Progress in Medical Physics 2024; 35(4): 135-144
Published online December 31, 2024
https://doi.org/10.14316/pmp.2024.35.4.135
Copyright © Korean Society of Medical Physics.
Jin Dong Cho , Su Chul Han , Jason Joon Bock Lee , Hyebin Lee , Heerim Nam
Correspondence to:Heerim Nam
(heerim.nam@gmail.com)
Tel: 82-2-2001-1171
Fax: 82-2-2001-1170
This is an Open-Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Purpose: This study compares the dosimetric properties of EBT3 and EBT4 GAFchromic films in transmission and reflection scanning modes, focusing on dose response, sensitivity, and postirradiation stability.
Methods: The EBT3 and EBT4 films were irradiated at doses of 0–10 Gy using a Varian TrueBeam linear accelerator at 6 MV. The films were scanned at intervals between 1 and 336 hours after irradiation in both transmission and reflection modes. Net optical density (NetOD) values from each scan were used to evaluate dose response and sensitivity, with calibration curves created for each film and scan mode. Dose differences between calculated and delivered doses were assessed over time.
Results: The EBT3 and EBT4 films exhibited similar dose–response curves and stable NetOD values across both scanning modes. However, EBT4 exhibited reduced sensitivity variability in response to dose changes. After irradiation, NetOD values increased up to 24 hours before stabilizing, suggesting that a 24-hour scan time is sufficient for consistent measurements. Dose differences between films and modes remained within ±4%.
Conclusions: EBT4 offers comparable dosimetric performance to EBT3, with additional benefits, such as improved dose–response linearity and reduced sensitivity fluctuations. The findings indicate that EBT4 can serve as a reliable successor to EBT3.
KeywordsRadiochromic film, Scan timing, Postirradiation stability, Film dosimetry, Dose sensitivity