Progress in Medical Physics 2021; 32(4): 122-129
Published online December 31, 2021
https://doi.org/10.14316/pmp.2021.32.4.122
Copyright © Korean Society of Medical Physics.
Correspondence to:Sangwook Lim
(medicalphysics@hotmail.com)
Tel: 82-51-990-6393
Purpose: This study aimed to design a multipurpose dose verification phantom for external audits to secure safe and optimal radiation therapy.
Methods: In this s tudy, we u sed I nternational Atomic Energy Agency (IAEA) LiF p owder thermoluminescence dosimeter (TLD), which is generally used in the therapeutic radiation dose assurance project. The newly designed multipurpose phantom (MPP) consists of a container filled with water, a TLD holder, and two water-pressing covers. The size of the phantom was designed to be sufficient (30×30×30 cm3). The water container was filled with water and pressed with the cover for normal incidence to be fixed. The surface of the MPP was devised to maintain the same distance from the source at all times, even in the case of oblique incidence regardless of the water level. The MPP was irradiated with 6, 10, and 15 MV photon beams from Varian Linear Accelerator and measured by a 1.25 cm3 ionization chamber to get the correction factors. Monte Carlo (MC) simulation was also used to compare the measurements.
Results: The result obtained by MC had a relatively high uncertainty of 1% at the dosimetry point, but it showed a correction factor value of 1.3% at the 5 cm point. The energy dependence was large at 6 MV and small at 15 MV. Various dosimetric parameters for external audits can be performed within an hour.
Conclusions: The results allow an objective comparison of the quality assurance (QA) of individual hospitals. Therefore, this can be employed for external audits or QA systems in radiation therapy institutions.
KeywordsExternal audit, Radiation therapy, Phantom, Photon, Thermoluminescence dosimeter
It is fundamental to ensure the accuracy of the output of the treatment device and the basic quality of the radiation used for treatment, and the treatment method implemented by each treatment institution must be considered. Therefore, it is necessary to prepare a program for securing clinical dose that reflects the high-quality treatment methods implemented in domestic medical institutions.
External audit on radiotherapy program is settled in many countries to increase radiotherapy quality. It can be performed by visiting [1-4] participant centers or by mailing [5-11]. Various international organizations, such as the International Atomic Energy Agency (IAEA), Radiological Physics Center, and ESTRO-QUALity assurance network (EQUAL), operate mailed dosimetry services [12-14].
The goal of these programs was to check only the output parameters of radiotherapy units. However, accuracy is needed in radiotherapy, and complexity is growing in the treatment process [15,16]. Therefore, the requirement for an effective quality assurance (QA) program has to be emphasized.
The concept of the multipurpose phantom (MPP) designed in this study is to measure various parameters at once for complex radiotherapy audit.
Referring to the recommendation of ESTRO, AAPM, and routine QA parameters of hospitals, the essential parameters were chosen for radiotherapy audit [15]. The parameters were for 4 to 20 MV photon beams. The MPP was constructed to check on- and off-axis in three irradiation conditions in a vertical and oblique beam setup, as shown in Table 1.
Table 1 Beam setup and parameters to be measured
Beam setup | Parameter |
---|---|
Square field - 10×10 cm2 of field size - Reference depth of 5 cm | Output Symmetry Flatness PDD Inhomogeneity |
Oblique field - 10×10 cm2 of field size - 30° rotation of gantry | Profile PDD |
Wedged field - 30° wedge - 10×10 cm2 of field size - 30° rotation of gantry | Wedge factor Profile PDD |
PDD, percent depth dose.
The phantom was developed under the following conditions. First, the phantom should be water equivalent materials. Second, it should be possible to measure multiple factors at once. Third, all measurements must be completed within an hour. Fourth, it should be simple to set up and easy to use. Fifth, it should be light in weight and convenient to deliver. Sixth, it should be possible to use the thermoluminescence dosimeter (TLD) previously used by IAEA for postal audit.
