Reference Dosimetry and Calibration of Glass Dosimeters for Cs-137 Gamma-rays

연구용 세슘-137 조사기에 대한 흡수선량 측정과 유리선량계 교정에 관한 연구

Young Min Moon, Dong Joo Rhee*, Jung Ki Kim*, Yeong-Rok Kang*, Man Woo Lee*, Heuijin Lim*, Dong Hyeok Jeong*

문영민ㆍ이동주*ㆍ김정기*ㆍ강영록*ㆍ이만우*ㆍ임희진*ㆍ정동혁*

*Research Center, Dongnam Inst. of Radiological & Medical Sciences, Department of Materials Physics, Dong-A University, Busan, Korea

*동남권원자력의학원 연구센터, 동아대학교 신소재 물리학과

Abstract

In this research, the glass dosimeter was calibrated to measure the standard absorbed dose of the Cs-137 irradiator and absorbed dose in a biological sample. Absorbed dose in water for Cs-137 gamma ray was determined by the IAEA TRS-277 protocol. The PTW-TM30013 ion chamber and the PTW-TM41023 water phantom were utilized for measuring absorbed dose and the value was compared with the reading from DoseAce GD-302M glass dosimeter from Asahi Techno Glass Corporation for its calibration. The uncertainty of measurement (1Փ) of the calibrated glass dosimeter was 2.7% and this result would be applied to improve the accuracy in measurement of absorbed dose in a biological sample.

KeywordsCs-137 irradiator, TRS-277 protocol, Glass dosimeter