Progress in Medical Physics 2020; 31(4): 179-188
Published online December 31, 2020
https://doi.org/10.14316/pmp.2020.31.4.179
Copyright © Korean Society of Medical Physics.
Correspondence to:So-Yeon Park
(vsoyounv@gmail.com)
Tel: 82-2-2225-4648
Fax: 82-2-2225-4640
Purpose: We compared noncoplanar volumetric modulated arc therapy (ncVMAT) plans to coplanar VMAT (cVMAT) plans by evaluating the dosimetric quality of each for esophageal cancer.
Methods: Twenty patients treated for esophageal cancer with the cVMAT technique were retrospectively selected. The cVMAT plans consisted of three coplanar full arc beams. The ncVMAT plans consisted of two coplanar full arc beams and one noncoplanar partial arc beam ranging from 45° to 315° with a couch rotation angle of 315°±5°. For dosimetric evaluation, the dose-volumetric (DV) parameters of the planning target volume (PTV) and organs at risk (OARs) were calculated for all VMAT plans.
Results: No clinically noticeable differences between the cVMAT and ncVMAT plans were observed in the DV parameters of the PTV. For the lungs, the VM13 Gy and mean dose for ncVMAT plans were smaller than those for cVMAT plans, showing statistically significant differences. For the heart, the values of the maximum dose for cVMAT and ncVMAT plans were 53.8±2.9 and 50.9±3.3 Gy, respectively (P=0.004). For the spinal cord, the values of the maximum dose for cVMAT and ncVMAT plans were 37.1±5.1 and 34.7±5.7 Gy, respectively (P<0.001).
Conclusions: The use of ncVMAT plans provides better PTV coverage and sparing of OARs compared to that of cVMAT plans for long, tube-like esophageal cancer. For esophageal cancer, the ncVMAT plans showed a more favorable plan quality than the cVMAT plans.
KeywordsDosimetric evaluation, Noncoplanar, Volumetric modulated arc therapy, Esophageal cancer
Esophageal cancer, which mainly occurs in patients aged from 50 to 80 years, is the 11th most common type of cancer in Korea, accounting for approximately 1.7% of all cancers. An estimated 1,600 new patients are diagnosed with esophageal cancer every year [1]. Treatment options for esophageal cancer are surgery, chemotherapy, or radiotherapy [2
-8]. Radiotherapy for inoperable or unresectable esophageal cancer plays a crucial role in achieving optimal tumor control along with reducing normal tissue complications. However, a planning target volume (PTV) is a long, tube-like structure that connects the throat and stomach, as shown in Fig. 1, which creates challenges when planning a treatment. As there are several organs at risk (OARs) in the vicinity of the esophageal cancer, which include the lungs, heart, liver, spinal cord, and stomach, incorrect treatment may cause serious side effects such as radiation pneumonitis and fibrosis in the lungs as well as pericardial effusion, pericarditis, and heart failure [9].
Several studies have compared various radiotherapy techniques for esophageal cancer, including 3-dimensional conformal radiation therapy (3D CRT), intensity modulated radiation therapy (IMRT), and volumetric modulated arc therapy (VMAT) [2,9
-12]. These studies have reported that VMAT plans provide better dose conformity to the PTV while reducing the doses affecting OARs, compared to 3D CRT and IMRT plans [2,9
-12]. Furthermore, hybrid IMRT plans that combine static and IMRT beams, and hybrid VMAT plans that combine static beams and VMAT arcs concurrently, have been proposed by several studies to reduce the dose volume applied to normal tissues [9,13
-15]. These techniques were found to be superior in dosimetric outcomes when compared to conventional IMRT and VMAT plans.
Li et al. [16] proposed noncoplanar IMRT plans to reduce doses to surrounding OARs while improving PTV coverage for esophageal cancer. They showed that compared to coplanar IMRT plans, noncoplanar IMRT plans reduce the low dose exposure to the lungs, while providing a similar PTV coverage. The disadvantage of noncoplanar IMRT plans is that higher doses are delivered to the heart and spinal cord, making it inadvisable to use noncoplanar IMRT plans for patients with potential cardiovascular and neurological problems. Martini et al. [17] studied noncoplanar VMAT (ncVMAT) plans for patients with esophageal cancer who underwent either definitive or preoperative radiotherapy. Martini et al. [17] evaluated the feasibility, dosimetric results, and clinical outcomes. They showed that, in patients treated with ncVMAT plans using two coplanar and one noncoplanar arc beams, the side effects observed included esophagitis, anorexia, fatigue, nausea, and vomiting. The dosimetric results showed adequate PTV coverage and sparing of normal tissue. They concluded that the use of ncVMAT plans was a feasible, safe, and effective strategy for delivering optimal dose distributions for esophageal cancer patients. Although dosimetric results for ncVMAT plans were investigated in that study, it was a simple feasibility study to confirm the clinical outcome of ncVMAT plans for esophageal cancer, and dosimetric comparisons to other plans were not performed. Therefore, in this study, we compared ncVMAT plans to cVMAT plans by evaluating the dosimetric quality for esophageal cancer with a long, tube-like PTV.
Twenty patients with esophageal cancer were retrospectively selected. All patients had been previously treated with radiotherapy at our institution. An approval for this study was obtained from the Institutional Review Board of the Veterans Health Service Medical Center (IRB No. 2020-07-011). For selected patients, the mean PTV was 388.3±163.8 cm3. The mean PTV craniocaudal extent was 19.4±3.2 cm, traversing critical normal organs such as lungs, heart, liver, spinal cord, and stomach. All patients underwent computed tomography (CT) scans using Brilliance CT Big BoreTM (Philips, Cleveland, OH, USA) with a slice thickness of 3.0 mm.
