATHIPSPETGPLAFLEX
Tensile strength (MPa)30.532.616.940.228.89.9
Bending stress (MPa)46.354.029.355.358.6174.3
Specific density (g/cm3)1.201.181.141.301.431.19
Shore hardness (D)69.273.473.271.477.631.0

ABS, acrylonitrile butadiene styrene; ULTRAT, ABS plastic blend unique manufactured by Zortrax; HIPS, high impact polystyrene; PETG, polyethylene terephthalate glycol; PLA, polylactic acid; FLEX, thermoplastic polyester elastomer manufactured by Zortrax.


Table 2 Specifications for 3-dimensional printer

SpecificationSpecification value
Build volume (mm3)300×300×300
Layer resolution (μm)90–290
Minimal wall thickness (μm)400
Material diameter (nm)1.75
Nozzle diameter (mm)1.4
Maximum printing temperature (°F)554
Supported input file types.stl, .obj, .dxf, .3mf

Table 3 Hounsfield unit (HU) values of materials used in 3-dimensional printing with infill density of 60%

ABSULTRATHIPSPETGPLAFLEX
HU values−535±12−557±10−542±7−508±20−530±25−633±15

ABS, acrylonitrile butadiene styrene; ULTRAT, ABS plastic blend unique manufactured by Zortrax; HIPS, high impact polystyrene; PETG, polyethylene terephthalate glycol; PLA, polylactic acid; FLEX, thermoplastic polyester elastomer manufactured by Zortrax.


Table 4 Values of fitting parameters and R-squared values for second polynomial fits of HIPS and PLA

MaterialParameterR-squared value

abc
HIPS0.0335.125−962.6070.994
PLA0.0791.286−892.8790.976

HIPS, high impact polystyrene; PLA; polylactic acid.


References

  1. Jones EL, Tonino Baldion A, Thomas C, Burrows T, Byrne N, and Newton V, et al. Introduction of novel 3D-printed superficial applicators for high-dose-rate skin brachytherapy. Brachytherapy 2017;16:409-414.
    Pubmed CrossRef
  2. Ricotti R, Vavassori A, Bazani A, Ciardo D, Pansini F, and Spoto R, et al. 3D-printed applicators for high dose rate brachytherapy: dosimetric assessment at different infill percentage. Phys Med 2016;32:1698-1706.
    Pubmed CrossRef
  3. Zhao Y, Moran K, Yewondwossen M, Allan J, Clarke S, and Rajaraman M, et al. Clinical applications of 3-dimensional printing in radiation therapy. Med Dosim 2017;42:150-155.
    Pubmed CrossRef
  4. Park SY, Kang S, Park JM, An HJ, Oh DH, and Kim JI. Development and dosimetric assessment of a patient-specific elastic skin applicator for high-dose-rate brachytherapy. Brachytherapy 2019;18:224-232.
    Pubmed CrossRef
  5. Madamesila J, McGeachy P, Villarreal Barajas JE, and Khan R. Characterizing 3D printing in the fabrication of variable density phantoms for quality assurance of radiotherapy. Phys Med 2016;32:242-247.
    Pubmed CrossRef
  6. Park SY, Kim JI, Joo YH, Lee JC, and Park JM. Total body irradiation with a compensator fabricated using a 3D optical scanner and a 3D printer. Phys Med Biol 2017;62:3735-3756.
    Pubmed CrossRef
  7. Park SY, Choi CH, Park JM, Chun M, Han JH, and Kim JI. A patient-specific polylactic acid bolus made by a 3D printer for breast cancer radiation therapy. PLoS One 2016;11:e0168063.
    Pubmed KoreaMed CrossRef
  8. Choi CH, Kim JI, and Park JM. A 3D-printed patient-specific applicator guide for use in high-dose-rate interstitial brachytherapy for tongue cancer: a phantom study. Phys Med Biol 2019;64 135002.
    Pubmed CrossRef
  9. Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos AA, Cai T, and Kumamaru KK, et al. Medical 3D printing for the radiologist. Radiographics 2015;35:1965-1988.
    Pubmed KoreaMed CrossRef
  10. Kadoya N, Abe K, Nemoto H, Sato K, Ieko Y, and Ito K, et al. Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy. Radiol Phys Technol 2019;12:351-356.
    Pubmed CrossRef
  11. Ehler ED, Barney BM, Higgins PD, and Dusenbery KE. Patient specific 3D printed phantom for IMRT quality assurance. Phys Med Biol 2014;59:5763-5773.
    Pubmed CrossRef
  12. Yea JW, Park JW, Kim SK, Kim DY, Kim JG, and Seo CY, et al. Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy. PLoS One 2017;12:e0181560.
    Pubmed KoreaMed CrossRef
  13. Lee CL, Dietrich MC, Desai UG, Das A, Yu S, and Xiang HF, et al. A 3D-printed patient-specific phantom for external beam radiation therapy of prostate cancer. J Eng Sci Med Diagn Ther 2018;1 041004.
    Pubmed KoreaMed CrossRef
  14. Oh SA, Kim MJ, Kang JS, Hwang HS, Kim YJ, and Kim SH, et al. Feasibility of fabricating variable density phantoms using 3D printing for quality assurance (QA) in radiotherapy. Prog Med Phys 2017;28:106-110.
    CrossRef
Korean Society of Medical Physics

Vol.35 No.3
September 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line