Real Time On-line Quality Assurance System for HDR Brachytherapy

고선량률 근접 방사선 치료를 위한 실시간 온-라인 정도 관리(QA) 시스템 개발

Su Jin Lee*, Re Na Lee*, Byong Yong Yi, Sang Wook Lim, and Jin Ho Choi

이수진*·이레나*·이병용·임상욱·최진호

*Department of Radiation Oncology, College of Medicine, Ewha Womans University, Seoul, Korea, Department of Radiation Oncology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea, Department of Radiation Oncology, Gachon Medical School, Gil Medical Center, Inchon, Korea

*이화여자대학교 의과대학 목동병원 방사선종양학과, 울산대학교 의과대학 아산병원 방사선종양학과 가천 의과대학 길병원 방사선종양학과

Abstract

An essential quality assurance (QA) procedure in high dose rate (HDR) remote after-loading brachytherapy is that of the verification of the Ir-192 HDR source positioning accuracy. A number of methods using mechanical rulers or autoradiograph and video cameras have been reported to check the positional error of the Ir-192 source. In this study, the feasibility of a CMOS (Complementary Metal Oxide Semiconductor) PC camera, with a fluorescent screen, was investigated. The agreement between the planned and measured dwell position was better than 1 mm and dwell times better than 0.4 sec. Our results indicate that the CMOS PC camera system could be used as a QA tool for the on-line determination of the source position and dwell time.

KeywordsHigh dose rate brachytherapy, QA, Source positioning accuracy