검색
검색 팝업 닫기

Ex) Article Title, Author, Keywords

Article

Original Article

Korean Journal of Medical Physics 2005; 16(1): 39-46

Published online March 25, 2005

Copyright © Korean Society of Medical Physics.

The Study on the Head and Neck Phantom for QualityAssurance of Intensity Modulated Radiotherapy

세기변조방사선치료의 정도관리를 위한 두경부 팬톰 제작에 관한 연구

Dongho Shin, Sung-Yong Park, Joo Young Kim, Se Byeong Lee, Jung Keun Cho, Dae Yong Kim, Kwan Ho Cho

신동호·박성용·김주영·이세병·조정근·김대용·조관호

Research Institute and Hospital, National Cancer Center

국립암센터 양성자치료센터

Abstract

For the QA of IMRT treatment of head and neck cancer by using M3 (BrainLAB Inc. Germany), it is not easy to measure delivery dose exactly because the dose attenuation appears by the couch according to the position of table and gantry. In order to solve this problem, we fabricated head and neck phantom which would be implemented on the couch mount of Brain Lab Inc. We investigated dose attenuation by the couch and found the difference of dose distribution by the couch, in the applying this phantom to the clinic. After measurement, we found that point dose attenuation was 35% at maximum and dose difference was 5.4% for a point dose measurement of actual patient quality assurance plan.

KeywordsIntensity modulated radiotherapy, Head and neck phantom, Dose attenuation

Korean Society of Medical Physics

Vol.35 No.3
September 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line