The prescription dose to the PTV was 60 Gy in 30 fractions for all patients. All VMAT plans were generated with 6-MV photon beams from a Clinac iX with a Millennium 120 multileaf collimator (Varian Medical Systems, Palo Alto, CA, USA). The coplanar VMAT (cVMAT) plans consisted of three coplanar full arc beams (one arc was rotated from 179° to 181°, whereas the other two arcs were rotated from 181° to 179°). For the cVMAT plans, the collimator angles were set to 3° or 357°. The ncVMAT plans consisted of two coplanar full arc beams (one arc was rotated from 179° to 181°, and the other was rotated from 181° to 179°), and one noncoplanar partial arc beam ranging from 45° to 315° with a couch rotation angle of 315°. Based on the positional relationship between the PTV and OARs, the original couch angle (315°) was adjusted within ±5°. For the ncVMAT plans, the collimator angles were set to 3° or 357° for the two coplanar full arcs and to 45°±5° for the single noncoplanar partial arc. The arc beam geometry for the cVMAT and ncVMAT plans is shown in Fig. 2. Both the cVMAT and ncVMAT plans were optimized with the progressive resolution optimizer 2 (PRO2, ver.8.9; Varian Medical Systems, Palo Alto, CA, USA). The dose distributions were calculated using the anisotropic analytic algorithm (AAA, ver. 8.9; Varian Medical Systems, Palo Alto, CA, USA) with a calculation grid of 1 mm. All plans were normalized so that 100% of the prescription dose covered 95% of the PTV.
The dose-volumetric (DV) parameters were used to assess the dosimetric quality with respect to the PTV coverage and volume of OARs irradiated. For the PTVs, the evaluated DV parameters were the dose received by at least 98% of the PTV D98%, D95%, D5%, D2%, the absolute volume of the PTV irradiated by at least 105% of the prescription dose V105%, the minimum dose, the maximum dose, the mean dose, the conformity index (
where the
For both (left and right) lungs, the absolute volume of the normal organ irradiated by up to 30 Gy, V30 Gy, and the V20 Gy, the V13 Gy, the V5 Gy, the V2.5 Gy, and the mean dose were all calculated. The DV parameters were also evaluated for whole lungs. For the heart, the dose received by at least 0.1 cm3 volume of the normal organ, D0.1 cm3, V40 Gy, V20 Gy, V10 Gy, the minimum dose, the maximum dose, and the mean dose were calculated. For the spinal cord, the D0.1 cm3 and the maximum dose were calculated. For the stomach, V27.2 Gy, V17.6 Gy, the minimum dose, the maximum dose, and the mean dose were calculated. For the liver, V32 Gy, V28 Gy, V15 Gy, the minimum dose, the maximum dose, and the mean dose were calculated.
For the whole body of each patient, V100%, V90%, V70%, V50%, V30%, V10%, the mean dose, and the gradient index (
The mean total monitor units (MUs) were compared for both the cVMAT and ncVMAT plans.
All statistical analyses were performed with paired t-tests to assess whether the observed differences between the cVMAT and ncVMAT plans were statistically significant. A
The mean total MUs of the cVMAT and the ncVMAT plans were 471.0±44.2 and 469.3±47.6, respectively (
Table 1 summarizes the DV parameters of the PTV for both the cVMAT and ncVMAT plans. The values of D98% for the two plans were almost identical with no statistical significance. The values of D95% were also identical for both plans. Except for these two DV parameters (D98% and D95%), all other DV parameters for the cVMAT plans were slightly lower than those for the ncVMAT plans. The differences in values of D5%, D2%, V105%, and the mean dose were statistically significant. The PTV conformity and homogeneity for the cVMAT plans were slightly better (lower) than those for the ncVMAT plans; however, the differences were not statistically significant (1.026 vs. 1.028 for CI with
Table 1 Average dose-volumetric (DV) parameters of the planning target volume
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
D98% (Gy) | 58.6±0.3 | 58.6±0.4 | 0.070 |
D95% (Gy) | 60.0±0.0 | 60.0±0.0 | - |
D5% (Gy) | 64.4±1.0 | 64.6±1.0 | 0.002 |
D2% (Gy) | 64.6±1.1 | 64.9±1.0 | 0.002 |
V105% (cm3) | 175.9±118.5 | 208.5±115.0 | 0.029 |
Minimum dose (Gy) | 50.7±3.2 | 50.8±2.5 | 0.737 |
Maximum dose (Gy) | 66.4±1.9 | 66.7±1.6 | 0.403 |
Mean dose (Gy) | 62.8±0.7 | 63.0±0.7 | 0.005 |
CI | 1.026±0.045 | 1.028±0.042 | 0.330 |
HI | 0.069±0.016 | 0.071±0.014 | 0.102 |
Values are presented as mean±standard deviation.
VMAT, volumetric modulated arc therapy; Dn%, dose received by at least n% volume of the planning target volume; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; CI, conformity index; HI, homogeneity index.
For dosimetric evaluation, the dose distributions of the cVMAT and ncVMAT plans from representative patient cases (patients 3 and 9) are shown in Fig. 3. The dose-volume histograms for these two patient cases are shown in Fig. 4.
