Jin Sung Kim, Eunhyuk Shin, Jung Suk Shin, Sang Gyu Ju, Youngyih Han, Hee-Chul Park, Doo Ho Choi
김진성ㆍ신은혁ㆍ신정석ㆍ주상규ㆍ한영이ㆍ박희철ㆍ최두호
Emerging technologies such as four-dimensional computed tomography (4D CT) is expected to allow clinicians to accurately model interfractional motion and to quantitatively estimate internal target volumes (ITVs) for radiation therapy involving moving targets. A need exists for a 4D radiation therapy quality assurance (QA) device that can incorporate and analyze the patient specific intrafractional motion as it relate to dose delivery and respiratory gating. We built a 4D RT prototype device and analyzed the patient-specific 4D radiation therapy QA for 2D dose distributions successfully. With more improvements, the 4D RT QA prototype device could be an integral part of a 4D RT decision process to confirm the dose delivery.
KeywordsRespiratory-gated radiotherapy, Quality assurance, Dose distribution