The TLDs used in the IAEA QA program were capsules filled with approximately 155 mg of LiF powder. The capsule was a black polyethylene cylinder of 20 mm inner length, 3 mm inner diameter, and 1 mm wall thickness (Fig. 1) [13]. The MPP designed in this study used 7 or 9 TLDs at a time.
The MPP was composed of (a) a water container, (b) a TLD holder, (c) a cover for normal incidence, and (d) a cover for oblique incidence, as shown in Fig. 2. The water container and support rod were made of acrylic, which has a density similar to that of water, and were designed to be filled with water in use to reduce weight during delivery. The AAPM TG-51 protocol recommended the use of water [17]. The size of the phantom was designed to be sufficient (30×30×30 cm3), considering the back and side scatters. The TLD holder was designed to be inserted into the water phantom after the TLDs were applied, making it easy to replace the TLD during measurement. In this study, the ionization chamber can be inserted at the same locations as the TLDs to evaluate the reliability of the MPP. The water container was filled with water and pressed with the cover for normal incidence to be fixed. The surface of the MPP was devised to maintain the same distance from the source at all times, even in the case of oblique incidence regardless of the water level (Fig. 3). Additionally, the covers prevented the water from sloshing due to the movement of the computed tomography (CT) table during the CT scan, and there was no sloshing artifact (Fig. 4). To minimize the attenuation of photon beams, the thickness of the acrylic window of the lid was designed to be the minimum thickness (2 mm) that can withstand the pressure of water.
The locations of the TLDs in the phantom are shown in Fig. 5, where one TLD is located at a depth of 5 cm from the surface, and 5 TLDs are located at a depth of 10 cm to determine the dose, flatness, and symmetry. At a depth of 20 cm, one TLD was used for depth dose, and two TLDs were for heterogeneity measurement. The dimensions of inhomogeneous materials were 3×3×3 cm3, Teflon (2.35 g/cm3), which has a density similar to that of human bone, and cork (0.32 g/cm3), which has a density similar to the lungs, were used, and the TLDs were designed to be inserted. The density of thermoluminescence was 2.65 g/cm3, equivalent to 8 mm depth of water.
When not in use, the two inhomogeneous materials can be pushed to either end. The total number of TLDs required for all measurements was 23 per energy.
The first method used to evaluate the perturbation at each point in an MPP was a measurement using an ionization chamber in the reference water phantom used to obtain beam data in a treatment institution.
The MPP was irradiated with 6, 10, and 15 MV photon beams from Varian Linear Accelerator and measured by a 1.25 cm3 ionization chamber (PTW) to evaluate the MPP. All energies were measured by an ionization chamber in the same geometric setup as the TLD to compare it with TLD measurements.
First, the doses were measured using the ionization chamber installed in a reference water phantom at the same location where the TLDs were installed in the MPP. Next, the dose was measured using the ionization chamber in 9 locations in the MPP.
To convert the doses measured in the MPP into the dose value measured in the water phantom, the following dose correction factors are defined as follows:
where
If the dose measured using the TLD in the MPP is
The NRC (National Research Council Canada) has calculated and announced photon spectra from 9 machines. The spectrum presented by the NRC for Varian 6, 10, and 15 MV and Siemens 6 and 10 MV was used in this study.
The MPP was modeled as shown in Fig. 6, and all points were modeled assuming water. MC simulation was performed in Monte Carlo N-Particle Transport Code System-extended (MCNPX, version 2.4.0), photons and electrons were traced to 1 keV and 100 keV, respectively, and F8 tally, a pulse height tally, was used to calculate energy absorption. In the case of the F8 tally, it can be used as a direct energy absorption because the energy transferred within the volume of the measurement point can be tracked by tracking the movement of particles.
Fig. 6 shows the doses measured in the water phantom and the MPP at 9 points, and Table 2 shows the correction factors obtained using the above equation. The geometrical positions of the ionization chamber in the MPP and the 3D water phantom are the same.