The DV parameters of the OARs for both the cVMAT and the ncVMAT plans are shown in Table 2. The values of V13 Gy and the mean dose values of the left and right lungs for the ncVMAT plans were smaller than those for the cVMAT plans, showing statistically significant differences. The value of V13 Gy of the right lung for the ncVMAT plans was much lower than that for the cVMAT plans (957.7 cm3 for cVMAT plans vs. 900.6 cm3 for ncVMAT plans with
Table 2 Average dose-volumetric (DV) parameters of normal organs
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
Left lung | |||
V30 Gy (cm3) | 111.3±63.1 | 112.9±72.2 | 0.688 |
V20 Gy (cm3) | 414.6±205.3 | 396.2±202.1 | 0.159 |
V13 Gy (cm3) | 790.6±293.2 | 736.6±291.4 | 0.001 |
V5 Gy (cm3) | 1,080.0±390.8 | 1,082.6±385.5 | 0.501 |
V2.5 Gy (cm3) | 1,160.6±409.8 | 1,167.0±405.2 | 0.113 |
Mean dose (Gy) | 16.2±4.1 | 15.8±4.2 | 0.017 |
Right lung | |||
V30 Gy (cm3) | 136.0±86.2 | 133.1±86.2 | 0.353 |
V20 Gy (cm3) | 434.4±237.5 | 417.2±219.5 | 0.189 |
V13 Gy (cm3) | 957.7±364.9 | 900.6±361.8 | 0.017 |
V5 Gy (cm3) | 1,396.5±416.6 | 1,406.5±425.8 | 0.095 |
V2.5 Gy (cm3) | 1,505.2±409.8 | 1,518.4±416.5 | 0.088 |
Mean dose (Gy) | 15.5±3.6 | 15.2±3.5 | 0.026 |
Both lungs | |||
V30 Gy (cm3) | 246.7±134.1 | 245.3±139.7 | 0.785 |
V20 Gy (cm3) | 848.2±424.7 | 812.7±391.5 | 0.113 |
V13 Gy (cm3) | 1,747.4±636.0 | 1,636.4±631.6 | 0.002 |
V5 Gy (cm3) | 2,475.5±752.9 | 2,488.2±754.7 | 0.008 |
V2.5 Gy (cm3) | 2,664.9±754.5 | 2,684.4±752.6 | 0.008 |
Mean dose (Gy) | 15.7±3.6 | 15.3±3.6 | 0.006 |
Heart | |||
D0.1 cm3 (Gy) | 50.3±4.8 | 50.4±5.0 | 0.633 |
V40 Gy (cm3) | 50.6±38.1 | 51.5±40.1 | 0.602 |
V20 Gy (cm3) | 283.1±171.9 | 271.5±160.1 | 0.257 |
V10 Gy (cm3) | 519.2±252.2 | 501.4±253.6 | 0.013 |
Minimum dose (Gy) | 3.2±3.7 | 2.8±3.2 | 0.275 |
Maximum dose (Gy) | 53.8±2.9 | 50.9±3.3 | 0.004 |
Mean dose (Gy) | 18.0±6.6 | 17.8±6.6 | 0.473 |
Spinal cord | |||
D0.1 cm3 (Gy) | 35.9±4.9 | 33.6±5.4 | <0.001 |
Maximum dose (Gy) | 37.1±5.1 | 34.7±5.7 | <0.001 |
Stomach | |||
V27.2 Gy (cm3) | 67.7±28.5 | 66.8±34.1 | 0.852 |
V17.6 Gy (cm3) | 96.4±39.8 | 90.3±44.6 | 0.184 |
Minimum dose (Gy) | 8.1±10.5 | 8.0±9.9 | 0.843 |
Maximum dose (Gy) | 61.3±5.6 | 61.4±5.5 | 0.699 |
Mean dose (Gy) | 27.0±15.0 | 26.8±15.3 | 0.687 |
Liver | |||
V32 Gy (cm3) | 68.7±52.7 | 57.8±53.6 | 0.027 |
V28 Gy (cm3) | 100.6±73.1 | 100.8±71.5 | 0.080 |
V15 Gy (cm3) | 399.6±267.6 | 348.7±245.5 | 0.015 |
Minimum dose (Gy) | 0.4±0.3 | 0.4±0.3 | 0.694 |
Maximum dose (Gy) | 57.3±17.2 | 58.2±15.5 | 0.193 |
Mean dose (Gy) | 12.5±6.6 | 12.2±6.2 | 0.226 |
VMAT, volumetric modulated arc therapy; Vn Gy, absolute volume of a structure irradiated by at least n Gy; Dn cm3, dose received by at least n cm3 volume of a structure.
For the heart, the values of V10 Gy and maximum dose values for the ncVMAT plans were lower than those for the cVMAT plans with the differences being statistically significant (
The DV parameters of the whole body for both the cVMAT and ncVMAT plans are shown in Table 3. The value of V100% for the ncVMAT plans was slightly higher than that for the cVMAT plans (394.4 cm3 for cVMAT plans vs. 396.1 cm3 for ncVMAT plans with
Table 3 Average dose-volumetric (DV) parameters of the body
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
V100% (cm3) | 394.4±161.1 | 396.1±163.2 | 0.015 |
V90% (cm3) | 507.2±26.2 | 509.8±211.0 | 0.137 |
V70% (cm3) | 765.5±309.6 | 780.6±327.0 | 0.118 |
V50% (cm3) | 1,431.0±578.1 | 1,458.7±609.9 | 0.422 |
V30% (cm3) | 3,728.7±1,540.8 | 3,640.0±1,530.4 | 0.001 |
V10% (cm3) | 9,803.8±3,382.6 | 9,754.9±3,431.2 | 0.320 |
Mean dose (Gy) | 8.8±2.3 | 8.7±2.3 | 0.004 |
GI | 3.7±0.3 | 3.7±0.2 | 0.257 |
VMAT, volumetric modulated arc therapy; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; GI, gradient index.
In this study, we compared the potential dosimetric improvements of the ncVMAT plans with those of the cVMAT plans, for the long, tube-like PTV of esophageal cancer. The cVMAT plans with three coplanar full arc beams and the ncVMAT plans with two coplanar full arc beams and one noncoplanar partial beam were used for planning. For the noncoplanar beam, the couch rotation angles were adjusted to 315°±5° to avoid collision between the patient on the couch and the gantry. For dosimetric evaluation, the DV parameters of the PTV and OARs and total MUs for each VMAT plan were obtained.
To date, this is the first study that attempted to evaluate the dosimetric superiority of ncVMAT plans in patients with long, tube-like esophageal targets for a large group of patients (n=20). In a similar study, Li et al. [16] investigated the dose distributions around esophageal cancer and the surrounding OARs using coplanar IMRT and noncoplanar IMRT. They reported that the conformity and mean dose of the PTV were comparable between the coplanar IMRT and noncoplanar IMRT plans while the values of V5%, V10%, V20%, and the mean dose for lungs using the noncoplanar IMRT plans decreased by 16.3, 10.6, 5.8, and 248.7 cm3, respectively, compared to those for the coplanar IMRT plans. The results of the present study are consistent with those found in earlier studies. The ncVMAT plans reduce the doses to OARs such as the lungs, heart, and spinal cord, as shown in this study. We found that the maximum doses to both the spinal cord and heart were dramatically reduced in the ncVMAT plans. For representative patient cases (patients 4 and 7) in this study, when using the ncVMAT plans, the maximum dose differences of 6.4 and 3.0 Gy were observed in volumes in the spinal cord and heart, respectively. We also found that low doses to the lungs and liver were significantly reduced for the ncVMAT plans. However, the values of V105% of the PTV and V100% of the whole body were increased for the ncVMAT plans. Nevertheless, these findings show some potential benefit of using the ncVMAT plans.
