Korean Journal of Medical Physics 2010; 21(3): 298-303

Published online September 25, 2010

Copyright © Korean Society of Medical Physics.

Dose Distribution of Intensity Modulated Radiation Therapy in Prostate Cancer

전립선암에서 세기조절방사선치료의 선량분포 특성

Sung Kyu Kim, Ji Hoon Choi, Sang Mo Yun

김성규ㆍ최지훈ㆍ윤상모

Department of Therapeutic Radiology and Oncology, College of Medicine, Yeungnam University, Daegu, Korea

영남대학교 의과대학 방사선종양학교실

Abstract

The aim of this study was to compare the dose distribution of intensity modulated radiation therapy (IMRT) with 3 dimensional conformal radiation therapy (3DCRT) in prostate cancer. The IMRT plan and the 3DCRT plan used the 9 fields technique, respectively. In IMRT, tumor dose was a total dose of 66 Gy at 2.0 Gy per day, 5 days a week for 5 weeks. All cases were following the dose volume histogram (DVH) constraints. The maximum and minimum tumor dose constraints were 6,700 cGy and 6,500 cGy, respectively. The rectum dose constraints were <35% over 50 Gy. The bladder dose constraints were <35% over 40 Gy. The femur head dose constraints were <15% over 20 Gy. Tumor dose in the 3DCRT were 66 Gy. In IMRT, the maximum dose of PTV was 104.4% and minimum dose was 89.5% for given dose. In 3DCRT, the maximum dose of PTV was 105.3% and minimum dose was 85.5% for given dose. The rectum dose was 34.0% over 50 Gy in IMRT compared with 63.3% in 3DCRT. The bladder dose was 30.1% over 40 Gy in IMRT compared with 30.6% in 3DCRT. The right femur head dose was 9.5% over 20 Gy in IMRT compared with 17.5% in 3DCRT. The left femur head dose was 10.6% over 20 Gy in IMRT compared with 18.3% in 3 DCRT. The dose of critical organs (rectum, bladder, and femur head) in IMRT showed to reduce than dose of 3DCRT. The rectum dose over 50 Gy in IMRT was reduced 29.3% than 3DCRT. The bladder dose over 40 Gy in IMRT was similar to 3DCRT. The femur head dose over 20 Gy in IMRT was reduced about 7∼8% than 3DCRT.

KeywordsIMRT, 3DCRT, Prostate, Dose distribution, Dose volume histogram

Korean Society of Medical Physics

Vol.35 No.4
December 2024

pISSN 2508-4445
eISSN 2508-4453
Formerly ISSN 1226-5829

Frequency: Quarterly

Current Issue   |   Archives

Stats or Metrics

Share this article on :

  • line