Table 2 Correction factors obtained from measurements of ionization chamber in MPP and water phantom for normal incidence
Depth (cm) | Location number | Correction factor (MPP/water phantom) | ||
---|---|---|---|---|
6 MV | 10 MV | 15 MV | ||
5 | 1 | 0.9888 | 0.9765 | 0.9873 |
10 | 2 | 0.9814 | 0.9715 | 0.9911 |
3 | 0.9925 | 0.9746 | 0.9960 | |
4 | 0.9896 | 0.9823 | 0.9853 | |
5 | 0.9883 | 0.9754 | 0.9828 | |
6 | 0.9767 | 0.9909 | 0.9799 | |
20 | 7 | 0.9822 | 0.9828 | 0.9914 |
8 | 0.9895 | 0.9815 | 0.9915 | |
9 | 0.9860 | 0.9854 | 0.9885 |
MPP, multipurpose phantom.
Fig. 7a shows the measured value in the case of normal incident, Fig. 7b shows the measured value in the case of oblique incident, and Fig. 7c shows the measured value in the case of oblique incident with a 30° wedge. In Fig. 7b, positions 2 to 6 are measured values at a depth of 10 cm.
The correction factors obtained from MC simulations are shown in Table 3 and plotted in Fig. 8.
Table 3 Correction factors calculated by MCNPX 2.4.0 at each point
Location number | Varian 6 MV | Siemens 6 MV | Siemens 10 MV | Varian 10 MV | Varian 15 MV |
---|---|---|---|---|---|
1 | 0.9875 | 0.9874 | 0.9897 | 0.9828 | 0.9904 |
2 | 0.9693 | 0.9702 | 0.9846 | 0.9790 | 0.9779 |
3 | 0.9793 | 0.9783 | 0.9867 | 0.9859 | 0.9809 |
4 | 0.9875 | 0.9912 | 0.9864 | 0.9807 | 0.9834 |
5 | 0.9817 | 0.9697 | 0.9768 | 0.9809 | 0.9711 |
6 | 0.9817 | 0.9683 | 0.9806 | 0.9730 | 0.9729 |
7 | 0.9653 | 0.9679 | 0.9740 | 0.9663 | 0.9683 |
8 | 0.9825 | 0.9796 | 0.9835 | 0.9863 | 0.9785 |
9 | 0.9790 | 0.9839 | 0.9882 | 0.9943 | 0.9878 |
The Type A standard uncertainties were taken from the deviation of TLD measurements based on ionization chamber measurements in MPP. The measurements were repeated 18 for normal and 14 for oblique and wedged incidences. The Type A standard uncertainties of the MPP for each energy and setup are shown in Table 4. The Type A standard uncertainties were calculated as follows:
Table 4 Type A standard uncertainties according to setup and energies
Energy | Normal (%) | Oblique (%) | Wedged (%) |
---|---|---|---|
6 MV | 0.54 | 0.66 | 0.43 |
10 MV | 0.48 | 0.69 | 0.43 |
15 MV | 0.54 | 0.66 | 0.43 |
where SD is the standard deviation of the sample and N is the number of samples.
Since the values of positions 2 to 6 in Fig. 7b were measured for oblique incidents, they differ depending on the depth reached. However, in Fig. 7c, since the wedge is compensated, it can be seen that the values are almost the same as in Fig. 7a.
The result obtained by MC had a relatively high uncertainty of 1% at the dosimetry point, and it showed a correction factor value of 1.3% at the 5 cm point. This result was similar to that of Mobit et al.’s experiment [18], which agrees within 1.5% between measurements and MC calculations.
The energy dependence was large at 6 MV and small at 15 MV. This is thought to be because the higher the energy of the photon beam, the stronger the penetrating power.
With the MPP developed in this study, the measurement of various dosimetric parameters for external audit can be performed within an hour.
The MPP is not limited to measuring dosimetric data in perpendicular incidence to the water surface but performs additional measurements at oblique incidence and close to clinical cases such as flatness, symmetry, oblique beam, and inhomogeneity without gantry rotations.