Except for the spinal cord and heart, there were minor improvements in dose distribution from the ncVMAT plans over the cVMAT plans in this study. This is mainly because the noncoplanar arc beam used in this study was limited. When using large noncoplanar angles, the probability of collision between the patient and the gantry increased. Theoretically, the 4π technique could provide significant OAR sparing compared to both the cVMAT and ncVMAT plans and then allow substantial dose escalation to the PTV [22,23]. Woods et al. [22] demonstrated that the 4π technique reduced 50% of the prescription doses by ~23%, compared to both cVMAT and ncVMAT plans. However, this technique significantly increases both the treatment time and the probability of collision between the patient and the gantry. In the clinic, we need to determine the noncoplanar arc beam configuration for effective treatment that represents a compromise to account for the combination of the gantry rotation angle and the couch angle.
Noncoplanar techniques have been widely examined for spherical targets for brain tumor radiotherapy, lung stereotactic ablative radiotherapy (SABR), and liver SABR patients, showing dosimetric benefits to either the PTV or the OARs [9,22,24
-27]. However, there are few dosimetric studies for large, irregularly shaped PTVs such as head, neck, and esophageal cancer. Gayen et al. [28] compared the dosimetric plan quality of the ncVMAT and cVMAT plans for head and neck malignancy and then demonstrated that ncVMAT plans provide better PTV coverage and OAR sparing, compared to cVMAT plans. Nevertheless, treatment of large, irregularly shaped head, neck, and esophageal cancer with noncoplanar beams is challenging due to the complicated positional relationship between the PTVs and the surrounding OARs. Further studies to prove the superiority of ncVMAT plans over cVMAT plans according to the size and shape of the PTV will be performed in the future.
The use of ncVMAT plans with two coplanar full arc beams and one noncoplanar partial beam provides better PTV coverage and OARs sparing, compared to that of cVMAT plans with three full arc beams for long, tube-like esophageal cancer. Furthermore, this technique can be adopted in the clinic for large, irregularly shaped targets with caution.
The study was approved by the Institutional Review Board of the Veterans Health Service Medical Center (IRB No. 2020-07-011).
This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Republic of Korea (grant number: VHSMC 20056).
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
Progress in Medical Physics 2020; 31(4): 179-188
Published online December 31, 2020 https://doi.org/10.14316/pmp.2020.31.4.179
Copyright © Korean Society of Medical Physics.
Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Korea
Correspondence to:So-Yeon Park
(vsoyounv@gmail.com)
Tel: 82-2-2225-4648
Fax: 82-2-2225-4640
Purpose: We compared noncoplanar volumetric modulated arc therapy (ncVMAT) plans to coplanar VMAT (cVMAT) plans by evaluating the dosimetric quality of each for esophageal cancer.
Methods: Twenty patients treated for esophageal cancer with the cVMAT technique were retrospectively selected. The cVMAT plans consisted of three coplanar full arc beams. The ncVMAT plans consisted of two coplanar full arc beams and one noncoplanar partial arc beam ranging from 45° to 315° with a couch rotation angle of 315°±5°. For dosimetric evaluation, the dose-volumetric (DV) parameters of the planning target volume (PTV) and organs at risk (OARs) were calculated for all VMAT plans.
Results: No clinically noticeable differences between the cVMAT and ncVMAT plans were observed in the DV parameters of the PTV. For the lungs, the VM13 Gy and mean dose for ncVMAT plans were smaller than those for cVMAT plans, showing statistically significant differences. For the heart, the values of the maximum dose for cVMAT and ncVMAT plans were 53.8±2.9 and 50.9±3.3 Gy, respectively (P=0.004). For the spinal cord, the values of the maximum dose for cVMAT and ncVMAT plans were 37.1±5.1 and 34.7±5.7 Gy, respectively (P<0.001).
Conclusions: The use of ncVMAT plans provides better PTV coverage and sparing of OARs compared to that of cVMAT plans for long, tube-like esophageal cancer. For esophageal cancer, the ncVMAT plans showed a more favorable plan quality than the cVMAT plans.
Keywords: Dosimetric evaluation, Noncoplanar, Volumetric modulated arc therapy, Esophageal cancer
Esophageal cancer, which mainly occurs in patients aged from 50 to 80 years, is the 11th most common type of cancer in Korea, accounting for approximately 1.7% of all cancers. An estimated 1,600 new patients are diagnosed with esophageal cancer every year [1]. Treatment options for esophageal cancer are surgery, chemotherapy, or radiotherapy [2
-8]. Radiotherapy for inoperable or unresectable esophageal cancer plays a crucial role in achieving optimal tumor control along with reducing normal tissue complications. However, a planning target volume (PTV) is a long, tube-like structure that connects the throat and stomach, as shown in Fig. 1, which creates challenges when planning a treatment. As there are several organs at risk (OARs) in the vicinity of the esophageal cancer, which include the lungs, heart, liver, spinal cord, and stomach, incorrect treatment may cause serious side effects such as radiation pneumonitis and fibrosis in the lungs as well as pericardial effusion, pericarditis, and heart failure [9].
Several studies have compared various radiotherapy techniques for esophageal cancer, including 3-dimensional conformal radiation therapy (3D CRT), intensity modulated radiation therapy (IMRT), and volumetric modulated arc therapy (VMAT) [2,9
-12]. These studies have reported that VMAT plans provide better dose conformity to the PTV while reducing the doses affecting OARs, compared to 3D CRT and IMRT plans [2,9
-12]. Furthermore, hybrid IMRT plans that combine static and IMRT beams, and hybrid VMAT plans that combine static beams and VMAT arcs concurrently, have been proposed by several studies to reduce the dose volume applied to normal tissues [9,13
-15]. These techniques were found to be superior in dosimetric outcomes when compared to conventional IMRT and VMAT plans.
