Korean Journal of Medical Physics 2009; 20(4): 253-259
Published online December 25, 2009
Copyright © Korean Society of Medical Physics.
Jun Haeng Lee*, Sang Bock Lee*, Kun Jo Chang†, Kwang Yong Lee‡, Hyun Koo Lee‡, Hyeog Ju Kim‡, Gye Hwan Jin*
이준행*ㆍ이상복*ㆍ장근조†ㆍ이광용‡ㆍ이현구‡ㆍ김혁주‡ㆍ진계환*
We have developed standards based on international criterions for the quality control of dose tested by the measurement institutions of individual exposure doses through improving the reliability of data on the exposure dose of individuals working in radioactive environment and securing the accuracy and reliability of individual dose measurements. Laws related to radiation dose applied to domestic institutions refer to ANSI N13.11‐1993, but currently , in U.S. and some other countries the measurement of radiation doses is based on ANSI N13.11‐2001 that reduced test categories and tightened the standards. We made efforts to simplify the standards and to reduce the number of dosimeters required in experiment, and avoided preventing or hindering the use of future technologies not approved under the current law such as glass dosimeter and optical stimulation dosimeter. The Quality Management Manual of Radiation Dosimetry Service, Assessment Manual of Radiation Dosimetry Service Accreditation Program, and the Personnel Dosimetry Performance – Criteria for Testing are documents applicable in supervising laboratories.
KeywordsRadiation dosimetry service, Quality control, Radiation worker, ISO/IEC 17025, ANSI N13.11