According to MC calculation and ionization chamber measurements, the energy correction factor should be considered.
The MPP has been demonstrated to be useful in accessing the overall accuracy of the treatment units by dosimetric checking of photon beams in reference and nonreference conditions and in checking the TPS by verifying the calculated data with the measured data. The MPP developed in this study is considered useful for external audits or QA systems of radiation therapy institutions.
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
Progress in Medical Physics 2021; 32(4): 122-129
Published online December 31, 2021 https://doi.org/10.14316/pmp.2021.32.4.122
Copyright © Korean Society of Medical Physics.
Department of Radiation Oncology, Kosin University College of Medicine, Busan, Korea
Correspondence to:Sangwook Lim
(medicalphysics@hotmail.com)
Tel: 82-51-990-6393
Purpose: This study aimed to design a multipurpose dose verification phantom for external audits to secure safe and optimal radiation therapy.
Methods: In this s tudy, we u sed I nternational Atomic Energy Agency (IAEA) LiF p owder thermoluminescence dosimeter (TLD), which is generally used in the therapeutic radiation dose assurance project. The newly designed multipurpose phantom (MPP) consists of a container filled with water, a TLD holder, and two water-pressing covers. The size of the phantom was designed to be sufficient (30×30×30 cm3). The water container was filled with water and pressed with the cover for normal incidence to be fixed. The surface of the MPP was devised to maintain the same distance from the source at all times, even in the case of oblique incidence regardless of the water level. The MPP was irradiated with 6, 10, and 15 MV photon beams from Varian Linear Accelerator and measured by a 1.25 cm3 ionization chamber to get the correction factors. Monte Carlo (MC) simulation was also used to compare the measurements.
Results: The result obtained by MC had a relatively high uncertainty of 1% at the dosimetry point, but it showed a correction factor value of 1.3% at the 5 cm point. The energy dependence was large at 6 MV and small at 15 MV. Various dosimetric parameters for external audits can be performed within an hour.
Conclusions: The results allow an objective comparison of the quality assurance (QA) of individual hospitals. Therefore, this can be employed for external audits or QA systems in radiation therapy institutions.
Keywords: External audit, Radiation therapy, Phantom, Photon, Thermoluminescence dosimeter
It is fundamental to ensure the accuracy of the output of the treatment device and the basic quality of the radiation used for treatment, and the treatment method implemented by each treatment institution must be considered. Therefore, it is necessary to prepare a program for securing clinical dose that reflects the high-quality treatment methods implemented in domestic medical institutions.
External audit on radiotherapy program is settled in many countries to increase radiotherapy quality. It can be performed by visiting [1-4] participant centers or by mailing [5-11]. Various international organizations, such as the International Atomic Energy Agency (IAEA), Radiological Physics Center, and ESTRO-QUALity assurance network (EQUAL), operate mailed dosimetry services [12-14].
The goal of these programs was to check only the output parameters of radiotherapy units. However, accuracy is needed in radiotherapy, and complexity is growing in the treatment process [15,16]. Therefore, the requirement for an effective quality assurance (QA) program has to be emphasized.
The concept of the multipurpose phantom (MPP) designed in this study is to measure various parameters at once for complex radiotherapy audit.
Referring to the recommendation of ESTRO, AAPM, and routine QA parameters of hospitals, the essential parameters were chosen for radiotherapy audit [15]. The parameters were for 4 to 20 MV photon beams. The MPP was constructed to check on- and off-axis in three irradiation conditions in a vertical and oblique beam setup, as shown in Table 1.
Table 1 . Beam setup and parameters to be measured.
Beam setup | Parameter |
---|---|
Square field - 10×10 cm2 of field size - Reference depth of 5 cm | Output Symmetry Flatness PDD Inhomogeneity |
Oblique field - 10×10 cm2 of field size - 30° rotation of gantry | Profile PDD |
Wedged field - 30° wedge - 10×10 cm2 of field size - 30° rotation of gantry | Wedge factor Profile PDD |
PDD, percent depth dose..