Li et al. [16] proposed noncoplanar IMRT plans to reduce doses to surrounding OARs while improving PTV coverage for esophageal cancer. They showed that compared to coplanar IMRT plans, noncoplanar IMRT plans reduce the low dose exposure to the lungs, while providing a similar PTV coverage. The disadvantage of noncoplanar IMRT plans is that higher doses are delivered to the heart and spinal cord, making it inadvisable to use noncoplanar IMRT plans for patients with potential cardiovascular and neurological problems. Martini et al. [17] studied noncoplanar VMAT (ncVMAT) plans for patients with esophageal cancer who underwent either definitive or preoperative radiotherapy. Martini et al. [17] evaluated the feasibility, dosimetric results, and clinical outcomes. They showed that, in patients treated with ncVMAT plans using two coplanar and one noncoplanar arc beams, the side effects observed included esophagitis, anorexia, fatigue, nausea, and vomiting. The dosimetric results showed adequate PTV coverage and sparing of normal tissue. They concluded that the use of ncVMAT plans was a feasible, safe, and effective strategy for delivering optimal dose distributions for esophageal cancer patients. Although dosimetric results for ncVMAT plans were investigated in that study, it was a simple feasibility study to confirm the clinical outcome of ncVMAT plans for esophageal cancer, and dosimetric comparisons to other plans were not performed. Therefore, in this study, we compared ncVMAT plans to cVMAT plans by evaluating the dosimetric quality for esophageal cancer with a long, tube-like PTV.
Twenty patients with esophageal cancer were retrospectively selected. All patients had been previously treated with radiotherapy at our institution. An approval for this study was obtained from the Institutional Review Board of the Veterans Health Service Medical Center (IRB No. 2020-07-011). For selected patients, the mean PTV was 388.3±163.8 cm3. The mean PTV craniocaudal extent was 19.4±3.2 cm, traversing critical normal organs such as lungs, heart, liver, spinal cord, and stomach. All patients underwent computed tomography (CT) scans using Brilliance CT Big BoreTM (Philips, Cleveland, OH, USA) with a slice thickness of 3.0 mm.
The prescription dose to the PTV was 60 Gy in 30 fractions for all patients. All VMAT plans were generated with 6-MV photon beams from a Clinac iX with a Millennium 120 multileaf collimator (Varian Medical Systems, Palo Alto, CA, USA). The coplanar VMAT (cVMAT) plans consisted of three coplanar full arc beams (one arc was rotated from 179° to 181°, whereas the other two arcs were rotated from 181° to 179°). For the cVMAT plans, the collimator angles were set to 3° or 357°. The ncVMAT plans consisted of two coplanar full arc beams (one arc was rotated from 179° to 181°, and the other was rotated from 181° to 179°), and one noncoplanar partial arc beam ranging from 45° to 315° with a couch rotation angle of 315°. Based on the positional relationship between the PTV and OARs, the original couch angle (315°) was adjusted within ±5°. For the ncVMAT plans, the collimator angles were set to 3° or 357° for the two coplanar full arcs and to 45°±5° for the single noncoplanar partial arc. The arc beam geometry for the cVMAT and ncVMAT plans is shown in Fig. 2. Both the cVMAT and ncVMAT plans were optimized with the progressive resolution optimizer 2 (PRO2, ver.8.9; Varian Medical Systems, Palo Alto, CA, USA). The dose distributions were calculated using the anisotropic analytic algorithm (AAA, ver. 8.9; Varian Medical Systems, Palo Alto, CA, USA) with a calculation grid of 1 mm. All plans were normalized so that 100% of the prescription dose covered 95% of the PTV.
The dose-volumetric (DV) parameters were used to assess the dosimetric quality with respect to the PTV coverage and volume of OARs irradiated. For the PTVs, the evaluated DV parameters were the dose received by at least 98% of the PTV D98%, D95%, D5%, D2%, the absolute volume of the PTV irradiated by at least 105% of the prescription dose V105%, the minimum dose, the maximum dose, the mean dose, the conformity index (
where the
For both (left and right) lungs, the absolute volume of the normal organ irradiated by up to 30 Gy, V30 Gy, and the V20 Gy, the V13 Gy, the V5 Gy, the V2.5 Gy, and the mean dose were all calculated. The DV parameters were also evaluated for whole lungs. For the heart, the dose received by at least 0.1 cm3 volume of the normal organ, D0.1 cm3, V40 Gy, V20 Gy, V10 Gy, the minimum dose, the maximum dose, and the mean dose were calculated. For the spinal cord, the D0.1 cm3 and the maximum dose were calculated. For the stomach, V27.2 Gy, V17.6 Gy, the minimum dose, the maximum dose, and the mean dose were calculated. For the liver, V32 Gy, V28 Gy, V15 Gy, the minimum dose, the maximum dose, and the mean dose were calculated.
For the whole body of each patient, V100%, V90%, V70%, V50%, V30%, V10%, the mean dose, and the gradient index (
The mean total monitor units (MUs) were compared for both the cVMAT and ncVMAT plans.
All statistical analyses were performed with paired t-tests to assess whether the observed differences between the cVMAT and ncVMAT plans were statistically significant. A
The mean total MUs of the cVMAT and the ncVMAT plans were 471.0±44.2 and 469.3±47.6, respectively (
Table 1 summarizes the DV parameters of the PTV for both the cVMAT and ncVMAT plans. The values of D98% for the two plans were almost identical with no statistical significance. The values of D95% were also identical for both plans. Except for these two DV parameters (D98% and D95%), all other DV parameters for the cVMAT plans were slightly lower than those for the ncVMAT plans. The differences in values of D5%, D2%, V105%, and the mean dose were statistically significant. The PTV conformity and homogeneity for the cVMAT plans were slightly better (lower) than those for the ncVMAT plans; however, the differences were not statistically significant (1.026 vs. 1.028 for CI with
Table 1 . Average dose-volumetric (DV) parameters of the planning target volume.
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
D98% (Gy) | 58.6±0.3 | 58.6±0.4 | 0.070 |
D95% (Gy) | 60.0±0.0 | 60.0±0.0 | - |
D5% (Gy) | 64.4±1.0 | 64.6±1.0 | 0.002 |
D2% (Gy) | 64.6±1.1 | 64.9±1.0 | 0.002 |
V105% (cm3) | 175.9±118.5 | 208.5±115.0 | 0.029 |
Minimum dose (Gy) | 50.7±3.2 | 50.8±2.5 | 0.737 |
Maximum dose (Gy) | 66.4±1.9 | 66.7±1.6 | 0.403 |
Mean dose (Gy) | 62.8±0.7 | 63.0±0.7 | 0.005 |
CI | 1.026±0.045 | 1.028±0.042 | 0.330 |
HI | 0.069±0.016 | 0.071±0.014 | 0.102 |
Values are presented as mean±standard deviation..
VMAT, volumetric modulated arc therapy; Dn%, dose received by at least n% volume of the planning target volume; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; CI, conformity index; HI, homogeneity index..
For dosimetric evaluation, the dose distributions of the cVMAT and ncVMAT plans from representative patient cases (patients 3 and 9) are shown in Fig. 3. The dose-volume histograms for these two patient cases are shown in Fig. 4.
The DV parameters of the OARs for both the cVMAT and the ncVMAT plans are shown in Table 2. The values of V13 Gy and the mean dose values of the left and right lungs for the ncVMAT plans were smaller than those for the cVMAT plans, showing statistically significant differences. The value of V13 Gy of the right lung for the ncVMAT plans was much lower than that for the cVMAT plans (957.7 cm3 for cVMAT plans vs. 900.6 cm3 for ncVMAT plans with
Table 2 . Average dose-volumetric (DV) parameters of normal organs.
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
Left lung | |||
V30 Gy (cm3) | 111.3±63.1 | 112.9±72.2 | 0.688 |
V20 Gy (cm3) | 414.6±205.3 | 396.2±202.1 | 0.159 |
V13 Gy (cm3) | 790.6±293.2 | 736.6±291.4 | 0.001 |
V5 Gy (cm3) | 1,080.0±390.8 | 1,082.6±385.5 | 0.501 |
V2.5 Gy (cm3) | 1,160.6±409.8 | 1,167.0±405.2 | 0.113 |
Mean dose (Gy) | 16.2±4.1 | 15.8±4.2 | 0.017 |
Right lung | |||
V30 Gy (cm3) | 136.0±86.2 | 133.1±86.2 | 0.353 |
V20 Gy (cm3) | 434.4±237.5 | 417.2±219.5 | 0.189 |
V13 Gy (cm3) | 957.7±364.9 | 900.6±361.8 | 0.017 |
V5 Gy (cm3) | 1,396.5±416.6 | 1,406.5±425.8 | 0.095 |
V2.5 Gy (cm3) | 1,505.2±409.8 | 1,518.4±416.5 | 0.088 |
Mean dose (Gy) | 15.5±3.6 | 15.2±3.5 | 0.026 |
Both lungs | |||
V30 Gy (cm3) | 246.7±134.1 | 245.3±139.7 | 0.785 |
V20 Gy (cm3) | 848.2±424.7 | 812.7±391.5 | 0.113 |
V13 Gy (cm3) | 1,747.4±636.0 | 1,636.4±631.6 | 0.002 |
V5 Gy (cm3) | 2,475.5±752.9 | 2,488.2±754.7 | 0.008 |
V2.5 Gy (cm3) | 2,664.9±754.5 | 2,684.4±752.6 | 0.008 |
Mean dose (Gy) | 15.7±3.6 | 15.3±3.6 | 0.006 |
Heart | |||
D0.1 cm3 (Gy) | 50.3±4.8 | 50.4±5.0 | 0.633 |
V40 Gy (cm3) | 50.6±38.1 | 51.5±40.1 | 0.602 |
V20 Gy (cm3) | 283.1±171.9 | 271.5±160.1 | 0.257 |
V10 Gy (cm3) | 519.2±252.2 | 501.4±253.6 | 0.013 |
Minimum dose (Gy) | 3.2±3.7 | 2.8±3.2 | 0.275 |
Maximum dose (Gy) | 53.8±2.9 | 50.9±3.3 | 0.004 |
Mean dose (Gy) | 18.0±6.6 | 17.8±6.6 | 0.473 |
Spinal cord | |||
D0.1 cm3 (Gy) | 35.9±4.9 | 33.6±5.4 | <0.001 |
Maximum dose (Gy) | 37.1±5.1 | 34.7±5.7 | <0.001 |
Stomach | |||
V27.2 Gy (cm3) | 67.7±28.5 | 66.8±34.1 | 0.852 |
V17.6 Gy (cm3) | 96.4±39.8 | 90.3±44.6 | 0.184 |
Minimum dose (Gy) | 8.1±10.5 | 8.0±9.9 | 0.843 |
Maximum dose (Gy) | 61.3±5.6 | 61.4±5.5 | 0.699 |
Mean dose (Gy) | 27.0±15.0 | 26.8±15.3 | 0.687 |
Liver | |||
V32 Gy (cm3) | 68.7±52.7 | 57.8±53.6 | 0.027 |
V28 Gy (cm3) | 100.6±73.1 | 100.8±71.5 | 0.080 |
V15 Gy (cm3) | 399.6±267.6 | 348.7±245.5 | 0.015 |
Minimum dose (Gy) | 0.4±0.3 | 0.4±0.3 | 0.694 |
Maximum dose (Gy) | 57.3±17.2 | 58.2±15.5 | 0.193 |
Mean dose (Gy) | 12.5±6.6 | 12.2±6.2 | 0.226 |
VMAT, volumetric modulated arc therapy; Vn Gy, absolute volume of a structure irradiated by at least n Gy; Dn cm3, dose received by at least n cm3 volume of a structure..