The phantom was developed under the following conditions. First, the phantom should be water equivalent materials. Second, it should be possible to measure multiple factors at once. Third, all measurements must be completed within an hour. Fourth, it should be simple to set up and easy to use. Fifth, it should be light in weight and convenient to deliver. Sixth, it should be possible to use the thermoluminescence dosimeter (TLD) previously used by IAEA for postal audit.
The TLDs used in the IAEA QA program were capsules filled with approximately 155 mg of LiF powder. The capsule was a black polyethylene cylinder of 20 mm inner length, 3 mm inner diameter, and 1 mm wall thickness (Fig. 1) [13]. The MPP designed in this study used 7 or 9 TLDs at a time.
The MPP was composed of (a) a water container, (b) a TLD holder, (c) a cover for normal incidence, and (d) a cover for oblique incidence, as shown in Fig. 2. The water container and support rod were made of acrylic, which has a density similar to that of water, and were designed to be filled with water in use to reduce weight during delivery. The AAPM TG-51 protocol recommended the use of water [17]. The size of the phantom was designed to be sufficient (30×30×30 cm3), considering the back and side scatters. The TLD holder was designed to be inserted into the water phantom after the TLDs were applied, making it easy to replace the TLD during measurement. In this study, the ionization chamber can be inserted at the same locations as the TLDs to evaluate the reliability of the MPP. The water container was filled with water and pressed with the cover for normal incidence to be fixed. The surface of the MPP was devised to maintain the same distance from the source at all times, even in the case of oblique incidence regardless of the water level (Fig. 3). Additionally, the covers prevented the water from sloshing due to the movement of the computed tomography (CT) table during the CT scan, and there was no sloshing artifact (Fig. 4). To minimize the attenuation of photon beams, the thickness of the acrylic window of the lid was designed to be the minimum thickness (2 mm) that can withstand the pressure of water.
The locations of the TLDs in the phantom are shown in Fig. 5, where one TLD is located at a depth of 5 cm from the surface, and 5 TLDs are located at a depth of 10 cm to determine the dose, flatness, and symmetry. At a depth of 20 cm, one TLD was used for depth dose, and two TLDs were for heterogeneity measurement. The dimensions of inhomogeneous materials were 3×3×3 cm3, Teflon (2.35 g/cm3), which has a density similar to that of human bone, and cork (0.32 g/cm3), which has a density similar to the lungs, were used, and the TLDs were designed to be inserted. The density of thermoluminescence was 2.65 g/cm3, equivalent to 8 mm depth of water.
When not in use, the two inhomogeneous materials can be pushed to either end. The total number of TLDs required for all measurements was 23 per energy.
The first method used to evaluate the perturbation at each point in an MPP was a measurement using an ionization chamber in the reference water phantom used to obtain beam data in a treatment institution.
The MPP was irradiated with 6, 10, and 15 MV photon beams from Varian Linear Accelerator and measured by a 1.25 cm3 ionization chamber (PTW) to evaluate the MPP. All energies were measured by an ionization chamber in the same geometric setup as the TLD to compare it with TLD measurements.
First, the doses were measured using the ionization chamber installed in a reference water phantom at the same location where the TLDs were installed in the MPP. Next, the dose was measured using the ionization chamber in 9 locations in the MPP.
To convert the doses measured in the MPP into the dose value measured in the water phantom, the following dose correction factors are defined as follows:
where
If the dose measured using the TLD in the MPP is
The NRC (National Research Council Canada) has calculated and announced photon spectra from 9 machines. The spectrum presented by the NRC for Varian 6, 10, and 15 MV and Siemens 6 and 10 MV was used in this study.