For the heart, the values of V10 Gy and maximum dose values for the ncVMAT plans were lower than those for the cVMAT plans with the differences being statistically significant (
The DV parameters of the whole body for both the cVMAT and ncVMAT plans are shown in Table 3. The value of V100% for the ncVMAT plans was slightly higher than that for the cVMAT plans (394.4 cm3 for cVMAT plans vs. 396.1 cm3 for ncVMAT plans with
Table 3 . Average dose-volumetric (DV) parameters of the body.
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
V100% (cm3) | 394.4±161.1 | 396.1±163.2 | 0.015 |
V90% (cm3) | 507.2±26.2 | 509.8±211.0 | 0.137 |
V70% (cm3) | 765.5±309.6 | 780.6±327.0 | 0.118 |
V50% (cm3) | 1,431.0±578.1 | 1,458.7±609.9 | 0.422 |
V30% (cm3) | 3,728.7±1,540.8 | 3,640.0±1,530.4 | 0.001 |
V10% (cm3) | 9,803.8±3,382.6 | 9,754.9±3,431.2 | 0.320 |
Mean dose (Gy) | 8.8±2.3 | 8.7±2.3 | 0.004 |
GI | 3.7±0.3 | 3.7±0.2 | 0.257 |
VMAT, volumetric modulated arc therapy; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; GI, gradient index..
In this study, we compared the potential dosimetric improvements of the ncVMAT plans with those of the cVMAT plans, for the long, tube-like PTV of esophageal cancer. The cVMAT plans with three coplanar full arc beams and the ncVMAT plans with two coplanar full arc beams and one noncoplanar partial beam were used for planning. For the noncoplanar beam, the couch rotation angles were adjusted to 315°±5° to avoid collision between the patient on the couch and the gantry. For dosimetric evaluation, the DV parameters of the PTV and OARs and total MUs for each VMAT plan were obtained.
To date, this is the first study that attempted to evaluate the dosimetric superiority of ncVMAT plans in patients with long, tube-like esophageal targets for a large group of patients (n=20). In a similar study, Li et al. [16] investigated the dose distributions around esophageal cancer and the surrounding OARs using coplanar IMRT and noncoplanar IMRT. They reported that the conformity and mean dose of the PTV were comparable between the coplanar IMRT and noncoplanar IMRT plans while the values of V5%, V10%, V20%, and the mean dose for lungs using the noncoplanar IMRT plans decreased by 16.3, 10.6, 5.8, and 248.7 cm3, respectively, compared to those for the coplanar IMRT plans. The results of the present study are consistent with those found in earlier studies. The ncVMAT plans reduce the doses to OARs such as the lungs, heart, and spinal cord, as shown in this study. We found that the maximum doses to both the spinal cord and heart were dramatically reduced in the ncVMAT plans. For representative patient cases (patients 4 and 7) in this study, when using the ncVMAT plans, the maximum dose differences of 6.4 and 3.0 Gy were observed in volumes in the spinal cord and heart, respectively. We also found that low doses to the lungs and liver were significantly reduced for the ncVMAT plans. However, the values of V105% of the PTV and V100% of the whole body were increased for the ncVMAT plans. Nevertheless, these findings show some potential benefit of using the ncVMAT plans.
Except for the spinal cord and heart, there were minor improvements in dose distribution from the ncVMAT plans over the cVMAT plans in this study. This is mainly because the noncoplanar arc beam used in this study was limited. When using large noncoplanar angles, the probability of collision between the patient and the gantry increased. Theoretically, the 4π technique could provide significant OAR sparing compared to both the cVMAT and ncVMAT plans and then allow substantial dose escalation to the PTV [22,23]. Woods et al. [22] demonstrated that the 4π technique reduced 50% of the prescription doses by ~23%, compared to both cVMAT and ncVMAT plans. However, this technique significantly increases both the treatment time and the probability of collision between the patient and the gantry. In the clinic, we need to determine the noncoplanar arc beam configuration for effective treatment that represents a compromise to account for the combination of the gantry rotation angle and the couch angle.
Noncoplanar techniques have been widely examined for spherical targets for brain tumor radiotherapy, lung stereotactic ablative radiotherapy (SABR), and liver SABR patients, showing dosimetric benefits to either the PTV or the OARs [9,22,24
-27]. However, there are few dosimetric studies for large, irregularly shaped PTVs such as head, neck, and esophageal cancer. Gayen et al. [28] compared the dosimetric plan quality of the ncVMAT and cVMAT plans for head and neck malignancy and then demonstrated that ncVMAT plans provide better PTV coverage and OAR sparing, compared to cVMAT plans. Nevertheless, treatment of large, irregularly shaped head, neck, and esophageal cancer with noncoplanar beams is challenging due to the complicated positional relationship between the PTVs and the surrounding OARs. Further studies to prove the superiority of ncVMAT plans over cVMAT plans according to the size and shape of the PTV will be performed in the future.
The use of ncVMAT plans with two coplanar full arc beams and one noncoplanar partial beam provides better PTV coverage and OARs sparing, compared to that of cVMAT plans with three full arc beams for long, tube-like esophageal cancer. Furthermore, this technique can be adopted in the clinic for large, irregularly shaped targets with caution.
The study was approved by the Institutional Review Board of the Veterans Health Service Medical Center (IRB No. 2020-07-011).
This study was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (NRF-2017R1D1A1B03036093) and by a VHS Medical Center Research Grant, Republic of Korea (grant number: VHSMC 20056).