The MPP was modeled as shown in Fig. 6, and all points were modeled assuming water. MC simulation was performed in Monte Carlo N-Particle Transport Code System-extended (MCNPX, version 2.4.0), photons and electrons were traced to 1 keV and 100 keV, respectively, and F8 tally, a pulse height tally, was used to calculate energy absorption. In the case of the F8 tally, it can be used as a direct energy absorption because the energy transferred within the volume of the measurement point can be tracked by tracking the movement of particles.
Fig. 6 shows the doses measured in the water phantom and the MPP at 9 points, and Table 2 shows the correction factors obtained using the above equation. The geometrical positions of the ionization chamber in the MPP and the 3D water phantom are the same.
Table 2 . Correction factors obtained from measurements of ionization chamber in MPP and water phantom for normal incidence.
Depth (cm) | Location number | Correction factor (MPP/water phantom) | ||
---|---|---|---|---|
6 MV | 10 MV | 15 MV | ||
5 | 1 | 0.9888 | 0.9765 | 0.9873 |
10 | 2 | 0.9814 | 0.9715 | 0.9911 |
3 | 0.9925 | 0.9746 | 0.9960 | |
4 | 0.9896 | 0.9823 | 0.9853 | |
5 | 0.9883 | 0.9754 | 0.9828 | |
6 | 0.9767 | 0.9909 | 0.9799 | |
20 | 7 | 0.9822 | 0.9828 | 0.9914 |
8 | 0.9895 | 0.9815 | 0.9915 | |
9 | 0.9860 | 0.9854 | 0.9885 |
MPP, multipurpose phantom..
Fig. 7a shows the measured value in the case of normal incident, Fig. 7b shows the measured value in the case of oblique incident, and Fig. 7c shows the measured value in the case of oblique incident with a 30° wedge. In Fig. 7b, positions 2 to 6 are measured values at a depth of 10 cm.
The correction factors obtained from MC simulations are shown in Table 3 and plotted in Fig. 8.
Table 3 . Correction factors calculated by MCNPX 2.4.0 at each point.
Location number | Varian 6 MV | Siemens 6 MV | Siemens 10 MV | Varian 10 MV | Varian 15 MV |
---|---|---|---|---|---|
1 | 0.9875 | 0.9874 | 0.9897 | 0.9828 | 0.9904 |
2 | 0.9693 | 0.9702 | 0.9846 | 0.9790 | 0.9779 |
3 | 0.9793 | 0.9783 | 0.9867 | 0.9859 | 0.9809 |
4 | 0.9875 | 0.9912 | 0.9864 | 0.9807 | 0.9834 |
5 | 0.9817 | 0.9697 | 0.9768 | 0.9809 | 0.9711 |
6 | 0.9817 | 0.9683 | 0.9806 | 0.9730 | 0.9729 |
7 | 0.9653 | 0.9679 | 0.9740 | 0.9663 | 0.9683 |
8 | 0.9825 | 0.9796 | 0.9835 | 0.9863 | 0.9785 |
9 | 0.9790 | 0.9839 | 0.9882 | 0.9943 | 0.9878 |
The Type A standard uncertainties were taken from the deviation of TLD measurements based on ionization chamber measurements in MPP. The measurements were repeated 18 for normal and 14 for oblique and wedged incidences. The Type A standard uncertainties of the MPP for each energy and setup are shown in Table 4. The Type A standard uncertainties were calculated as follows:
Table 4 . Type A standard uncertainties according to setup and energies.
Energy | Normal (%) | Oblique (%) | Wedged (%) |
---|---|---|---|
6 MV | 0.54 | 0.66 | 0.43 |
10 MV | 0.48 | 0.69 | 0.43 |
15 MV | 0.54 | 0.66 | 0.43 |
where SD is the standard deviation of the sample and N is the number of samples.
Since the values of positions 2 to 6 in Fig. 7b were measured for oblique incidents, they differ depending on the depth reached. However, in Fig. 7c, since the wedge is compensated, it can be seen that the values are almost the same as in Fig. 7a.
The result obtained by MC had a relatively high uncertainty of 1% at the dosimetry point, and it showed a correction factor value of 1.3% at the 5 cm point. This result was similar to that of Mobit et al.’s experiment [18], which agrees within 1.5% between measurements and MC calculations.
The energy dependence was large at 6 MV and small at 15 MV. This is thought to be because the higher the energy of the photon beam, the stronger the penetrating power.
With the MPP developed in this study, the measurement of various dosimetric parameters for external audit can be performed within an hour.
The MPP is not limited to measuring dosimetric data in perpendicular incidence to the water surface but performs additional measurements at oblique incidence and close to clinical cases such as flatness, symmetry, oblique beam, and inhomogeneity without gantry rotations.
According to MC calculation and ionization chamber measurements, the energy correction factor should be considered.
The MPP has been demonstrated to be useful in accessing the overall accuracy of the treatment units by dosimetric checking of photon beams in reference and nonreference conditions and in checking the TPS by verifying the calculated data with the measured data. The MPP developed in this study is considered useful for external audits or QA systems of radiation therapy institutions.
The authors have nothing to disclose.
All relevant data are within the paper and its Supporting Information files.
Table 1 Beam setup and parameters to be measured
Beam setup | Parameter |
---|---|
Square field - 10×10 cm2 of field size - Reference depth of 5 cm | Output Symmetry Flatness PDD Inhomogeneity |
Oblique field - 10×10 cm2 of field size - 30° rotation of gantry | Profile PDD |
Wedged field - 30° wedge - 10×10 cm2 of field size - 30° rotation of gantry | Wedge factor Profile PDD |
PDD, percent depth dose.
Table 2 Correction factors obtained from measurements of ionization chamber in MPP and water phantom for normal incidence
Depth (cm) | Location number | Correction factor (MPP/water phantom) | ||
---|---|---|---|---|
6 MV | 10 MV | 15 MV | ||
5 | 1 | 0.9888 | 0.9765 | 0.9873 |
10 | 2 | 0.9814 | 0.9715 | 0.9911 |
3 | 0.9925 | 0.9746 | 0.9960 | |
4 | 0.9896 | 0.9823 | 0.9853 | |
5 | 0.9883 | 0.9754 | 0.9828 | |
6 | 0.9767 | 0.9909 | 0.9799 | |
20 | 7 | 0.9822 | 0.9828 | 0.9914 |
8 | 0.9895 | 0.9815 | 0.9915 | |
9 | 0.9860 | 0.9854 | 0.9885 |
MPP, multipurpose phantom.
Table 3 Correction factors calculated by MCNPX 2.4.0 at each point
Location number | Varian 6 MV | Siemens 6 MV | Siemens 10 MV | Varian 10 MV | Varian 15 MV |
---|---|---|---|---|---|
1 | 0.9875 | 0.9874 | 0.9897 | 0.9828 | 0.9904 |
2 | 0.9693 | 0.9702 | 0.9846 | 0.9790 | 0.9779 |
3 | 0.9793 | 0.9783 | 0.9867 | 0.9859 | 0.9809 |
4 | 0.9875 | 0.9912 | 0.9864 | 0.9807 | 0.9834 |
5 | 0.9817 | 0.9697 | 0.9768 | 0.9809 | 0.9711 |
6 | 0.9817 | 0.9683 | 0.9806 | 0.9730 | 0.9729 |
7 | 0.9653 | 0.9679 | 0.9740 | 0.9663 | 0.9683 |
8 | 0.9825 | 0.9796 | 0.9835 | 0.9863 | 0.9785 |
9 | 0.9790 | 0.9839 | 0.9882 | 0.9943 | 0.9878 |
Table 4 Type A standard uncertainties according to setup and energies
Energy | Normal (%) | Oblique (%) | Wedged (%) |
---|---|---|---|
6 MV | 0.54 | 0.66 | 0.43 |
10 MV | 0.48 | 0.69 | 0.43 |
15 MV | 0.54 | 0.66 | 0.43 |