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
Table 1 Average dose-volumetric (DV) parameters of the planning target volume
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
D98% (Gy) | 58.6±0.3 | 58.6±0.4 | 0.070 |
D95% (Gy) | 60.0±0.0 | 60.0±0.0 | - |
D5% (Gy) | 64.4±1.0 | 64.6±1.0 | 0.002 |
D2% (Gy) | 64.6±1.1 | 64.9±1.0 | 0.002 |
V105% (cm3) | 175.9±118.5 | 208.5±115.0 | 0.029 |
Minimum dose (Gy) | 50.7±3.2 | 50.8±2.5 | 0.737 |
Maximum dose (Gy) | 66.4±1.9 | 66.7±1.6 | 0.403 |
Mean dose (Gy) | 62.8±0.7 | 63.0±0.7 | 0.005 |
CI | 1.026±0.045 | 1.028±0.042 | 0.330 |
HI | 0.069±0.016 | 0.071±0.014 | 0.102 |
Values are presented as mean±standard deviation.
VMAT, volumetric modulated arc therapy; Dn%, dose received by at least n% volume of the planning target volume; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; CI, conformity index; HI, homogeneity index.
Table 2 Average dose-volumetric (DV) parameters of normal organs
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
Left lung | |||
V30 Gy (cm3) | 111.3±63.1 | 112.9±72.2 | 0.688 |
V20 Gy (cm3) | 414.6±205.3 | 396.2±202.1 | 0.159 |
V13 Gy (cm3) | 790.6±293.2 | 736.6±291.4 | 0.001 |
V5 Gy (cm3) | 1,080.0±390.8 | 1,082.6±385.5 | 0.501 |
V2.5 Gy (cm3) | 1,160.6±409.8 | 1,167.0±405.2 | 0.113 |
Mean dose (Gy) | 16.2±4.1 | 15.8±4.2 | 0.017 |
Right lung | |||
V30 Gy (cm3) | 136.0±86.2 | 133.1±86.2 | 0.353 |
V20 Gy (cm3) | 434.4±237.5 | 417.2±219.5 | 0.189 |
V13 Gy (cm3) | 957.7±364.9 | 900.6±361.8 | 0.017 |
V5 Gy (cm3) | 1,396.5±416.6 | 1,406.5±425.8 | 0.095 |
V2.5 Gy (cm3) | 1,505.2±409.8 | 1,518.4±416.5 | 0.088 |
Mean dose (Gy) | 15.5±3.6 | 15.2±3.5 | 0.026 |
Both lungs | |||
V30 Gy (cm3) | 246.7±134.1 | 245.3±139.7 | 0.785 |
V20 Gy (cm3) | 848.2±424.7 | 812.7±391.5 | 0.113 |
V13 Gy (cm3) | 1,747.4±636.0 | 1,636.4±631.6 | 0.002 |
V5 Gy (cm3) | 2,475.5±752.9 | 2,488.2±754.7 | 0.008 |
V2.5 Gy (cm3) | 2,664.9±754.5 | 2,684.4±752.6 | 0.008 |
Mean dose (Gy) | 15.7±3.6 | 15.3±3.6 | 0.006 |
Heart | |||
D0.1 cm3 (Gy) | 50.3±4.8 | 50.4±5.0 | 0.633 |
V40 Gy (cm3) | 50.6±38.1 | 51.5±40.1 | 0.602 |
V20 Gy (cm3) | 283.1±171.9 | 271.5±160.1 | 0.257 |
V10 Gy (cm3) | 519.2±252.2 | 501.4±253.6 | 0.013 |
Minimum dose (Gy) | 3.2±3.7 | 2.8±3.2 | 0.275 |
Maximum dose (Gy) | 53.8±2.9 | 50.9±3.3 | 0.004 |
Mean dose (Gy) | 18.0±6.6 | 17.8±6.6 | 0.473 |
Spinal cord | |||
D0.1 cm3 (Gy) | 35.9±4.9 | 33.6±5.4 | <0.001 |
Maximum dose (Gy) | 37.1±5.1 | 34.7±5.7 | <0.001 |
Stomach | |||
V27.2 Gy (cm3) | 67.7±28.5 | 66.8±34.1 | 0.852 |
V17.6 Gy (cm3) | 96.4±39.8 | 90.3±44.6 | 0.184 |
Minimum dose (Gy) | 8.1±10.5 | 8.0±9.9 | 0.843 |
Maximum dose (Gy) | 61.3±5.6 | 61.4±5.5 | 0.699 |
Mean dose (Gy) | 27.0±15.0 | 26.8±15.3 | 0.687 |
Liver | |||
V32 Gy (cm3) | 68.7±52.7 | 57.8±53.6 | 0.027 |
V28 Gy (cm3) | 100.6±73.1 | 100.8±71.5 | 0.080 |
V15 Gy (cm3) | 399.6±267.6 | 348.7±245.5 | 0.015 |
Minimum dose (Gy) | 0.4±0.3 | 0.4±0.3 | 0.694 |
Maximum dose (Gy) | 57.3±17.2 | 58.2±15.5 | 0.193 |
Mean dose (Gy) | 12.5±6.6 | 12.2±6.2 | 0.226 |
VMAT, volumetric modulated arc therapy; Vn Gy, absolute volume of a structure irradiated by at least n Gy; Dn cm3, dose received by at least n cm3 volume of a structure.
Table 3 Average dose-volumetric (DV) parameters of the body
DV parameter | Coplanar VMAT | Noncoplanar VMAT | P-value |
---|---|---|---|
V100% (cm3) | 394.4±161.1 | 396.1±163.2 | 0.015 |
V90% (cm3) | 507.2±26.2 | 509.8±211.0 | 0.137 |
V70% (cm3) | 765.5±309.6 | 780.6±327.0 | 0.118 |
V50% (cm3) | 1,431.0±578.1 | 1,458.7±609.9 | 0.422 |
V30% (cm3) | 3,728.7±1,540.8 | 3,640.0±1,530.4 | 0.001 |
V10% (cm3) | 9,803.8±3,382.6 | 9,754.9±3,431.2 | 0.320 |
Mean dose (Gy) | 8.8±2.3 | 8.7±2.3 | 0.004 |
GI | 3.7±0.3 | 3.7±0.2 | 0.257 |
VMAT, volumetric modulated arc therapy; Vn%, absolute volume of a structure irradiated by at least n% of the prescription dose; GI, gradient index.
pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829
Frequency: Quarterly